Novel detector technology for clinical PET

Author: Roger Lecomte

Published in: European Journal of Nuclear Medicine and Molecular Imaging | Special Issue 1/2009

Login to get access

Abstract

Introduction

Positron emission tomography (PET) is the most sensitive of all medical imaging modalities for quantitatively probing biologic processes at the molecular level. However, spatial resolution in PET is significantly inferior to that of other imaging modalities that can provide exquisite images of the anatomy, such as X-ray computed tomography (CT) or magnetic resonance (MR) imaging.

Objective

It has been one of the outstanding challenges of the last decade to combine PET with these complementary imaging modalities in order to synergistically exploit the benefits of each modality and to enhance the role of PET in pre-clinical research as well as in clinical routine and research.

Discussion

The simple juxtaposition of tomographs around a common axial bed, such as with current PET/CT technology, is very successful in allowing sequential acquisition of PET and anatomical data. However, novel imaging combinations are being considered that would enable simultaneous, or at least concurrent, dual-modality imaging through combined PET/MR or PET/CT. The development of these new integrated instruments creates new bewildering challenges for PET detection systems, which, in addition to the ability to measure annihilation radiation in PET, must satisfy several other critical requirements.
Literature
1.
go back to reference Derenzo SE, Moses WW, Huesman RH, Budinger TF. Critical instrumentation issues for resolution smaller than 2 mm, high sensitivity brain PET. In: Uemura K, Lassen NA, Jones T, Kanno I, editors. Quantification of brain function, tracer kinetics and image analysis in brain PET. Amsterdam: Excerpta Medica; 1993. p. 25–40. Derenzo SE, Moses WW, Huesman RH, Budinger TF. Critical instrumentation issues for resolution smaller than 2 mm, high sensitivity brain PET. In: Uemura K, Lassen NA, Jones T, Kanno I, editors. Quantification of brain function, tracer kinetics and image analysis in brain PET. Amsterdam: Excerpta Medica; 1993. p. 25–40.
2.
go back to reference Levin CS, Zaidi H. Current trends in preclinical PET system design. PET Clin 2007;2:125–60.CrossRef Levin CS, Zaidi H. Current trends in preclinical PET system design. PET Clin 2007;2:125–60.CrossRef
3.
go back to reference St.James S, Thompson CJ. Image blurring due to light-sharing in PET block detectors. Med Phys 2006;33 2:405–10.PubMedCrossRef St.James S, Thompson CJ. Image blurring due to light-sharing in PET block detectors. Med Phys 2006;33 2:405–10.PubMedCrossRef
4.
go back to reference Lecomte R. Technology challenges in small animal PET imaging. Nucl Instrum Meth Phys Res A 2004;527:157–65.CrossRef Lecomte R. Technology challenges in small animal PET imaging. Nucl Instrum Meth Phys Res A 2004;527:157–65.CrossRef
5.
go back to reference Derenzo SE. Mathematical removal of positron range blurring in high resolution tomography. IEEE Trans Nucl Sci 1986;33 1:565–9.CrossRef Derenzo SE. Mathematical removal of positron range blurring in high resolution tomography. IEEE Trans Nucl Sci 1986;33 1:565–9.CrossRef
6.
go back to reference Levin CS, Hoffman EJ. Calculation of positron range and its effect on the fundamental limit of positron emission tomography system spatial resolution. Phys Med Biol 1999;44 3:781–99.PubMedCrossRef Levin CS, Hoffman EJ. Calculation of positron range and its effect on the fundamental limit of positron emission tomography system spatial resolution. Phys Med Biol 1999;44 3:781–99.PubMedCrossRef
7.
go back to reference Bentourkia M, Msaki P, Cadorette J, Lecomte R. Assessment of scatter components in high-resolution PET: correction by nonstationary convolution subtraction. J Nucl Med 1995;36 1:121–30.PubMed Bentourkia M, Msaki P, Cadorette J, Lecomte R. Assessment of scatter components in high-resolution PET: correction by nonstationary convolution subtraction. J Nucl Med 1995;36 1:121–30.PubMed
8.
go back to reference Hoffman EJ, Phelps ME. Positron emission tomography: principles and quantitation. In: Phelps ME, Mazziotta JC, Shelbert HR, editors. Positron emission tomography and autoradiography: principles and applications for the brain and heart. Chap. 6. New York: Raven; 1986. p. 237–86. Hoffman EJ, Phelps ME. Positron emission tomography: principles and quantitation. In: Phelps ME, Mazziotta JC, Shelbert HR, editors. Positron emission tomography and autoradiography: principles and applications for the brain and heart. Chap. 6. New York: Raven; 1986. p. 237–86.
9.
go back to reference Strother SC, Casey ME, Hoffman EJ. Measuring PET scanner sensitivity: relating countrates to image signal-to-noise ratios using noise equivalent counts. IEEE Trans Nucl Sci 1990;37 1:783–8.CrossRef Strother SC, Casey ME, Hoffman EJ. Measuring PET scanner sensitivity: relating countrates to image signal-to-noise ratios using noise equivalent counts. IEEE Trans Nucl Sci 1990;37 1:783–8.CrossRef
10.
go back to reference Budinger TF. Time-of-flight positron emission tomography: status relative to conventional PET. J Nucl Med 1983;24:73–8.PubMed Budinger TF. Time-of-flight positron emission tomography: status relative to conventional PET. J Nucl Med 1983;24:73–8.PubMed
11.
12.
go back to reference Kinahan PE, Townsend DW, Beyer T, Sashin D. Attenuation correction for a combined 3D PET/CT scanner. Med Phys 1998;25:2046–53.PubMedCrossRef Kinahan PE, Townsend DW, Beyer T, Sashin D. Attenuation correction for a combined 3D PET/CT scanner. Med Phys 1998;25:2046–53.PubMedCrossRef
13.
go back to reference Soret M, Bacharach SL, Buvat I. Partial-volume effect in PET tumor imaging. J Nucl Med 2007;48:932–45.PubMedCrossRef Soret M, Bacharach SL, Buvat I. Partial-volume effect in PET tumor imaging. J Nucl Med 2007;48:932–45.PubMedCrossRef
14.
go back to reference Townsend DW, Isoardi RA, Bendriem B. Volume imaging tomographs. In: Bendriem B, Townsend DW, editors. The theory and practice of 3D PET. Dordrecht: Kluwer; 1998. p. 111–32. Townsend DW, Isoardi RA, Bendriem B. Volume imaging tomographs. In: Bendriem B, Townsend DW, editors. The theory and practice of 3D PET. Dordrecht: Kluwer; 1998. p. 111–32.
15.
go back to reference Strobel K, Rudy M, Treyer V, Veit-Haibach P, Burger C, Hany TF. Objective and subjective comparison of standard 2-D and fully 3-D reconstructed data on a PET/CT system. Nucl Med Commun 2007;28:555–9.PubMedCrossRef Strobel K, Rudy M, Treyer V, Veit-Haibach P, Burger C, Hany TF. Objective and subjective comparison of standard 2-D and fully 3-D reconstructed data on a PET/CT system. Nucl Med Commun 2007;28:555–9.PubMedCrossRef
16.
go back to reference Eriksson L, Townsend D, Conti M, Eriksson M, Rothfuss H, Schmand M, et al. An investigation of sensitivity limits in PET scanners. Nucl Instrum Meth Phys Res A 2007;580 2:836–42.CrossRef Eriksson L, Townsend D, Conti M, Eriksson M, Rothfuss H, Schmand M, et al. An investigation of sensitivity limits in PET scanners. Nucl Instrum Meth Phys Res A 2007;580 2:836–42.CrossRef
17.
go back to reference Moses WW. Time of flight in PET revisited. IEEE Trans Nucl Sci 2003;50 5:1325–30.CrossRef Moses WW. Time of flight in PET revisited. IEEE Trans Nucl Sci 2003;50 5:1325–30.CrossRef
18.
go back to reference Surti S, Karp S, Popescu LM, Daube-Witherspoon E, Werner M. Investigation of time-of-flight benefit for fully 3-D PET. IEEE Trans Med Imaging 2006;25 5:529–38.PubMedCrossRef Surti S, Karp S, Popescu LM, Daube-Witherspoon E, Werner M. Investigation of time-of-flight benefit for fully 3-D PET. IEEE Trans Med Imaging 2006;25 5:529–38.PubMedCrossRef
19.
go back to reference Kuhn A, Surti S, Karp JS, Muehllehner G, Newcomer FM, VanBerg R. Performance assessment of pixelated LaBr3 detector modules for time-of-flight PET. IEEE Trans Nucl Sci 2006;53 3:1090–5.CrossRef Kuhn A, Surti S, Karp JS, Muehllehner G, Newcomer FM, VanBerg R. Performance assessment of pixelated LaBr3 detector modules for time-of-flight PET. IEEE Trans Nucl Sci 2006;53 3:1090–5.CrossRef
20.
go back to reference Moses WW. Recent advances and future advances in time-of-flight PET. Nucl Instrum Meth Phys Res A 2007;580 2:919–24.CrossRef Moses WW. Recent advances and future advances in time-of-flight PET. Nucl Instrum Meth Phys Res A 2007;580 2:919–24.CrossRef
21.
go back to reference Surti S, Kuhn A, Werner ME, Perkins AE, Kolthammer J, Karp JS. Performance of Philips Gemini TF PET/CT scanner with special consideration for its time-of-flight imaging capabilities. J Nucl Med 2007;48:471–80.PubMed Surti S, Kuhn A, Werner ME, Perkins AE, Kolthammer J, Karp JS. Performance of Philips Gemini TF PET/CT scanner with special consideration for its time-of-flight imaging capabilities. J Nucl Med 2007;48:471–80.PubMed
22.
go back to reference Surti S, Karp JS, Adam LE, Muehllchner G. Performance measurements for the GSO-based brain PET camera (G-PET). IEEE Nucl Sci Symp Conf Rec 2001;2:1109–14. Surti S, Karp JS, Adam LE, Muehllchner G. Performance measurements for the GSO-based brain PET camera (G-PET). IEEE Nucl Sci Symp Conf Rec 2001;2:1109–14.
23.
go back to reference Watanabe M, Shimizu K, Omura T, Takahashi M, Kosugi T, Yoshikawa E, et al. A new high-resolution PET scanner dedicated to brain research. IEEE Trans Nucl Sci 2002;49 3:634–9.CrossRef Watanabe M, Shimizu K, Omura T, Takahashi M, Kosugi T, Yoshikawa E, et al. A new high-resolution PET scanner dedicated to brain research. IEEE Trans Nucl Sci 2002;49 3:634–9.CrossRef
24.
go back to reference Wienhard K, Schmand M, Casey ME, Baker K, Bao J, Eriksson L, et al. The ECAT HRRT: performance and first clinical application of the new high resolution research tomograph. IEEE Trans Nucl Sci 2002;49 1:104–10.CrossRef Wienhard K, Schmand M, Casey ME, Baker K, Bao J, Eriksson L, et al. The ECAT HRRT: performance and first clinical application of the new high resolution research tomograph. IEEE Trans Nucl Sci 2002;49 1:104–10.CrossRef
25.
go back to reference de Jong HW, van Velden FH, Kloet RW, Buijs FL, Boellaard R, Lammertsma AA. Performance evaluation of the ECAT HRRT: an LSO–LYSO double layer high resolution, high sensitivity scanner. Phys Med Biol 2007;52 5:1505–26.PubMedCrossRef de Jong HW, van Velden FH, Kloet RW, Buijs FL, Boellaard R, Lammertsma AA. Performance evaluation of the ECAT HRRT: an LSO–LYSO double layer high resolution, high sensitivity scanner. Phys Med Biol 2007;52 5:1505–26.PubMedCrossRef
26.
go back to reference Thompson CJ, Murthy K, Picard Y, Weinberg IN, Mako R. Positron emission mammography (PEM): a promising technique for detecting breast cancer. IEEE Trans Nucl Sci 1995;42 4:1012–7.CrossRef Thompson CJ, Murthy K, Picard Y, Weinberg IN, Mako R. Positron emission mammography (PEM): a promising technique for detecting breast cancer. IEEE Trans Nucl Sci 1995;42 4:1012–7.CrossRef
27.
go back to reference Doshi NK, Silverman RW, Shao Y, Cherry SR. maxPET, a dedicated mammary and axillary region PET imaging system for breast cancer. IEEE Trans Nucl Sci 2001;48 3:811–5.CrossRef Doshi NK, Silverman RW, Shao Y, Cherry SR. maxPET, a dedicated mammary and axillary region PET imaging system for breast cancer. IEEE Trans Nucl Sci 2001;48 3:811–5.CrossRef
28.
go back to reference Abreu MC, Aguiar JD, Almeida FG, Almeida P, Bento P, Carrico B, et al. Design and evaluation of the Clear-PEM scanner for positron emission mammography. IEEE Trans Nucl Sci 2006;53 1:71–7.CrossRef Abreu MC, Aguiar JD, Almeida FG, Almeida P, Bento P, Carrico B, et al. Design and evaluation of the Clear-PEM scanner for positron emission mammography. IEEE Trans Nucl Sci 2006;53 1:71–7.CrossRef
29.
go back to reference Raylman RR, Majewski S, Smith MF, Proffitt J, Hammond W, Srinivasan A, et al. The positron emission mammography/tomography breast imaging and biopsy system (PEM/PET): design, construction and phantom-based measurements. Phys Med Biol 2008;53 3:637–53.PubMedCrossRef Raylman RR, Majewski S, Smith MF, Proffitt J, Hammond W, Srinivasan A, et al. The positron emission mammography/tomography breast imaging and biopsy system (PEM/PET): design, construction and phantom-based measurements. Phys Med Biol 2008;53 3:637–53.PubMedCrossRef
30.
go back to reference Beyer T, Townsend DW, Brun T, Kinahan PE, Charron M, Roddy R, et al. A combined PET/CT scanner for clinical oncology. J Nucl Med 2000;41:1369–79.PubMed Beyer T, Townsend DW, Brun T, Kinahan PE, Charron M, Roddy R, et al. A combined PET/CT scanner for clinical oncology. J Nucl Med 2000;41:1369–79.PubMed
31.
32.
go back to reference Goertzen AL, Meadors AK, Silverman RW, Cherry SR. Simultaneous molecular and anatomical imaging of the mouse in vivo. Phys Med Biol 2002;47 24:4315–28.PubMedCrossRef Goertzen AL, Meadors AK, Silverman RW, Cherry SR. Simultaneous molecular and anatomical imaging of the mouse in vivo. Phys Med Biol 2002;47 24:4315–28.PubMedCrossRef
33.
go back to reference Domenico GD, Cesca N, Zavattini G, Auricchio N, Gambaccini M. CT with a CMOS flat panel detector integrated on the YAP-(S)PET scanner for in vivo small animal imaging. Nucl Instrum Meth Phys Res A 2007;571:110–3.CrossRef Domenico GD, Cesca N, Zavattini G, Auricchio N, Gambaccini M. CT with a CMOS flat panel detector integrated on the YAP-(S)PET scanner for in vivo small animal imaging. Nucl Instrum Meth Phys Res A 2007;571:110–3.CrossRef
34.
go back to reference Khodaverdi M, Nicol S, Loess J, Cassol Brunner F, Karkar S, Morel C. Design study for the ClearPET/XPAD small animal PET/CT scanner. IEEE Nucl Sci Symp Conf Rec 2007;6:4300–2. Khodaverdi M, Nicol S, Loess J, Cassol Brunner F, Karkar S, Morel C. Design study for the ClearPET/XPAD small animal PET/CT scanner. IEEE Nucl Sci Symp Conf Rec 2007;6:4300–2.
35.
go back to reference Hasegawa BH, Gingold EL, Reilly SM, Liew SC, Cann CE. Description of a simultaneous emission–transmission CT system. Proc SPIE 1990;1231:50–60.CrossRef Hasegawa BH, Gingold EL, Reilly SM, Liew SC, Cann CE. Description of a simultaneous emission–transmission CT system. Proc SPIE 1990;1231:50–60.CrossRef
36.
go back to reference Lang TF, Hasegawa BH, Liew SC, Brown JK, Blankespoor SC, Reilly SM, et al. Description of a prototype emission–transmission computed tomography imaging system. J Nucl Med 1992;33 10:1881–7.PubMed Lang TF, Hasegawa BH, Liew SC, Brown JK, Blankespoor SC, Reilly SM, et al. Description of a prototype emission–transmission computed tomography imaging system. J Nucl Med 1992;33 10:1881–7.PubMed
37.
go back to reference Saoudi A, Lecomte R. A novel APD-based detector module for multi-modality PET/SPECT/CT scanners. IEEE Trans Nucl Sci 1999;46 3:479–84.CrossRef Saoudi A, Lecomte R. A novel APD-based detector module for multi-modality PET/SPECT/CT scanners. IEEE Trans Nucl Sci 1999;46 3:479–84.CrossRef
38.
go back to reference Cirignano L, Shah KS, Bennett PR, Klugerman M, Dmitryev Y, Squillante MR, et al. Pixellated CdZnTe detector for emission/transmission computed tomography. Nucl Instrum Meth Phys Res A 1999;422:216–20.CrossRef Cirignano L, Shah KS, Bennett PR, Klugerman M, Dmitryev Y, Squillante MR, et al. Pixellated CdZnTe detector for emission/transmission computed tomography. Nucl Instrum Meth Phys Res A 1999;422:216–20.CrossRef
39.
go back to reference Bérard P, Pepin CM, Rouleau D, Cadorette J, Lecomte R. CT acquisition using PET detectors and electronics. IEEE Trans Nucl Sci 2005;52 3:634–7.CrossRef Bérard P, Pepin CM, Rouleau D, Cadorette J, Lecomte R. CT acquisition using PET detectors and electronics. IEEE Trans Nucl Sci 2005;52 3:634–7.CrossRef
40.
go back to reference Nassalski A, Moszynski M, Szczesniak T, et al. The road to the common PET/CT detector. IEEE Trans Nucl Sci 2007;54 5:1459–63.CrossRef Nassalski A, Moszynski M, Szczesniak T, et al. The road to the common PET/CT detector. IEEE Trans Nucl Sci 2007;54 5:1459–63.CrossRef
41.
go back to reference Nassalski A, Kapusta M, Batsch T, Wolski D, Mockel D, Enghardt W, et al. Comparative study of scintillators for PET/CT detectors. IEEE Trans Nucl Sci 2007;54 1:3–10.CrossRef Nassalski A, Kapusta M, Batsch T, Wolski D, Mockel D, Enghardt W, et al. Comparative study of scintillators for PET/CT detectors. IEEE Trans Nucl Sci 2007;54 1:3–10.CrossRef
42.
go back to reference Powolny F, Auffray E, Hillemanns H, Jarron P, Lecoq P, Meyer TC, et al. Development of a new photo-detector readout technique for PET and CT. AIP Conf Proc 2007;958 1:260–1.CrossRef Powolny F, Auffray E, Hillemanns H, Jarron P, Lecoq P, Meyer TC, et al. Development of a new photo-detector readout technique for PET and CT. AIP Conf Proc 2007;958 1:260–1.CrossRef
43.
go back to reference Kruger H, Fink J, Kraft E, Wermes N, Fischer P, Peric I. CIX: a detector for spectrally enhanced X-ray imaging by simultaneous counting and integrating. Proc SPIE 2008;6913:69130P–1–12. Kruger H, Fink J, Kraft E, Wermes N, Fischer P, Peric I. CIX: a detector for spectrally enhanced X-ray imaging by simultaneous counting and integrating. Proc SPIE 2008;6913:69130P–1–12.
44.
go back to reference Bérard P, Bergeron M, Pepin CM, Cadorette J, Tétrault M-A, Viscogliosi N, et al. LabPET II, a novel 64-channel APD-based PET detector module with individual pixel readout achieving submillimetric spatial resolution IEEE Nucl Sci Symp Conf Rec 2008, M12-6; in press. Bérard P, Bergeron M, Pepin CM, Cadorette J, Tétrault M-A, Viscogliosi N, et al. LabPET II, a novel 64-channel APD-based PET detector module with individual pixel readout achieving submillimetric spatial resolution IEEE Nucl Sci Symp Conf Rec 2008, M12-6; in press.
45.
go back to reference Fontaine R, Bélanger F, Cadorette J, Leroux JD, Martin JP, Michaud JB, et al. Architecture of a dual-modality, high-resolution, fully digital positron emission tomography/computed tomography (PET/CT) scanner for small animal imaging. IEEE Trans Nucl Sci 2005;52 3:691–6.CrossRef Fontaine R, Bélanger F, Cadorette J, Leroux JD, Martin JP, Michaud JB, et al. Architecture of a dual-modality, high-resolution, fully digital positron emission tomography/computed tomography (PET/CT) scanner for small animal imaging. IEEE Trans Nucl Sci 2005;52 3:691–6.CrossRef
46.
go back to reference Meyer TC, Powolny F, Anghinolfi F, Auffray E, Dosanjh M, Hillemanns H, et al. A time-based front end readout system for PET & CT. IEEE Nucl Sci Symp Conf Rec 2006;4:2494–8. Meyer TC, Powolny F, Anghinolfi F, Auffray E, Dosanjh M, Hillemanns H, et al. A time-based front end readout system for PET & CT. IEEE Nucl Sci Symp Conf Rec 2006;4:2494–8.
47.
go back to reference Riendeau J, Bérard P, Viscogliosi N, Tétrault MA, Lemieux F, Lecomte R, et al. High rate photon counting CT using parallel digital PET electronics. IEEE Trans Nucl Sci 2008;55 1:40–7.CrossRef Riendeau J, Bérard P, Viscogliosi N, Tétrault MA, Lemieux F, Lecomte R, et al. High rate photon counting CT using parallel digital PET electronics. IEEE Trans Nucl Sci 2008;55 1:40–7.CrossRef
48.
go back to reference Thibaudeau C, Bérard P, Tétrault M-A, Leroux J-D, Fontaine R, Lecomte R. Iterative CT reconstruction using LabPET™ detector modules. IEEE Nucl Sci Symp Conf Rec 2008, M06-361; in press. Thibaudeau C, Bérard P, Tétrault M-A, Leroux J-D, Fontaine R, Lecomte R. Iterative CT reconstruction using LabPET™ detector modules. IEEE Nucl Sci Symp Conf Rec 2008, M06-361; in press.
49.
go back to reference Francke T, Eklund M, Ericsson L, Kristoffersson T, Peskov V, Rantanen J, et al. Dose reduction in medical X-ray imaging using noise free photon counting. Nucl Instrum Meth Phys Res A 2001;471:85–7.CrossRef Francke T, Eklund M, Ericsson L, Kristoffersson T, Peskov V, Rantanen J, et al. Dose reduction in medical X-ray imaging using noise free photon counting. Nucl Instrum Meth Phys Res A 2001;471:85–7.CrossRef
50.
go back to reference Bérard P, Riendeau J, Pepin CM, Rouleau D, Cadorette J, Fontaine R, et al. Investigation of the LabPET™ detector and electronics for photon-counting CT imaging. Nucl Instrum Meth Phys Res A 2007;571:114–7.CrossRef Bérard P, Riendeau J, Pepin CM, Rouleau D, Cadorette J, Fontaine R, et al. Investigation of the LabPET™ detector and electronics for photon-counting CT imaging. Nucl Instrum Meth Phys Res A 2007;571:114–7.CrossRef
51.
go back to reference Schlomka JP, Roessl E, Dorscheid R, Dill S, Martens G, Istel T, et al. Experimental feasibility of multi-energy photon-counting K-edge imaging in pre-clinical computed tomography. Phys Med Biol 2008;53:4031–47.PubMedCrossRef Schlomka JP, Roessl E, Dorscheid R, Dill S, Martens G, Istel T, et al. Experimental feasibility of multi-energy photon-counting K-edge imaging in pre-clinical computed tomography. Phys Med Biol 2008;53:4031–47.PubMedCrossRef
52.
go back to reference Frey EC, Wang X, Du Y, Taguchi K, Xu J, Tsui BMW. Investigation of the use of photon counting x-ray detectors with energy discrimination capability for material decomposition in micro-computed tomography. Proc SPIE 2007;6510:65100A.CrossRef Frey EC, Wang X, Du Y, Taguchi K, Xu J, Tsui BMW. Investigation of the use of photon counting x-ray detectors with energy discrimination capability for material decomposition in micro-computed tomography. Proc SPIE 2007;6510:65100A.CrossRef
53.
go back to reference Cardi CA, Cao Z, Thakur ML, Karp JS, Acton PD. Pinhole PET (pPET): a multi-pinhole collimator insert for small animal SPECT imaging on PET cameras. IEEE Nucl Sci Symp Conf Rec 2005;4:4. Cardi CA, Cao Z, Thakur ML, Karp JS, Acton PD. Pinhole PET (pPET): a multi-pinhole collimator insert for small animal SPECT imaging on PET cameras. IEEE Nucl Sci Symp Conf Rec 2005;4:4.
54.
go back to reference Cao Z, Bal G, Accorsi R, Acton PD. Optimal number of pinholes in multi-pinhole SPECT for mouse brain imaging—a simulation study. Phys Med Biol 2005;50:4609–24.PubMedCrossRef Cao Z, Bal G, Accorsi R, Acton PD. Optimal number of pinholes in multi-pinhole SPECT for mouse brain imaging—a simulation study. Phys Med Biol 2005;50:4609–24.PubMedCrossRef
55.
go back to reference Del Guerra A, Bartoli A, Belcari N, Herbert D, Motta A, Vaiano A, et al. Performance evaluation of the fully engineered YAP-(S)PET scanner for small animal imaging. IEEE Trans Nucl Sci 2006;53 3:1078–83.CrossRef Del Guerra A, Bartoli A, Belcari N, Herbert D, Motta A, Vaiano A, et al. Performance evaluation of the fully engineered YAP-(S)PET scanner for small animal imaging. IEEE Trans Nucl Sci 2006;53 3:1078–83.CrossRef
56.
go back to reference Shao Y, Yao R, Ma T, Manchiraju P. Initial studies of PET–SPECT dual-tracer imaging. IEEE Nucl Sci Symp Conf Rec 2007;6:4198–204. Shao Y, Yao R, Ma T, Manchiraju P. Initial studies of PET–SPECT dual-tracer imaging. IEEE Nucl Sci Symp Conf Rec 2007;6:4198–204.
57.
go back to reference Thompson CJ, Murthy K, Weinberg IN, Mako F. Feasibility study for positron emission mammography. Med Phys 1994;21 4:529–38.PubMedCrossRef Thompson CJ, Murthy K, Weinberg IN, Mako F. Feasibility study for positron emission mammography. Med Phys 1994;21 4:529–38.PubMedCrossRef
58.
go back to reference Bergman AM, Thompson CJ, Murthy K, Robar JL, Clancy RL, English MJ, et al. Technique to obtain positron emission mammography images in registration with X-ray mammograms. Med Phys 1998;25 11:2119–29.PubMedCrossRef Bergman AM, Thompson CJ, Murthy K, Robar JL, Clancy RL, English MJ, et al. Technique to obtain positron emission mammography images in registration with X-ray mammograms. Med Phys 1998;25 11:2119–29.PubMedCrossRef
59.
go back to reference Huber JS, Moses WW, Pouliot J, Hsu IC. Dual-modality PET/ultrasound imaging of the prostate. IEEE Nucl Sci Symp Conf Rec 2005;4:2187–90.CrossRef Huber JS, Moses WW, Pouliot J, Hsu IC. Dual-modality PET/ultrasound imaging of the prostate. IEEE Nucl Sci Symp Conf Rec 2005;4:2187–90.CrossRef
60.
go back to reference Huber JS, Moses WW, Peng Q, Huesman RH, Reutter BW, Wilson DS, et al. Dual-modality prostate imaging with PET and transrectal ultrasound. IEEE Nucl Sci Symp 2008. Huber JS, Moses WW, Peng Q, Huesman RH, Reutter BW, Wilson DS, et al. Dual-modality prostate imaging with PET and transrectal ultrasound. IEEE Nucl Sci Symp 2008.
61.
go back to reference Lecoq P, Mundler O. The ClearPEM-Sonic project, a dual modality approach of breast cancer. Eur J Nucl Med. 2008; in press. Lecoq P, Mundler O. The ClearPEM-Sonic project, a dual modality approach of breast cancer. Eur J Nucl Med. 2008; in press.
62.
go back to reference Prout DL, Silverman RW, Chatziioannou A. Detector concept for OPET-a combined PET and optical imaging system. IEEE Trans Nucl Sci 2004;51 3:752–6.PubMedCrossRef Prout DL, Silverman RW, Chatziioannou A. Detector concept for OPET-a combined PET and optical imaging system. IEEE Trans Nucl Sci 2004;51 3:752–6.PubMedCrossRef
63.
go back to reference Takahashi K, Inadama N, Murayama H, Yamaya T, Yoshida E, Nishikido F, et al. Preliminary study of a DOI-PET detector with optical imaging capability. IEEE Nucl Sci Symp Conf Rec 2007;5:3318–21. Takahashi K, Inadama N, Murayama H, Yamaya T, Yoshida E, Nishikido F, et al. Preliminary study of a DOI-PET detector with optical imaging capability. IEEE Nucl Sci Symp Conf Rec 2007;5:3318–21.
64.
go back to reference Culver J, Akers W, Achilefu S. Multimodality molecular imaging with combined optical and SPECT/PET modalities. J Nucl Med 2008;49 2:169–72.PubMedCrossRef Culver J, Akers W, Achilefu S. Multimodality molecular imaging with combined optical and SPECT/PET modalities. J Nucl Med 2008;49 2:169–72.PubMedCrossRef
65.
go back to reference van Eijk CWE. Inorganic scintillators in medical imaging. Phys Med Biol 2002;47 8:85–106.CrossRef van Eijk CWE. Inorganic scintillators in medical imaging. Phys Med Biol 2002;47 8:85–106.CrossRef
66.
go back to reference Korzhika M, Fedorova A, Annenkovb A, Borissevitcha A, Dossovitskic A, Missevitcha O, et al. Development of scintillation materials for PET scanners. Nucl Instrum Meth Phys Res A 2007;571:122–5.CrossRef Korzhika M, Fedorova A, Annenkovb A, Borissevitcha A, Dossovitskic A, Missevitcha O, et al. Development of scintillation materials for PET scanners. Nucl Instrum Meth Phys Res A 2007;571:122–5.CrossRef
67.
go back to reference Derenzo SE, Weber MJ, Bourret-Courchesne E, Klintenberg MK. The quest for the ideal inorganic scintillator. Nucl Instrum Meth Phys Res A 2003;505:111–7.CrossRef Derenzo SE, Weber MJ, Bourret-Courchesne E, Klintenberg MK. The quest for the ideal inorganic scintillator. Nucl Instrum Meth Phys Res A 2003;505:111–7.CrossRef
68.
go back to reference Cho ZH, Farukhi MR. Bismuth germanate as a potential scintillation detector in positron cameras. J Nucl Med 1977;18 8:840–4.PubMed Cho ZH, Farukhi MR. Bismuth germanate as a potential scintillation detector in positron cameras. J Nucl Med 1977;18 8:840–4.PubMed
69.
70.
go back to reference Melcher CL, Friedrich S, Cramer SP, Spurrier MA, Szupryczynski P, Nutt R. Cerium oxidation state in LSO:Ce scintillators. IEEE Trans Nucl Sci 2005;52 5:1809–12.CrossRef Melcher CL, Friedrich S, Cramer SP, Spurrier MA, Szupryczynski P, Nutt R. Cerium oxidation state in LSO:Ce scintillators. IEEE Trans Nucl Sci 2005;52 5:1809–12.CrossRef
71.
go back to reference Melcher CL. Perspectives on the future development of new scintillators. Nucl Instrum Meth Phys Res A 2005;537:6–14.CrossRef Melcher CL. Perspectives on the future development of new scintillators. Nucl Instrum Meth Phys Res A 2005;537:6–14.CrossRef
72.
go back to reference Karp JS, Surti S, Daube-Witherspoon ME, Muehllehner G. Benefit of time-of-flight in PET: experimental and clinical results. J Nucl Med 2008;49 3:462–70.PubMedCrossRef Karp JS, Surti S, Daube-Witherspoon ME, Muehllehner G. Benefit of time-of-flight in PET: experimental and clinical results. J Nucl Med 2008;49 3:462–70.PubMedCrossRef
73.
go back to reference Moses WW, Ullisch M. Factors influencing timing resolution in a commercial LSO PET camera. IEEE Trans Nucl Sci 2006;53 1:78–85.CrossRef Moses WW, Ullisch M. Factors influencing timing resolution in a commercial LSO PET camera. IEEE Trans Nucl Sci 2006;53 1:78–85.CrossRef
74.
go back to reference van Loef EVD, Dorenbos P, van Eijk CWE, Kramer K, Gudel HU. High-energy-resolution scintillator: Ce3+ activated LaBr3. Appl Phys Lett 2001;79 10:1573–5.CrossRef van Loef EVD, Dorenbos P, van Eijk CWE, Kramer K, Gudel HU. High-energy-resolution scintillator: Ce3+ activated LaBr3. Appl Phys Lett 2001;79 10:1573–5.CrossRef
75.
go back to reference Shah KS, Glodo J, Klugerman M, Moses WW, Derenzo SE, Weber MJ. LaBr3:Ce scintillators for gamma-ray spectroscopy. IEEE Trans Nucl Sci 2003;50 6:2410–3.CrossRef Shah KS, Glodo J, Klugerman M, Moses WW, Derenzo SE, Weber MJ. LaBr3:Ce scintillators for gamma-ray spectroscopy. IEEE Trans Nucl Sci 2003;50 6:2410–3.CrossRef
76.
go back to reference Kuhn A, Surti S, Karp JS, Raby PS, Shah KS, Perkins AE, et al. Design of a lanthanum bromide detector for time-of-flight PET. IEEE Trans Nucl Sci 2004;51 5:2550–7.CrossRef Kuhn A, Surti S, Karp JS, Raby PS, Shah KS, Perkins AE, et al. Design of a lanthanum bromide detector for time-of-flight PET. IEEE Trans Nucl Sci 2004;51 5:2550–7.CrossRef
77.
go back to reference Yamamoto S, Kuroda K, Senda M. Scintillator selection for MR-compatible gamma detectors. IEEE Trans Nucl Sci 2003;50:1683–5.CrossRef Yamamoto S, Kuroda K, Senda M. Scintillator selection for MR-compatible gamma detectors. IEEE Trans Nucl Sci 2003;50:1683–5.CrossRef
78.
go back to reference Derenzo SE, Huesman RH, Cahoon JL, Geyer AB, Moses WW, Uber DC, et al. A positron tomograph with 600 BGO crystals and 2.6 mm resolution. IEEE Trans Nucl Sci 1988;35 1:659–64.CrossRef Derenzo SE, Huesman RH, Cahoon JL, Geyer AB, Moses WW, Uber DC, et al. A positron tomograph with 600 BGO crystals and 2.6 mm resolution. IEEE Trans Nucl Sci 1988;35 1:659–64.CrossRef
79.
go back to reference Muehllehner G, Colsher JG, Lewitt RM. Hexagonal bar positron camera: problems and solutions. IEEE Trans Nucl Sci 1982;30 1:652–60.CrossRef Muehllehner G, Colsher JG, Lewitt RM. Hexagonal bar positron camera: problems and solutions. IEEE Trans Nucl Sci 1982;30 1:652–60.CrossRef
80.
go back to reference Freifelder R, Karp JS, Geagan M, Muehllehner G. Design and performance of the head PENN-PET scanner. IEEE Trans Nucl Sci 1994;41 4:1436–40.CrossRef Freifelder R, Karp JS, Geagan M, Muehllehner G. Design and performance of the head PENN-PET scanner. IEEE Trans Nucl Sci 1994;41 4:1436–40.CrossRef
81.
go back to reference Burnham CA, Bradshaw J, Kaufman D, Chesler D, Brownell GL. A stationary positron emission ring tomograph using BGO detector and analog readout. IEEE Trans Nucl Sci 1984;31 1:632–6.CrossRef Burnham CA, Bradshaw J, Kaufman D, Chesler D, Brownell GL. A stationary positron emission ring tomograph using BGO detector and analog readout. IEEE Trans Nucl Sci 1984;31 1:632–6.CrossRef
82.
go back to reference Surti S, Karp JS, Freifelder R, Liu F. Optimizing the performance of a PET detector using discrete GSO crystals on a continuous lightguide. IEEE Trans Nucl Sci 2000;47 3:1030–6.CrossRef Surti S, Karp JS, Freifelder R, Liu F. Optimizing the performance of a PET detector using discrete GSO crystals on a continuous lightguide. IEEE Trans Nucl Sci 2000;47 3:1030–6.CrossRef
83.
go back to reference Casey ME, Nutt R. A multicrystal two dimensional BGO detector system for positron emission tomography. IEEE Trans Nucl Sci 1986;33 1:460–3.CrossRef Casey ME, Nutt R. A multicrystal two dimensional BGO detector system for positron emission tomography. IEEE Trans Nucl Sci 1986;33 1:460–3.CrossRef
84.
go back to reference Uribe J, Aykac M, Baghaei H, Li H, Wang Y, Liu Y. Inexpensive position sensitive detector block for dedicated PET cameras using 40-mm diameter PMT in quadrant sharing configuration. IEEE Trans Nucl Sci 2003;50 3:367–2. Uribe J, Aykac M, Baghaei H, Li H, Wang Y, Liu Y. Inexpensive position sensitive detector block for dedicated PET cameras using 40-mm diameter PMT in quadrant sharing configuration. IEEE Trans Nucl Sci 2003;50 3:367–2.
85.
go back to reference Shao Y, Cherry SR, Siegel S, Silverman RW, Majewski S. Evaluation of multi-channel PMTs for readout of scintillator arrays. Nucl Instrum Meth Phys Res A 1997;390:209–18.CrossRef Shao Y, Cherry SR, Siegel S, Silverman RW, Majewski S. Evaluation of multi-channel PMTs for readout of scintillator arrays. Nucl Instrum Meth Phys Res A 1997;390:209–18.CrossRef
86.
go back to reference Shao Y, Cherry SR, Chatziioannou AF. Design and development of 1 mm resolution PET detectors with position-sensitive PMTs. Nucl Instrum Meth Phys Res A 2002;477:486–90.CrossRef Shao Y, Cherry SR, Chatziioannou AF. Design and development of 1 mm resolution PET detectors with position-sensitive PMTs. Nucl Instrum Meth Phys Res A 2002;477:486–90.CrossRef
87.
go back to reference Rouze NC, Schmand M, Siegel S, Hutchins GD. Design of a small animal PET imaging system with 1 microliter volume resolution. IEEE Trans Nucl Sci 2004;51 3:757–63.CrossRef Rouze NC, Schmand M, Siegel S, Hutchins GD. Design of a small animal PET imaging system with 1 microliter volume resolution. IEEE Trans Nucl Sci 2004;51 3:757–63.CrossRef
88.
go back to reference Lee K, Miyaoka RS, Janes ML, Lewellen TK. Detector characteristics of the micro crystal element scanner (MiCES). IEEE Trans Nucl Sci 2005;52 5:1428–33.CrossRef Lee K, Miyaoka RS, Janes ML, Lewellen TK. Detector characteristics of the micro crystal element scanner (MiCES). IEEE Trans Nucl Sci 2005;52 5:1428–33.CrossRef
89.
go back to reference Stickel JR, Qi J, Cherry SR. Fabrication and characterization of a 0.5-mm lutetium oxyorthosilicate detector array for high-resolution PET applications. J Nucl Med 2007;48 1:115–21.PubMed Stickel JR, Qi J, Cherry SR. Fabrication and characterization of a 0.5-mm lutetium oxyorthosilicate detector array for high-resolution PET applications. J Nucl Med 2007;48 1:115–21.PubMed
90.
go back to reference Rouze NC, Soon VC, Young JW, Siegel S, Hutchins GD. Initial evaluation of the Indiana small animal PET scanner. IEEE Nucl Sci Symp Conf Rec 2005;4:2394–8.CrossRef Rouze NC, Soon VC, Young JW, Siegel S, Hutchins GD. Initial evaluation of the Indiana small animal PET scanner. IEEE Nucl Sci Symp Conf Rec 2005;4:2394–8.CrossRef
91.
go back to reference Stickel JR, Cherry SR. High-resolution PET detector design: modelling components of intrinsic spatial resolution. Phys Med Biol 2005;50:179–95.PubMedCrossRef Stickel JR, Cherry SR. High-resolution PET detector design: modelling components of intrinsic spatial resolution. Phys Med Biol 2005;50:179–95.PubMedCrossRef
92.
go back to reference Rafecas M, Böning G, Pichler BJ, Lorenz E, Schwaiger M, Ziegler SI. Inter-crystal scatter in a dual layer, high resolution LSO–APD positron emission tomograph. Phys Med Biol 2003;48:821–48.PubMedCrossRef Rafecas M, Böning G, Pichler BJ, Lorenz E, Schwaiger M, Ziegler SI. Inter-crystal scatter in a dual layer, high resolution LSO–APD positron emission tomograph. Phys Med Biol 2003;48:821–48.PubMedCrossRef
93.
go back to reference Michaud JB, Brunet CA, Rafecas M, Lecomte R, Fontaine R. Sensitivity in PET: neural networks as an alternative to Compton photons LOR analysis. IEEE Nucl Sci Symp Conf Rec 2007;5:3594–600. Michaud JB, Brunet CA, Rafecas M, Lecomte R, Fontaine R. Sensitivity in PET: neural networks as an alternative to Compton photons LOR analysis. IEEE Nucl Sci Symp Conf Rec 2007;5:3594–600.
94.
go back to reference Levin CS. New imaging technologies to enhance the molecular sensitivity of positron emission tomography. Proc IEEE 2008;96 3:439–67.CrossRef Levin CS. New imaging technologies to enhance the molecular sensitivity of positron emission tomography. Proc IEEE 2008;96 3:439–67.CrossRef
95.
go back to reference Moses WW, Derenzo SE, Melcher CL, Manente RA. Gamma ray spectroscopy and timing using LSO and PIN photodiodes. IEEE Trans Nucl Sci 1995;42 4:597–600.CrossRef Moses WW, Derenzo SE, Melcher CL, Manente RA. Gamma ray spectroscopy and timing using LSO and PIN photodiodes. IEEE Trans Nucl Sci 1995;42 4:597–600.CrossRef
96.
go back to reference Levin CS, Hoffman EJ. Investigation of a new readout scheme for high resolution scintillation crystal arrays using photodiodes. IEEE Trans Nucl Sci 1997;44 3:1208–13.CrossRef Levin CS, Hoffman EJ. Investigation of a new readout scheme for high resolution scintillation crystal arrays using photodiodes. IEEE Trans Nucl Sci 1997;44 3:1208–13.CrossRef
97.
go back to reference Moses WW, Derenzo SE, Nutt R, Digby WM, Williams CW, Andreaco M. Performance of a PET detector module utilizing an array of silicon photodiodes to identify the crystal of interaction. IEEE Trans Nucl Sci 1993;40 4:1036–40.CrossRef Moses WW, Derenzo SE, Nutt R, Digby WM, Williams CW, Andreaco M. Performance of a PET detector module utilizing an array of silicon photodiodes to identify the crystal of interaction. IEEE Trans Nucl Sci 1993;40 4:1036–40.CrossRef
98.
go back to reference Huber JS, Moses WW, Andreaco MS, Petterson O. An LSO scintillator array for a PET detector module with depth of interaction measurement. IEEE Trans Nucl Sci 2001;48 3:684–8.CrossRef Huber JS, Moses WW, Andreaco MS, Petterson O. An LSO scintillator array for a PET detector module with depth of interaction measurement. IEEE Trans Nucl Sci 2001;48 3:684–8.CrossRef
99.
go back to reference McIntyre RJ. Recent developments in silicon avalanche photodiodes. Measurement 1985;3 4:146–52.CrossRef McIntyre RJ. Recent developments in silicon avalanche photodiodes. Measurement 1985;3 4:146–52.CrossRef
100.
go back to reference Lecomte R, Pepin C, Rouleau D, Dautet H, McIntyre RJ, McSween D, et al. Radiation detection measurements with a new ‘buried junction’ silicon avalanche photodiode. Nucl Instrum Meth Phys Res A 1999;423 1:92–102.CrossRef Lecomte R, Pepin C, Rouleau D, Dautet H, McIntyre RJ, McSween D, et al. Radiation detection measurements with a new ‘buried junction’ silicon avalanche photodiode. Nucl Instrum Meth Phys Res A 1999;423 1:92–102.CrossRef
101.
go back to reference Farrell R, Shah K, Vanderpuye K, Grazioso R, Myers R, Entine G. APD arrays and large-area APDs via a new planar process. Nucl Instrum Meth Phys Res A 2000;442:171–8.CrossRef Farrell R, Shah K, Vanderpuye K, Grazioso R, Myers R, Entine G. APD arrays and large-area APDs via a new planar process. Nucl Instrum Meth Phys Res A 2000;442:171–8.CrossRef
102.
go back to reference Webb PP, McIntyre RJ, Conradi J. Properties of avalanche photodiodes. RCA Rev 1974;35 2:234–78. Webb PP, McIntyre RJ, Conradi J. Properties of avalanche photodiodes. RCA Rev 1974;35 2:234–78.
103.
go back to reference Lecomte R, Schmitt D, Lightstone AW, McIntyre RJ. Performance characteristics of BGO-silicon avalanche photodiode detectors for PET. IEEE Trans Nucl Sci 1985;32 1:482–6.CrossRef Lecomte R, Schmitt D, Lightstone AW, McIntyre RJ. Performance characteristics of BGO-silicon avalanche photodiode detectors for PET. IEEE Trans Nucl Sci 1985;32 1:482–6.CrossRef
104.
go back to reference Grazioso R, Aykac M, Casey ME, Givens G, Schmand M. APD performance in light sharing PET applications. IEEE Trans Nucl Sci 2005;52 5:1413–6.CrossRef Grazioso R, Aykac M, Casey ME, Givens G, Schmand M. APD performance in light sharing PET applications. IEEE Trans Nucl Sci 2005;52 5:1413–6.CrossRef
105.
go back to reference Pichler BJ, Bernecker F, Boning G, Rafecas M, Pimpl W, Schwaiger M, et al. A 4×8 APD array, consisting of two monolithic silicon wafers, coupled to a 32-channel LSO matrix for high-resolution PET. IEEE Trans Nucl Sci 2001;48 4:1391–6.CrossRef Pichler BJ, Bernecker F, Boning G, Rafecas M, Pimpl W, Schwaiger M, et al. A 4×8 APD array, consisting of two monolithic silicon wafers, coupled to a 32-channel LSO matrix for high-resolution PET. IEEE Trans Nucl Sci 2001;48 4:1391–6.CrossRef
106.
go back to reference Pichler BJ, Swann BK, Rochelle J, Nutt RE, Cherry SR, Siegel SB. Lutetium oxyorthosilicate block detector readout by avalanche photodiode arrays for high resolution animal PET. Phys Med Biol 2004;49:4305–19.PubMedCrossRef Pichler BJ, Swann BK, Rochelle J, Nutt RE, Cherry SR, Siegel SB. Lutetium oxyorthosilicate block detector readout by avalanche photodiode arrays for high resolution animal PET. Phys Med Biol 2004;49:4305–19.PubMedCrossRef
107.
go back to reference Maas MC, van der Laan DJ, Schaart DR, Huizenga J, Brouwer JC, Bruyndonckx P, et al. Experimental characterization of monolithic-crystal small animal PET detectors read out by APD arrays. IEEE Trans Nucl Sci 2006;53 3:1071–7.CrossRef Maas MC, van der Laan DJ, Schaart DR, Huizenga J, Brouwer JC, Bruyndonckx P, et al. Experimental characterization of monolithic-crystal small animal PET detectors read out by APD arrays. IEEE Trans Nucl Sci 2006;53 3:1071–7.CrossRef
108.
go back to reference Bérard P, Bergeron M, Pepin C, Cadorette J, Tétrault MA, Viscogliosi N, et al. Development of a 64-channel APD detector module with individual pixel readout for submillimeter spatial resolution in PET. Nucl Instrum Meth Phys Res A. 2008; in press. Bérard P, Bergeron M, Pepin C, Cadorette J, Tétrault MA, Viscogliosi N, et al. Development of a 64-channel APD detector module with individual pixel readout for submillimeter spatial resolution in PET. Nucl Instrum Meth Phys Res A. 2008; in press.
109.
go back to reference Casey ME, Dautet H, Waechter D, Lecomte R, Eriksson L, Schmand M. An LSO block detector for PET using an avalanche photodiode array. IEEE Nucl Sci Symp Conf Rec 1998;2:1105–8. Casey ME, Dautet H, Waechter D, Lecomte R, Eriksson L, Schmand M. An LSO block detector for PET using an avalanche photodiode array. IEEE Nucl Sci Symp Conf Rec 1998;2:1105–8.
110.
go back to reference Lecomte R, Saoudi A, Rouleau D, Dautet H, Waechter D, Andreaco M, et al. An APD-based quad scintillator detector module with pulse shape discrimination coding for PET. IEEE Nucl Sci Symp Conf Rec 1998;3:1445–7. Lecomte R, Saoudi A, Rouleau D, Dautet H, Waechter D, Andreaco M, et al. An APD-based quad scintillator detector module with pulse shape discrimination coding for PET. IEEE Nucl Sci Symp Conf Rec 1998;3:1445–7.
111.
go back to reference Krishnamoorthy S, Vaska P, Stoll S, Purschke M, Pratte JF, Woody CL, et al. A prototype Anger-type detector for PET using LSO and large-area APDs. IEEE Nucl Sci Symp Conf Rec 2005;5:2845–8.CrossRef Krishnamoorthy S, Vaska P, Stoll S, Purschke M, Pratte JF, Woody CL, et al. A prototype Anger-type detector for PET using LSO and large-area APDs. IEEE Nucl Sci Symp Conf Rec 2005;5:2845–8.CrossRef
112.
go back to reference Schmitt D, Lecomte R, Lapointe M, Martel C, Carrier C, Karuta B, et al. Ultra-low noise charge sensitive preamplifier for scintillation detection with avalanche photodiodes in PET applications. IEEE Trans Nucl Sci 1987;34 1:91–6.CrossRef Schmitt D, Lecomte R, Lapointe M, Martel C, Carrier C, Karuta B, et al. Ultra-low noise charge sensitive preamplifier for scintillation detection with avalanche photodiodes in PET applications. IEEE Trans Nucl Sci 1987;34 1:91–6.CrossRef
113.
go back to reference Binkley DM, Puckett BS, Casey ME, Lecomte R, Saoudi A. A power-efficient, low-noise, wideband, integrated CMOS preamplifier for LSO/APD PET systems. IEEE Trans Nucl Sci 2000;47 3:810–7.CrossRef Binkley DM, Puckett BS, Casey ME, Lecomte R, Saoudi A. A power-efficient, low-noise, wideband, integrated CMOS preamplifier for LSO/APD PET systems. IEEE Trans Nucl Sci 2000;47 3:810–7.CrossRef
114.
go back to reference Pichler BJ, Pimpl W, Buttler W, Kotoulas L, Boning G, Rafecas M, et al. Integrated low-noise low-power fast charge-sensitive preamplifier for avalanche photodiodes in JFET–CMOS technology. IEEE Trans Nucl Sci 2001;48 6:2370–4.CrossRef Pichler BJ, Pimpl W, Buttler W, Kotoulas L, Boning G, Rafecas M, et al. Integrated low-noise low-power fast charge-sensitive preamplifier for avalanche photodiodes in JFET–CMOS technology. IEEE Trans Nucl Sci 2001;48 6:2370–4.CrossRef
115.
go back to reference Pratte JF, Robert S, De Geronimo G, O'Connor P, Stoll S, Pepin CM, et al. Preamplifiers for APD-based PET scanners. IEEE Trans Nucl Sci 2004;51 5:1979–85.CrossRef Pratte JF, Robert S, De Geronimo G, O'Connor P, Stoll S, Pepin CM, et al. Preamplifiers for APD-based PET scanners. IEEE Trans Nucl Sci 2004;51 5:1979–85.CrossRef
116.
go back to reference Habte F, Levin CS. Study of low noise multichannel readout electronics for high sensitivity PET systems based on avalanche photodiode arrays. IEEE Trans Nucl Sci 2004;51 3:764–9.CrossRef Habte F, Levin CS. Study of low noise multichannel readout electronics for high sensitivity PET systems based on avalanche photodiode arrays. IEEE Trans Nucl Sci 2004;51 3:764–9.CrossRef
117.
go back to reference Spanoudaki VC, McElroy DP, Ziegler SI. An analog signal processing ASIC for a small animal LSO–APD PET tomograph. Nucl Instrum Meth Phys Res A 2006;564 1:451–62.CrossRef Spanoudaki VC, McElroy DP, Ziegler SI. An analog signal processing ASIC for a small animal LSO–APD PET tomograph. Nucl Instrum Meth Phys Res A 2006;564 1:451–62.CrossRef
118.
go back to reference Burr KC, Ivan A, LeBlanc J, Zelakiewicz S, McDaniel DL, Kim CL, et al. Evaluation of a position sensitive avalanche photodiode for PET. IEEE Trans Nucl Sci 2003;50 4:792–6.CrossRef Burr KC, Ivan A, LeBlanc J, Zelakiewicz S, McDaniel DL, Kim CL, et al. Evaluation of a position sensitive avalanche photodiode for PET. IEEE Trans Nucl Sci 2003;50 4:792–6.CrossRef
119.
go back to reference Levin CS, Foudray AMK, Olcott PD, Habte F. Investigation of position sensitive avalanche photodiodes for a new high-resolution PET detector design. IEEE Trans Nucl Sci 2004;51 3:805–10.CrossRef Levin CS, Foudray AMK, Olcott PD, Habte F. Investigation of position sensitive avalanche photodiodes for a new high-resolution PET detector design. IEEE Trans Nucl Sci 2004;51 3:805–10.CrossRef
120.
go back to reference Catana C, Wu Y, Judenhofer MS, Qi J, Pichler BJ, Cherry SR. Simultaneous acquisition of multislice PET and MR images: initial results with a MR-compatible PET scanner. J Nucl Med 2006;47 12:1968–76.PubMed Catana C, Wu Y, Judenhofer MS, Qi J, Pichler BJ, Cherry SR. Simultaneous acquisition of multislice PET and MR images: initial results with a MR-compatible PET scanner. J Nucl Med 2006;47 12:1968–76.PubMed
121.
go back to reference Renker D. New trends on photodetectors. Nucl Instrum Meth Phys Res A 2007;571:1–6.CrossRef Renker D. New trends on photodetectors. Nucl Instrum Meth Phys Res A 2007;571:1–6.CrossRef
122.
go back to reference Hughes PJ, Herbert D, Stewart A, Jackson JC. Tiled silicon photomultipliers for large area, low light sensing applications. Proc SPIE 2007;6471:647112–1–8. Hughes PJ, Herbert D, Stewart A, Jackson JC. Tiled silicon photomultipliers for large area, low light sensing applications. Proc SPIE 2007;6471:647112–1–8.
123.
go back to reference Stewart AG, Greene-O'Sullivan E, Herbert DJ, Saveliev V, Quinlan F, Wall L, et al. Study of the properties of new SPM detectors. Proc SPIE 2006;6119:61190A–1–10. Stewart AG, Greene-O'Sullivan E, Herbert DJ, Saveliev V, Quinlan F, Wall L, et al. Study of the properties of new SPM detectors. Proc SPIE 2006;6119:61190A–1–10.
124.
go back to reference Renker D. Geiger-mode avalanche photodiodes, history, properties and problems. Nucl Instrum Meth Phys Res A 2006;567 1:48–56.CrossRef Renker D. Geiger-mode avalanche photodiodes, history, properties and problems. Nucl Instrum Meth Phys Res A 2006;567 1:48–56.CrossRef
125.
go back to reference Renker D. Progress in the development of Geiger-mode avalanche photodiodes. IEEE Nucl Sci Symp Conf Rec, 2008. Renker D. Progress in the development of Geiger-mode avalanche photodiodes. IEEE Nucl Sci Symp Conf Rec, 2008.
126.
go back to reference Moehrs S, Del Guerra A, Herbert DJ, Mandelkern MA. A detector head design for small-animal PET with silicon photomultipliers (SiPM). Phys Med Biol 2006;51 5:1113–27.PubMedCrossRef Moehrs S, Del Guerra A, Herbert DJ, Mandelkern MA. A detector head design for small-animal PET with silicon photomultipliers (SiPM). Phys Med Biol 2006;51 5:1113–27.PubMedCrossRef
127.
go back to reference Schaart DR, van Dam HT, Seifert S, Vinke R, Dendooven P, Löhner H, Beekman FJ. SiPM-array based PET detectors with depth-of-interaction correction. IEEE Nucl Sci Symp Conf Rec. 2008, M02-2; in press. Schaart DR, van Dam HT, Seifert S, Vinke R, Dendooven P, Löhner H, Beekman FJ. SiPM-array based PET detectors with depth-of-interaction correction. IEEE Nucl Sci Symp Conf Rec. 2008, M02-2; in press.
128.
go back to reference Spanoudaki VC, Mann AB, Ottei AN, Konorov I, Torres-Espallardo I, Paul S, et al. Use of single photon counting detector arrays in combined PET/MR: characterization of LYSO-SiPM detector modules and comparison with a LSO–APD detector. J Instrum 2007;2 12:0–16.CrossRef Spanoudaki VC, Mann AB, Ottei AN, Konorov I, Torres-Espallardo I, Paul S, et al. Use of single photon counting detector arrays in combined PET/MR: characterization of LYSO-SiPM detector modules and comparison with a LSO–APD detector. J Instrum 2007;2 12:0–16.CrossRef
129.
go back to reference España S, Tapias G, Fraile LM, Herraiz JL, Vicente E, Udias JM, et al. Performance evaluation of SiPM detectors for PET imaging in the presence of magnetic fields. IEEE Nucl Sci Symp Conf Rec. 2008, M02-4; in press. España S, Tapias G, Fraile LM, Herraiz JL, Vicente E, Udias JM, et al. Performance evaluation of SiPM detectors for PET imaging in the presence of magnetic fields. IEEE Nucl Sci Symp Conf Rec. 2008, M02-4; in press.
130.
go back to reference Kim CL, Wang GC, Dolinsky S. Multi-pixel photon counters for TOF PET detector and its challenges. IEEE Nucl Sci Symp Conf Rec 2008, M02-3; in press. Kim CL, Wang GC, Dolinsky S. Multi-pixel photon counters for TOF PET detector and its challenges. IEEE Nucl Sci Symp Conf Rec 2008, M02-3; in press.
131.
go back to reference Vinke R. Optimizing timing resolution of SiPM sensors for use in TOF-PET detectors. 5th Conference on New Developments in Photodetection, Aix-les-Bains, France, June 15–20, 2008. Vinke R. Optimizing timing resolution of SiPM sensors for use in TOF-PET detectors. 5th Conference on New Developments in Photodetection, Aix-les-Bains, France, June 15–20, 2008.
132.
go back to reference Schulz V. Simultaneous time-of-flight PET/MR, MR-PET Workshop, Jülich, Germany, October 27–28, 2008. Schulz V. Simultaneous time-of-flight PET/MR, MR-PET Workshop, Jülich, Germany, October 27–28, 2008.
133.
go back to reference Kikuchi Y, Ishii K, Yamazaki H, Matsuyama S, Yamaguchi T, et al. Preliminary report on the development of a high resolution PET camera using semiconductor detectors. Nucl Instrum Meth Phys Res B 2005;241:727–31.CrossRef Kikuchi Y, Ishii K, Yamazaki H, Matsuyama S, Yamaguchi T, et al. Preliminary report on the development of a high resolution PET camera using semiconductor detectors. Nucl Instrum Meth Phys Res B 2005;241:727–31.CrossRef
134.
go back to reference Vaska P, Bolotnikov A, Carini G, Camarda G, Pratte JF, Dilmanian FA, et al. Studies of CZT for PET applications. IEEE Nucl Sci Symp Conf Rec 2005;5:2799–802.CrossRef Vaska P, Bolotnikov A, Carini G, Camarda G, Pratte JF, Dilmanian FA, et al. Studies of CZT for PET applications. IEEE Nucl Sci Symp Conf Rec 2005;5:2799–802.CrossRef
135.
go back to reference Verger L, Gros d’Aillon E, Monnet O, Montémont G, Pelliciari B. New trends in γ-ray imaging with CdZnTe/CdTe at CEA-Leti. Nucl Instrum Meth Phys Res A 2007;571:33–43.CrossRef Verger L, Gros d’Aillon E, Monnet O, Montémont G, Pelliciari B. New trends in γ-ray imaging with CdZnTe/CdTe at CEA-Leti. Nucl Instrum Meth Phys Res A 2007;571:33–43.CrossRef
136.
go back to reference Drezet A, Monnet O, Mathy F, Montemont G, Verger L. CdZnTe detectors for small field of view positron emission tomographic imaging. Nucl Instrum Meth Phys Res A 2007;571:465–70.CrossRef Drezet A, Monnet O, Mathy F, Montemont G, Verger L. CdZnTe detectors for small field of view positron emission tomographic imaging. Nucl Instrum Meth Phys Res A 2007;571:465–70.CrossRef
137.
go back to reference Ishii K, Kikuchia Y, Matsuyamaa S, Kanaia Y, Kotanib K, Ito T, et al. First achievement of less than 1 mm FWHM resolution in practical semiconductor animal PET scanner. Nucl Instrum Meth Phys Res A 2007;576:435–40.CrossRef Ishii K, Kikuchia Y, Matsuyamaa S, Kanaia Y, Kotanib K, Ito T, et al. First achievement of less than 1 mm FWHM resolution in practical semiconductor animal PET scanner. Nucl Instrum Meth Phys Res A 2007;576:435–40.CrossRef
138.
go back to reference Vaska P, Dragone A, Lee W, Kim DH, Pratte JF, Cui YG, et al. A prototype CZT-based PET scanner for high resolution mouse brain imaging. Nucl Sci Symp Conf Rec 2007;5:3816–9. Vaska P, Dragone A, Lee W, Kim DH, Pratte JF, Cui YG, et al. A prototype CZT-based PET scanner for high resolution mouse brain imaging. Nucl Sci Symp Conf Rec 2007;5:3816–9.
139.
go back to reference Peng H, Olcott PD, Pratx G, Foudray AMK, Chinn G, Levin CS. Design study of a high-resolution breast-dedicated PET system built from cadmium zinc telluride detectors. IEEE Nucl Sci Symp Conf Rec 2007;5:3700–4. Peng H, Olcott PD, Pratx G, Foudray AMK, Chinn G, Levin CS. Design study of a high-resolution breast-dedicated PET system built from cadmium zinc telluride detectors. IEEE Nucl Sci Symp Conf Rec 2007;5:3700–4.
140.
go back to reference Jeavons AP, Townsend DW, Ford NL, Kull K, Manuel A, Fischer O, et al. A high-resolution proportional chamber positron camera and its applications. IEEE Trans Nucl Sci 1978;25 1:164–73.CrossRef Jeavons AP, Townsend DW, Ford NL, Kull K, Manuel A, Fischer O, et al. A high-resolution proportional chamber positron camera and its applications. IEEE Trans Nucl Sci 1978;25 1:164–73.CrossRef
141.
go back to reference Bateman JE, Connolly JF, Stephenson R, Tappern GJ, Flesher AC. The Rutherford Appleton Laboratory’s Mark I multiwire proportional counter positron camera. Nucl Instrum Meth Phys Res A 1984;225:209–31.CrossRef Bateman JE, Connolly JF, Stephenson R, Tappern GJ, Flesher AC. The Rutherford Appleton Laboratory’s Mark I multiwire proportional counter positron camera. Nucl Instrum Meth Phys Res A 1984;225:209–31.CrossRef
142.
go back to reference Jeavons AP, Chandler RA, Dettmar CAR. A 3D HIDAC-PET camera with sub-millimetre resolution for imaging small animals. IEEE Trans Nucl Sci 1999;46 3:468–73.CrossRef Jeavons AP, Chandler RA, Dettmar CAR. A 3D HIDAC-PET camera with sub-millimetre resolution for imaging small animals. IEEE Trans Nucl Sci 1999;46 3:468–73.CrossRef
143.
go back to reference Missimer J, Madi Z, Honer M, Keller C, Schubiger A, Ametamey SM. Performance evaluation of the 16-module quad-HIDAC small animal PET camera. Phys Med Biol 2004;49 10:2069–81.PubMedCrossRef Missimer J, Madi Z, Honer M, Keller C, Schubiger A, Ametamey SM. Performance evaluation of the 16-module quad-HIDAC small animal PET camera. Phys Med Biol 2004;49 10:2069–81.PubMedCrossRef
144.
go back to reference Honer M, Brühlmeier M, Missimer J, Schubiger AP, Ametamey SM. Dynamic imaging of striatal D2 receptors in mice using quad-HIDAC PET. J Nucl Med 2004;45 3:464–70.PubMed Honer M, Brühlmeier M, Missimer J, Schubiger AP, Ametamey SM. Dynamic imaging of striatal D2 receptors in mice using quad-HIDAC PET. J Nucl Med 2004;45 3:464–70.PubMed
145.
go back to reference Blanco A, Carolino N, Correia CMBA, Fazendeiro L, Ferreira NC, Ferreira Marques MF, et al. RPC-PET: a new very high resolution PET technology. IEEE Trans Nucl Sci 2006;53:2489–94.CrossRef Blanco A, Carolino N, Correia CMBA, Fazendeiro L, Ferreira NC, Ferreira Marques MF, et al. RPC-PET: a new very high resolution PET technology. IEEE Trans Nucl Sci 2006;53:2489–94.CrossRef
146.
go back to reference Eriksson LA, Townsend DW, Conti M, Melcher CL, Eriksson M, Jakoby BW, et al. Potentials for large axial field of view positron camera systems. IEEE Nucl Sci Symp Conf Rec. 2008, NM1-5; in press. Eriksson LA, Townsend DW, Conti M, Melcher CL, Eriksson M, Jakoby BW, et al. Potentials for large axial field of view positron camera systems. IEEE Nucl Sci Symp Conf Rec. 2008, NM1-5; in press.
147.
go back to reference Couceiro M, Blanco A, Ferreira NC, Marques RF, Fonte P, Lopes L. RPC-PET: status and perspectives. Nucl Instrum Meth Phys Res A 2007;580:915–8.CrossRef Couceiro M, Blanco A, Ferreira NC, Marques RF, Fonte P, Lopes L. RPC-PET: status and perspectives. Nucl Instrum Meth Phys Res A 2007;580:915–8.CrossRef
148.
go back to reference Zaklad H, Derenzo SE, Miller RA, Smadja G, Smits RG, Alvarez LW. A liquid xenon radioisotope camera. IEEE Trans Nucl Sci 1972;19:206–13.CrossRef Zaklad H, Derenzo SE, Miller RA, Smadja G, Smits RG, Alvarez LW. A liquid xenon radioisotope camera. IEEE Trans Nucl Sci 1972;19:206–13.CrossRef
149.
go back to reference Chepel V, Lopes MI, Kuchenkov A, Marques RF, Policarpo AJPL. Performance study of liquid xenon detector for PET. Nucl Instrum Meth Phys Res A 1997;392:427–32.CrossRef Chepel V, Lopes MI, Kuchenkov A, Marques RF, Policarpo AJPL. Performance study of liquid xenon detector for PET. Nucl Instrum Meth Phys Res A 1997;392:427–32.CrossRef
150.
go back to reference Gallin-Martel ML, Martin P, Mayet F, Ballon J, Barbier G, Barnoux C, et al. Experimental study of a liquid Xenon PET prototype module. Nucl Instrum Meth Phys Res A 2006;563 1:225–8.CrossRef Gallin-Martel ML, Martin P, Mayet F, Ballon J, Barbier G, Barnoux C, et al. Experimental study of a liquid Xenon PET prototype module. Nucl Instrum Meth Phys Res A 2006;563 1:225–8.CrossRef
151.
go back to reference Amaudruz P, Bryman D, Kurchaninov L, Lu P, Marshall C, Martin JP. Investigation of liquid xenon detectors for PET: simultaneous reconstruction of light and charge signals from 511 keV photons. IEEE Nucl Sci Symp Conf Rec 2007;4:2889–91. Amaudruz P, Bryman D, Kurchaninov L, Lu P, Marshall C, Martin JP. Investigation of liquid xenon detectors for PET: simultaneous reconstruction of light and charge signals from 511 keV photons. IEEE Nucl Sci Symp Conf Rec 2007;4:2889–91.
152.
go back to reference Lecomte R. Molecular PET Instrumentation and Imaging Techniques. In: Pomper M, Gelovani J, editors. Molecular imaging in oncology. New York: Informa Healthcare; 2008; p. 65–90. Lecomte R. Molecular PET Instrumentation and Imaging Techniques. In: Pomper M, Gelovani J, editors. Molecular imaging in oncology. New York: Informa Healthcare; 2008; p. 65–90.
153.
go back to reference H. Peng, Investigation of new approaches to combined positron emission tomography and magnetic resonance imaging systems. Ph.D. thesis. University of Western Ontario, 2006. H. Peng, Investigation of new approaches to combined positron emission tomography and magnetic resonance imaging systems. Ph.D. thesis. University of Western Ontario, 2006.
154.
go back to reference Pichler BJ, Wehrl HF, Kolb A, Judenhofer MS. Positron emission tomography/magnetic resonance imaging: the next generation of multimodality imaging? Sem Nucl Med 2008;38 3:199–208.CrossRef Pichler BJ, Wehrl HF, Kolb A, Judenhofer MS. Positron emission tomography/magnetic resonance imaging: the next generation of multimodality imaging? Sem Nucl Med 2008;38 3:199–208.CrossRef
155.
go back to reference McIntyre RJ, Webb PP, Dautet H. A short-wavelength selective reach-through avalanche photodiode. IEEE Trans Nucl Sci 1996;43 3:1341–6.CrossRef McIntyre RJ, Webb PP, Dautet H. A short-wavelength selective reach-through avalanche photodiode. IEEE Trans Nucl Sci 1996;43 3:1341–6.CrossRef
156.
go back to reference Pepin CM, Bérard P, Perrot AL, Pépin C, Houde D, Lecomte R, et al. Properties of LYSO and recent LSO scintillators for phoswich PET detectors. IEEE Trans Nucl Sci 2004;51 3:789–95.CrossRef Pepin CM, Bérard P, Perrot AL, Pépin C, Houde D, Lecomte R, et al. Properties of LYSO and recent LSO scintillators for phoswich PET detectors. IEEE Trans Nucl Sci 2004;51 3:789–95.CrossRef
157.
go back to reference Llosá G, Belcari N, Bisogni MG, Collazuol G, Marcatili S, Boscardin M, et al. First results in the application of silicon photomultiplier matrices to small animal PET. Nucl Instrum Meth Phys Res A 2008; in press. Llosá G, Belcari N, Bisogni MG, Collazuol G, Marcatili S, Boscardin M, et al. First results in the application of silicon photomultiplier matrices to small animal PET. Nucl Instrum Meth Phys Res A 2008; in press.
158.
go back to reference Szawlowski M, Meier D, Maehlum G, Wagenaar DJ, Patt BE. Spectroscopy and timing with multi-pixel photon counters (MPPC) and LYSO scintillators. IEEE Nucl Sci Symp Conf Rec 2007;6:4591–6. Szawlowski M, Meier D, Maehlum G, Wagenaar DJ, Patt BE. Spectroscopy and timing with multi-pixel photon counters (MPPC) and LYSO scintillators. IEEE Nucl Sci Symp Conf Rec 2007;6:4591–6.
159.
go back to reference Nassalski A, Moszyński M, Syntfeld-Każuch A, Szczęśniak T, Świderski L, Wolski D. Silicon photomultiplier as an alternative for APD in PET applications. IEEE Nucl Sci Symp Conf Rec 2008, NM1-3; in press. Nassalski A, Moszyński M, Syntfeld-Każuch A, Szczęśniak T, Świderski L, Wolski D. Silicon photomultiplier as an alternative for APD in PET applications. IEEE Nucl Sci Symp Conf Rec 2008, NM1-3; in press.
160.
go back to reference Verger L, Mathy F, Monnet O, Montémont G. Gamma-ray imaging with CdZnTe/CdTe detectors. Conférences Frédéric Joliot 2007: Imagerie multi modalités incluant la Tomographie par Émission de Positons, Orsay, 26–27 March 2007. Verger L, Mathy F, Monnet O, Montémont G. Gamma-ray imaging with CdZnTe/CdTe detectors. Conférences Frédéric Joliot 2007: Imagerie multi modalités incluant la Tomographie par Émission de Positons, Orsay, 26–27 March 2007.
Metadata
Title
Novel detector technology for clinical PET
Author
Roger Lecomte
Publication date
01-03-2009
Publisher
Springer-Verlag
Published in
European Journal of Nuclear Medicine and Molecular Imaging / Issue Special Issue 1/2009
Print ISSN: 1619-7070
Electronic ISSN: 1619-7089
DOI
https://doi.org/10.1007/s00259-008-1054-0

Other articles of this Special Issue 1/2009

European Journal of Nuclear Medicine and Molecular Imaging 1/2009 Go to the issue