Skip to main content
Top
Published in: BMC Emergency Medicine 1/2020

Open Access 01-12-2020 | Laryngoscopy | Research article

Lower maximum forces on oral structures when using gum-elastic bougie than when using endotracheal tube and stylet during both direct and indirect laryngoscopy by novices: a crossover study using a high-fidelity simulator

Authors: Yuko Ono, Kazuaki Shinohara, Jiro Shimada, Shigeaki Inoue, Joji Kotani

Published in: BMC Emergency Medicine | Issue 1/2020

Login to get access

Abstract

Background

Applying excessive force during endotracheal intubation (ETI) is associated with several complications, including dental trauma and hemodynamic alterations. A gum-elastic bougie (GEB), a type of tracheal tube introducer, is a useful airway adjunct for patients with poor laryngoscopic views. However, how the use of a GEB affects the force applied during laryngoscopy is unclear. We compared the force applied on the oral structures during ETI performed by novices using the GEB versus an endotracheal tube + stylet.

Methods

This prospective crossover study was conducted from April 2017 to March 2019 in a public medical university in Japan. In total, 209 medical students (4th and 5th grade, mean age of 23.7 ± 2.0 years) without clinical ETI experience were recruited. The participants used either a Macintosh direct laryngoscope (DL) or C-MAC video laryngoscope (VL) in combination with a GEB or stylet to perform ETI on a high-fidelity airway management simulator. The order of the first ETI method was randomized to minimize the learning curve effect. The outcomes of interest were the maximum forces applied on the maxillary incisors and tongue during laryngoscopy. The implanted sensors in the simulator quantified these forces automatically.

Results

The maximum force applied on the maxillary incisors was significantly lower when using a GEB than when using an endotracheal tube + stylet both with the Macintosh DL (39.0 ± 23.3 vs. 47.4 ± 32.6 N, P < 0.001) and C-MAC VL (38.9 ± 18.6 vs. 42.0 ± 22.1 N, P < 0.001). Similarly, the force applied on the tongue was significantly lower when using a GEB than when using an endotracheal tube + stylet both with the Macintosh DL (31.9 ± 20.8 vs. 37.8 ± 22.2 N, P < 0.001) and C-MAC VL (35.2 ± 17.5 vs. 38.4 ± 17.5 N, P < 0.001).

Conclusions

Compared with the use of an endotracheal tube + stylet, the use of a GEB was associated with lower maximum forces on the oral structures during both direct and indirect laryngoscopy performed by novices. Our results suggest the expanded role of a GEB beyond an airway adjunct for difficult airways.
Literature
1.
go back to reference Brown CA 3rd, Bair AE, Pallin DJ, Walls RM. NEAR III Investigators. Techniques, success, and adverse events of emergency department adult intubations. Ann Emerg Med. 2015;65:363–70.PubMedCrossRef Brown CA 3rd, Bair AE, Pallin DJ, Walls RM. NEAR III Investigators. Techniques, success, and adverse events of emergency department adult intubations. Ann Emerg Med. 2015;65:363–70.PubMedCrossRef
2.
go back to reference Hasegawa K, Shigemitsu K, Hagiwara Y, Chiba T, Watase H, Brown CA 3rd, et al. Association between repeated intubation attempts and adverse events in emergency departments: an analysis of a multicenter prospective observational study. Ann Emerg Med. 2012;60:749–54.PubMedCrossRef Hasegawa K, Shigemitsu K, Hagiwara Y, Chiba T, Watase H, Brown CA 3rd, et al. Association between repeated intubation attempts and adverse events in emergency departments: an analysis of a multicenter prospective observational study. Ann Emerg Med. 2012;60:749–54.PubMedCrossRef
3.
go back to reference Mort TC. Emergency tracheal intubation: complications associated with repeated laryngoscopic attempts. Anesth Analg. 2004;99:607–13.PubMedCrossRef Mort TC. Emergency tracheal intubation: complications associated with repeated laryngoscopic attempts. Anesth Analg. 2004;99:607–13.PubMedCrossRef
4.
go back to reference Martin LD, Mhyre JM, Shanks AM, Tremper KK, Kheterpal S. 3,423 emergency tracheal intubations at a university hospital: airway outcomes and complications. Anesthesiology. 2011;114:42–8.PubMedCrossRef Martin LD, Mhyre JM, Shanks AM, Tremper KK, Kheterpal S. 3,423 emergency tracheal intubations at a university hospital: airway outcomes and complications. Anesthesiology. 2011;114:42–8.PubMedCrossRef
5.
go back to reference Ono Y, Kakamu T, Kikuchi H, Mori Y, Watanabe Y, Shinohara K. Expert-performed endotracheal intubation-related complications in trauma patients: incidence, possible risk factors, and outcomes in the Prehospital setting and emergency department. Emerg Med Int. 2018;2018:5649476.PubMedPubMedCentralCrossRef Ono Y, Kakamu T, Kikuchi H, Mori Y, Watanabe Y, Shinohara K. Expert-performed endotracheal intubation-related complications in trauma patients: incidence, possible risk factors, and outcomes in the Prehospital setting and emergency department. Emerg Med Int. 2018;2018:5649476.PubMedPubMedCentralCrossRef
6.
go back to reference Dashti M, Amini S, Azarfarin R, Totonchi Z, Hatami M. Hemodynamic changes following endotracheal intubation with glidescope® video-laryngoscope in patients with untreated hypertension. Res Cardiovasc Med. 2014;3:e17598.PubMedPubMedCentral Dashti M, Amini S, Azarfarin R, Totonchi Z, Hatami M. Hemodynamic changes following endotracheal intubation with glidescope® video-laryngoscope in patients with untreated hypertension. Res Cardiovasc Med. 2014;3:e17598.PubMedPubMedCentral
7.
go back to reference Lee H. The Pentax airway scope versus the Macintosh laryngoscope: comparison of hemodynamic responses and concentrations of plasma norepinephrine to tracheal intubation. Korean J Anesthesiol. 2013;64:315–20.PubMedPubMedCentralCrossRef Lee H. The Pentax airway scope versus the Macintosh laryngoscope: comparison of hemodynamic responses and concentrations of plasma norepinephrine to tracheal intubation. Korean J Anesthesiol. 2013;64:315–20.PubMedPubMedCentralCrossRef
8.
go back to reference Takeuchi S, Shiga T, Koyama Y, Nakanishi T, Honma Y, Morita H, et al. Longitudinal acquisition of endotracheal intubation skills in novice physicians. PLoS One. 2017;12:e0188224.PubMedPubMedCentralCrossRef Takeuchi S, Shiga T, Koyama Y, Nakanishi T, Honma Y, Morita H, et al. Longitudinal acquisition of endotracheal intubation skills in novice physicians. PLoS One. 2017;12:e0188224.PubMedPubMedCentralCrossRef
9.
go back to reference Kim C, Kang HG, Lim TH, Choi BY, Shin YJ, Choi HJ. What factors affect the success rate of the first attempt at endotracheal intubation in emergency departments? Emerg Med J. 2013;30:888–92.PubMedCrossRef Kim C, Kang HG, Lim TH, Choi BY, Shin YJ, Choi HJ. What factors affect the success rate of the first attempt at endotracheal intubation in emergency departments? Emerg Med J. 2013;30:888–92.PubMedCrossRef
10.
go back to reference Goto Y, Goto T, Hagiwara Y, Tsugawa Y, Watase H, Okamoto H, et al. Techniques and outcomes of emergency airway management in Japan: an analysis of two multicentre prospective observational studies, 2010-2016. Resuscitation. 2017;114:14–20.PubMedCrossRef Goto Y, Goto T, Hagiwara Y, Tsugawa Y, Watase H, Okamoto H, et al. Techniques and outcomes of emergency airway management in Japan: an analysis of two multicentre prospective observational studies, 2010-2016. Resuscitation. 2017;114:14–20.PubMedCrossRef
11.
go back to reference Goto Y, Watase H, Brown CA 3rd, Tsuboi S, Kondo T, Brown DFM, et al. Emergency airway management by resident physicians in Japan: an analysis of multicentre prospective observational study. Acute Med Surg. 2014;1:214–21.PubMedPubMedCentralCrossRef Goto Y, Watase H, Brown CA 3rd, Tsuboi S, Kondo T, Brown DFM, et al. Emergency airway management by resident physicians in Japan: an analysis of multicentre prospective observational study. Acute Med Surg. 2014;1:214–21.PubMedPubMedCentralCrossRef
12.
go back to reference Driver B, Dodd K, Klein LR, Buckley R, Robinson A, McGill JW, et al. The Bougie and first-pass success in the emergency department. Ann Emerg Med. 2017;70:473–8.PubMedCrossRef Driver B, Dodd K, Klein LR, Buckley R, Robinson A, McGill JW, et al. The Bougie and first-pass success in the emergency department. Ann Emerg Med. 2017;70:473–8.PubMedCrossRef
13.
go back to reference Driver BE, Prekker ME, Klein LR, Reardon RF, Miner JR, Fagerstrom ET, et al. Effect of use of a Bougie vs endotracheal tube and Stylet on first-attempt intubation success among patients with difficult airways undergoing emergency intubation: a randomized clinical trial. JAMA. 2018;319:2179–89.PubMedPubMedCentralCrossRef Driver BE, Prekker ME, Klein LR, Reardon RF, Miner JR, Fagerstrom ET, et al. Effect of use of a Bougie vs endotracheal tube and Stylet on first-attempt intubation success among patients with difficult airways undergoing emergency intubation: a randomized clinical trial. JAMA. 2018;319:2179–89.PubMedPubMedCentralCrossRef
14.
go back to reference Combes X, Jabre P, Margenet A, Merle JC, Leroux B, Dru M, et al. Unanticipated difficult airway management in the prehospital emergency setting: prospective validation of an algorithm. Anesthesiology. 2011;114:105–10.PubMedCrossRef Combes X, Jabre P, Margenet A, Merle JC, Leroux B, Dru M, et al. Unanticipated difficult airway management in the prehospital emergency setting: prospective validation of an algorithm. Anesthesiology. 2011;114:105–10.PubMedCrossRef
15.
go back to reference Combes X, Le Roux B, Suen P, Dumerat M, Motamed C, Sauvat S, et al. Unanticipated difficult airway in anesthetized patients: prospective validation of a management algorithm. Anesthesiology. 2004;100:1146–50.PubMedCrossRef Combes X, Le Roux B, Suen P, Dumerat M, Motamed C, Sauvat S, et al. Unanticipated difficult airway in anesthetized patients: prospective validation of a management algorithm. Anesthesiology. 2004;100:1146–50.PubMedCrossRef
16.
go back to reference Hung RK, Lewinsohn A, Jovaisa T, Wijayatilake DS, Sherren PB. Does the use of a bougie reduce the force of laryngoscopy in a difficult airway with manual in-line stabilisation?: a randomised crossover simulation study. Eur J Anaesthesiol. 2013;30:563–6.PubMedCrossRef Hung RK, Lewinsohn A, Jovaisa T, Wijayatilake DS, Sherren PB. Does the use of a bougie reduce the force of laryngoscopy in a difficult airway with manual in-line stabilisation?: a randomised crossover simulation study. Eur J Anaesthesiol. 2013;30:563–6.PubMedCrossRef
17.
go back to reference Maclure M, Mittleman MA. Should we use a case-crossover design? Annu Rev Public Health. 2000;21:193–221.PubMedCrossRef Maclure M, Mittleman MA. Should we use a case-crossover design? Annu Rev Public Health. 2000;21:193–221.PubMedCrossRef
18.
go back to reference Liu L, Tanigawa K, Kusunoki S, Tamura T, Ota K, Yamaga S, et al. Tracheal intubation of a difficult airway using airway scope, Airtraq, and Macintosh laryngoscope: a comparative manikin study of inexperienced personnel. Anesth Analg. 2010;110:1049–55.PubMedCrossRef Liu L, Tanigawa K, Kusunoki S, Tamura T, Ota K, Yamaga S, et al. Tracheal intubation of a difficult airway using airway scope, Airtraq, and Macintosh laryngoscope: a comparative manikin study of inexperienced personnel. Anesth Analg. 2010;110:1049–55.PubMedCrossRef
19.
go back to reference De Oliveira GS Jr, Glassenberg R, Chang R, Fitzgerald P, McCarthy RJ. Virtual airway simulation to improve dexterity among novices performing fibreoptic intubation. Anaesthesia. 2013;68:1053–8.PubMedCrossRef De Oliveira GS Jr, Glassenberg R, Chang R, Fitzgerald P, McCarthy RJ. Virtual airway simulation to improve dexterity among novices performing fibreoptic intubation. Anaesthesia. 2013;68:1053–8.PubMedCrossRef
20.
go back to reference Silva R, Minim V, Silva A, Simiqueli A, Della Lucia S, Minim L. Balanced incomplete block design: an alternative for data collection in the optimized descriptive profile. Food Res Int. 2014;64:289–97.PubMedCrossRef Silva R, Minim V, Silva A, Simiqueli A, Della Lucia S, Minim L. Balanced incomplete block design: an alternative for data collection in the optimized descriptive profile. Food Res Int. 2014;64:289–97.PubMedCrossRef
21.
go back to reference Truszewski Z, Czyzewski L, Smereka J, Krajewski P, Fudalej M, Madziala M, Szarpak L. Ability of paramedics to perform endotracheal intubation during continuous chest compressions: a randomized cadaver study comparing Pentax AWS and Macintosh laryngoscopes. Am J Emerg Med. 2016;34:1835–9.PubMedCrossRef Truszewski Z, Czyzewski L, Smereka J, Krajewski P, Fudalej M, Madziala M, Szarpak L. Ability of paramedics to perform endotracheal intubation during continuous chest compressions: a randomized cadaver study comparing Pentax AWS and Macintosh laryngoscopes. Am J Emerg Med. 2016;34:1835–9.PubMedCrossRef
22.
go back to reference Ono Y, Sugiyama T, Chida Y, Sato T, Kikuchi H, et al. Association between off-hour presentation and endotracheal-intubation-related adverse events in trauma patients with a predicted difficult airway: a historical cohort study at a community emergency department in Japan. Scand J Trauma Resusc Emerg Med. 2016;24:106.PubMedPubMedCentralCrossRef Ono Y, Sugiyama T, Chida Y, Sato T, Kikuchi H, et al. Association between off-hour presentation and endotracheal-intubation-related adverse events in trauma patients with a predicted difficult airway: a historical cohort study at a community emergency department in Japan. Scand J Trauma Resusc Emerg Med. 2016;24:106.PubMedPubMedCentralCrossRef
23.
24.
go back to reference Nakanishi T, Shiga T, Homma Y, Koyama Y, Goto T. Comparison of the force applied on oral structures during intubation attempts by novice physicians between the Macintosh direct laryngoscope, Airway Scope and C-MAC PM: a high-fidelity simulator-based study. BMJ Open. 2016;6:e011039.PubMedPubMedCentralCrossRef Nakanishi T, Shiga T, Homma Y, Koyama Y, Goto T. Comparison of the force applied on oral structures during intubation attempts by novice physicians between the Macintosh direct laryngoscope, Airway Scope and C-MAC PM: a high-fidelity simulator-based study. BMJ Open. 2016;6:e011039.PubMedPubMedCentralCrossRef
25.
go back to reference Goto T, Koyama Y, Kondo T, Tsugawa Y, Hasegawa K. A comparison of the force applied on oral structures during intubation attempts between the Pentax-AWS airwayscope and the Macintosh laryngoscope: a high-fidelity simulator-based study. BMJ Open. 2014;4:e006416.PubMedPubMedCentralCrossRef Goto T, Koyama Y, Kondo T, Tsugawa Y, Hasegawa K. A comparison of the force applied on oral structures during intubation attempts between the Pentax-AWS airwayscope and the Macintosh laryngoscope: a high-fidelity simulator-based study. BMJ Open. 2014;4:e006416.PubMedPubMedCentralCrossRef
26.
go back to reference Evans A, Vaughan RS, Hall JE, Mecklenburgh J, Wilkes AR. A comparison of the forces exerted during laryngoscopy using disposable and non-disposable laryngoscope blades. Anaesthesia. 2003;58:869–73.PubMedCrossRef Evans A, Vaughan RS, Hall JE, Mecklenburgh J, Wilkes AR. A comparison of the forces exerted during laryngoscopy using disposable and non-disposable laryngoscope blades. Anaesthesia. 2003;58:869–73.PubMedCrossRef
27.
go back to reference Lee RA, van Zundert AA, Maassen RL, Wieringa PA. Forces applied to the maxillary incisors by video laryngoscopes and the Macintosh laryngoscope. Acta Anaesthesiol Scand. 2012;56:224–9.PubMedCrossRef Lee RA, van Zundert AA, Maassen RL, Wieringa PA. Forces applied to the maxillary incisors by video laryngoscopes and the Macintosh laryngoscope. Acta Anaesthesiol Scand. 2012;56:224–9.PubMedCrossRef
28.
go back to reference Lee RA, van Zundert AA, Maassen RL, Willems RJ, Beeke LP, Schaaper JN, et al. Forces applied to the maxillary incisors during video-assisted intubation. Anesth Analg. 2009;108:187–91.PubMedCrossRef Lee RA, van Zundert AA, Maassen RL, Willems RJ, Beeke LP, Schaaper JN, et al. Forces applied to the maxillary incisors during video-assisted intubation. Anesth Analg. 2009;108:187–91.PubMedCrossRef
29.
go back to reference Fukuda T, Sugimoto Y, Yamashita S, Toyooka H, Tanaka M. Forces applied to the maxillary incisors during tracheal intubation and dental injury risks of intubation by beginners: a manikin study. Acta Anaesthesiol Taiwanica. 2011;49:12–5.CrossRef Fukuda T, Sugimoto Y, Yamashita S, Toyooka H, Tanaka M. Forces applied to the maxillary incisors during tracheal intubation and dental injury risks of intubation by beginners: a manikin study. Acta Anaesthesiol Taiwanica. 2011;49:12–5.CrossRef
30.
go back to reference Waddington MS, Paech MJ, Kurowski IH, Reed CJ, Nicholls GJ, Guy DT, et al. The influence of gender and experience on intubation ability and technique: a manikin study. Anaesth Intensive Care. 2009;37:791–801.PubMedCrossRef Waddington MS, Paech MJ, Kurowski IH, Reed CJ, Nicholls GJ, Guy DT, et al. The influence of gender and experience on intubation ability and technique: a manikin study. Anaesth Intensive Care. 2009;37:791–801.PubMedCrossRef
31.
go back to reference Pieters B, Maassen R, Van Eig E, Maathuis B, Van Den Dobbelsteen J, Van Zundert A. Indirect video laryngoscopy using Macintosh blades in patients with non-anticipated difficult airways results in significantly lower forces exerted on teeth relative to classic direct laryngoscopy: a randomized crossover trial. Minerva Anestesiol. 2015;81:846–54.PubMed Pieters B, Maassen R, Van Eig E, Maathuis B, Van Den Dobbelsteen J, Van Zundert A. Indirect video laryngoscopy using Macintosh blades in patients with non-anticipated difficult airways results in significantly lower forces exerted on teeth relative to classic direct laryngoscopy: a randomized crossover trial. Minerva Anestesiol. 2015;81:846–54.PubMed
32.
go back to reference D'Amico EJ, Neilands TB, Zambarano R. Power analysis for multivariate and repeated measures designs: a flexible approach using the SPSS MANOVA procedure. Behav Res Methods Instrum Comput. 2001;33:479–84.PubMedCrossRef D'Amico EJ, Neilands TB, Zambarano R. Power analysis for multivariate and repeated measures designs: a flexible approach using the SPSS MANOVA procedure. Behav Res Methods Instrum Comput. 2001;33:479–84.PubMedCrossRef
34.
go back to reference Owen H, Waddell-Smith I. Dental trauma associated with anaesthesia. Anaesth Intensive Care. 2000;28:133e45.CrossRef Owen H, Waddell-Smith I. Dental trauma associated with anaesthesia. Anaesth Intensive Care. 2000;28:133e45.CrossRef
35.
go back to reference Warner ME, Benenfeld SM, Warner MA, Schroeder DR, Maxson PM. Perianesthetic dental injuries: frequency, outcomes, and risk factors. Anesthesiology. 1999;90:1302–5.PubMedCrossRef Warner ME, Benenfeld SM, Warner MA, Schroeder DR, Maxson PM. Perianesthetic dental injuries: frequency, outcomes, and risk factors. Anesthesiology. 1999;90:1302–5.PubMedCrossRef
36.
go back to reference Derbyshire DR, Chmielewski A, Fell D, Vater M, Achola K, Smith G. Plasma catecholamine responses to tracheal intubation. Br J Anaesth. 1983;55:855–60.PubMedCrossRef Derbyshire DR, Chmielewski A, Fell D, Vater M, Achola K, Smith G. Plasma catecholamine responses to tracheal intubation. Br J Anaesth. 1983;55:855–60.PubMedCrossRef
37.
go back to reference Koyama Y, Nishihama M, Inagawa G, Kamiya Y, Miki T, Kurihara R, Goto T. Comparison of haemodynamic responses to tracheal intubation using the airway scope® and Macintosh laryngoscope in normotensive and hypertensive patients. Anaesthesia. 2011;66:895–900.PubMedCrossRef Koyama Y, Nishihama M, Inagawa G, Kamiya Y, Miki T, Kurihara R, Goto T. Comparison of haemodynamic responses to tracheal intubation using the airway scope® and Macintosh laryngoscope in normotensive and hypertensive patients. Anaesthesia. 2011;66:895–900.PubMedCrossRef
38.
go back to reference Fontijn-Tekamp FA, Slagter AP, Van Der Bilt A, Van’T Hof MA, Witter DJ, Kalk W, et al. Biting and chewing in overdentures, full dentures, and natural dentitions. J Dent Res. 2000;79:1519–24.PubMedCrossRef Fontijn-Tekamp FA, Slagter AP, Van Der Bilt A, Van’T Hof MA, Witter DJ, Kalk W, et al. Biting and chewing in overdentures, full dentures, and natural dentitions. J Dent Res. 2000;79:1519–24.PubMedCrossRef
39.
go back to reference Noguchi T, Koga K, Shiga Y, Shigematsu A. The gum elastic bougie eases tracheal intubation while applying cricoid pressure compared to a stylet. Can J Anaesth. 2003;50:712–7.PubMedCrossRef Noguchi T, Koga K, Shiga Y, Shigematsu A. The gum elastic bougie eases tracheal intubation while applying cricoid pressure compared to a stylet. Can J Anaesth. 2003;50:712–7.PubMedCrossRef
40.
go back to reference Komasawa N, Cho T, Mihara R, Minami T. Utility of gum-elastic bougie for tracheal intubation during chest compressions in a manikin: a randomized crossover trial. Am J Emerg Med. 2016;34:54–6.PubMedCrossRef Komasawa N, Cho T, Mihara R, Minami T. Utility of gum-elastic bougie for tracheal intubation during chest compressions in a manikin: a randomized crossover trial. Am J Emerg Med. 2016;34:54–6.PubMedCrossRef
41.
go back to reference Komasawa N, Hyoda A, Matsunami S, Majima N, Minami T. Utility of a gum-elastic bougie for difficult airway management in infants: a simulation-based crossover analysis. Biomed Res Int. 2015;2015:617805.PubMedPubMedCentral Komasawa N, Hyoda A, Matsunami S, Majima N, Minami T. Utility of a gum-elastic bougie for difficult airway management in infants: a simulation-based crossover analysis. Biomed Res Int. 2015;2015:617805.PubMedPubMedCentral
42.
go back to reference Sakles JC, Mosier J, Chiu S, Cosentino M, Kalin L. A comparison of the C-MAC video laryngoscope to the Macintosh direct laryngoscope for intubation in the emergency department. Ann Emerg Med. 2012;60:739–48.PubMedPubMedCentralCrossRef Sakles JC, Mosier J, Chiu S, Cosentino M, Kalin L. A comparison of the C-MAC video laryngoscope to the Macintosh direct laryngoscope for intubation in the emergency department. Ann Emerg Med. 2012;60:739–48.PubMedPubMedCentralCrossRef
43.
go back to reference Brown CA 3rd, Kaji AH, Fantegrossi A, Carlson JN, April MD, Kilgo RW, et al. Video laryngoscopy compared to augmented direct laryngoscopy in adult emergency department tracheal intubations: a National Emergency Airway Registry (NEAR) study. Acad Emerg Med. 2020;27:100–8.PubMedCrossRef Brown CA 3rd, Kaji AH, Fantegrossi A, Carlson JN, April MD, Kilgo RW, et al. Video laryngoscopy compared to augmented direct laryngoscopy in adult emergency department tracheal intubations: a National Emergency Airway Registry (NEAR) study. Acad Emerg Med. 2020;27:100–8.PubMedCrossRef
44.
go back to reference Hwang S, Lee SU, Lee TR, Yoon H, Park JH, Cha WC, et al. Usefulness of C-MAC video laryngoscope in direct laryngoscopy training in the emergency department: a propensity score matching analysis. PLoS One. 2018;13:e0208077.PubMedPubMedCentralCrossRef Hwang S, Lee SU, Lee TR, Yoon H, Park JH, Cha WC, et al. Usefulness of C-MAC video laryngoscope in direct laryngoscopy training in the emergency department: a propensity score matching analysis. PLoS One. 2018;13:e0208077.PubMedPubMedCentralCrossRef
45.
go back to reference Ångerman S, Kirves H, Nurmi J. A before-and-after observational study of a protocol for use of the C-MAC videolaryngoscope with a Frova introducer in pre-hospital rapid sequence intubation. Anaesthesia. 2018;73:348–55.PubMedCrossRef Ångerman S, Kirves H, Nurmi J. A before-and-after observational study of a protocol for use of the C-MAC videolaryngoscope with a Frova introducer in pre-hospital rapid sequence intubation. Anaesthesia. 2018;73:348–55.PubMedCrossRef
46.
go back to reference Elattar H, Abdel-Rahman I, Ibrahim M, Kocz R, Raczka M, Kumar A, et al. A randomized trial of the glottic views with the classic miller, Wis-Hipple and C-MAC (videolaryngoscope and direct views) straight size 1 blades in young children. J Clin Anesth. 2020;60:57–61.PubMedCrossRef Elattar H, Abdel-Rahman I, Ibrahim M, Kocz R, Raczka M, Kumar A, et al. A randomized trial of the glottic views with the classic miller, Wis-Hipple and C-MAC (videolaryngoscope and direct views) straight size 1 blades in young children. J Clin Anesth. 2020;60:57–61.PubMedCrossRef
47.
go back to reference Risse J, Volberg C, Kratz T, Plöger B, Jerrentrup A, Pabst D, et al. Comparison of videolaryngoscopy and direct laryngoscopy by German paramedics during out-of-hospital cardiopulmonary resuscitation; an observational prospective study. BMC Emerg Med. 2020;20:22.PubMedPubMedCentralCrossRef Risse J, Volberg C, Kratz T, Plöger B, Jerrentrup A, Pabst D, et al. Comparison of videolaryngoscopy and direct laryngoscopy by German paramedics during out-of-hospital cardiopulmonary resuscitation; an observational prospective study. BMC Emerg Med. 2020;20:22.PubMedPubMedCentralCrossRef
48.
go back to reference Asai T. Tracheal intubation with restricted access: a randomised comparison of the Pentax-airway scope and Macintosh laryngoscope in a manikin. Anaesthesia. 2009;64:1114–7.PubMedCrossRef Asai T. Tracheal intubation with restricted access: a randomised comparison of the Pentax-airway scope and Macintosh laryngoscope in a manikin. Anaesthesia. 2009;64:1114–7.PubMedCrossRef
Metadata
Title
Lower maximum forces on oral structures when using gum-elastic bougie than when using endotracheal tube and stylet during both direct and indirect laryngoscopy by novices: a crossover study using a high-fidelity simulator
Authors
Yuko Ono
Kazuaki Shinohara
Jiro Shimada
Shigeaki Inoue
Joji Kotani
Publication date
01-12-2020
Publisher
BioMed Central
Published in
BMC Emergency Medicine / Issue 1/2020
Electronic ISSN: 1471-227X
DOI
https://doi.org/10.1186/s12873-020-00328-9

Other articles of this Issue 1/2020

BMC Emergency Medicine 1/2020 Go to the issue