Abstract

Objective

This study compared the recently introduced Microcuff endotracheal tube HVLP ICU featuring an ultrathin (7-µm) polyurethane cuff membrane with endotracheal tubes from different manufacturers regarding fluid leakage past the tube cuff.

Design

In vitro setup.

Measurements and results

The following endotracheal tubes (ID 7.5 mm) were compared: Mallinckrodt HiLo, Microcuff HVLP ICU, Portex Profile Soft Seal, Rüsch Super Safety Clear, and Sheridan CF. A vertical PVC trachea model (ID 20 mm) was intubated, and cuffs were inflated to 10, 15, 20, 25, 30, and 60 cmH2O. Colored water (5 ml) was added to the top of the cuff. The amount of leaked fluid past the tube cuff within 5, 10, and 60 min was recorded. Experiments were performed four times using two examples of each tube brand. Fluid leakage past tube cuffs occurred in all conventional endotracheal tubes at cuff pressures from 10 to 60 cmH2O. In the Microcuff tube cuff pressure fluid leakage was observed within 10 min only at 10 cmH2O. Results with the Microcuff tube were significantly better than all other tube brands at cuff pressures of 10–30 cmH2O.

Conclusions

Within the acceptable upper limit for tracheal cuff pressure (25–30 cmH2O) the Microcuff endotracheal tube was the only one of the tested tubes to prevent fluid leakage in our in vitro setup. In vivo studies are required to confirm these findings.
Literature
1.
go back to reference Latto P (1997) The cuff. In: Latto IP, Vaughan RS (ed) Difficulties in tracheal intubation. Saunders, Philadelphia, pp 51–78 Latto P (1997) The cuff. In: Latto IP, Vaughan RS (ed) Difficulties in tracheal intubation. Saunders, Philadelphia, pp 51–78
2.
go back to reference Seegobin RD, van Hasselt GL (1984) Endotracheal cuff pressure and tracheal mucosal blood flow: endoscopic study of effects of four large volume cuffs. BMJ 288:965–968PubMed Seegobin RD, van Hasselt GL (1984) Endotracheal cuff pressure and tracheal mucosal blood flow: endoscopic study of effects of four large volume cuffs. BMJ 288:965–968PubMed
3.
go back to reference Asai T, Shingu K (2001) Leakage of fluid around high-volume, low-pressure cuffs. Anaesthesia 56:38–42CrossRefPubMed Asai T, Shingu K (2001) Leakage of fluid around high-volume, low-pressure cuffs. Anaesthesia 56:38–42CrossRefPubMed
4.
go back to reference Young PJ, Blunt MC (1999) Improving the shape and compliance characteristics of a high-volume, low-pressure cuff improves tracheal seal. Br J Anaesth 83:887–889PubMed Young PJ, Blunt MC (1999) Improving the shape and compliance characteristics of a high-volume, low-pressure cuff improves tracheal seal. Br J Anaesth 83:887–889PubMed
5.
go back to reference Young PJ, Basson C, Hamilton D, Ridley SA (1999) Prevention of tracheal aspiration using the pressure-limited tracheal tube cuff. Anaesthesia 54:559–563CrossRefPubMed Young PJ, Basson C, Hamilton D, Ridley SA (1999) Prevention of tracheal aspiration using the pressure-limited tracheal tube cuff. Anaesthesia 54:559–563CrossRefPubMed
6.
go back to reference Young PJ, Ridley SA, Downward G (1998) Evaluation of a new design of tracheal tube cuff to prevent leakage of fluid to the lungs. Br J Anaesth 80:796–799PubMed Young PJ, Ridley SA, Downward G (1998) Evaluation of a new design of tracheal tube cuff to prevent leakage of fluid to the lungs. Br J Anaesth 80:796–799PubMed
7.
go back to reference Young PJ, Rollinson M, Downward G, Henderson S (1997) Leakage of fluid past the tracheal tube cuff in a benchtop model. Br J Anaesth 78:557–562PubMed Young PJ, Rollinson M, Downward G, Henderson S (1997) Leakage of fluid past the tracheal tube cuff in a benchtop model. Br J Anaesth 78:557–562PubMed
8.
go back to reference Oikkonen M, Aromaa U (1997) Leakage of fluid around low-pressure tracheal tube cuffs. Anaesthesia 52:567–569PubMed Oikkonen M, Aromaa U (1997) Leakage of fluid around low-pressure tracheal tube cuffs. Anaesthesia 52:567–569PubMed
9.
go back to reference Reali-Forster C, Kolobow T, Giacomini M, Hayashi T, Horiba K, Ferrans V (1996) New ultrathin-walled endotracheal tube with a novel laryngeal seal design: long-term evaluation in sheep. Anesthesiology 84:162–172PubMed Reali-Forster C, Kolobow T, Giacomini M, Hayashi T, Horiba K, Ferrans V (1996) New ultrathin-walled endotracheal tube with a novel laryngeal seal design: long-term evaluation in sheep. Anesthesiology 84:162–172PubMed
10.
go back to reference Blunt MC, Young PJ, Patil A, Haddock A (2001) Gel lubrication of the tracheal tube cuff reduces pulmonary aspiration. Anesthesiology 95:377–381PubMed Blunt MC, Young PJ, Patil A, Haddock A (2001) Gel lubrication of the tracheal tube cuff reduces pulmonary aspiration. Anesthesiology 95:377–381PubMed
11.
go back to reference Metha S, Myat HM (1984) The cross sectional shape and circumference of the human trachea. Ann R Coll Surg Engl 66:356–358PubMed Metha S, Myat HM (1984) The cross sectional shape and circumference of the human trachea. Ann R Coll Surg Engl 66:356–358PubMed
12.
go back to reference Inglis TJJ, Sproat LJ, Sherrat MJ, Hawkey PM, Gibson JS, Shah MV (1992) Gastroduodenal dysfunction as a cause of gastric bacterial overgrowth in patients undergoing mechanical ventilation of the lungs. Br J Anaesth 68:499–502PubMed Inglis TJJ, Sproat LJ, Sherrat MJ, Hawkey PM, Gibson JS, Shah MV (1992) Gastroduodenal dysfunction as a cause of gastric bacterial overgrowth in patients undergoing mechanical ventilation of the lungs. Br J Anaesth 68:499–502PubMed
13.
go back to reference Ho AM, Aun CS, Karmakar MK (2002) The margin of safety associated with the use of cuffed paediatric tracheal tubes. Anaesthesia 57:173–175CrossRefPubMed Ho AM, Aun CS, Karmakar MK (2002) The margin of safety associated with the use of cuffed paediatric tracheal tubes. Anaesthesia 57:173–175CrossRefPubMed
15.
go back to reference Satishkumar S, Young P (2002) Tracheal cuff pressure–a survey of clinical practice. Br J Anaesth 88:456PubMed Satishkumar S, Young P (2002) Tracheal cuff pressure–a survey of clinical practice. Br J Anaesth 88:456PubMed
Metadata
Title
Fluid leakage past tracheal tube cuffs: evaluation of the new Microcuff endotracheal tube
Authors
Alexander Dullenkopf
Andreas Gerber
Markus Weiss
Publication date
01-10-2003
Publisher
Springer-Verlag
Published in
Intensive Care Medicine / Issue 10/2003
Print ISSN: 0342-4642
Electronic ISSN: 1432-1238
DOI
https://doi.org/10.1007/s00134-003-1933-6

Other articles of this Issue 10/2003

Intensive Care Medicine 10/2003 Go to the issue