01-07-2008 | Brief Report

The effect of different reference transducer positions on intra-abdominal pressure measurement: a multicenter analysis

Authors: Jan J. De Waele, Inneke De laet, Bart De Keulenaer, Sandy Widder, Andrew W. Kirkpatrick, Adrian B. Cresswell, Manu Malbrain, Zsolt Bodnar, Jorge H. Mejia-Mantilla, Richard Reis, Michael Parr, Robert Schulze, Sonia Compano, Michael Cheatham

Published in: Intensive Care Medicine | Issue 7/2008

Login to get access

Abstract

Objective

To investigate the effect of different reference transducer positions on intra-abdominal pressure (IAP) measurement. Three reference levels were studied: the symphysis pubis; the phlebostatic axis; and the midaxillary line at the level of the iliac crest.

Design

Prospective cohort study.

Setting

The intensive care units of participating hospitals

Patients and participants

One hundred thirty-two critically ill patients at risk for intra-abdominal hypertension (IAH).

Interventions

In each patient, three sets of IAP measurements were obtained in the supine position, using the different reference levels. The IAP measurements obtained at the different reference levels were compared using a paired t-test and Bland–Altman statistics were calculated.

Measurements and results

IAPphlebostatic (9.9 ± 4.67 mmHg) and IAPpubis (8.4 ± 4.60 mmHg) were significantly lower that IAPmidax (12.2 ± 4.66 mmHg; p < 0.0001 for both comparisons). The bias between the IAPmidax and IAPpubis was 3.8 mmHg (95% CI 3.5–4.1) and 2.3 mmHg (95% CI 1.9–2.6) between the IAPmidax and the IAPphlebostatic. The precision was 3.03 and 3.40, respectively.

Conclusions

In the supine position, IAPmidax is higher than both IAPphlebostatic and IAPpubis, differences found to be clinically significant; therefore, the symphysis pubis or phlebostatic axis reference lines are not interchangeable with the midaxillary level.
Literature
1.
go back to reference Malbrain ML, Chiumello D, Pelosi P, Bihari D, Innes R, Ranieri VM, del Turco M, Wilmer A, Brienza N, Malcangi V, Cohen J, Japiassu A, De Keulenaer BL, Daelemans R, Jacquet L, Laterre PF, Frank G, de Souza P, Cesana B, Gattinoni L (2005) Incidence and prognosis of intraabdominal hypertension in a mixed population of critically ill patients: a multiple-center epidemiological study. Crit Care Med 33:315–322PubMedCrossRef Malbrain ML, Chiumello D, Pelosi P, Bihari D, Innes R, Ranieri VM, del Turco M, Wilmer A, Brienza N, Malcangi V, Cohen J, Japiassu A, De Keulenaer BL, Daelemans R, Jacquet L, Laterre PF, Frank G, de Souza P, Cesana B, Gattinoni L (2005) Incidence and prognosis of intraabdominal hypertension in a mixed population of critically ill patients: a multiple-center epidemiological study. Crit Care Med 33:315–322PubMedCrossRef
2.
go back to reference Kirkpatrick AW, Brenneman FD, McLean RF, Rapanos T, Boulanger BR (2000) Is clinical examination an accurate indicator of raised intra-abdominal pressure in critically injured patients? Can J Surg 43:207–211PubMed Kirkpatrick AW, Brenneman FD, McLean RF, Rapanos T, Boulanger BR (2000) Is clinical examination an accurate indicator of raised intra-abdominal pressure in critically injured patients? Can J Surg 43:207–211PubMed
3.
go back to reference Sugrue M, Bauman A, Jones F, Bishop G, Flabouris A, Parr M, Stewart A, Hillman K, Deane SA (2002) Clinical examination is an inaccurate predictor of intraabdominal pressure. World J Surg 26:1428–1431PubMedCrossRef Sugrue M, Bauman A, Jones F, Bishop G, Flabouris A, Parr M, Stewart A, Hillman K, Deane SA (2002) Clinical examination is an inaccurate predictor of intraabdominal pressure. World J Surg 26:1428–1431PubMedCrossRef
4.
go back to reference De Waele JJ, De laet I, Malbrain ML (2007) Rational intraabdominal pressure monitoring: How to do it? Acta Clin Belg Suppl 16–25 De Waele JJ, De laet I, Malbrain ML (2007) Rational intraabdominal pressure monitoring: How to do it? Acta Clin Belg Suppl 16–25
5.
go back to reference Kirkpatrick AW, Balogh Z, Ball CG, Ahmed N, Chun R, McBeth P, Kirby A, Zygun DA (2006) The secondary abdominal compartment syndrome: Iatrogenic or unavoidable? J Am Coll Surg 202:668–679PubMedCrossRef Kirkpatrick AW, Balogh Z, Ball CG, Ahmed N, Chun R, McBeth P, Kirby A, Zygun DA (2006) The secondary abdominal compartment syndrome: Iatrogenic or unavoidable? J Am Coll Surg 202:668–679PubMedCrossRef
6.
go back to reference Kirkpatrick AW, De Waele JJ, Ball CG, Ranson K, Widder S, Laupland KB (2007) The secondary and recurrent abdominal compartment syndrome. Acta Clin Belg 1(Suppl):60–65 Kirkpatrick AW, De Waele JJ, Ball CG, Ranson K, Widder S, Laupland KB (2007) The secondary and recurrent abdominal compartment syndrome. Acta Clin Belg 1(Suppl):60–65
7.
go back to reference Malbrain ML, Cheatham ML, Kirkpatrick A, Sugrue M, Parr M, De Waele J, Balogh Z, Leppaniemi A, Olvera C, Ivatury R, D'Amours S, Wendon J, Hillman K, Johansson K, Kolkman K, Wilmer A (2006) Results from the International Conference of Experts on Intra-abdominal Hypertension and Abdominal Compartment Syndrome. I. Definitions. Intensive Care Med 32:1722–1732PubMedCrossRef Malbrain ML, Cheatham ML, Kirkpatrick A, Sugrue M, Parr M, De Waele J, Balogh Z, Leppaniemi A, Olvera C, Ivatury R, D'Amours S, Wendon J, Hillman K, Johansson K, Kolkman K, Wilmer A (2006) Results from the International Conference of Experts on Intra-abdominal Hypertension and Abdominal Compartment Syndrome. I. Definitions. Intensive Care Med 32:1722–1732PubMedCrossRef
8.
go back to reference Bland JM, Altman DG (1986) Statistical methods for assessing agreement between two methods of clinical measurement. Lancet 1:307–310PubMed Bland JM, Altman DG (1986) Statistical methods for assessing agreement between two methods of clinical measurement. Lancet 1:307–310PubMed
9.
go back to reference Malbrain ML, De laet I, Cheatham M (2007) Consensus conference definitions and recommendations on intra-abdominal hypertension (IAH) and the abdominal compartment syndrome (ACS): the long road to the final publications, how did we get there? Acta Clin Belg 1(Suppl):44–59 Malbrain ML, De laet I, Cheatham M (2007) Consensus conference definitions and recommendations on intra-abdominal hypertension (IAH) and the abdominal compartment syndrome (ACS): the long road to the final publications, how did we get there? Acta Clin Belg 1(Suppl):44–59
10.
go back to reference Malbrain ML (2004) Different techniques to measure intra-abdominal pressure (IAP): time for a critical re-appraisal. Intensive Care Med 30:357–371PubMedCrossRef Malbrain ML (2004) Different techniques to measure intra-abdominal pressure (IAP): time for a critical re-appraisal. Intensive Care Med 30:357–371PubMedCrossRef
11.
go back to reference De Waele J, Pletinckx P, Blot S, Hoste E (2006) Saline volume in transvesical intra-abdominal pressure measurement: enough is enough. Intensive Care Med 32:455–459PubMedCrossRef De Waele J, Pletinckx P, Blot S, Hoste E (2006) Saline volume in transvesical intra-abdominal pressure measurement: enough is enough. Intensive Care Med 32:455–459PubMedCrossRef
12.
go back to reference McBeth PB, Zygun DA, Widder S, Cheatham M, Zengerink I, Glowa J, Kirkpatrick AW (2007) Effect of patient positioning on intra-abdominal pressure monitoring. Am J Surg 193:644–647PubMedCrossRef McBeth PB, Zygun DA, Widder S, Cheatham M, Zengerink I, Glowa J, Kirkpatrick AW (2007) Effect of patient positioning on intra-abdominal pressure monitoring. Am J Surg 193:644–647PubMedCrossRef
13.
go back to reference Darovic GO, Vanriper S, Vanriper J (1995) Fluid-filled monitoring systems. In: Darovic GO (ed) Hemodynamic monitoring. Saunders, Philadelphia, pp 149–175 Darovic GO, Vanriper S, Vanriper J (1995) Fluid-filled monitoring systems. In: Darovic GO (ed) Hemodynamic monitoring. Saunders, Philadelphia, pp 149–175
Metadata
Title
The effect of different reference transducer positions on intra-abdominal pressure measurement: a multicenter analysis
Authors
Jan J. De Waele
Inneke De laet
Bart De Keulenaer
Sandy Widder
Andrew W. Kirkpatrick
Adrian B. Cresswell
Manu Malbrain
Zsolt Bodnar
Jorge H. Mejia-Mantilla
Richard Reis
Michael Parr
Robert Schulze
Sonia Compano
Michael Cheatham
Publication date
01-07-2008
Publisher
Springer-Verlag
Published in
Intensive Care Medicine / Issue 7/2008
Print ISSN: 0342-4642
Electronic ISSN: 1432-1238
DOI
https://doi.org/10.1007/s00134-008-1098-4

Other articles of this Issue 7/2008

Intensive Care Medicine 7/2008 Go to the issue