Precision and accuracy of measuring changes in bone mineral density by dual-energy X-ray absorptiometry
Authors:
P. Tothill, W. J. Hannan
Published in:
Osteoporosis International
|
Issue 11/2007
Login to get access
Abstract
Summary
Long-term precision of two Hologic DXA scanners was derived from repeated clinical measurements. With typical subjects, the long-term coefficients of variation were about twice the short-term. The accuracy of the measurement of changes was compromised by anomalies, but this did not seriously affect clinical conclusions.
Introduction
Long-term precision and accuracy of BMD measurements need review.
Methods
Long-term precision was examined by selecting, from Hologic databases, subjects who had had four scans over a period of 2 to 5 years and was calculated from the SEE of a regression of BMD against time. Accuracy was assessed from relationships between changes (Δ) in BA, BMD and BMC.
Results
For one group of subjects, the long-term precision was 2.4% for lumbar spine, 2.3% for total hip and 2.7% for femoral neck when expressed as CVs. These values were nearly twice the short-term CVs of 1.3% for spine, 1.2% for total hip and 1.4% for femoral neck. For another group, a negative exponential regression gave a better fit, leading to CVs of 1.3% for the spine, 1.4% for total hip and 2.1% for femoral neck. Significant correlations between ΔBA and ΔBMC were found. These led to an underestimate of ΔBMD in spine and hip by 25%.
Conclusions
The poorer long-term precision for typical patients should be borne in mind in monitoring progress. The underestimate of changes could account for only a part of the underestimation by BMD measurements of the anti-fracture effects of anti-resorptive drugs.