Extravascular lung water volume measurement by a novel lithium-thermal indicator dilution method: comparison of three techniques to post-mortem gravimetry

Authors: Benjamin Maddison, Riccardo Giudici, Enrico Calzia, Christopher Wolff, Charles Hinds, Peter Radermacher, Rupert M. Pearse

Published in: Intensive Care Medicine | Issue 11/2008

Login to get access

Abstract

Objective

To compare the lithium-thermal double indicator dilution (Li-thermal), indocyanine green-thermal double indicator dilution (ICG-thermal), single thermal indicator dilution (single-thermal) and gravimetric techniques of extravascular lung water volume (EVLW) measurement in porcine models of acute lung injury.

Design

Two animal models designed to invoke a systemic inflammatory response.

Setting

Laboratory study.

Subjects

A total of 12 immature Deutsches Landschwein pigs.

Interventions

Extravascular lung water volume was measured at four time points using Li-thermal, ICG-thermal and single-thermal techniques. Measurements were performed using existing technology according to manufacturer’s instructions. Post-mortem gravimetric EVLW measurements were performed by measuring wet and dry mass of lung tissue. Measurements were compared using the Bland–Altman method. Data are presented as mean (SD).

Measurements and main results

Data were collected in 12 animals and comparison between all 4 techniques was possible in 10 animals. EVLW measured by gravimetry was 9.2 (±3.0)ml kg−1. When compared to gravimetry, both Li-thermal and ICG-thermal techniques showed minimal bias but wide limits of agreement (LOA) [Li-thermal: bias −1.8 ml kg−1 (LOA ± 13.1); ICG-thermal bias −1.0 ml kg−1 (LOA ± 6.6)]. Comparison between the single-thermal and gravimetric methods identified both considerable bias and wide LOA [+8.5 ml kg−1 (LOA ± 14.5)].

Conclusion

Clinically significant differences between EVLW measurements obtained with the gravimetric method and three in vivo indicator dilution techniques were identified. While none of the techniques could be considered ideal, the ICG-thermal method appeared more reliable than either the Li-thermal or single thermal techniques. Further research is required to determine whether the accuracy of the prototype Li-thermal technique can be improved.
Appendix
Available only for authorised users
Literature
1.
go back to reference Sakka SG, Klein M, Reinhart K, Meier-Hellmann A (2002) Prognostic value of extravascular lung water in critically ill patients. Chest 122:2080–2086PubMedCrossRef Sakka SG, Klein M, Reinhart K, Meier-Hellmann A (2002) Prognostic value of extravascular lung water in critically ill patients. Chest 122:2080–2086PubMedCrossRef
2.
go back to reference Mitchell JP, Schuller D, Calandrino FS, Schuster DP (1992) Improved outcome based on fluid management in critically ill patients requiring pulmonary artery catheterization. Am Rev Respir Dis 145:990–998PubMed Mitchell JP, Schuller D, Calandrino FS, Schuster DP (1992) Improved outcome based on fluid management in critically ill patients requiring pulmonary artery catheterization. Am Rev Respir Dis 145:990–998PubMed
3.
go back to reference Laggner A, Kleinberger G, Haller J, Lenz K, Sommer G, Druml W (1984) Bedside estimation of extravascular lung water in critically ill patients: comparison of the chest radiograph and the thermal dye technique. Intensive Care Med 10:309–313PubMedCrossRef Laggner A, Kleinberger G, Haller J, Lenz K, Sommer G, Druml W (1984) Bedside estimation of extravascular lung water in critically ill patients: comparison of the chest radiograph and the thermal dye technique. Intensive Care Med 10:309–313PubMedCrossRef
4.
go back to reference Chinard FP (1975) Estimation of extravascular lung water by indicator-dilution techniques. Circ Res 37:137–145PubMed Chinard FP (1975) Estimation of extravascular lung water by indicator-dilution techniques. Circ Res 37:137–145PubMed
5.
go back to reference Stewart GN (1921) The pulmonary circulation time, the quantity of blood in the lungs and the output of the heart. Am J Physiol 58:20–44 Stewart GN (1921) The pulmonary circulation time, the quantity of blood in the lungs and the output of the heart. Am J Physiol 58:20–44
6.
go back to reference Slutsky RA, Higgins CB (1985) In vivo validation of the thermal-green dye technique for measuring extravascular lung water. Crit Care Med 13:432–435PubMedCrossRef Slutsky RA, Higgins CB (1985) In vivo validation of the thermal-green dye technique for measuring extravascular lung water. Crit Care Med 13:432–435PubMedCrossRef
7.
go back to reference Pfeiffer UJ, Lichtwarck-Aschoff M, Beale R (1994) Single thermodilution monitoring of global end-diastolic volume, intrathoracic blood volume and extravascular lung water (abstract). Clin Intensive Care 5(Suppl):38–39 Pfeiffer UJ, Lichtwarck-Aschoff M, Beale R (1994) Single thermodilution monitoring of global end-diastolic volume, intrathoracic blood volume and extravascular lung water (abstract). Clin Intensive Care 5(Suppl):38–39
8.
go back to reference Rossi P, Wanecek M, Rudehill A, Konrad D, Weitzberg E, Oldner A (2006) Comparison of a single indicator and gravimetric technique for estimation of extravascular lung water in endotoxemic pigs. Crit Care Med 34:1437–1443PubMedCrossRef Rossi P, Wanecek M, Rudehill A, Konrad D, Weitzberg E, Oldner A (2006) Comparison of a single indicator and gravimetric technique for estimation of extravascular lung water in endotoxemic pigs. Crit Care Med 34:1437–1443PubMedCrossRef
9.
go back to reference Band DM, Linton RA, O’Brien TK, Jonas MM, Linton NW (1997) The shape of indicator dilution curves used for cardiac output measurement in man. J Physiol 498:225–229PubMed Band DM, Linton RA, O’Brien TK, Jonas MM, Linton NW (1997) The shape of indicator dilution curves used for cardiac output measurement in man. J Physiol 498:225–229PubMed
10.
go back to reference Jonas M, Linton RAF, O’Brien TK, Band DM, Linton NWF, Kelly F, Burden TJ, Chevalier SFA, Thompson RPH, Birch NJ, Powell JJ (2001) The pharmacokinetics of intravenous lithium chloride in patients and normal volunteers. J Trace Microprobe Tech 19:313–320CrossRef Jonas M, Linton RAF, O’Brien TK, Band DM, Linton NWF, Kelly F, Burden TJ, Chevalier SFA, Thompson RPH, Birch NJ, Powell JJ (2001) The pharmacokinetics of intravenous lithium chloride in patients and normal volunteers. J Trace Microprobe Tech 19:313–320CrossRef
11.
go back to reference Renner LE, Morton MJ, Sakuma GY (1993) Indicator amount, temperature, and intrinsic cardiac output affect thermodilution cardiac output accuracy and reproducibility. Crit Care Med 21:586–597PubMedCrossRef Renner LE, Morton MJ, Sakuma GY (1993) Indicator amount, temperature, and intrinsic cardiac output affect thermodilution cardiac output accuracy and reproducibility. Crit Care Med 21:586–597PubMedCrossRef
12.
go back to reference Kurita T, Morita K, Kawasaki H, Fujii K, Kazama T, Sato S (2002) Lithium dilution cardiac output measurement in oleic acid-induced pulmonary edema. J Cardiothorac Vasc Anesth 16:334–337PubMedCrossRef Kurita T, Morita K, Kawasaki H, Fujii K, Kazama T, Sato S (2002) Lithium dilution cardiac output measurement in oleic acid-induced pulmonary edema. J Cardiothorac Vasc Anesth 16:334–337PubMedCrossRef
13.
go back to reference Pfeiffer UJ, Birk M, Aschenbrenner A, Blumel G (1982) The system for quantitating thermal-dye extravascular lung water. In: Prakash O (ed) Computers in critical care and pulmonary medicine. Plenum, New York, pp 123–125 Pfeiffer UJ, Birk M, Aschenbrenner A, Blumel G (1982) The system for quantitating thermal-dye extravascular lung water. In: Prakash O (ed) Computers in critical care and pulmonary medicine. Plenum, New York, pp 123–125
14.
go back to reference Pearse RM, Ikram K, Barry J (2004) Equipment review: an appraisal of the LiDCO plus method of measuring cardiac output. Crit Care 8:190–195PubMedCrossRef Pearse RM, Ikram K, Barry J (2004) Equipment review: an appraisal of the LiDCO plus method of measuring cardiac output. Crit Care 8:190–195PubMedCrossRef
15.
go back to reference von Spiegel T, Wietasch G, Bursch J, Hoeft A (1996) Cardiac output determination with transpulmonary thermodilution. An alternative to pulmonary catheterization? Anaesthesist 45:1045–1050CrossRef von Spiegel T, Wietasch G, Bursch J, Hoeft A (1996) Cardiac output determination with transpulmonary thermodilution. An alternative to pulmonary catheterization? Anaesthesist 45:1045–1050CrossRef
16.
go back to reference Hill SL, Elings VB, Lewis F (1981) Effect of cardiac output on extravascular lung water. Am Surg 47:522–528PubMed Hill SL, Elings VB, Lewis F (1981) Effect of cardiac output on extravascular lung water. Am Surg 47:522–528PubMed
17.
go back to reference Pohl T, Kozieras J, Sakka S (2008) Influence of extravascular lung water on transpulmonary thermodilution-derived cardiac output measurement. Intensive Care Med 34:533–537PubMedCrossRef Pohl T, Kozieras J, Sakka S (2008) Influence of extravascular lung water on transpulmonary thermodilution-derived cardiac output measurement. Intensive Care Med 34:533–537PubMedCrossRef
18.
go back to reference Fernandez-Mondejar E, Castano-Perez J, Rivera-Fernandez R, Colmenero-Ruiz M, Manzano F, Perez-Villares J, de la Chica R (2003) Quantification of lung water by transpulmonary thermodilution in normal and edematous lung. J Crit Care 18:253–258PubMedCrossRef Fernandez-Mondejar E, Castano-Perez J, Rivera-Fernandez R, Colmenero-Ruiz M, Manzano F, Perez-Villares J, de la Chica R (2003) Quantification of lung water by transpulmonary thermodilution in normal and edematous lung. J Crit Care 18:253–258PubMedCrossRef
19.
go back to reference Pearce ML, Yamashita J, Beazell J (1965) Measurement of pulmonary edema. Circ Res 16:482–488PubMed Pearce ML, Yamashita J, Beazell J (1965) Measurement of pulmonary edema. Circ Res 16:482–488PubMed
20.
go back to reference Rossi P, Oldner A, Wanecek M, Leksell LG, Rudehill A, Konrad D, Weitzberg E (2003) Comparison of gravimetric and a double-indicator dilution technique for assessment of extra-vascular lung water in endotoxaemia. Intensive Care Med 29:460–466PubMedCrossRef Rossi P, Oldner A, Wanecek M, Leksell LG, Rudehill A, Konrad D, Weitzberg E (2003) Comparison of gravimetric and a double-indicator dilution technique for assessment of extra-vascular lung water in endotoxaemia. Intensive Care Med 29:460–466PubMedCrossRef
21.
go back to reference Roch A, Michelet P, Lambert D, Delliaux S, Saby C, Perrin G, Ghez O, Bregeon F, Thomas P, Carpentier JP, Papazian L, Auffray JP (2004) Accuracy of the double indicator method for measurement of extravascular lung water depends on the type of acute lung injury. Crit Care Med 32:811–817PubMedCrossRef Roch A, Michelet P, Lambert D, Delliaux S, Saby C, Perrin G, Ghez O, Bregeon F, Thomas P, Carpentier JP, Papazian L, Auffray JP (2004) Accuracy of the double indicator method for measurement of extravascular lung water depends on the type of acute lung injury. Crit Care Med 32:811–817PubMedCrossRef
22.
go back to reference Carlile PV, Lowery DD, Gray BA (1986) Effect of PEEP and type of injury on thermal-dye estimation of pulmonary edema. J Appl Physiol 60:22–31PubMed Carlile PV, Lowery DD, Gray BA (1986) Effect of PEEP and type of injury on thermal-dye estimation of pulmonary edema. J Appl Physiol 60:22–31PubMed
23.
go back to reference Katzenelson R, Perel A, Berkenstadt H, Preisman S, Kogan S, Sternik L, Segal E (2004) Accuracy of transpulmonary thermodilution versus gravimetric measurement of extravascular lung water. Crit Care Med 32:1550–1554PubMedCrossRef Katzenelson R, Perel A, Berkenstadt H, Preisman S, Kogan S, Sternik L, Segal E (2004) Accuracy of transpulmonary thermodilution versus gravimetric measurement of extravascular lung water. Crit Care Med 32:1550–1554PubMedCrossRef
24.
go back to reference Kirov MY, Kuzkov VV, Kuklin VN, Waerhaug K, Bjertnaes LJ (2004) Extravascular lung water assessed by transpulmonary single thermodilution and postmortem gravimetry in sheep. Crit Care 8:R451–R458PubMedCrossRef Kirov MY, Kuzkov VV, Kuklin VN, Waerhaug K, Bjertnaes LJ (2004) Extravascular lung water assessed by transpulmonary single thermodilution and postmortem gravimetry in sheep. Crit Care 8:R451–R458PubMedCrossRef
25.
go back to reference Neumann P (1999) Extravascular lung water and intrathoracic blood volume: double versus single indicator dilution technique. Intensive Care Med 25:216–219PubMedCrossRef Neumann P (1999) Extravascular lung water and intrathoracic blood volume: double versus single indicator dilution technique. Intensive Care Med 25:216–219PubMedCrossRef
Metadata
Title
Extravascular lung water volume measurement by a novel lithium-thermal indicator dilution method: comparison of three techniques to post-mortem gravimetry
Authors
Benjamin Maddison
Riccardo Giudici
Enrico Calzia
Christopher Wolff
Charles Hinds
Peter Radermacher
Rupert M. Pearse
Publication date
01-11-2008
Publisher
Springer-Verlag
Published in
Intensive Care Medicine / Issue 11/2008
Print ISSN: 0342-4642
Electronic ISSN: 1432-1238
DOI
https://doi.org/10.1007/s00134-008-1207-4

Other articles of this Issue 11/2008

Intensive Care Medicine 11/2008 Go to the issue