The impact of electrosurgical devices on electromyography-based neuromuscular monitoring during abdominal laparotomy: a case series

Authors: Hajime Iwasaki, Kenichi Takahoko, Takuro Matsui, Yui Takada, Yukako Takahashi, Ami Sugawara, Atsushi Kurosawa

Published in: Journal of Clinical Monitoring and Computing | Issue 4/2023

Login to get access

Abstract

The present study aimed to evaluate the effect of electrosurgical devices on neuromuscular monitoring using an electromyography (EMG)-based neuromuscular monitor during abdominal laparotomy. Seventeen women (aged 32–64 years) undergoing gynecological laparotomy under total intravenous general anesthesia were enrolled in the study. A TetraGraph™ was placed to stimulate the ulnar nerve and to monitor the abductor digiti minimi muscle. After device calibration, train-of-four (TOF) measurements were repeated at intervals of 20 s. Rocuronium 0.6 to 0.9 mg/kg was administered for induction, and additional doses of 0.1 to 0.2 mg/kg were administered to maintain TOF counts ≤ 2 during the surgery. The primary outcome of the study was the ratio of measurement failure. The secondary outcomes of the study were the total number of measurements, the number of measurement failures, and the most extended consecutive number of measurement failures. The data are expressed as median (range). Of the 3091 (1480–8134) measurements, the number of measurement failures was 94 (60–200), resulting in a failure ratio of 3.5% (1.4–6.5%). The most extended consecutive number of measurement failures was 8 (4–13). All attending anesthesiologists were able to maintain and reverse neuromuscular blocks under EMG guidance. This prospective observational study demonstrated that the use of EMG-based neuromuscular monitoring does not seem to be heavily affected by electrical interference during lower abdominal laparotomic surgery.
Trial registration This trial was registered in the University Hospital Medical Information Network under registration number UMIN000048138 (registration date; June 23, 2022).
Literature
1.
go back to reference Iwasaki H, Yamamoto M, Sato H, Doshu-Kajiura A, Kitajima O, Takagi S, Luthe SK, Suzuki T. A comparison between the Adductor Pollicis muscle using TOF-Watch SX and the Abductor Digiti Minimi muscle using TetraGraph in Rocuronium-Induced Neuromuscular Block: a prospective observational study. Anesth Analg. 2022;135(2):370–5. https://doi.org/10.1213/ane.0000000000005897CrossRefPubMed Iwasaki H, Yamamoto M, Sato H, Doshu-Kajiura A, Kitajima O, Takagi S, Luthe SK, Suzuki T. A comparison between the Adductor Pollicis muscle using TOF-Watch SX and the Abductor Digiti Minimi muscle using TetraGraph in Rocuronium-Induced Neuromuscular Block: a prospective observational study. Anesth Analg. 2022;135(2):370–5. https://​doi.​org/​10.​1213/​ane.​0000000000005897​CrossRefPubMed
Metadata
Title
The impact of electrosurgical devices on electromyography-based neuromuscular monitoring during abdominal laparotomy: a case series
Authors
Hajime Iwasaki
Kenichi Takahoko
Takuro Matsui
Yui Takada
Yukako Takahashi
Ami Sugawara
Atsushi Kurosawa
Publication date
27-05-2023
Publisher
Springer Netherlands
Published in
Journal of Clinical Monitoring and Computing / Issue 4/2023
Print ISSN: 1387-1307
Electronic ISSN: 1573-2614
DOI
https://doi.org/10.1007/s10877-023-01039-9

Other articles of this Issue 4/2023

Journal of Clinical Monitoring and Computing 4/2023 Go to the issue