Incorporation of wavelet-based denoising in iterative deconvolution for partial volume correction in whole-body PET imaging

Authors: N. Boussion, C. Cheze Le Rest, M. Hatt, D. Visvikis

Published in: European Journal of Nuclear Medicine and Molecular Imaging | Issue 7/2009

Login to get access

Abstract

Purpose

Partial volume effects (PVEs) are consequences of the limited resolution of emission tomography. The aim of the present study was to compare two new voxel-wise PVE correction algorithms based on deconvolution and wavelet-based denoising.

Materials and methods

Deconvolution was performed using the Lucy-Richardson and the Van-Cittert algorithms. Both of these methods were tested using simulated and real FDG PET images. Wavelet-based denoising was incorporated into the process in order to eliminate the noise observed in classical deconvolution methods.

Results

Both deconvolution approaches led to significant intensity recovery, but the Van-Cittert algorithm provided images of inferior qualitative appearance. Furthermore, this method added massive levels of noise, even with the associated use of wavelet-denoising. On the other hand, the Lucy-Richardson algorithm combined with the same denoising process gave the best compromise between intensity recovery, noise attenuation and qualitative aspect of the images.

Conclusion

The appropriate combination of deconvolution and wavelet-based denoising is an efficient method for reducing PVEs in emission tomography.
Literature
1.
go back to reference Kato H, Shimosegawa E, Oku N, Kitagawa K, Kishima H, Saitoh Y, et al. MRI-based correction for partial volume effect improves detectability of intractable epileptogenic foci on 123I-iomazenil brain SPECT images. J Nucl Med 2008;49:383–9.PubMedCrossRef Kato H, Shimosegawa E, Oku N, Kitagawa K, Kishima H, Saitoh Y, et al. MRI-based correction for partial volume effect improves detectability of intractable epileptogenic foci on 123I-iomazenil brain SPECT images. J Nucl Med 2008;49:383–9.PubMedCrossRef
2.
go back to reference Kalpouzos G, Chetelat G, Baron JC, Landeau B, Mevel K, Godeau C, et al. Voxel-based mapping of brain gray matter volume and glucose metabolism profiles in normal aging. Neurobiol Aging 2009;30:112–24.PubMedCrossRef Kalpouzos G, Chetelat G, Baron JC, Landeau B, Mevel K, Godeau C, et al. Voxel-based mapping of brain gray matter volume and glucose metabolism profiles in normal aging. Neurobiol Aging 2009;30:112–24.PubMedCrossRef
3.
go back to reference Mevel K, Desgranges B, Baron JC, Landeau B, De la Sayette V, Viader F, et al. Detecting hippocampal hypometabolism in Mild Cognitive Impairment using automatic voxel-based approaches. Neuroimage 2007;37(1):18–25.PubMedCrossRef Mevel K, Desgranges B, Baron JC, Landeau B, De la Sayette V, Viader F, et al. Detecting hippocampal hypometabolism in Mild Cognitive Impairment using automatic voxel-based approaches. Neuroimage 2007;37(1):18–25.PubMedCrossRef
4.
go back to reference Samuraki M, Matsunari I, Chen WP, Yajima K, Yanase D, Fujikawa A, et al. Partial volume effect-corrected FDG PET and grey matter volume loss in patients with mild Alzheimer’s disease. Eur J Nucl Med Mol Imaging 2007;34(10):1658–69.PubMedCrossRef Samuraki M, Matsunari I, Chen WP, Yajima K, Yanase D, Fujikawa A, et al. Partial volume effect-corrected FDG PET and grey matter volume loss in patients with mild Alzheimer’s disease. Eur J Nucl Med Mol Imaging 2007;34(10):1658–69.PubMedCrossRef
5.
go back to reference Rousset OG, Ma Y, Evans AC. Correction for partial volume effects in PET: principle and validation. J Nucl Med 1998;39(5):904–11.PubMed Rousset OG, Ma Y, Evans AC. Correction for partial volume effects in PET: principle and validation. J Nucl Med 1998;39(5):904–11.PubMed
6.
go back to reference Rousset OG, Collins DL, Rahmin A, Wong DF. Design and implementation of an automated partial volume correction in PET: application to dopamine receptor quantification in the normal human striatum. J Nucl Med 2008;49:1097–106.PubMedCrossRef Rousset OG, Collins DL, Rahmin A, Wong DF. Design and implementation of an automated partial volume correction in PET: application to dopamine receptor quantification in the normal human striatum. J Nucl Med 2008;49:1097–106.PubMedCrossRef
7.
go back to reference Basu S, Alavi A. Feasibility of automated partial-volume correction of SUVs in current PET/CT scanners: can manufacturers provide integrated, ready-to-use software. J Nucl Med 2008;49:1031–2.PubMedCrossRef Basu S, Alavi A. Feasibility of automated partial-volume correction of SUVs in current PET/CT scanners: can manufacturers provide integrated, ready-to-use software. J Nucl Med 2008;49:1031–2.PubMedCrossRef
8.
go back to reference Soret M, Bacharach SL, Buvat I. Partial-volume effect in PET tumor imaging. J Nucl Med 2007;48(6):932–45.PubMedCrossRef Soret M, Bacharach SL, Buvat I. Partial-volume effect in PET tumor imaging. J Nucl Med 2007;48(6):932–45.PubMedCrossRef
9.
go back to reference Boussion N, Hatt M, Lamare F, Bizais Y, Turzo A, Cheze-Le Rest C, et al. A multiresolution image based approach for correction of partial volume effects in emission tomography. Phys Med Biol 2006;51(7):1857–76.PubMedCrossRef Boussion N, Hatt M, Lamare F, Bizais Y, Turzo A, Cheze-Le Rest C, et al. A multiresolution image based approach for correction of partial volume effects in emission tomography. Phys Med Biol 2006;51(7):1857–76.PubMedCrossRef
10.
go back to reference Tohka J, Reilhac A. Deconvolution-based partial volume correction in raclopride-PET and Monte Carlo comparison to MR-based method. Neuroimage 2008;39:1570–84.PubMedCrossRef Tohka J, Reilhac A. Deconvolution-based partial volume correction in raclopride-PET and Monte Carlo comparison to MR-based method. Neuroimage 2008;39:1570–84.PubMedCrossRef
11.
go back to reference Teo BK, Seo Y, Bacharach SL, Carrasquillo JA, Libutti SK, Shukla H, et al. Partial-volume correction in PET: validation of an iterative postreconstruction method with phantom and patient data. J Nucl Med 2007;48(5):802–10.PubMed Teo BK, Seo Y, Bacharach SL, Carrasquillo JA, Libutti SK, Shukla H, et al. Partial-volume correction in PET: validation of an iterative postreconstruction method with phantom and patient data. J Nucl Med 2007;48(5):802–10.PubMed
12.
go back to reference Kirov AS, Piao JZ, Schmidtlein CR. Partial volume effect correction in PET using regularized iterative deconvolution with variance control based on local topology. Phys Med Biol 2008;53:2577–91.PubMedCrossRef Kirov AS, Piao JZ, Schmidtlein CR. Partial volume effect correction in PET using regularized iterative deconvolution with variance control based on local topology. Phys Med Biol 2008;53:2577–91.PubMedCrossRef
13.
go back to reference Van Cittert PH. Zum einfluss der spaltbreite auf die intensita tsverteilung in spektrallinien. Z Physik 1931;69:298.CrossRef Van Cittert PH. Zum einfluss der spaltbreite auf die intensita tsverteilung in spektrallinien. Z Physik 1931;69:298.CrossRef
14.
go back to reference Lucy LB. An iteration technique for the rectification of observed distributions. Astron J 1974;79:745–54.CrossRef Lucy LB. An iteration technique for the rectification of observed distributions. Astron J 1974;79:745–54.CrossRef
15.
go back to reference Richardson WH. Bayesian-based Iterative Method of Image Restoration. J Opt Soc Am 1972;62(1):55–9.CrossRef Richardson WH. Bayesian-based Iterative Method of Image Restoration. J Opt Soc Am 1972;62(1):55–9.CrossRef
16.
go back to reference Chang SG, Yu B, Vetterli M. Adaptive wavelet thresholding for image denoising and compression. IEEE Trans Image Process 2000;9(9):1532–46.PubMedCrossRef Chang SG, Yu B, Vetterli M. Adaptive wavelet thresholding for image denoising and compression. IEEE Trans Image Process 2000;9(9):1532–46.PubMedCrossRef
17.
go back to reference Donoho DL. De-noising by soft-thresholding. IEEE Trans Inf Theory 1995;41(3):613–27.CrossRef Donoho DL. De-noising by soft-thresholding. IEEE Trans Inf Theory 1995;41(3):613–27.CrossRef
18.
go back to reference Donoho DL, Johnstone IM. Ideal spatial adaptation via wavelet shrinkage. Biometrika 1994;81:425–55.CrossRef Donoho DL, Johnstone IM. Ideal spatial adaptation via wavelet shrinkage. Biometrika 1994;81:425–55.CrossRef
19.
go back to reference Turkheimer FE, Aston JA, Asselin MC, Hinz R. Multi-resolution Bayesian regression in PET dynamic studies using wavelets. Neuroimage 2006;32(1):111–21.PubMedCrossRef Turkheimer FE, Aston JA, Asselin MC, Hinz R. Multi-resolution Bayesian regression in PET dynamic studies using wavelets. Neuroimage 2006;32(1):111–21.PubMedCrossRef
20.
go back to reference Kalifa J, Laine A, Esser PD. Regularization in tomographic reconstruction using thresholding estimators. IEEE Trans Image Proc 2003;22(3):351–9.CrossRef Kalifa J, Laine A, Esser PD. Regularization in tomographic reconstruction using thresholding estimators. IEEE Trans Image Proc 2003;22(3):351–9.CrossRef
21.
go back to reference Starck JL, Murtagh F, Bijaoui A. Image processing and data analysis: the multiscale approach. Cambridge: Cambridge University Press; 1998. Starck JL, Murtagh F, Bijaoui A. Image processing and data analysis: the multiscale approach. Cambridge: Cambridge University Press; 1998.
22.
go back to reference Mallat S. A theory for multiresolution signal decomposition: the wavelet representation. IEEE Trans Pattern Anal Mach Intell 1989;11:674–93.CrossRef Mallat S. A theory for multiresolution signal decomposition: the wavelet representation. IEEE Trans Pattern Anal Mach Intell 1989;11:674–93.CrossRef
23.
go back to reference Shensa MJ. Discrete wavelet transform: wedding the à trous and Mallat algorithms. IEEE Trans Signal Proc 1992;40(10):2464–82.CrossRef Shensa MJ. Discrete wavelet transform: wedding the à trous and Mallat algorithms. IEEE Trans Signal Proc 1992;40(10):2464–82.CrossRef
24.
go back to reference Starck JL, Fadili J, Murtagh F. The undecimated wavelet decomposition and its reconstruction. IEEE Trans Image Proc 2007;16(2):297–309.CrossRef Starck JL, Fadili J, Murtagh F. The undecimated wavelet decomposition and its reconstruction. IEEE Trans Image Proc 2007;16(2):297–309.CrossRef
25.
go back to reference Visvikis D, Turzo A, Gouret S, Damien P, Lamare F, Bizais Y, et al. Characterisation of SUV accuracy in FDG PET using 3D RAMLA and the Philips Allegro PET scanner. J Nucl Med 2004;45:103P. Visvikis D, Turzo A, Gouret S, Damien P, Lamare F, Bizais Y, et al. Characterisation of SUV accuracy in FDG PET using 3D RAMLA and the Philips Allegro PET scanner. J Nucl Med 2004;45:103P.
26.
go back to reference Ramani S, Blu T, Unser M. Monte-Carlo SURE: a black-box optimization of regularization parameters for general denoising algorithms. IEEE Trans Image Proc 2008;17:1540–54.CrossRef Ramani S, Blu T, Unser M. Monte-Carlo SURE: a black-box optimization of regularization parameters for general denoising algorithms. IEEE Trans Image Proc 2008;17:1540–54.CrossRef
Metadata
Title
Incorporation of wavelet-based denoising in iterative deconvolution for partial volume correction in whole-body PET imaging
Authors
N. Boussion
C. Cheze Le Rest
M. Hatt
D. Visvikis
Publication date
01-07-2009
Publisher
Springer-Verlag
Published in
European Journal of Nuclear Medicine and Molecular Imaging / Issue 7/2009
Print ISSN: 1619-7070
Electronic ISSN: 1619-7089
DOI
https://doi.org/10.1007/s00259-009-1065-5

Other articles of this Issue 7/2009

European Journal of Nuclear Medicine and Molecular Imaging 7/2009 Go to the issue