Published in: Knee Surgery, Sports Traumatology, Arthroscopy 3/2014

01-03-2014 | Knee

Biomechanical analysis of the tibial tray design in TKA: comparison between modular and offset tibial trays

Authors: Ângela Chan, João Gamelas, João Folgado, Paulo R. Fernandes

Published in: Knee Surgery, Sports Traumatology, Arthroscopy | Issue 3/2014

Login to get access

Abstract

Purpose

The aim of this work was to develop a computational biomechanical study to compare the performance of tibial trays with different offsets for a total knee arthroplasty. The goal was to investigate whether the offset tibial tray shifts the bone stress distribution, influencing the clinical outcome.

Methods

Three geometric models were developed for the intact tibia bone: one considering a standard tibia case and the other two reproducing tibias with a medial or a lateral offset of the metaphysis. Appropriate prosthetic components were assembled in the bone for the aforementioned cases. The finite element method was used to obtain the mechanical stress distribution for the models, and the stress shielding effect due to the prosthesis was analysed.

Results

The obtained results revealed that the offset cases are subjected to higher stresses than the standard case. These values can be two times superior to the ones verified in a standard case. The stress shielding effect was confirmed along all the analysed paths, except near the stem’s end in some areas.

Conclusion

The higher stresses registered can originate lower clinical outcomes in the offset cases. These findings can be an important beginning to understand whether better bone stress distribution could be achieved in deformity correction with associated osteotomies instead of offsetting.
Literature
1.
go back to reference Abraham R, Malkani AL, Lewis J, Beck D (2007) An anatomical study of tibial metaphyseal/diaphyseal mismatch during revision total knee arthroplasty. J Arthroplasty 22(2):241–244PubMedCrossRef Abraham R, Malkani AL, Lewis J, Beck D (2007) An anatomical study of tibial metaphyseal/diaphyseal mismatch during revision total knee arthroplasty. J Arthroplasty 22(2):241–244PubMedCrossRef
2.
go back to reference Barrack RL, Rorabeck C, Burt M, Sawhney J (1999) Pain at the end of the stem after revision total knee arthroplasty. Clin Orthop Relat Res 367:216–225PubMedCrossRef Barrack RL, Rorabeck C, Burt M, Sawhney J (1999) Pain at the end of the stem after revision total knee arthroplasty. Clin Orthop Relat Res 367:216–225PubMedCrossRef
3.
go back to reference Barrack RL, Stanley T, Burt M (2004) The effect of stem design on end-of-stem pain in revision total knee arthroplasty. J Arthroplasty 19:119–120PubMedCrossRef Barrack RL, Stanley T, Burt M (2004) The effect of stem design on end-of-stem pain in revision total knee arthroplasty. J Arthroplasty 19:119–120PubMedCrossRef
4.
go back to reference Bendjaballah MZ, Shirazi-Adl A, Zukor DJ (1995) Biomechanics of the human knee joint in compression: reconstruction, mesh generation and finite element analysis. Knee 2:69–79CrossRef Bendjaballah MZ, Shirazi-Adl A, Zukor DJ (1995) Biomechanics of the human knee joint in compression: reconstruction, mesh generation and finite element analysis. Knee 2:69–79CrossRef
5.
go back to reference Bourne RB, Finlay JB (1986) The influence of tibial component intramedullary stems and implant-cortex contact on the strain distribution of the proximal tibia following total knee arthroplasty. An in vitro study. Clin Orthop Relat Res 208:95–99PubMed Bourne RB, Finlay JB (1986) The influence of tibial component intramedullary stems and implant-cortex contact on the strain distribution of the proximal tibia following total knee arthroplasty. An in vitro study. Clin Orthop Relat Res 208:95–99PubMed
6.
go back to reference Carr BC, Goswami T (2009) Knee implants—review of models and biomechanics. Mater Des 30:398–413CrossRef Carr BC, Goswami T (2009) Knee implants—review of models and biomechanics. Mater Des 30:398–413CrossRef
7.
go back to reference Completo A, Fonseca F, Simões JA (2007) Finite element and experimental cortex strains of the intact and implanted tibia. J Biomech Eng 129:791–797PubMedCrossRef Completo A, Fonseca F, Simões JA (2007) Finite element and experimental cortex strains of the intact and implanted tibia. J Biomech Eng 129:791–797PubMedCrossRef
8.
go back to reference Completo A, Simões JA, Fonseca F, Oliveira M (2008) The influence of different tibial stem designs in load sharing and stability at the cement-bone interface in revision TKA. Knee 15:227–232PubMedCrossRef Completo A, Simões JA, Fonseca F, Oliveira M (2008) The influence of different tibial stem designs in load sharing and stability at the cement-bone interface in revision TKA. Knee 15:227–232PubMedCrossRef
9.
go back to reference Completo A, Talaia P, Fonseca F, Simões JA (2009) Relationship of design features of stemmed tibial knee prosthesis with stress shielding and end-of-stem pain. Mater Des 30:1391–1397CrossRef Completo A, Talaia P, Fonseca F, Simões JA (2009) Relationship of design features of stemmed tibial knee prosthesis with stress shielding and end-of-stem pain. Mater Des 30:1391–1397CrossRef
10.
go back to reference Darwish SM, Al-Samhan A (2008) The effect of cement stiffness and tibia tray material on the stresses developed in artificial knee. Int J Adhes Adhes 28:120–125CrossRef Darwish SM, Al-Samhan A (2008) The effect of cement stiffness and tibia tray material on the stresses developed in artificial knee. Int J Adhes Adhes 28:120–125CrossRef
11.
go back to reference Doblaré M, García JM, Gómez MJ (2004) Modelling bone tissue fracture and healing: a review. Eng Fract Mech 71:1809–1840CrossRef Doblaré M, García JM, Gómez MJ (2004) Modelling bone tissue fracture and healing: a review. Eng Fract Mech 71:1809–1840CrossRef
12.
go back to reference Donahue TL, Hull ML, Rashid MM, Jacobs CR (2002) A finite element model of the human knee joint for the study of tibio-femoral contact. J Biomech Eng 124(3):273–280PubMedCrossRef Donahue TL, Hull ML, Rashid MM, Jacobs CR (2002) A finite element model of the human knee joint for the study of tibio-femoral contact. J Biomech Eng 124(3):273–280PubMedCrossRef
13.
go back to reference Espinoza Orias AA, Deuerling JM, Landrigan MD, Renaud JE, Roeder RK (2009) Anatomic variation in the elastic anisotropy of cortical bone tissue in the human femur. J Mech Behav Biomed Mater 2(3):255–263PubMedCentralPubMedCrossRef Espinoza Orias AA, Deuerling JM, Landrigan MD, Renaud JE, Roeder RK (2009) Anatomic variation in the elastic anisotropy of cortical bone tissue in the human femur. J Mech Behav Biomed Mater 2(3):255–263PubMedCentralPubMedCrossRef
14.
go back to reference Grupp TM, Kaddick C, Schwiesau J, Maas A, Stulberg SD (2009) Fixed and mobile bearing total knee arthroplasty—influence on wear generation, corresponding wear areas, knee kinematics and particle composition. Clin Biomech (Bristol, Avon) 24(2):210–217CrossRef Grupp TM, Kaddick C, Schwiesau J, Maas A, Stulberg SD (2009) Fixed and mobile bearing total knee arthroplasty—influence on wear generation, corresponding wear areas, knee kinematics and particle composition. Clin Biomech (Bristol, Avon) 24(2):210–217CrossRef
15.
go back to reference Heinlein B, Kutzner I, Graichen F, Bender A, Rohlmann A, Halder AM, Beier A, Bergmann G (2009) ESB clinical biomechanics award 2008: complete data of total knee replacement loading for level walking and stair climbing measured in vivo with a follow-up of 6–10 months. Clin Biomech (Bristol, Avon) 24(4):315–326CrossRef Heinlein B, Kutzner I, Graichen F, Bender A, Rohlmann A, Halder AM, Beier A, Bergmann G (2009) ESB clinical biomechanics award 2008: complete data of total knee replacement loading for level walking and stair climbing measured in vivo with a follow-up of 6–10 months. Clin Biomech (Bristol, Avon) 24(4):315–326CrossRef
16.
go back to reference Lavernia CJ, Sierra RJ, Hungerford DS, Krackow K (2001) Activity level and wear in total knee arthroplasty: a study of autopsy retrieved specimens. J Arthroplasty 16(4):446–453PubMedCrossRef Lavernia CJ, Sierra RJ, Hungerford DS, Krackow K (2001) Activity level and wear in total knee arthroplasty: a study of autopsy retrieved specimens. J Arthroplasty 16(4):446–453PubMedCrossRef
17.
go back to reference Mann KA, Bartel DL, Wright TM, Ingraffea AR (1991) Mechanical characteristics of the stem-cement interface. J Orthop Res 9(6):798–808PubMedCrossRef Mann KA, Bartel DL, Wright TM, Ingraffea AR (1991) Mechanical characteristics of the stem-cement interface. J Orthop Res 9(6):798–808PubMedCrossRef
18.
go back to reference Martin RB, Burr David B, Sharkey NA (1998) Skeletal tissue mechanics. Springer, New YorkCrossRef Martin RB, Burr David B, Sharkey NA (1998) Skeletal tissue mechanics. Springer, New YorkCrossRef
19.
go back to reference Miyoshi S, Takahashi T, Ohtani M, Yamamoto H, Kameyama K (2002) Analysis of the shape of the tibial tray in total knee arthroplasty using a three dimension finite element model. Clin Biomech (Bristol, Avon) 17(7):521–525CrossRef Miyoshi S, Takahashi T, Ohtani M, Yamamoto H, Kameyama K (2002) Analysis of the shape of the tibial tray in total knee arthroplasty using a three dimension finite element model. Clin Biomech (Bristol, Avon) 17(7):521–525CrossRef
20.
go back to reference Rancourt D, Shirazi-Adl A, Drouin G, Paiement G (1990) Friction properties of the interface between porous-surfaced metals and tibial cancellous bone. J Biomed Mater Res 24(11):1503–1519PubMedCrossRef Rancourt D, Shirazi-Adl A, Drouin G, Paiement G (1990) Friction properties of the interface between porous-surfaced metals and tibial cancellous bone. J Biomed Mater Res 24(11):1503–1519PubMedCrossRef
21.
go back to reference Rand JA, Coventry MB (1980) Stress fractures after total knee arthroplasty. J Bone Joint Surg Am 62(2):226–233PubMed Rand JA, Coventry MB (1980) Stress fractures after total knee arthroplasty. J Bone Joint Surg Am 62(2):226–233PubMed
22.
go back to reference Shirazi-Adl A, Dammak M, Paiement G (1993) Experimental determination of friction characteristics at the trabecular bone/porous-coated metal interface in cementless implants. J Biomed Mater Res 27(2):167–175PubMedCrossRef Shirazi-Adl A, Dammak M, Paiement G (1993) Experimental determination of friction characteristics at the trabecular bone/porous-coated metal interface in cementless implants. J Biomed Mater Res 27(2):167–175PubMedCrossRef
23.
go back to reference Taddei F, Pancanti A, Viceconti M (2004) An improved method for the automatic mapping of computed tomography numbers onto finite element models. Med Eng Phys 26(1):61–69PubMedCrossRef Taddei F, Pancanti A, Viceconti M (2004) An improved method for the automatic mapping of computed tomography numbers onto finite element models. Med Eng Phys 26(1):61–69PubMedCrossRef
24.
go back to reference Tharani R, Nakasone C, Vince KG (2005) Periprosthetic fractures after total knee arthroplasty. J Arthroplasty 20:27–32PubMedCrossRef Tharani R, Nakasone C, Vince KG (2005) Periprosthetic fractures after total knee arthroplasty. J Arthroplasty 20:27–32PubMedCrossRef
25.
go back to reference Viceconti M, Muccini R, Bernakiewicz M, Baleani M, Cristofolini L (2000) Large-sliding contact elements accurately predict levels of bone-implant micromotion relevant to osseointegration. J Biomech 33(12):1611–1618PubMedCrossRef Viceconti M, Muccini R, Bernakiewicz M, Baleani M, Cristofolini L (2000) Large-sliding contact elements accurately predict levels of bone-implant micromotion relevant to osseointegration. J Biomech 33(12):1611–1618PubMedCrossRef
26.
go back to reference Villa T, Migliavacca F, Gastaldi D, Colombo M, Pietrabissa R (2004) Contact stresses and fatigue life in a knee prosthesis: comparison between in vitro measurements and computational simulation. J Biomech 37:45–53PubMedCrossRef Villa T, Migliavacca F, Gastaldi D, Colombo M, Pietrabissa R (2004) Contact stresses and fatigue life in a knee prosthesis: comparison between in vitro measurements and computational simulation. J Biomech 37:45–53PubMedCrossRef
Metadata
Title
Biomechanical analysis of the tibial tray design in TKA: comparison between modular and offset tibial trays
Authors
Ângela Chan
João Gamelas
João Folgado
Paulo R. Fernandes
Publication date
01-03-2014
Publisher
Springer Berlin Heidelberg
Published in
Knee Surgery, Sports Traumatology, Arthroscopy / Issue 3/2014
Print ISSN: 0942-2056
Electronic ISSN: 1433-7347
DOI
https://doi.org/10.1007/s00167-013-2628-3

Other articles of this Issue 3/2014

Knee Surgery, Sports Traumatology, Arthroscopy 3/2014 Go to the issue