Skip to main content
Top
Published in: Knee Surgery, Sports Traumatology, Arthroscopy 3/2012

Open Access 01-03-2012 | Knee

No differences in in vivo kinematics between six different types of knee prostheses

Authors: N. Wolterbeek, R. G. H. H. Nelissen, E. R. Valstar

Published in: Knee Surgery, Sports Traumatology, Arthroscopy | Issue 3/2012

Login to get access

Abstract

Purpose

The aim of this study was to compare a broad range of total knee prostheses with different design parameters to determine whether in vivo kinematics was consistently related to design. The hypothesis was that there are no clear recognizable differences in in vivo kinematics between different design parameters or prostheses.

Methods

At two sites, data were collected by a single observer on 52 knees (49 subjects with rheumatoid arthritis or osteoarthritis). Six different total knee prostheses were used: multi-radius, single-radius, fixed-bearing, mobile-bearing, posterior-stabilized, cruciate retaining and cruciate sacrificing. Knee kinematics was recorded using fluoroscopy as the patients performed a step-up motion.

Results

There was a significant effect of prosthetic design on all outcome parameters; however, post hoc tests showed that the NexGen group was responsible for 80% of the significant values. The range of knee flexion was much smaller in this group, resulting in smaller anterior-posterior translations and rotations.

Conclusion

Despite kinematics being generally consistent with the kinematics intended by their design, there were no clear recognizable differences in in vivo kinematics between different design parameters or prostheses. Hence, the differences in design parameters or prostheses are not distinct enough to have an effect on clinical outcome of patients.

Level of evidence

Therapeutic study, Level III.
Literature
1.
go back to reference Andriacchi TP (1994) Dynamics of knee malalignment. Orthop Clin N Am 25:395–403 Andriacchi TP (1994) Dynamics of knee malalignment. Orthop Clin N Am 25:395–403
2.
go back to reference Andriacchi TP, Galante JO, Fermier RW (1982) The influence of total knee-replacement design on walking and stair-climbing. J Bone Joint Surg Am 64:1328–1335PubMed Andriacchi TP, Galante JO, Fermier RW (1982) The influence of total knee-replacement design on walking and stair-climbing. J Bone Joint Surg Am 64:1328–1335PubMed
3.
go back to reference Banks SA, Harman MK, Bellemans J, Hodge WA (2003) Making sense of knee arthroplasty kinematics: news you can use. J Bone Joint Surg Am 85-A(Suppl 4):64–72 Banks SA, Harman MK, Bellemans J, Hodge WA (2003) Making sense of knee arthroplasty kinematics: news you can use. J Bone Joint Surg Am 85-A(Suppl 4):64–72
4.
go back to reference Banks SA, Hodge WA (2004) 2003 Hap Paul Award Paper of the International Society for Technology in Arthroplasty. Design and activity dependence of kinematics in fixed and mobile-bearing knee arthroplasties. J Arthroplast 19:809–816CrossRef Banks SA, Hodge WA (2004) 2003 Hap Paul Award Paper of the International Society for Technology in Arthroplasty. Design and activity dependence of kinematics in fixed and mobile-bearing knee arthroplasties. J Arthroplast 19:809–816CrossRef
5.
go back to reference Banks SA, Hodge WA (2004) Implant design affects knee arthroplasty kinematics during stair-stepping. Clin Orthop Relat Res 426:187–193PubMedCrossRef Banks SA, Hodge WA (2004) Implant design affects knee arthroplasty kinematics during stair-stepping. Clin Orthop Relat Res 426:187–193PubMedCrossRef
6.
go back to reference Blunn GW, Joshi AB, Minns RJ, Lidgren L, Lilley P, Ryd L, Engelbrecht E, Walker PS (1997) Wear in retrieved condylar knee arthroplasties. A comparison of wear in different designs of 280 retrieved condylar knee prostheses. J Arthroplast 12:281–290CrossRef Blunn GW, Joshi AB, Minns RJ, Lidgren L, Lilley P, Ryd L, Engelbrecht E, Walker PS (1997) Wear in retrieved condylar knee arthroplasties. A comparison of wear in different designs of 280 retrieved condylar knee prostheses. J Arthroplast 12:281–290CrossRef
7.
go back to reference Bottlang M, Erne OK, Lacatusu E, Sommers MB, Kessler O (2006) A mobile-bearing knee prosthesis can reduce strain at the proximal tibia. Clin Orthop Relat Res 447:105–111PubMedCrossRef Bottlang M, Erne OK, Lacatusu E, Sommers MB, Kessler O (2006) A mobile-bearing knee prosthesis can reduce strain at the proximal tibia. Clin Orthop Relat Res 447:105–111PubMedCrossRef
8.
go back to reference Callaghan JJ (2001) Mobile-bearing knee replacement: clinical results: a review of the literature. Clin Orthop Relat Res 392:221–225PubMedCrossRef Callaghan JJ (2001) Mobile-bearing knee replacement: clinical results: a review of the literature. Clin Orthop Relat Res 392:221–225PubMedCrossRef
9.
go back to reference Delport HP, Banks SA, De Schepper J, Bellemans J (2006) A kinematic comparison of fixed- and mobile-bearing knee replacements. J Bone Joint Surg Br 88:1016–1021PubMedCrossRef Delport HP, Banks SA, De Schepper J, Bellemans J (2006) A kinematic comparison of fixed- and mobile-bearing knee replacements. J Bone Joint Surg Br 88:1016–1021PubMedCrossRef
10.
go back to reference Dennis DA, Komistek RD, Mahfouz MR, Outten JT, Sharma A (2005) Mobile-bearing total knee arthroplasty: do the polyethylene bearings rotate? Clin Orthop Relat Res 440:88–95PubMedCrossRef Dennis DA, Komistek RD, Mahfouz MR, Outten JT, Sharma A (2005) Mobile-bearing total knee arthroplasty: do the polyethylene bearings rotate? Clin Orthop Relat Res 440:88–95PubMedCrossRef
11.
go back to reference Ewald FC (1989) The Knee Society total knee arthroplasty roentgenographic evaluation and scoring system. Clin Orthop Relat Res 248:9–12PubMed Ewald FC (1989) The Knee Society total knee arthroplasty roentgenographic evaluation and scoring system. Clin Orthop Relat Res 248:9–12PubMed
12.
go back to reference Garling EH, Valstar ER, Nelissen RGHH (2005) Comparison of micromotion in mobile bearing and posterior stabilized total knee prostheses—A randomized RSA study of 40 knees followed for 2 years. Acta Orthop 76:353–361PubMed Garling EH, Valstar ER, Nelissen RGHH (2005) Comparison of micromotion in mobile bearing and posterior stabilized total knee prostheses—A randomized RSA study of 40 knees followed for 2 years. Acta Orthop 76:353–361PubMed
13.
go back to reference Garling EH, Kaptein BL, Valstar ER, Nelissen RGHH (2007) Limited rotation of the mobile-bearing in a rotating platform total knee prosthesis. J Biomech 40(Suppl 1):S25–S30 Garling EH, Kaptein BL, Valstar ER, Nelissen RGHH (2007) Limited rotation of the mobile-bearing in a rotating platform total knee prosthesis. J Biomech 40(Suppl 1):S25–S30
14.
go back to reference Hall J, Copp SN, Adelson WS, D’Lima DD, Colwell CW Jr (2008) Extensor mechanism function in single-radius vs multiradius femoral components for total knee arthroplasty. J Arthroplast 23:216–219CrossRef Hall J, Copp SN, Adelson WS, D’Lima DD, Colwell CW Jr (2008) Extensor mechanism function in single-radius vs multiradius femoral components for total knee arthroplasty. J Arthroplast 23:216–219CrossRef
15.
go back to reference Hilding MB, Lanshammar H, Ryd L (1996) Knee joint loading and tibial component loosening. RSA and gait analysis in 45 osteoarthritic patients before and after TKA. J Bone Joint Surg Br 78:66–73PubMed Hilding MB, Lanshammar H, Ryd L (1996) Knee joint loading and tibial component loosening. RSA and gait analysis in 45 osteoarthritic patients before and after TKA. J Bone Joint Surg Br 78:66–73PubMed
16.
go back to reference Insall JN, Lachiewicz PF, Burstein AH (1982) The posterior stabilized condylar prosthesis: a modification of the total condylar design. Two to four-year clinical experience. J Bone Joint Surg Am 64:1317–1323PubMed Insall JN, Lachiewicz PF, Burstein AH (1982) The posterior stabilized condylar prosthesis: a modification of the total condylar design. Two to four-year clinical experience. J Bone Joint Surg Am 64:1317–1323PubMed
17.
go back to reference Kaptein BL, Valstar ER, Stoel BC, Rozing PM, Reiber JH (2003) A new model-based RSA method validated using CAD models and models from reversed engineering. J Biomech 36:873–882PubMedCrossRef Kaptein BL, Valstar ER, Stoel BC, Rozing PM, Reiber JH (2003) A new model-based RSA method validated using CAD models and models from reversed engineering. J Biomech 36:873–882PubMedCrossRef
18.
go back to reference McEwen HM, Barnett PI, Bell CJ, Farrar R, Auger DD, Stone MH, Fisher J (2005) The influence of design, materials and kinematics on the in vitro wear of total knee replacements. J Biomech 38:357–365PubMedCrossRef McEwen HM, Barnett PI, Bell CJ, Farrar R, Auger DD, Stone MH, Fisher J (2005) The influence of design, materials and kinematics on the in vitro wear of total knee replacements. J Biomech 38:357–365PubMedCrossRef
19.
go back to reference McEwen HM, Fisher J, Goldsmith AA, Auger DD, Hardaker C, Stone MH (2001) Wear of fixed bearing and rotating platform mobile bearing knees subjected to high levels of internal and external tibial rotation. J Mater Sci Mater Med 12:1049–1052PubMedCrossRef McEwen HM, Fisher J, Goldsmith AA, Auger DD, Hardaker C, Stone MH (2001) Wear of fixed bearing and rotating platform mobile bearing knees subjected to high levels of internal and external tibial rotation. J Mater Sci Mater Med 12:1049–1052PubMedCrossRef
20.
go back to reference Pandit H, Ward T, Hollinghurst D, Beard DJ, Gill HS, Thomas NP, Murray DW (2005) Influence of surface geometry and the cam-post mechanism on the kinematics of total knee replacement. J Bone Joint Surg Br 87:940–945PubMedCrossRef Pandit H, Ward T, Hollinghurst D, Beard DJ, Gill HS, Thomas NP, Murray DW (2005) Influence of surface geometry and the cam-post mechanism on the kinematics of total knee replacement. J Bone Joint Surg Br 87:940–945PubMedCrossRef
21.
go back to reference Rousseau MA, Lazennec JY, Catonne Y (2008) Early mechanical failure in total knee arthroplasty. Int Orthop 32:53–56PubMedCrossRef Rousseau MA, Lazennec JY, Catonne Y (2008) Early mechanical failure in total knee arthroplasty. Int Orthop 32:53–56PubMedCrossRef
22.
go back to reference Saari T, Tranberg R, Zugner R, Uvehammer J, Karrholm J (2005) Changed gait pattern in patients with total knee arthroplasty but minimal influence of tibial insert design: gait analysis during level walking in 39 TKR patients and 18 healthy controls. Acta Orthop 76:253–260PubMedCrossRef Saari T, Tranberg R, Zugner R, Uvehammer J, Karrholm J (2005) Changed gait pattern in patients with total knee arthroplasty but minimal influence of tibial insert design: gait analysis during level walking in 39 TKR patients and 18 healthy controls. Acta Orthop 76:253–260PubMedCrossRef
23.
go back to reference Saari T, Uvehammer J, Carlsson LV, Regner L, Karrholm J (2006) Influence of polyethylene constraint on tibial component fixation in total knee arthroplasty: follow-up report after 5 years. J Arthroplast 21:1032–1037CrossRef Saari T, Uvehammer J, Carlsson LV, Regner L, Karrholm J (2006) Influence of polyethylene constraint on tibial component fixation in total knee arthroplasty: follow-up report after 5 years. J Arthroplast 21:1032–1037CrossRef
24.
go back to reference Snider MG, Macdonald SJ (2009) The influence of the posterior cruciate ligament and component design on joint line position after primary total knee arthroplasty. J Arthroplast 24:1093–1098CrossRef Snider MG, Macdonald SJ (2009) The influence of the posterior cruciate ligament and component design on joint line position after primary total knee arthroplasty. J Arthroplast 24:1093–1098CrossRef
25.
go back to reference Stiehl JB, Dennis DA, Komistek RD, Keblish PA (1997) In vivo kinematic analysis of a mobile bearing total knee prosthesis. Clin Orthop Relat Res 345:60–66PubMedCrossRef Stiehl JB, Dennis DA, Komistek RD, Keblish PA (1997) In vivo kinematic analysis of a mobile bearing total knee prosthesis. Clin Orthop Relat Res 345:60–66PubMedCrossRef
26.
go back to reference Uvehammer J, Karrholm J, Carlsson L (2007) Influence of joint area design on tibial component migration: comparison among a fixed symmetrical, asymmetrical, and moveable bearing. J Knee Surg 20:20–26PubMed Uvehammer J, Karrholm J, Carlsson L (2007) Influence of joint area design on tibial component migration: comparison among a fixed symmetrical, asymmetrical, and moveable bearing. J Knee Surg 20:20–26PubMed
27.
go back to reference Wang H, Simpson KJ, Ferrara MS, Chamnongkich S, Kinsey T, Mahoney OM (2006) Biomechanical differences exhibited during sit-to-stand between total knee arthroplasty designs of varying radii. J Arthroplast 21:1193–1199CrossRef Wang H, Simpson KJ, Ferrara MS, Chamnongkich S, Kinsey T, Mahoney OM (2006) Biomechanical differences exhibited during sit-to-stand between total knee arthroplasty designs of varying radii. J Arthroplast 21:1193–1199CrossRef
Metadata
Title
No differences in in vivo kinematics between six different types of knee prostheses
Authors
N. Wolterbeek
R. G. H. H. Nelissen
E. R. Valstar
Publication date
01-03-2012
Publisher
Springer-Verlag
Published in
Knee Surgery, Sports Traumatology, Arthroscopy / Issue 3/2012
Print ISSN: 0942-2056
Electronic ISSN: 1433-7347
DOI
https://doi.org/10.1007/s00167-011-1605-y

Other articles of this Issue 3/2012

Knee Surgery, Sports Traumatology, Arthroscopy 3/2012 Go to the issue