Published in: Neuroradiology 12/2014

01-12-2014 | Interventional Neuroradiology

Thromboembolic events associated with single balloon-, double balloon-, and stent-assisted coil embolization of asymptomatic unruptured cerebral aneurysms: evaluation with diffusion-weighted MR imaging

Authors: Tomoji Takigawa, Kensuke Suzuki, Yoshiki Sugiura, Ryotaro Suzuki, Issei Takano, Nobuyuki Shimizu, Yoshihiro Tanaka, Akio Hyodo

Published in: Neuroradiology | Issue 12/2014

Login to get access

Abstract

Introduction

The introduction of the balloon remodeling and stent-assisted technique has revolutionized the approach to coil embolization for wide-neck aneurysms. The purpose of this study was to determine the frequency of thromboembolic events associated with single balloon-assisted, double balloon-assisted, and stent-assisted coil embolization for asymptomatic unruptured aneurysms.

Methods

A retrospective review was undertaken by 119 patients undergoing coiling with an adjunctive technique for unruptured saccular aneurysms (64 single balloon, 12 double balloon, 43 stent assisted). All underwent diffusion-weighted imaging (DWI) within 24 h after the procedure.

Results

DWI showed hyperintense lesions in 48 (40 %) patients, and ten (21 %) of these patients incurred neurological deterioration (permanent, two; transient, eight). Hyperintense lesions were detected significantly more often in procedures with the double balloon-assisted technique (7/12, 58 %) than with the single balloon-assisted technique (16/64, 25 %, p = 0.05). Occurrence of new lesions was significantly higher with the use of stent-assisted technique (25/43, 58 %) than with the single balloon-assisted technique (p = 0.001). Symptomatic ischemic rates were similar between the three groups. The increased number of microcatheters was significantly related to the DWI abnormalities (two microcatheters, 15/63 (23.8 %); three microcatheters, 20/41 (48.8 %) (p = 0.008); four microcatheters, 12/15 (80 %) (p = 0.001)).

Conclusion

Thromboembolic events detected on DWI related to coil embolization for unruptured aneurysms are relatively common, especially in association with the double balloon-assisted and stent-assisted techniques. Furthermore, the number of microcatheters is highly correlated with DWI abnormalities. The high rate of thromboembolic events suggests the need for evaluation of platelet reactivity and the addition or change of antiplatelet agents.
Literature
1.
go back to reference Guglielmi G, Viñuela F, Sepetka I, Macellari V (1991) Electrothrombosis of saccular aneurysms via endovascular approach. Part 1: electrochemical basis, technique, and experimental results. J Neurosurg 75:1–7PubMedCrossRef Guglielmi G, Viñuela F, Sepetka I, Macellari V (1991) Electrothrombosis of saccular aneurysms via endovascular approach. Part 1: electrochemical basis, technique, and experimental results. J Neurosurg 75:1–7PubMedCrossRef
2.
go back to reference Geremia G, Haklin M, Brennecke L (1994) Embolization of experimentally created aneurysms with intravascular stent devices. AJNR Am J Neuroradiol 15:1223–1231PubMed Geremia G, Haklin M, Brennecke L (1994) Embolization of experimentally created aneurysms with intravascular stent devices. AJNR Am J Neuroradiol 15:1223–1231PubMed
3.
go back to reference Moret J, Cognard C, Weill A, Castaings L, Rey A (1997) Reconstruction technic in the treatment of wide-neck intracranial aneurysms. Long-term angiographic and clinical results. Apropos of 56 cases. J Neuroradiol 24:30–44PubMed Moret J, Cognard C, Weill A, Castaings L, Rey A (1997) Reconstruction technic in the treatment of wide-neck intracranial aneurysms. Long-term angiographic and clinical results. Apropos of 56 cases. J Neuroradiol 24:30–44PubMed
4.
go back to reference Albayram S, Selcuk H, Kara B, Bozdag E, Uzma O, Kocer N, Islak C (2004) Thromboembolic events associated with balloon-assisted coil embolization: evaluation with diffusion-weighted MR imaging. AJNR Am J Neuroradiol 25:1768–1777PubMed Albayram S, Selcuk H, Kara B, Bozdag E, Uzma O, Kocer N, Islak C (2004) Thromboembolic events associated with balloon-assisted coil embolization: evaluation with diffusion-weighted MR imaging. AJNR Am J Neuroradiol 25:1768–1777PubMed
5.
go back to reference Cottier JP, Pasco A, Gallas S, Gabrillargues J, Cognard C, Drouineau J, Brunereau L, Herbreteau D (2001) Utility of balloon-assisted Guglielmi detachable coiling in the treatment of 49 cerebral aneurysms: a retrospective, multicenter study. AJNR Am J Neuroradiol 22:345–351PubMed Cottier JP, Pasco A, Gallas S, Gabrillargues J, Cognard C, Drouineau J, Brunereau L, Herbreteau D (2001) Utility of balloon-assisted Guglielmi detachable coiling in the treatment of 49 cerebral aneurysms: a retrospective, multicenter study. AJNR Am J Neuroradiol 22:345–351PubMed
6.
go back to reference Lefkowitz MA, Gobin YP, Akiba Y, Duckwiler GR, Murayama Y, Guglielmi G, Martin NA, Viñuela F (1999) Balloon-assisted Guglielmi detachable coiling of wide-necked aneurysma: part II—clinical results. Neurosurgery 45:531–538PubMedCrossRef Lefkowitz MA, Gobin YP, Akiba Y, Duckwiler GR, Murayama Y, Guglielmi G, Martin NA, Viñuela F (1999) Balloon-assisted Guglielmi detachable coiling of wide-necked aneurysma: part II—clinical results. Neurosurgery 45:531–538PubMedCrossRef
7.
go back to reference Malek AM, Halbach VV, Phatouros CC, Lempert TE, Meyers PM, Dowd CF, Higashida RT (2000) Balloon-assist technique for endovascular coil embolization of geometrically difficult intracranial aneurysms. Neurosurgery 46:1397–1407PubMedCrossRef Malek AM, Halbach VV, Phatouros CC, Lempert TE, Meyers PM, Dowd CF, Higashida RT (2000) Balloon-assist technique for endovascular coil embolization of geometrically difficult intracranial aneurysms. Neurosurgery 46:1397–1407PubMedCrossRef
8.
go back to reference Nelson PK, Levy DI (2001) Balloon-assisted coil embolization of wide-necked aneurysms of the internal carotid artery: medium-term angiographic and clinical follow-up in 22 patients. AJNR Am J Neuroradiol 22:19–26PubMed Nelson PK, Levy DI (2001) Balloon-assisted coil embolization of wide-necked aneurysms of the internal carotid artery: medium-term angiographic and clinical follow-up in 22 patients. AJNR Am J Neuroradiol 22:19–26PubMed
9.
go back to reference Lanterna LA, Tredici G, Dimitrov BD, Biroli F (2004) Treatment of unruptured cerebral aneurysms by embolization with Guglielmi detachable coils: case-fatality, morbidity, and effectiveness in preventing bleeding—a systematic review of the literature. Neurosurgery 55:767–775PubMedCrossRef Lanterna LA, Tredici G, Dimitrov BD, Biroli F (2004) Treatment of unruptured cerebral aneurysms by embolization with Guglielmi detachable coils: case-fatality, morbidity, and effectiveness in preventing bleeding—a systematic review of the literature. Neurosurgery 55:767–775PubMedCrossRef
10.
go back to reference Molyneux A, Kerr R, Stratton I, Sandercock P, Clarke M, Shrimpton J, Holman R, International Subarachnoid Aneurysm Trial (ISAT) Collaborative Group (2002) International subarachnoid aneurysm trial (ISAT) of neurosurgical clipping versus endovascular coiling in 2143 patients with ruptured intracranial aneurysms: a randomised trial. Lancet 360:1267–1274PubMedCrossRef Molyneux A, Kerr R, Stratton I, Sandercock P, Clarke M, Shrimpton J, Holman R, International Subarachnoid Aneurysm Trial (ISAT) Collaborative Group (2002) International subarachnoid aneurysm trial (ISAT) of neurosurgical clipping versus endovascular coiling in 2143 patients with ruptured intracranial aneurysms: a randomised trial. Lancet 360:1267–1274PubMedCrossRef
11.
go back to reference Wiebers DO, Whisnant JP, Huston J 3rd, Meissner I, Brown RD Jr, Piepgras DG, Forbes GS, Thielen K, Nichols D, O’Fallon WM, Peacock J, Jaeger L, Kassell NF, Kongable-Beckman GL, Torner JC, International Study of Unruptured Intracranial Aneurysms Investigators (2003) Unruptured intracranial aneurysms: natural history, clinical outcome, and risks of surgical and endovascular treatment. Lancet 362:103–110PubMedCrossRef Wiebers DO, Whisnant JP, Huston J 3rd, Meissner I, Brown RD Jr, Piepgras DG, Forbes GS, Thielen K, Nichols D, O’Fallon WM, Peacock J, Jaeger L, Kassell NF, Kongable-Beckman GL, Torner JC, International Study of Unruptured Intracranial Aneurysms Investigators (2003) Unruptured intracranial aneurysms: natural history, clinical outcome, and risks of surgical and endovascular treatment. Lancet 362:103–110PubMedCrossRef
12.
go back to reference Pelz DM, Lownie SP, Fox AJ (1998) Thromboembolic events associated with the treatment of cerebral aneurysms with Guglielmi detachable coils. AJNR Am J Neuroradiol 19:1541–1547PubMed Pelz DM, Lownie SP, Fox AJ (1998) Thromboembolic events associated with the treatment of cerebral aneurysms with Guglielmi detachable coils. AJNR Am J Neuroradiol 19:1541–1547PubMed
13.
go back to reference Qureshi AI, Luft AR, Sharma M, Guterman LR, Hopkins LN (2000) Prevention and treatment of thromboembolic and ischemic complications associated with endovascular procedures: part II—clinical aspects and recommendations. Neurosurgery 46:1360–1376PubMedCrossRef Qureshi AI, Luft AR, Sharma M, Guterman LR, Hopkins LN (2000) Prevention and treatment of thromboembolic and ischemic complications associated with endovascular procedures: part II—clinical aspects and recommendations. Neurosurgery 46:1360–1376PubMedCrossRef
14.
go back to reference Pierot L, Spelle L, Vitry F, ATENA Investigators (2008) Immediate clinical outcome of patients harboring unruptured intracranial aneurysms treated by endovascular approach: results of the ATENA study. Stroke 39:2497–2504PubMedCrossRef Pierot L, Spelle L, Vitry F, ATENA Investigators (2008) Immediate clinical outcome of patients harboring unruptured intracranial aneurysms treated by endovascular approach: results of the ATENA study. Stroke 39:2497–2504PubMedCrossRef
15.
go back to reference Soeda A, Sakai N, Sakai H, Iihara K, Yamada N, Imakita S, Nagata I (2003) Thromboembolic events associated with Guglielmi detachable coil embolization of asymptomatic cerebral aneurysms: evaluation of 66 consecutive cases with use of diffusion-weighted MR imaging. AJNR Am J Neuroradiol 24:127–132PubMed Soeda A, Sakai N, Sakai H, Iihara K, Yamada N, Imakita S, Nagata I (2003) Thromboembolic events associated with Guglielmi detachable coil embolization of asymptomatic cerebral aneurysms: evaluation of 66 consecutive cases with use of diffusion-weighted MR imaging. AJNR Am J Neuroradiol 24:127–132PubMed
16.
go back to reference Rordorf G, Bellon RJ, Budzik RE Jr, Farkas J, Reinking GF, Pergolizzi RS, Ezzeddine M, Norbash AM, Gonzalez RG, Putman CM (2001) Silent thromboembolic events associated with the treatment of unruptured cerebral aneurysms by use of Guglielmi detachable coils: prospective study applying diffusion-weighted imaging. AJNR Am J Neuroradiol 22:5–10PubMed Rordorf G, Bellon RJ, Budzik RE Jr, Farkas J, Reinking GF, Pergolizzi RS, Ezzeddine M, Norbash AM, Gonzalez RG, Putman CM (2001) Silent thromboembolic events associated with the treatment of unruptured cerebral aneurysms by use of Guglielmi detachable coils: prospective study applying diffusion-weighted imaging. AJNR Am J Neuroradiol 22:5–10PubMed
17.
go back to reference Cronqvist M, Wirestam R, Ramgren B, Brandt L, Nilsson O, Säveland H, Holtås S, Larsson EM (2005) Diffusion and perfusion MRI in patients with ruptured and unruptured intracranial aneurysms treated by endovascular coiling: complications, procedural results, MR findings and clinical outcome. Neuroradiology 47:855–873PubMedCrossRef Cronqvist M, Wirestam R, Ramgren B, Brandt L, Nilsson O, Säveland H, Holtås S, Larsson EM (2005) Diffusion and perfusion MRI in patients with ruptured and unruptured intracranial aneurysms treated by endovascular coiling: complications, procedural results, MR findings and clinical outcome. Neuroradiology 47:855–873PubMedCrossRef
18.
go back to reference Derdeyn CP, Cross DT 3rd, Moran CJ, Brown GW, Pilgram TK, Diringer MN, Grubb RL Jr, Rich KM, Chicoine MR, Dacey RG Jr (2002) Postprocedure ischemic events after treatment of intracranial aneurysms with Guglielmi detachable coils. J Neurosurg 96:837–843PubMedCrossRef Derdeyn CP, Cross DT 3rd, Moran CJ, Brown GW, Pilgram TK, Diringer MN, Grubb RL Jr, Rich KM, Chicoine MR, Dacey RG Jr (2002) Postprocedure ischemic events after treatment of intracranial aneurysms with Guglielmi detachable coils. J Neurosurg 96:837–843PubMedCrossRef
19.
go back to reference Shapiro M, Babb J, Becske T, Nelson PK (2008) Safety and efficacy of adjunctive balloon remodeling during endovascular treatment of intracranial aneurysms: a literature review. AJNR Am J Neuroradiol 29:1777–1781PubMedCrossRef Shapiro M, Babb J, Becske T, Nelson PK (2008) Safety and efficacy of adjunctive balloon remodeling during endovascular treatment of intracranial aneurysms: a literature review. AJNR Am J Neuroradiol 29:1777–1781PubMedCrossRef
20.
go back to reference Brooks NP, Turk AS, Niemann DB, Aagaard-Kienitz B, Pulfer K, Cook T (2008) Frequency of thromboembolic events associated with endovascular aneurysm treatment: retrospective case series. J Neurosurg 108:1095–1100PubMedCrossRef Brooks NP, Turk AS, Niemann DB, Aagaard-Kienitz B, Pulfer K, Cook T (2008) Frequency of thromboembolic events associated with endovascular aneurysm treatment: retrospective case series. J Neurosurg 108:1095–1100PubMedCrossRef
21.
go back to reference Ross IB, Dhillon GS (2005) Complications of endovascular treatment of cerebral aneurysms. Surg Neurol 64:12–19PubMedCrossRef Ross IB, Dhillon GS (2005) Complications of endovascular treatment of cerebral aneurysms. Surg Neurol 64:12–19PubMedCrossRef
22.
go back to reference Lubicz B, Lefranc F, Bruneau M, Balériaux D, De Witte O (2008) Balloon-assisted coiling of intracranial aneurysms is not associated with a higher complication rate. Neuroradiology 50:769–776PubMedCrossRef Lubicz B, Lefranc F, Bruneau M, Balériaux D, De Witte O (2008) Balloon-assisted coiling of intracranial aneurysms is not associated with a higher complication rate. Neuroradiology 50:769–776PubMedCrossRef
23.
go back to reference Spiotta AM, Bhalla T, Hussain MS, Sivapatham T, Batra A, Hui F, Rasmussen PA, Moskowitz SI (2011) An analysis of inflation times during balloon-assisted aneurysm coil embolization and ischemic complications. Stroke 42:1051–1055PubMedCrossRef Spiotta AM, Bhalla T, Hussain MS, Sivapatham T, Batra A, Hui F, Rasmussen PA, Moskowitz SI (2011) An analysis of inflation times during balloon-assisted aneurysm coil embolization and ischemic complications. Stroke 42:1051–1055PubMedCrossRef
24.
go back to reference Ishibashi T, Murayama Y, Saguchi T, Ebara M, Irie K, Takao H, Abe T (2006) Thromboembolic events during endovascular coil embolization of cerebral aneurysms. Interv Neuroradiol 12(Suppl 1):112–116PubMedCentralPubMed Ishibashi T, Murayama Y, Saguchi T, Ebara M, Irie K, Takao H, Abe T (2006) Thromboembolic events during endovascular coil embolization of cerebral aneurysms. Interv Neuroradiol 12(Suppl 1):112–116PubMedCentralPubMed
25.
go back to reference Altay T, Kang HI, Woo HH, Masaryk TJ, Rasmussen PA, Fiorella DJ, Moskowitz SI (2011) Thromboembolic events associated with endovascular treatment of cerebral aneurysms. J Neurointerv Surg 3:147–150PubMedCrossRef Altay T, Kang HI, Woo HH, Masaryk TJ, Rasmussen PA, Fiorella DJ, Moskowitz SI (2011) Thromboembolic events associated with endovascular treatment of cerebral aneurysms. J Neurointerv Surg 3:147–150PubMedCrossRef
26.
go back to reference Earnest F 4th, Forbes G, Sandok BA, Piepgras DG, Faust RJ, Ilstrup DM, Arndt LJ (1984) Complications of cerebral angiography: prospective assessment of risk. AJR Am J Roentgenol 142:247–253PubMedCrossRef Earnest F 4th, Forbes G, Sandok BA, Piepgras DG, Faust RJ, Ilstrup DM, Arndt LJ (1984) Complications of cerebral angiography: prospective assessment of risk. AJR Am J Roentgenol 142:247–253PubMedCrossRef
27.
go back to reference Raymond J, Guilbert F, Weill A, Georganos SA, Juravsky L, Lambert A, Lamoureux J, Chagnon M, Roy D (2003) Long-term angiographic recurrences after selective endovascular treatment of aneurysms with detachable coils. Stroke 34:1398–1403PubMedCrossRef Raymond J, Guilbert F, Weill A, Georganos SA, Juravsky L, Lambert A, Lamoureux J, Chagnon M, Roy D (2003) Long-term angiographic recurrences after selective endovascular treatment of aneurysms with detachable coils. Stroke 34:1398–1403PubMedCrossRef
28.
go back to reference Schneider PA, Silva MB Jr, Bohannon WT, Kasirajan K, Caps MT, Nelken N, Santana D (2005) Safety and efficacy of carotid arteriography in vascular surgery practice. J Vasc Surg 41:238–245PubMedCrossRef Schneider PA, Silva MB Jr, Bohannon WT, Kasirajan K, Caps MT, Nelken N, Santana D (2005) Safety and efficacy of carotid arteriography in vascular surgery practice. J Vasc Surg 41:238–245PubMedCrossRef
29.
go back to reference Kerber CW, Cromwell LD, Drayer BP, Bank WO (1978) Cerebral ischemia. I. Current angiographic techniques, complications, and safety. AJR Am J Roentgenol 130:1097–1103PubMedCrossRef Kerber CW, Cromwell LD, Drayer BP, Bank WO (1978) Cerebral ischemia. I. Current angiographic techniques, complications, and safety. AJR Am J Roentgenol 130:1097–1103PubMedCrossRef
30.
go back to reference Mani RL, Eisenberg RL, McDonald EJ Jr, Pollock JA, Mani JR (1978) Complications of catheter cerebral arteriography: analysis of 5,000 procedures. I. Criteria and incidence. AJR Am J Roentgenol 131:861–865PubMedCrossRef Mani RL, Eisenberg RL, McDonald EJ Jr, Pollock JA, Mani JR (1978) Complications of catheter cerebral arteriography: analysis of 5,000 procedures. I. Criteria and incidence. AJR Am J Roentgenol 131:861–865PubMedCrossRef
31.
go back to reference Eisenberg RL, Bank WO, Hedgcock MW (1980) Neurologic complications of angiography for cerebrovascular disease. Neurology 30:895–897PubMedCrossRef Eisenberg RL, Bank WO, Hedgcock MW (1980) Neurologic complications of angiography for cerebrovascular disease. Neurology 30:895–897PubMedCrossRef
32.
go back to reference Wolfel DA, Lovett BP, Ortenburger AI, Johnson LS, Sommerville DL (1984) Outpatient arteriography: its safety and cost effectiveness. Radiology 153:363–364PubMedCrossRef Wolfel DA, Lovett BP, Ortenburger AI, Johnson LS, Sommerville DL (1984) Outpatient arteriography: its safety and cost effectiveness. Radiology 153:363–364PubMedCrossRef
33.
go back to reference Saint-Georges G, Aube M (1985) Safety of outpatient angiography: a prospective study. AJR Am J Roentgenol 144:235–236PubMedCrossRef Saint-Georges G, Aube M (1985) Safety of outpatient angiography: a prospective study. AJR Am J Roentgenol 144:235–236PubMedCrossRef
34.
go back to reference Gasparyan AY, Watson T, Lip GY (2008) The role of aspirin in cardiovascular prevention: implications of aspirin resistance. J Am Coll Cardiol 51:1829–1843PubMedCrossRef Gasparyan AY, Watson T, Lip GY (2008) The role of aspirin in cardiovascular prevention: implications of aspirin resistance. J Am Coll Cardiol 51:1829–1843PubMedCrossRef
35.
go back to reference Krasopoulos G, Brister SJ, Beattie WS, Buchanan MR (2008) Aspirin “resistance” and risk of cardiovascular morbidity: systematic review and meta-analysis. BMJ 336:195–198PubMedCentralPubMedCrossRef Krasopoulos G, Brister SJ, Beattie WS, Buchanan MR (2008) Aspirin “resistance” and risk of cardiovascular morbidity: systematic review and meta-analysis. BMJ 336:195–198PubMedCentralPubMedCrossRef
36.
go back to reference Patrono C, García Rodríguez LA, Landolfi R, Baigent C (2005) Low-dose aspirin for the prevention of atherothrombosis. N Engl J Med 353:2373–2383PubMedCrossRef Patrono C, García Rodríguez LA, Landolfi R, Baigent C (2005) Low-dose aspirin for the prevention of atherothrombosis. N Engl J Med 353:2373–2383PubMedCrossRef
37.
38.
go back to reference Feher G, Feher A, Pusch G, Koltai K, Tibold A, Gasztonyi B, Papp E, Szapary L, Kesmarky G, Toth K (2010) Clinical importance of aspirin and clopidogrel resistance. World J Cardiol 2:171–186PubMedCentralPubMedCrossRef Feher G, Feher A, Pusch G, Koltai K, Tibold A, Gasztonyi B, Papp E, Szapary L, Kesmarky G, Toth K (2010) Clinical importance of aspirin and clopidogrel resistance. World J Cardiol 2:171–186PubMedCentralPubMedCrossRef
39.
go back to reference Lee DH, Arat A, Morsi H, Shaltoni H, Harris JR, Mawad ME (2008) Dual antiplatelet therapy monitoring for neurointerventional procedures using a point-of-care platelet function test: a single-center experience. AJNR Am J Neuroradiol 29:1389–1394PubMedCrossRef Lee DH, Arat A, Morsi H, Shaltoni H, Harris JR, Mawad ME (2008) Dual antiplatelet therapy monitoring for neurointerventional procedures using a point-of-care platelet function test: a single-center experience. AJNR Am J Neuroradiol 29:1389–1394PubMedCrossRef
40.
go back to reference Snoep JD, Hovens MM, Eikenboom JC, van der Bom JG, Jukema JW, Huisman MV (2007) Clopidogrel nonresponsiveness in patients undergoing percutaneous coronary intervention with stenting: a systematic review and meta-analysis. Am Heart J 154:221–231PubMedCrossRef Snoep JD, Hovens MM, Eikenboom JC, van der Bom JG, Jukema JW, Huisman MV (2007) Clopidogrel nonresponsiveness in patients undergoing percutaneous coronary intervention with stenting: a systematic review and meta-analysis. Am Heart J 154:221–231PubMedCrossRef
41.
go back to reference Delgado Almandoz JE, Crandall BM, Scholz JM, Fease JL, Anderson RE, Kadkhodayan Y, Tubman DE (2013) Pre-procedure P2Y12 reaction units value predicts perioperative thromboembolic and hemorrhagic complications in patients with cerebral aneurysms treated with the Pipeline Embolization Device. J Neurointerv Surg 5(Suppl 3):iii3–iii10PubMedCrossRef Delgado Almandoz JE, Crandall BM, Scholz JM, Fease JL, Anderson RE, Kadkhodayan Y, Tubman DE (2013) Pre-procedure P2Y12 reaction units value predicts perioperative thromboembolic and hemorrhagic complications in patients with cerebral aneurysms treated with the Pipeline Embolization Device. J Neurointerv Surg 5(Suppl 3):iii3–iii10PubMedCrossRef
42.
go back to reference Kang HS, Kwon BJ, Kim JE, Han MH (2010) Preinterventional clopidogrel response variability for coil embolization of intracranial aneurysms: clinical implications. AJNR Am J Neuroradiol 31:1206–1210PubMedCrossRef Kang HS, Kwon BJ, Kim JE, Han MH (2010) Preinterventional clopidogrel response variability for coil embolization of intracranial aneurysms: clinical implications. AJNR Am J Neuroradiol 31:1206–1210PubMedCrossRef
43.
go back to reference Murthy SB, Shah S, Venkatasubba Rao CP, Bershad EM, Suarez JI (2014) Treatment of unruptured intracranial aneurysms with the pipeline embolization device. J Clin Neurosci 21:6–11PubMedCrossRef Murthy SB, Shah S, Venkatasubba Rao CP, Bershad EM, Suarez JI (2014) Treatment of unruptured intracranial aneurysms with the pipeline embolization device. J Clin Neurosci 21:6–11PubMedCrossRef
44.
go back to reference Tan LA, Keigher KM, Munich SA, Moftakhar R, Lopes DK (2014) Thromboembolic complications with Pipeline Embolization Device placement: impact of procedure time, number of stents and pre-procedure P2Y12 reaction unit (PRU) value. J Neurointerv Surg 19. doi:10.1136/neurintsurg-2014-011111 Tan LA, Keigher KM, Munich SA, Moftakhar R, Lopes DK (2014) Thromboembolic complications with Pipeline Embolization Device placement: impact of procedure time, number of stents and pre-procedure P2Y12 reaction unit (PRU) value. J Neurointerv Surg 19. doi:10.​1136/​neurintsurg-2014-011111
45.
go back to reference Heller RS, Dandamudi V, Lanfranchi M, Malek AM (2013) Effect of antiplatelet therapy on thromboembolism after flow diversion with the pipeline embolization device. J Neurosurg 119:1603–1610PubMedCrossRef Heller RS, Dandamudi V, Lanfranchi M, Malek AM (2013) Effect of antiplatelet therapy on thromboembolism after flow diversion with the pipeline embolization device. J Neurosurg 119:1603–1610PubMedCrossRef
Metadata
Title
Thromboembolic events associated with single balloon-, double balloon-, and stent-assisted coil embolization of asymptomatic unruptured cerebral aneurysms: evaluation with diffusion-weighted MR imaging
Authors
Tomoji Takigawa
Kensuke Suzuki
Yoshiki Sugiura
Ryotaro Suzuki
Issei Takano
Nobuyuki Shimizu
Yoshihiro Tanaka
Akio Hyodo
Publication date
01-12-2014
Publisher
Springer Berlin Heidelberg
Published in
Neuroradiology / Issue 12/2014
Print ISSN: 0028-3940
Electronic ISSN: 1432-1920
DOI
https://doi.org/10.1007/s00234-014-1421-4

Other articles of this Issue 12/2014

Neuroradiology 12/2014 Go to the issue