Published in:
01-12-2006 | Original Article
Using multidetector-row CT in neonates with complex congenital heart disease to replace diagnostic cardiac catheterization for anatomical investigation: initial experiences in technical and clinical feasibility
Authors:
Tain Lee, I-Chen Tsai, Yun-Ching Fu, Sheng-Lin Jan, Chung-Chi Wang, Yen Chang, Min-Chi Chen
Published in:
Pediatric Radiology
|
Issue 12/2006
Login to get access
Abstract
Background
Echocardiography is the first-line modality for the investigation of neonatal congenital heart disease. Diagnostic cardiac catheterization, which has a small but recognized risk, is usually performed if echocardiography fails to provide a confident evaluation of the lesions.
Objective
To verify the technical and clinical feasibilities of replacing diagnostic cardiac catheterization with multidetector-row CT (MDCT) in neonatal complex congenital heart disease.
Materials and methods
Over a 1-year period we prospectively enrolled all neonates with complex congenital heart disease referred for diagnostic cardiac catheterization after initial assessment by echocardiography. MDCT was performed using a 40-detector-row CT scanner with dual syringe injection. A multidisciplinary congenital heart disease team evaluated the MDCT images and decided if further diagnostic cardiac catheterization was necessary. The accuracy of MDCT in detecting separate cardiovascular anomalies and bolus geometry of contrast enhancement were calculated.
Results
A total of 14 neonates were included in the study. No further diagnostic cardiac catheterization was needed in any neonate. The accuracy of MDCT in diagnosing separate cardiovascular anomalies was 98% (53/54) with only one atrial septal defect missed in a patient with coarctation syndrome. The average cardiovascular enhancement in evaluated chambers was 471 HU. No obvious beam-hardening artefact was observed.
Conclusion
The technical and clinical feasibility of MDCT in complex congenital heart disease in neonates is confirmed. After initial assessment with echocardiography, MDCT could probably replace diagnostic cardiac catheterization for further anatomical clarification in neonates.