Published in: Knee Surgery, Sports Traumatology, Arthroscopy 3/2014

01-03-2014 | Knee

No difference in in vivo polyethylene wear particles between oxidized zirconium and cobalt–chromium femoral component in total knee arthroplasty

Authors: Yukihide Minoda, Kanako Hata, Hiroyoshi Iwaki, Mitsuhiko Ikebuchi, Yusuke Hashimoto, Fumiaki Inori, Hiroaki Nakamura

Published in: Knee Surgery, Sports Traumatology, Arthroscopy | Issue 3/2014

Login to get access

Abstract

Purpose

Polyethylene wear particle generation is one of the most important factors affecting mid- to long-term results of total knee arthroplasties. Oxidized zirconium was introduced as a material for femoral components to reduce polyethylene wear generation. However, an in vivo advantage of oxidized zirconium on polyethylene wear particle generation is still controversial. The purpose of this study was to compare in vivo polyethylene wear particles between oxidized zirconium total knee prosthesis and conventional cobalt–chromium (Co–Cr) total knee prosthesis.

Methods

Synovial fluid was obtained from the knees of 6 patients with oxidized zirconium total knee prosthesis and from 6 patients with conventional cobalt–chromium (Co–Cr) total knee prosthesis 12 months after the operation. Polyethylene particles were isolated and examined using a scanning electron microscope and image analyser.

Results

Total number of particles in each knee was 3.3 ± 1.3 × 107 in the case of oxidized zirconium (mean ± SD) and 3.4 ± 1.2 × 107 in that of Co–Cr (n.s.). The particle size (equivalent circle diameter) was 0.8 ± 0.3 μm in the case of oxidized zirconium and 0.6 ± 0.1 μm in that of Co–Cr (n.s.). The particle shape (aspect ratio) was 1.4 ± 0.0 in the case of oxidized zirconium and 1.4 ± 0.0 in that of metal Co–Cr (n.s).

Conclusions

Although newly introduced oxidized zirconium femoral component did not reduce the in vivo polyethylene wear particles in early clinical stage, there was no adverse effect of newly introduced material. At this moment, there is no need to abandon oxidized zirconium femoral component. However, further follow-up of polyethylene wear particle generation should be performed to confirm the advantage of the oxidized zirconium femoral component.

Level of evidence

Therapeutic study, Level III.
Literature
1.
go back to reference Amstutz HC, Thomas BJ, Jinnah R et al (1984) Treatment of primary osteoarthritis of the hip. A comparison of total joint and surface replacement arthroplasty. J Bone Joint Surg 66A:228–241 Amstutz HC, Thomas BJ, Jinnah R et al (1984) Treatment of primary osteoarthritis of the hip. A comparison of total joint and surface replacement arthroplasty. J Bone Joint Surg 66A:228–241
2.
go back to reference Benevenia J, Lee FY, Buechel F et al (1988) Pathologic supracondylar fracture due to osteolytic pseudotumor of knee following cementless total knee replacement. J Biomed Mater Res 43:473–477CrossRef Benevenia J, Lee FY, Buechel F et al (1988) Pathologic supracondylar fracture due to osteolytic pseudotumor of knee following cementless total knee replacement. J Biomed Mater Res 43:473–477CrossRef
3.
go back to reference Berry DJ, Currier JH, Mayor MB et al (2012) Knee wear measured in retrievals: a polished tray reduces insert wear. Clin Orthop Relat Res 470:1860–1868PubMedCrossRef Berry DJ, Currier JH, Mayor MB et al (2012) Knee wear measured in retrievals: a polished tray reduces insert wear. Clin Orthop Relat Res 470:1860–1868PubMedCrossRef
4.
go back to reference DeHeer DH, Engels JA, DeVries AS et al (2001) In situ complement activation by polyethylene wear debris. J Biomed Mater Res 54:12–19PubMedCrossRef DeHeer DH, Engels JA, DeVries AS et al (2001) In situ complement activation by polyethylene wear debris. J Biomed Mater Res 54:12–19PubMedCrossRef
5.
go back to reference Engh GA, Dwyer KA, Hanes CK (1988) Polyethylene wear of metal-backed tibial components in total and unicompartmental knee prosthesis. J Bone Joint Surg 74B:9–17 Engh GA, Dwyer KA, Hanes CK (1988) Polyethylene wear of metal-backed tibial components in total and unicompartmental knee prosthesis. J Bone Joint Surg 74B:9–17
6.
go back to reference Ezzet KA, Hermida JC, Colwell CW Jr et al (2004) Oxidized zirconium femoral components reduce polyethylene wear in a knee wear simulator. Clin Orthop Relat Res 428:120–124PubMedCrossRef Ezzet KA, Hermida JC, Colwell CW Jr et al (2004) Oxidized zirconium femoral components reduce polyethylene wear in a knee wear simulator. Clin Orthop Relat Res 428:120–124PubMedCrossRef
7.
go back to reference Heyse TJ, Chen DX, Kelly N, Boettner F et al (2011) Matched-pair total knee arthroplasty retrieval analysis: oxidized zirconium vs. CoCrMo. Knee 18:448–452PubMedCrossRef Heyse TJ, Chen DX, Kelly N, Boettner F et al (2011) Matched-pair total knee arthroplasty retrieval analysis: oxidized zirconium vs. CoCrMo. Knee 18:448–452PubMedCrossRef
8.
go back to reference Innocenti M, Civinini R, Carulli C et al (2010) The 5-year results of an oxidized zirconium femoral component for TKA. Clin Orthop Relat Res 468:1258–1263PubMedCrossRef Innocenti M, Civinini R, Carulli C et al (2010) The 5-year results of an oxidized zirconium femoral component for TKA. Clin Orthop Relat Res 468:1258–1263PubMedCrossRef
9.
go back to reference Insall J, Dorr L, Scott R, Scott W (1989) Rationale of the knee society clinical rating system. Clin Orthop Relat Res 248:13–14PubMed Insall J, Dorr L, Scott R, Scott W (1989) Rationale of the knee society clinical rating system. Clin Orthop Relat Res 248:13–14PubMed
10.
go back to reference Iwakiri K, Minoda Y, Kobayashi A et al (2009) In vivo comparison of wear particles between highly crosslinked polyethylene and conventional polyethylene in the same design of total knee arthroplasties. J Biomed Mater Res B Appl Biomater 91:799–804PubMedCrossRef Iwakiri K, Minoda Y, Kobayashi A et al (2009) In vivo comparison of wear particles between highly crosslinked polyethylene and conventional polyethylene in the same design of total knee arthroplasties. J Biomed Mater Res B Appl Biomater 91:799–804PubMedCrossRef
11.
go back to reference Kadoya Y, Revell PA, al-Saffar N et al (1996) Bone formation and bone resorption in failed total joint arthroplasties: histomorphometric analysis with histochemical and immunohistochemical technique. J Orthop Res 14:473–482PubMedCrossRef Kadoya Y, Revell PA, al-Saffar N et al (1996) Bone formation and bone resorption in failed total joint arthroplasties: histomorphometric analysis with histochemical and immunohistochemical technique. J Orthop Res 14:473–482PubMedCrossRef
12.
go back to reference Kadoya Y, Revell PA, Kobayashi A et al (1997) Wear particulate species and bone loss in failed total joint arthroplasties. Clin Orthop Relat Res 340:118–129PubMedCrossRef Kadoya Y, Revell PA, Kobayashi A et al (1997) Wear particulate species and bone loss in failed total joint arthroplasties. Clin Orthop Relat Res 340:118–129PubMedCrossRef
13.
go back to reference Kobayashi A, Freeman MA, Bonfield W et al (1997) Number of polyethylene particles and osteolysis in total joint replacements. A quantitative study using a tissue-digestion method. J Bone Joint Surg 79B:844–848CrossRef Kobayashi A, Freeman MA, Bonfield W et al (1997) Number of polyethylene particles and osteolysis in total joint replacements. A quantitative study using a tissue-digestion method. J Bone Joint Surg 79B:844–848CrossRef
14.
go back to reference Laskin RS (2003) An oxidized Zr ceramic surfaced femoral component for total knee arthroplasty. Clin Orthop Relat Res 416:191–196PubMedCrossRef Laskin RS (2003) An oxidized Zr ceramic surfaced femoral component for total knee arthroplasty. Clin Orthop Relat Res 416:191–196PubMedCrossRef
15.
go back to reference Laskin RS, Davis J (2005) Total knee replacement using the Genesis II prosthesis: a 5-year follow up study of the first 100 consecutive cases. Knee 12:163–167PubMedCrossRef Laskin RS, Davis J (2005) Total knee replacement using the Genesis II prosthesis: a 5-year follow up study of the first 100 consecutive cases. Knee 12:163–167PubMedCrossRef
16.
go back to reference Matthews JB, Besong AA, Green TR et al (2000) Evaluation of the response of primary human peripheral blood mononuclear phagocytes to challenge with in vitro generated clinically relevant UHMWPE particles of known size and dose. J Biomed Mater Res 52:296–307PubMedCrossRef Matthews JB, Besong AA, Green TR et al (2000) Evaluation of the response of primary human peripheral blood mononuclear phagocytes to challenge with in vitro generated clinically relevant UHMWPE particles of known size and dose. J Biomed Mater Res 52:296–307PubMedCrossRef
17.
go back to reference Minoda Y, Kobayashi A, Iwaki H et al (2003) Polyethylene wear particles in synovial fluid after total knee arthroplasty. Clin Orthop Relat Res 410:165–172PubMedCrossRef Minoda Y, Kobayashi A, Iwaki H et al (2003) Polyethylene wear particles in synovial fluid after total knee arthroplasty. Clin Orthop Relat Res 410:165–172PubMedCrossRef
18.
go back to reference Minoda Y, Kobayashi A, Iwaki H et al (2004) Characteristics of polyethylene wear particles isolated from synovial fluid after mobile-bearing and posterior-stabilized total knee arthroplasties. J Biomed Mater Res 71B:1–6CrossRef Minoda Y, Kobayashi A, Iwaki H et al (2004) Characteristics of polyethylene wear particles isolated from synovial fluid after mobile-bearing and posterior-stabilized total knee arthroplasties. J Biomed Mater Res 71B:1–6CrossRef
19.
go back to reference Minoda Y, Kobayashi A, Iwaki H et al (2005) Polyethylene wear particle generation in vivo in an alumina medial pivot total knee prosthesis. Biomaterials 26:6034–6040PubMedCrossRef Minoda Y, Kobayashi A, Iwaki H et al (2005) Polyethylene wear particle generation in vivo in an alumina medial pivot total knee prosthesis. Biomaterials 26:6034–6040PubMedCrossRef
20.
go back to reference Minoda Y, Kobayashi A, Iwaki H et al (2009) In vivo analysis of polyethylene wear particles after total knee arthroplasty: the influence of improved materials and designs. J Bone Joint Surg 91A(Suppl 6):67–73CrossRef Minoda Y, Kobayashi A, Iwaki H et al (2009) In vivo analysis of polyethylene wear particles after total knee arthroplasty: the influence of improved materials and designs. J Bone Joint Surg 91A(Suppl 6):67–73CrossRef
21.
22.
go back to reference Schmalzried TP, Szuszczewicz ES, Northfield MR et al (1998) Quantitative assessment of walking activity after total hip or knee replacement. J Bone Joint Surg 80A:54–59 Schmalzried TP, Szuszczewicz ES, Northfield MR et al (1998) Quantitative assessment of walking activity after total hip or knee replacement. J Bone Joint Surg 80A:54–59
23.
go back to reference Spector BM, Ries MD, Bourne RB et al (2001) Wear performance of ultra-high molecular weight polyethylene on oxidized zirconium total knee femoral components. J Bone Joint Surg 83A(Suppl 2 Pt 2):80–86 Spector BM, Ries MD, Bourne RB et al (2001) Wear performance of ultra-high molecular weight polyethylene on oxidized zirconium total knee femoral components. J Bone Joint Surg 83A(Suppl 2 Pt 2):80–86
24.
go back to reference White SE, Whiteside LA, McCarthy DS et al (1944) Simulated knee wear with cobalt chromium and oxidized zirconium knee femoral components. Clin Orthop Relat Res 309:176–184 White SE, Whiteside LA, McCarthy DS et al (1944) Simulated knee wear with cobalt chromium and oxidized zirconium knee femoral components. Clin Orthop Relat Res 309:176–184
25.
go back to reference Willert HG (1977) Reactions of the articular capsule to wear products of artificial joint prostheses. J Biomed Mater Res 11:157–164PubMedCrossRef Willert HG (1977) Reactions of the articular capsule to wear products of artificial joint prostheses. J Biomed Mater Res 11:157–164PubMedCrossRef
26.
go back to reference Yang SY, Ren W, Park Y et al (2002) Diverse cellular and apoptotic responses to variant shapes of UHMWPE particles in a murine model of inflammation. Biomaterials 23:3535–3543PubMedCrossRef Yang SY, Ren W, Park Y et al (2002) Diverse cellular and apoptotic responses to variant shapes of UHMWPE particles in a murine model of inflammation. Biomaterials 23:3535–3543PubMedCrossRef
Metadata
Title
No difference in in vivo polyethylene wear particles between oxidized zirconium and cobalt–chromium femoral component in total knee arthroplasty
Authors
Yukihide Minoda
Kanako Hata
Hiroyoshi Iwaki
Mitsuhiko Ikebuchi
Yusuke Hashimoto
Fumiaki Inori
Hiroaki Nakamura
Publication date
01-03-2014
Publisher
Springer Berlin Heidelberg
Published in
Knee Surgery, Sports Traumatology, Arthroscopy / Issue 3/2014
Print ISSN: 0942-2056
Electronic ISSN: 1433-7347
DOI
https://doi.org/10.1007/s00167-013-2724-4

Other articles of this Issue 3/2014

Knee Surgery, Sports Traumatology, Arthroscopy 3/2014 Go to the issue