01-09-2011 | Interventional Neuroradiology

An in vitro study of silk stent morphology

Authors: Thaweesak Aurboonyawat, Raphaël Blanc, Paul Schmidt, Michel Piotin, Laurent Spelle, Amir Nakib, Jacques Moret

Published in: Neuroradiology | Issue 9/2011

Login to get access

Abstract

Introduction

Morphology of the Silk stent (Balt, Montmorency, France) after deployment is not fully understood, especially in tortuous vessels. An in vitro study was conducted to study morphology and flow-diverting parameters of this stent.

Methods

Two sets of different-sized and curved polytetrafluoroethylene tubes were studied. To simulate the aneurysm neck, a small hole was created in a tube. A stent was placed in each of the different tubes. Angiographic computerized tomography and macroscopic photography were then obtained. The images were analyzed to calculate a Percentage of Area Coverage (PAC).

Results

Good stent conformability was observed. The PAC was 21% in the straight model with matched stent and vessel diameter. In the straight model with an oversized stent, the PAC was increased. In the curved models, dynamic wire repositioning occurred. The repositioning was affected by the size of the stent and the angle of the vessel curve. Compared to the straight model, this increased the PAC in two instances: on the convexity (oversized stent), and on the concavity (matched stent and vessel diameter). The PAC did not significantly change at the sides of the curve.

Conclusions

By design, the wires of the silk stent move relative to each other. In a curved model, the PAC is different at the convexity, concavity, and lateral walls. The stent diameter affects the PAC. These results are clinically relevant because it is desirable to maximize and minimize the PAC across the aneurysm neck and branch vessel orifice, respectively.
Literature
1.
go back to reference Fiorella D, Kelly ME, Albuquerque FC et al (2009) Curative reconstruction of a giant midbasilar trunk aneurysm with the pipeline embolization device. Neurosurgery 64:212–217, discussion 217PubMedCrossRef Fiorella D, Kelly ME, Albuquerque FC et al (2009) Curative reconstruction of a giant midbasilar trunk aneurysm with the pipeline embolization device. Neurosurgery 64:212–217, discussion 217PubMedCrossRef
2.
go back to reference Fiorella D, Woo HH, Albuquerque FC et al (2008) Definitive reconstruction of circumferential, fusiform intracranial aneurysms with the pipeline embolization device. Neurosurgery 62:1115–1120, discussion 1120-1111PubMedCrossRef Fiorella D, Woo HH, Albuquerque FC et al (2008) Definitive reconstruction of circumferential, fusiform intracranial aneurysms with the pipeline embolization device. Neurosurgery 62:1115–1120, discussion 1120-1111PubMedCrossRef
3.
go back to reference Lylyk P, Miranda C, Ceratto R et al (2009) Curative endovascular reconstruction of cerebral aneurysms with the pipeline embolization device: the Buenos Aires experience. Neurosurgery 64:632–642, discussion 642-633; quiz N636PubMedCrossRef Lylyk P, Miranda C, Ceratto R et al (2009) Curative endovascular reconstruction of cerebral aneurysms with the pipeline embolization device: the Buenos Aires experience. Neurosurgery 64:632–642, discussion 642-633; quiz N636PubMedCrossRef
4.
go back to reference Szikora I, Berentei Z, Kulcsar Z et al (2010) Treatment of intracranial aneurysms by functional reconstruction of the parent artery: the Budapest experience with the pipeline embolization device. AJNR Am J Neuroradiol 31:1139–1147PubMedCrossRef Szikora I, Berentei Z, Kulcsar Z et al (2010) Treatment of intracranial aneurysms by functional reconstruction of the parent artery: the Budapest experience with the pipeline embolization device. AJNR Am J Neuroradiol 31:1139–1147PubMedCrossRef
5.
go back to reference Ebrahimi N, Claus B, Lee CY et al (2007) Stent conformity in curved vascular models with simulated aneurysm necks using flat-panel CT: an in vitro study. AJNR Am J Neuroradiol 28:823–829PubMed Ebrahimi N, Claus B, Lee CY et al (2007) Stent conformity in curved vascular models with simulated aneurysm necks using flat-panel CT: an in vitro study. AJNR Am J Neuroradiol 28:823–829PubMed
6.
go back to reference Nakib A, Oulhadja H, Siarry P (2009) A thresholding method based on two-dimensional fractional differentiation. Image Vis Comput 27:1343–1357CrossRef Nakib A, Oulhadja H, Siarry P (2009) A thresholding method based on two-dimensional fractional differentiation. Image Vis Comput 27:1343–1357CrossRef
7.
go back to reference Alvarado MVY, Ebrahimi N, Benndorf G (2009) Study of conformability of the new LEO PLUS stent to a curved vascular mode using flat-panel detector computed tomography (DynaCT). Neurosurgery 64:S130–S134, discussion 134 Alvarado MVY, Ebrahimi N, Benndorf G (2009) Study of conformability of the new LEO PLUS stent to a curved vascular mode using flat-panel detector computed tomography (DynaCT). Neurosurgery 64:S130–S134, discussion 134
8.
go back to reference Steiger HJ (2007) Pathophysiology of development and rupture of cerebral aneurysms. Acta Neurochir Suppl (Wien) 48:1–57CrossRef Steiger HJ (2007) Pathophysiology of development and rupture of cerebral aneurysms. Acta Neurochir Suppl (Wien) 48:1–57CrossRef
10.
go back to reference Lieber BB, Stancampiano AP, Wakhloo AK (1997) Alteration of hemodynamics in aneurysm models by stenting: influence of stent porosity. Ann Biomed Eng 25:460–469PubMedCrossRef Lieber BB, Stancampiano AP, Wakhloo AK (1997) Alteration of hemodynamics in aneurysm models by stenting: influence of stent porosity. Ann Biomed Eng 25:460–469PubMedCrossRef
11.
go back to reference Liou TM, Li YC (2008) Effects of stent porosity on hemodynamics in a sidewall aneurysm model. J Biomech 41:1174–1183PubMedCrossRef Liou TM, Li YC (2008) Effects of stent porosity on hemodynamics in a sidewall aneurysm model. J Biomech 41:1174–1183PubMedCrossRef
12.
go back to reference Kim JH, Kang TJ, Yu WR (2008) Mechanical modeling of self-expandable stent fabricated using braiding technology. J Biomech 41:3202–3212PubMedCrossRef Kim JH, Kang TJ, Yu WR (2008) Mechanical modeling of self-expandable stent fabricated using braiding technology. J Biomech 41:3202–3212PubMedCrossRef
13.
go back to reference Kallmes DF, Ding YH, Dai D et al (2007) A new endoluminal, flow-disrupting device for treatment of saccular aneurysms. Stroke 38:2346–2352PubMedCrossRef Kallmes DF, Ding YH, Dai D et al (2007) A new endoluminal, flow-disrupting device for treatment of saccular aneurysms. Stroke 38:2346–2352PubMedCrossRef
Metadata
Title
An in vitro study of silk stent morphology
Authors
Thaweesak Aurboonyawat
Raphaël Blanc
Paul Schmidt
Michel Piotin
Laurent Spelle
Amir Nakib
Jacques Moret
Publication date
01-09-2011
Publisher
Springer-Verlag
Published in
Neuroradiology / Issue 9/2011
Print ISSN: 0028-3940
Electronic ISSN: 1432-1920
DOI
https://doi.org/10.1007/s00234-010-0784-4

Other articles of this Issue 9/2011

Neuroradiology 9/2011 Go to the issue