Published in: Canadian Journal of Anesthesia/Journal canadien d'anesthésie 7/2012

01-07-2012 | Reports of Original Investigations

Less leakage and dislodgement with a catheter-over-needle versus a catheter-through-needle approach for peripheral nerve block: an ex vivo study

Authors: Ban C. H. Tsui, MD, Jenkin Tsui

Published in: Canadian Journal of Anesthesia/Journal canadien d'anesthésie | Issue 7/2012

Login to get access

Abstract

Purpose

The objective of this study was to compare the catheter-through-needle (CTN) and catheter-over-needle (CON) catheterization techniques ex vivo by measuring leak pressure around the catheter and the catheter’s resistance to pulling force.

Methods

Using an ex vivo porcine limb model, we compared the conventional CTN design with the CON design with respect to the ability to resist leakage at the catheter insertion site under high injection pressure and the force required to withdraw the catheter from tissue. One CON assembly (MultiSet, Pajunk) and three CTN assemblies (Contiplex, B.Braun; StimuCath, Arrow; Stimulong Sono, Pajunk) were studied. Ten porcine hind limbs were used to test leakage and another ten were used to measure withdrawal force. Catheters were placed at angles of 15° and 30° at depths of 3 cm and 5 cm. Leakage was assessed visually at the insertion site, and pressure was measured at the moment leakage occurred. Withdrawal force was measured by pulling the catheter from the tissue.

Results

No evidence of leakage was detected at the CON catheter insertion site at the highest pressure applied (1,000 mmHg) (n = 40). The CON assembly withstood significantly higher injection pressure than the CTN catheters without causing leaks at the catheter insertion site [CON, mean (standard deviation) > 1,000 (0) mmHg; B.Braun, 596 (92) mmHg; Pajunk Stimulong, 615 (107) mmHg; and Arrow, 422 (104) mmHg; P < 0.001 CON vs CTN]. The force required to withdraw the catheter from the porcine limb was greater with CON catheters [3.8 (0.8) N] than with any of the CTN catheters [range, 0.4 (0.2) - 0.8 (0.2) N], depending on depth, angle, and manufacturer (P < 0.001 CON vs CTN).

Conclusion

In the porcine leg model, CON catheterization provides greater resistance to leakage under high injection pressure and greater holding force in tissue than traditional CTN catheters.
Literature
1.
go back to reference Ilfeld BM. Continuous peripheral nerve blocks: a review of the published evidence. Anesth Analg 2011; 113: 904-25.PubMedCrossRef Ilfeld BM. Continuous peripheral nerve blocks: a review of the published evidence. Anesth Analg 2011; 113: 904-25.PubMedCrossRef
2.
go back to reference Capdevila X, Ponrouch M, Choquet O. Continuous peripheral nerve blocks in clinical practice. Curr Opin Anaesthesiol 2008; 21: 619-23.PubMedCrossRef Capdevila X, Ponrouch M, Choquet O. Continuous peripheral nerve blocks in clinical practice. Curr Opin Anaesthesiol 2008; 21: 619-23.PubMedCrossRef
3.
go back to reference Chelly JE, Ghisi D, Fanelli A. Continuous peripheral nerve blocks in acute pain management. Br J Anaesth 2010; 105: i86-96.PubMedCrossRef Chelly JE, Ghisi D, Fanelli A. Continuous peripheral nerve blocks in acute pain management. Br J Anaesth 2010; 105: i86-96.PubMedCrossRef
4.
go back to reference Ganapathy S, Wasserman RA, Watson JT, et al. Modified continuous femoral three-in-one block for postoperative pain after total knee arthroplasty. Anesth Analg 1999; 89: 1197-202.PubMedCrossRef Ganapathy S, Wasserman RA, Watson JT, et al. Modified continuous femoral three-in-one block for postoperative pain after total knee arthroplasty. Anesth Analg 1999; 89: 1197-202.PubMedCrossRef
5.
go back to reference Pham-Dang C, Kick O, Collet T, Gouin F, Pinaud M. Continuous peripheral nerve blocks with stimulating catheters. Reg Anesth Pain Med 2003; 28: 83-8.PubMedCrossRef Pham-Dang C, Kick O, Collet T, Gouin F, Pinaud M. Continuous peripheral nerve blocks with stimulating catheters. Reg Anesth Pain Med 2003; 28: 83-8.PubMedCrossRef
6.
go back to reference Boezaart AP, de Beer JF, du Toit C, van Rooyen K. A new technique of continuous interscalene nerve block. Can J Anesth 1999; 46: 275-81.PubMedCrossRef Boezaart AP, de Beer JF, du Toit C, van Rooyen K. A new technique of continuous interscalene nerve block. Can J Anesth 1999; 46: 275-81.PubMedCrossRef
7.
go back to reference Kick O, Blanche E, Pham-Dang C, Pinaud M, Estebe JP. A new stimulating stylet for immediate control of catheter tip position in continuous peripheral nerve blocks. Anesth Analg 1999; 89: 533-4.PubMed Kick O, Blanche E, Pham-Dang C, Pinaud M, Estebe JP. A new stimulating stylet for immediate control of catheter tip position in continuous peripheral nerve blocks. Anesth Analg 1999; 89: 533-4.PubMed
8.
go back to reference Antonakakis JG, Ting PH, Sites B. Ultrasound-guided regional anesthesia for peripheral nerve blocks: an evidence-based outcome review. Anesthesiol Clin 2011; 29: 179-91.PubMedCrossRef Antonakakis JG, Ting PH, Sites B. Ultrasound-guided regional anesthesia for peripheral nerve blocks: an evidence-based outcome review. Anesthesiol Clin 2011; 29: 179-91.PubMedCrossRef
9.
go back to reference Liu SS, Ngeow JE, Yadeau JT. Ultrasound-guided regional anesthesia and analgesia: a qualitative systematic review. Reg Anesth Pain Med 2009; 34: 47-59.PubMedCrossRef Liu SS, Ngeow JE, Yadeau JT. Ultrasound-guided regional anesthesia and analgesia: a qualitative systematic review. Reg Anesth Pain Med 2009; 34: 47-59.PubMedCrossRef
10.
go back to reference Ilfeld BM, Morey TE, Enneking FK. Continuous infraclavicular brachial plexus block for postoperative pain control at home: a randomized, double-blinded, placebo-controlled study. Anesthesiology 2002; 96: 1297-304.PubMedCrossRef Ilfeld BM, Morey TE, Enneking FK. Continuous infraclavicular brachial plexus block for postoperative pain control at home: a randomized, double-blinded, placebo-controlled study. Anesthesiology 2002; 96: 1297-304.PubMedCrossRef
11.
go back to reference Gurnaney H, Kraemer FW, Ganesh A. Dermabond decreases pericatheter local anesthetic leakage after continuous perineural infusions. Anesth Analg 2011; 113: 206.PubMedCrossRef Gurnaney H, Kraemer FW, Ganesh A. Dermabond decreases pericatheter local anesthetic leakage after continuous perineural infusions. Anesth Analg 2011; 113: 206.PubMedCrossRef
12.
go back to reference Klein SM, Nielsen KC, Buckenmaier CC III, Kamal AS, Rubin Y, Steele SM. 2-Octyl cyanoacrylate glue for the fixation of continuous peripheral nerve catheters. Anesthesiology 2003; 98: 590-1.PubMedCrossRef Klein SM, Nielsen KC, Buckenmaier CC III, Kamal AS, Rubin Y, Steele SM. 2-Octyl cyanoacrylate glue for the fixation of continuous peripheral nerve catheters. Anesthesiology 2003; 98: 590-1.PubMedCrossRef
13.
go back to reference Claudio R, Hadzic A, Shih H, et al. Injection pressures by anesthesiologists during simulated peripheral nerve block. Reg Anesth Pain Med 2004; 29: 201-5.PubMed Claudio R, Hadzic A, Shih H, et al. Injection pressures by anesthesiologists during simulated peripheral nerve block. Reg Anesth Pain Med 2004; 29: 201-5.PubMed
14.
go back to reference Holst D, Mollmann M, Scheuch E, Meissner K, Wendt M. Intrathecal local anesthetic distribution with the new spinocath catheter. Reg Anesth Pain Med 1998; 23: 463-8.PubMed Holst D, Mollmann M, Scheuch E, Meissner K, Wendt M. Intrathecal local anesthetic distribution with the new spinocath catheter. Reg Anesth Pain Med 1998; 23: 463-8.PubMed
15.
go back to reference Borgeat A, Blumenthal S, Lambert M, Theodorou P, Vienne P. The feasibility and complications of the continuous popliteal nerve block: a 1001-case survey. Anesth Analg 2006; 103: 229-33.PubMedCrossRef Borgeat A, Blumenthal S, Lambert M, Theodorou P, Vienne P. The feasibility and complications of the continuous popliteal nerve block: a 1001-case survey. Anesth Analg 2006; 103: 229-33.PubMedCrossRef
16.
go back to reference Gaertner E, Lascurain P, Venet C, et al. Continuous parasacral sciatic block: a radiographic study. Anesth Analg 2004; 98: 831-4.PubMedCrossRef Gaertner E, Lascurain P, Venet C, et al. Continuous parasacral sciatic block: a radiographic study. Anesth Analg 2004; 98: 831-4.PubMedCrossRef
17.
go back to reference Ilfeld BM, Sandhu NS, Loland VJ, et al. Ultrasound-guided (needle-in-plane) perineural catheter insertion: the effect of catheter-insertion distance on postoperative analgesia. Reg Anesth Pain Med 2011; 36: 261-5.PubMedCrossRef Ilfeld BM, Sandhu NS, Loland VJ, et al. Ultrasound-guided (needle-in-plane) perineural catheter insertion: the effect of catheter-insertion distance on postoperative analgesia. Reg Anesth Pain Med 2011; 36: 261-5.PubMedCrossRef
18.
go back to reference Chawathe MS, Jones RM, Gildersleve CD, Harrison SK, Morris SJ, Eickmann C. Detection of epidural catheters with ultrasound in children. Pediatr Anesth 2003; 13: 681-4.CrossRef Chawathe MS, Jones RM, Gildersleve CD, Harrison SK, Morris SJ, Eickmann C. Detection of epidural catheters with ultrasound in children. Pediatr Anesth 2003; 13: 681-4.CrossRef
19.
go back to reference Swenson JD, Davis JJ, DeCou JA. A novel approach for assessing catheter position after ultrasound-guided placement of continuous interscalene block. Anesth Analg 2008; 106: 1015-6.PubMedCrossRef Swenson JD, Davis JJ, DeCou JA. A novel approach for assessing catheter position after ultrasound-guided placement of continuous interscalene block. Anesth Analg 2008; 106: 1015-6.PubMedCrossRef
Metadata
Title
Less leakage and dislodgement with a catheter-over-needle versus a catheter-through-needle approach for peripheral nerve block: an ex vivo study
Authors
Ban C. H. Tsui, MD
Jenkin Tsui
Publication date
01-07-2012
Publisher
Springer-Verlag
Published in
Canadian Journal of Anesthesia/Journal canadien d'anesthésie / Issue 7/2012
Print ISSN: 0832-610X
Electronic ISSN: 1496-8975
DOI
https://doi.org/10.1007/s12630-012-9713-9

Other articles of this Issue 7/2012

Canadian Journal of Anesthesia/Journal canadien d'anesthésie 7/2012 Go to the issue