01-08-2012 | Physiological and Technical Notes

Bench testing of a new hyperbaric chamber ventilator at different atmospheric pressures

Authors: Jean-Claude Lefebvre, Aissam Lyazidi, Miguel Parceiro, Giuseppe F. Sferrazza Papa, Evangelia Akoumianaki, Deborah Pugin, Didier Tassaux, Laurent Brochard, Jean-Christophe M. Richard

Published in: Intensive Care Medicine | Issue 8/2012

Login to get access

Abstract

Purpose

Providing mechanical ventilation is challenging at supra-atmospheric pressure. The higher gas density increases resistance, reducing the flow delivered by the ventilator. A new hyperbaric ventilator (Siaretron IPER 1000) is said to compensate for these effects automatically. The aim of this bench test study was to validate the compensation, define its limits and provide details on the ventilator’s output at varied atmospheric pressures.

Methods

Experiments were conducted inside a multiplace hyperbaric chamber at 1, 2.2, 2.8 and 4 atmospheres absolute (ATA), with the ventilator connected to a test lung. Transducers were recalibrated at each ATA level. Various ventilator settings were tested in volume and pressure control modes. Measured tidal volumes were compared with theoretical predictions based on gas laws.

Results

Results confirmed the ventilator’s ability to provide compensation, but also identified its limits. The compensation range could be predicted and depended on the maximal flow attainable, decreasing linearly with increasing atmospheric pressure. With settings inside the range, tidal volumes approximated set values (mean error 10 ± 5 %). With settings outside the range, the volume was limited to the predicted maximal value calculated from maximal flow. A practical guide for clinicians is provided.

Conclusion

The IPER 1000 ventilator attempted to deliver stable tidal volume by adjusting the opening of the inspiratory valve in proportion to atmospheric pressure. Adequate compensation was observed, albeit only within a predictable range, which can be reliably predicted for each setting and ATA level combination. Setting a tidal volume outside this range can result in an unwanted decrease in minute ventilation.
Appendix
Available only for authorised users
Literature
2.
go back to reference Skinner M (1998) Ventilator function under hyperbaric conditions. Diving Hyperb Med J 28:62–71 Skinner M (1998) Ventilator function under hyperbaric conditions. Diving Hyperb Med J 28:62–71
3.
go back to reference Workman WT (1999) Hyperbaric facility safety: a practical guide. Best Pub. Co., Flagstaff Workman WT (1999) Hyperbaric facility safety: a practical guide. Best Pub. Co., Flagstaff
4.
go back to reference Stahl W, Radermacher P, Calzia E (2000) Functioning of ICU ventilators under hyperbaric conditions–comparison of volume- and pressure-controlled modes. Intensive Care Med 26:442–448PubMedCrossRef Stahl W, Radermacher P, Calzia E (2000) Functioning of ICU ventilators under hyperbaric conditions–comparison of volume- and pressure-controlled modes. Intensive Care Med 26:442–448PubMedCrossRef
5.
go back to reference Stanga D, Beck G, Chimiak J (2003) Evaluation of respiratory support devices for use in the hyperbaric chamber. NEDU TR 03-18 Panama City, FL, Department of Defense, Navy Experimental Diving Unit Stanga D, Beck G, Chimiak J (2003) Evaluation of respiratory support devices for use in the hyperbaric chamber. NEDU TR 03-18 Panama City, FL, Department of Defense, Navy Experimental Diving Unit
6.
go back to reference Blanch PB, Desautels DA, Gallagher TJ (1989) Mechanical ventilator function under hyperbaric conditions. Undersea Hyperb Med 16 (Supp):17 Blanch PB, Desautels DA, Gallagher TJ (1989) Mechanical ventilator function under hyperbaric conditions. Undersea Hyperb Med 16 (Supp):17
7.
go back to reference Lyazidi A, Thille AW, Carteaux G, Galia F, Brochard L, Richard JC (2010) Bench test evaluation of volume delivered by modern ICU ventilators during volume-controlled ventilation. Intensive Care Med 36:2074–2080PubMedCrossRef Lyazidi A, Thille AW, Carteaux G, Galia F, Brochard L, Richard JC (2010) Bench test evaluation of volume delivered by modern ICU ventilators during volume-controlled ventilation. Intensive Care Med 36:2074–2080PubMedCrossRef
8.
go back to reference Thille AW, Lyazidi A, Richard JC, Galia F, Brochard L (2009) A bench study of intensive-care-unit ventilators: new versus old and turbine-based versus compressed gas-based ventilators. Intensive Care Med 35:1368–1376PubMedCrossRef Thille AW, Lyazidi A, Richard JC, Galia F, Brochard L (2009) A bench study of intensive-care-unit ventilators: new versus old and turbine-based versus compressed gas-based ventilators. Intensive Care Med 35:1368–1376PubMedCrossRef
9.
go back to reference West JB (1999) The original presentation of Boyle’s law. J Appl Physiol 87:1543–1545PubMed West JB (1999) The original presentation of Boyle’s law. J Appl Physiol 87:1543–1545PubMed
10.
go back to reference Tassaux D, Jolliet P, Thouret JM, Roeseler J, Dorne R, Chevrolet JC (1999) Calibration of seven ICU ventilators for mechanical ventilation with helium-oxygen mixtures. Am J Respir Crit Care Med 160:22–32PubMed Tassaux D, Jolliet P, Thouret JM, Roeseler J, Dorne R, Chevrolet JC (1999) Calibration of seven ICU ventilators for mechanical ventilation with helium-oxygen mixtures. Am J Respir Crit Care Med 160:22–32PubMed
11.
go back to reference Weaver LK, Greenway L, Elliot CG (1988) Performance of the Sechrist 500A hyperbaric ventilator in a monoplace hyperbaric chamber. J Hyperb Med 3:215–225 Weaver LK, Greenway L, Elliot CG (1988) Performance of the Sechrist 500A hyperbaric ventilator in a monoplace hyperbaric chamber. J Hyperb Med 3:215–225
12.
go back to reference Churchill S, Weaver LK, Haberstock D (1999) Performance of the Omni-vent mechanical ventilator for use with the monoplace hyperbaric chamber. Undersea Hyperb Med 26(Suppl):70–71 Churchill S, Weaver LK, Haberstock D (1999) Performance of the Omni-vent mechanical ventilator for use with the monoplace hyperbaric chamber. Undersea Hyperb Med 26(Suppl):70–71
13.
go back to reference Moon RE, Bergquist LV, Conklin B, Miller JN (1986) Monaghan 225 ventilator use under hyperbaric conditions. Chest 89:846–851PubMedCrossRef Moon RE, Bergquist LV, Conklin B, Miller JN (1986) Monaghan 225 ventilator use under hyperbaric conditions. Chest 89:846–851PubMedCrossRef
14.
go back to reference Hennessey PJ (2007) Evaluation of the UNI-VENT Eagle Model 754 ventilator for use in the surface recompression chambers (SRC) of the submarine rescue diving and recompression system (SRDRS). NEDU TR 07-03 Panama City, FL, Department of Defense, Navy Experimental Diving Unit Hennessey PJ (2007) Evaluation of the UNI-VENT Eagle Model 754 ventilator for use in the surface recompression chambers (SRC) of the submarine rescue diving and recompression system (SRDRS). NEDU TR 07-03 Panama City, FL, Department of Defense, Navy Experimental Diving Unit
Metadata
Title
Bench testing of a new hyperbaric chamber ventilator at different atmospheric pressures
Authors
Jean-Claude Lefebvre
Aissam Lyazidi
Miguel Parceiro
Giuseppe F. Sferrazza Papa
Evangelia Akoumianaki
Deborah Pugin
Didier Tassaux
Laurent Brochard
Jean-Christophe M. Richard
Publication date
01-08-2012
Publisher
Springer-Verlag
Published in
Intensive Care Medicine / Issue 8/2012
Print ISSN: 0342-4642
Electronic ISSN: 1432-1238
DOI
https://doi.org/10.1007/s00134-012-2590-4

Other articles of this Issue 8/2012

Intensive Care Medicine 8/2012 Go to the issue