Does lucency equate to revision? A five-year retrospective review of Attune and Triathlon total knee arthroplasty

Authors: Paul O’Donovan, Timothy McAleese, James Harty

Published in: Knee Surgery, Sports Traumatology, Arthroscopy | Issue 11/2023

Login to get access

Abstract

Purpose

The Attune® total knee arthroplasty system was introduced in 2013 to address lingering issues of patient dissatisfaction. However, recent literature reports concerns of early tibial tray debonding. The aim of this study was to compare the incidence of radiolucent lines, survivorship and patient reported outcome-measures between the Attune® system and the well-established Triathlon® system.

Methods

This retrospective database review was conducted at a single institution in Cork, Ireland. All primary Attune® (N = 445) and Triathlon® (N = 285) systems implanted between 2015 and 2016 were reviewed. Radiolucent lines were assessed for those with a minimum two-year radiological follow-up (Attune® = 338; Triathlon® = 231). X-rays were taken post op, at 6 months, 2 years and 5 years. Radiolucent lines were documented using the Modern Knee Society Radiographic System. Five-year survival was assessed using Kaplan–Meier analysis with the Log Rank method to determine statistical significance. The Oxford Knee Score and EQ-5D-5L, were collected pre-op, at 6 months, 2 years and 5 years post-operatively and compared using the Kruskal–Wallis Test.

Results

The Attune® had a higher proportion of radiolucent lines at the tibial tray [87.1% (54/62) vs 61.4% (27/44); p = 0.001] and at the implant–cement interface [62.9% (39/62) vs 43.2% (19/44); p = 0.02]. Conversely, the Triathlon® had a higher proportion AT the femur [38.6% (17/44) vs 12.9% (8/62); p = 0.001] and at the cement–bone interface [56.8% (25/44) vs 37.1% (23/62); p = 0.02]. The overall frequency of radiolucent lines was similar in both the Attune® and Triathlon® groups [17.8%, (60/338) vs 17.7%, (41/231); p = 0.49]. There was no difference in revision-free survival analysis at 5 years (Attune® 97.8% vs Triathlon® 95.8%; p = 0.129). The Attune® performed better at 5 years in the Oxford Knee Score [Attune® = 42.6 (SD 5.2) vs Triathlon® = 41 (SD 6.4); p = 0.001] and in the EQ-5D [Attune® = 0.773 (SD 0.187) vs Triathlon® = 0.729 (SD 0.218); p = 0.013]. There was no difference at 5 years in the EQ-VAS [Attune® = 80.4 (SD 13.7) vs Triathlon® = 78.5 (SD 15.3); p = 0.25].

Conclusion

The Attune® system exhibited a higher incidence of  radiolucent lines at the tibial tray. However, this did not lead to decreased survivorship at medium term follow-up compared to the Triathlon®. Furthermore, improvements in patient reported outcomes modestly favoured the Attune® system.

Level of evidence

III.
Literature
1.
go back to reference Araujo IL, Castro MC, Daltro C, Matos MA (2016) Quality of life and functional independence in patients with osteoarthritis of the knee. Knee Surg Relat Res 28:219–224CrossRefPubMedPubMedCentral Araujo IL, Castro MC, Daltro C, Matos MA (2016) Quality of life and functional independence in patients with osteoarthritis of the knee. Knee Surg Relat Res 28:219–224CrossRefPubMedPubMedCentral
2.
go back to reference Behrend H, Hochreiter B, Potocnik P, El Baz Y, Zdravkovic V, Tomazi T (2020) No difference in radiolucent lines after TKA: a matched-pair analysis of the classic implant and its evolutional design. Knee Surg Sports Traumatol Arthrosc 28:3962–3968CrossRefPubMed Behrend H, Hochreiter B, Potocnik P, El Baz Y, Zdravkovic V, Tomazi T (2020) No difference in radiolucent lines after TKA: a matched-pair analysis of the classic implant and its evolutional design. Knee Surg Sports Traumatol Arthrosc 28:3962–3968CrossRefPubMed
3.
go back to reference Billi F, Kavanaugh A, Schmalzried H, Schmalzried TP (2019) Techniques for improving the initial strength of the tibial tray-cement interface bond. Bone Joint J 101:53–58CrossRefPubMed Billi F, Kavanaugh A, Schmalzried H, Schmalzried TP (2019) Techniques for improving the initial strength of the tibial tray-cement interface bond. Bone Joint J 101:53–58CrossRefPubMed
4.
go back to reference Bonutti PM, Khlopas A, Chughtai M, Cole C, Gwam CU, Harwin SF et al (2017) Unusually high rate of early failure of tibial component in ATTUNE total knee arthroplasty system at implant-cement interface. J Knee Surg 30:435–439CrossRefPubMed Bonutti PM, Khlopas A, Chughtai M, Cole C, Gwam CU, Harwin SF et al (2017) Unusually high rate of early failure of tibial component in ATTUNE total knee arthroplasty system at implant-cement interface. J Knee Surg 30:435–439CrossRefPubMed
5.
go back to reference Bourne RB, Chesworth B, Davis A, Mahomed N, Charron K (2010) Comparing patient outcomes after THA and TKA: is there a difference? Clin Orthop Relat Res 468:542–546CrossRefPubMed Bourne RB, Chesworth B, Davis A, Mahomed N, Charron K (2010) Comparing patient outcomes after THA and TKA: is there a difference? Clin Orthop Relat Res 468:542–546CrossRefPubMed
6.
go back to reference Cerquiglini A, Henckel J, Hothi H, Allen P, Lewis J, Eskelinen A et al (2019) Analysis of the attune tibial tray backside: a comparat ive retrieval study. Bone Jt Res 8:136–145CrossRef Cerquiglini A, Henckel J, Hothi H, Allen P, Lewis J, Eskelinen A et al (2019) Analysis of the attune tibial tray backside: a comparat ive retrieval study. Bone Jt Res 8:136–145CrossRef
7.
go back to reference Cerquiglini A, Henckel J, Hothi H, Allen P, Lewis J, Eskelinen A et al (2019) Analysis of the attune tibial tray backside: a comparative retrieval study. Bone Jt Res 8:136–145CrossRef Cerquiglini A, Henckel J, Hothi H, Allen P, Lewis J, Eskelinen A et al (2019) Analysis of the attune tibial tray backside: a comparative retrieval study. Bone Jt Res 8:136–145CrossRef
8.
go back to reference Chua JL, Goh GS, Liow MHL, Tay DK, Lo NN, Yeo SJ (2019) Modern TKA implants are equivalent to traditional TKA implants in functional and patellofemoral joint-related outcomes. Knee Surg Sports Traumatol Arthrosc 27:1116–1123CrossRefPubMed Chua JL, Goh GS, Liow MHL, Tay DK, Lo NN, Yeo SJ (2019) Modern TKA implants are equivalent to traditional TKA implants in functional and patellofemoral joint-related outcomes. Knee Surg Sports Traumatol Arthrosc 27:1116–1123CrossRefPubMed
9.
go back to reference Clement ND, MacDonald D, Simpson AH (2014) The minimal clinically important difference in the Oxford knee score and Short Form 12 score after total knee arthroplasty. Knee Surg Sports Traumatol Arthrosc 22:1933–1939CrossRefPubMed Clement ND, MacDonald D, Simpson AH (2014) The minimal clinically important difference in the Oxford knee score and Short Form 12 score after total knee arthroplasty. Knee Surg Sports Traumatol Arthrosc 22:1933–1939CrossRefPubMed
10.
go back to reference Elmallah RK, Scuderi GR, Jauregui JJ, Meneghini RM, Dennis DA, Backstein DB et al (2015) Radiographic evaluations of revision total knee arthroplasty: a plea for uniform assessments. J Arthroplasty 30:1981–1984CrossRefPubMed Elmallah RK, Scuderi GR, Jauregui JJ, Meneghini RM, Dennis DA, Backstein DB et al (2015) Radiographic evaluations of revision total knee arthroplasty: a plea for uniform assessments. J Arthroplasty 30:1981–1984CrossRefPubMed
11.
go back to reference Ewald FC (1989) The Knee Society total knee arthroplasty roentgenographic evaluation and scoring system. Clin Orthop Relat Res 2:9–12 Ewald FC (1989) The Knee Society total knee arthroplasty roentgenographic evaluation and scoring system. Clin Orthop Relat Res 2:9–12
12.
go back to reference Giaretta S, Berti M, Micheloni GM, Ceccato A, Marangoni F, Momoli A (2019) Early experience with the ATTUNE total knee replacement system. Acta Biomed 90:98–103PubMedPubMedCentral Giaretta S, Berti M, Micheloni GM, Ceccato A, Marangoni F, Momoli A (2019) Early experience with the ATTUNE total knee replacement system. Acta Biomed 90:98–103PubMedPubMedCentral
13.
go back to reference Greco NJ, Anderson AF, Mann BJ, Cole BJ, Farr J, Nissen CW et al (2010) Responsiveness of the international knee documentation committee subjective knee form in comparison to the Western Ontario and McMaster Universities Osteoarthritis Index, modified Cincinnati Knee Rating System, and Short Form 36 in patients with focal articular cartilage defects. Am J Sports Med 38:891–902CrossRefPubMed Greco NJ, Anderson AF, Mann BJ, Cole BJ, Farr J, Nissen CW et al (2010) Responsiveness of the international knee documentation committee subjective knee form in comparison to the Western Ontario and McMaster Universities Osteoarthritis Index, modified Cincinnati Knee Rating System, and Short Form 36 in patients with focal articular cartilage defects. Am J Sports Med 38:891–902CrossRefPubMed
14.
go back to reference Gulati A, Chau R, Pandit HG, Gray H, Price AJ, Dodd CAF et al (2009) The incidence of physiological radiolucency following Oxford unicompartmental knee replacement and its relationship to outcome. J Bone Jt Surg Br 91:896–902CrossRef Gulati A, Chau R, Pandit HG, Gray H, Price AJ, Dodd CAF et al (2009) The incidence of physiological radiolucency following Oxford unicompartmental knee replacement and its relationship to outcome. J Bone Jt Surg Br 91:896–902CrossRef
15.
go back to reference Hamilton WG, Brenkel IJ, Barnett SL, Allen PW, Dwyer KA, Lesko JP et al (2021) Comparison of existing and new total knee arthroplasty implant systems from the same manufacturer: a prospective, multicenter study. JAAOS Glob Res Rev 5(e21):00136 Hamilton WG, Brenkel IJ, Barnett SL, Allen PW, Dwyer KA, Lesko JP et al (2021) Comparison of existing and new total knee arthroplasty implant systems from the same manufacturer: a prospective, multicenter study. JAAOS Glob Res Rev 5(e21):00136
16.
go back to reference Hamilton WG, Brenkel IJ, Barnett SL, Allen PW, Dwyer KA, Lesko JP et al (2021) Comparison of existing and new total knee arthroplasty implant systems from the same manufacturer: a prospective, multicenter study. J Am Acad Orthop Surg Glob Res Rev 5:2 Hamilton WG, Brenkel IJ, Barnett SL, Allen PW, Dwyer KA, Lesko JP et al (2021) Comparison of existing and new total knee arthroplasty implant systems from the same manufacturer: a prospective, multicenter study. J Am Acad Orthop Surg Glob Res Rev 5:2
17.
go back to reference Kaptein BL, Hollander Pd, Thomassen B, Fiocco M, Nelissen RGHH (2020) A randomized controlled trial comparing tibial migration of the ATTUNE cemented cruciate-retaining knee prosthesis with the PFC-sigma design. Bone J J 102:1158–1166CrossRef Kaptein BL, Hollander Pd, Thomassen B, Fiocco M, Nelissen RGHH (2020) A randomized controlled trial comparing tibial migration of the ATTUNE cemented cruciate-retaining knee prosthesis with the PFC-sigma design. Bone J J 102:1158–1166CrossRef
18.
go back to reference Keohane D, Sheridan GA, Masterson E (2022) High rate of tibial debonding and failure in a popular knee replacement : a follow-up review. Bone Jt Open 3:495–501CrossRefPubMedPubMedCentral Keohane D, Sheridan GA, Masterson E (2022) High rate of tibial debonding and failure in a popular knee replacement : a follow-up review. Bone Jt Open 3:495–501CrossRefPubMedPubMedCentral
19.
go back to reference Lachiewicz PF, Steele JR, Wellman SS (2021) Unexpected high rate of revision of a modern cemented fixed bearing modular posterior-stabilized knee arthroplasty. Bone Jt J 103:137–144CrossRef Lachiewicz PF, Steele JR, Wellman SS (2021) Unexpected high rate of revision of a modern cemented fixed bearing modular posterior-stabilized knee arthroplasty. Bone Jt J 103:137–144CrossRef
20.
go back to reference Meneghini RM, Mont MA, Backstein DB, Bourne RB, Dennis DA, Scuderi GR (2015) Development of a modern knee society radiographic evaluation system and methodology for total knee arthroplasty. J Arthroplasty 30:2311–2314CrossRefPubMed Meneghini RM, Mont MA, Backstein DB, Bourne RB, Dennis DA, Scuderi GR (2015) Development of a modern knee society radiographic evaluation system and methodology for total knee arthroplasty. J Arthroplasty 30:2311–2314CrossRefPubMed
21.
go back to reference Prodromidis AD, Chloros GD, Thivaios GC, Sutton PM, Pandit H, Giannoudis PV et al (2023) High rate of radiolucent lines following the cemented original design of the ATTUNE total knee arthroplasty. Bone Jt J 105:610–621CrossRef Prodromidis AD, Chloros GD, Thivaios GC, Sutton PM, Pandit H, Giannoudis PV et al (2023) High rate of radiolucent lines following the cemented original design of the ATTUNE total knee arthroplasty. Bone Jt J 105:610–621CrossRef
22.
go back to reference Ranawat CS, White PB, West S, Ranawat AS (2017) Clinical and radiographic results of attune and PFC sigma knee designs at 2-year follow-up: a prospective matched-pair analysis. J Arthroplasty 32:431–436CrossRefPubMed Ranawat CS, White PB, West S, Ranawat AS (2017) Clinical and radiographic results of attune and PFC sigma knee designs at 2-year follow-up: a prospective matched-pair analysis. J Arthroplasty 32:431–436CrossRefPubMed
24.
go back to reference Registry AOANJR, (AOANJRR) (2020) Hip, knee, & shoulder arthroplasty: 2020 annual report. Accessed 2020 Jan 1 Registry AOANJR, (AOANJRR) (2020) Hip, knee, & shoulder arthroplasty: 2020 annual report. Accessed 2020 Jan 1
26.
go back to reference Robinson T, King SW, Pilling RWD, Aderinto J, Veysi V, Wall O et al (2021) Attune total knee arthroplasty: is there evidence of early tibial component de-bonding? A prospective cohort study with a minimum two year follow-up. J Arthroscopy Jt Surg 8:139–147CrossRef Robinson T, King SW, Pilling RWD, Aderinto J, Veysi V, Wall O et al (2021) Attune total knee arthroplasty: is there evidence of early tibial component de-bonding? A prospective cohort study with a minimum two year follow-up. J Arthroscopy Jt Surg 8:139–147CrossRef
27.
go back to reference Rodríguez-Collell JR, Mifsut D, Ruiz-Sauri A, Rodríguez-Pino L, González-Soler EM, Valverde-Navarro AA (2021) Improving the cementation of the tibial component in knee arthroplasty : a study of four techniques in the cadaver. Bone Jt Res 10:467–473CrossRef Rodríguez-Collell JR, Mifsut D, Ruiz-Sauri A, Rodríguez-Pino L, González-Soler EM, Valverde-Navarro AA (2021) Improving the cementation of the tibial component in knee arthroplasty : a study of four techniques in the cadaver. Bone Jt Res 10:467–473CrossRef
28.
go back to reference Schlegel UJ, Bishop NE, Püschel K, Morlock MM, Nagel K (2015) Comparison of different cement application techniques for tibial component fixation in TKA. Int Orthop 39:47–54CrossRefPubMed Schlegel UJ, Bishop NE, Püschel K, Morlock MM, Nagel K (2015) Comparison of different cement application techniques for tibial component fixation in TKA. Int Orthop 39:47–54CrossRefPubMed
29.
go back to reference Staats K, Wannmacher T, Weihs V, Koller U, Kubista B, Windhager R (2019) Modern cemented total knee arthroplasty design shows a higher incidence of radiolucent lines compared to its predecessor. Knee Surg Sports Traumatol Arthrosc 27:1148–1155CrossRefPubMed Staats K, Wannmacher T, Weihs V, Koller U, Kubista B, Windhager R (2019) Modern cemented total knee arthroplasty design shows a higher incidence of radiolucent lines compared to its predecessor. Knee Surg Sports Traumatol Arthrosc 27:1148–1155CrossRefPubMed
30.
go back to reference Turgeon TR, Gascoyne TC, Laende EK, Dunbar MJ, Bohm ER, Richardson CG (2018) The assessment of the stability of the tibial component of a novel knee arthroplasty system using radiostereometric analysis. Bone Jt J 100:1579–1584CrossRef Turgeon TR, Gascoyne TC, Laende EK, Dunbar MJ, Bohm ER, Richardson CG (2018) The assessment of the stability of the tibial component of a novel knee arthroplasty system using radiostereometric analysis. Bone Jt J 100:1579–1584CrossRef
31.
go back to reference Vandenbroucke JP, von Elm E, Altman DG, Gøtzsche PC, Mulrow CD, Pocock SJ et al (2014) Strengthening the reporting of observational studies in epidemiology (STROBE): explanation and elaboration. Int J Surg 12:1500–1524CrossRefPubMed Vandenbroucke JP, von Elm E, Altman DG, Gøtzsche PC, Mulrow CD, Pocock SJ et al (2014) Strengthening the reporting of observational studies in epidemiology (STROBE): explanation and elaboration. Int J Surg 12:1500–1524CrossRefPubMed
32.
go back to reference Vanitcharoenkul E, Unnanuntana A (2021) Midterm functional recovery of Total knee arthroplasty patients compared between the ATTUNE knee system and the press fit condylar (PFC) SIGMA knee system. BMC Musculoskelet Disord 22:620CrossRefPubMedPubMedCentral Vanitcharoenkul E, Unnanuntana A (2021) Midterm functional recovery of Total knee arthroplasty patients compared between the ATTUNE knee system and the press fit condylar (PFC) SIGMA knee system. BMC Musculoskelet Disord 22:620CrossRefPubMedPubMedCentral
33.
go back to reference Webb JE, Yang HY, Collins JE, Losina E, Thornhill TS, Katz JN (2017) The evolution of implant design decreases the incidence of lateral release in primary total knee arthroplasty. J Arthroplasty 32:1505–1509CrossRefPubMed Webb JE, Yang HY, Collins JE, Losina E, Thornhill TS, Katz JN (2017) The evolution of implant design decreases the incidence of lateral release in primary total knee arthroplasty. J Arthroplasty 32:1505–1509CrossRefPubMed
34.
go back to reference White PB, Turcan S, Satalich JR, Ranawat AS, Ranawat CS (2020) Clinical outcomes of a modern total knee arthroplasty prosthesis compared to its predecessor at 5-year follow-up: matched pair analysis. J Arthroplasty 35:3150–3155CrossRefPubMed White PB, Turcan S, Satalich JR, Ranawat AS, Ranawat CS (2020) Clinical outcomes of a modern total knee arthroplasty prosthesis compared to its predecessor at 5-year follow-up: matched pair analysis. J Arthroplasty 35:3150–3155CrossRefPubMed
35.
go back to reference Willburger RE, Oberberg S (2022) Early and mid-term results with the ATTUNE total knee replacement system compared to PFC Sigma: a prospective comparative study. J Orthop Surg Res 17:509CrossRefPubMedPubMedCentral Willburger RE, Oberberg S (2022) Early and mid-term results with the ATTUNE total knee replacement system compared to PFC Sigma: a prospective comparative study. J Orthop Surg Res 17:509CrossRefPubMedPubMedCentral
Metadata
Title
Does lucency equate to revision? A five-year retrospective review of Attune and Triathlon total knee arthroplasty
Authors
Paul O’Donovan
Timothy McAleese
James Harty
Publication date
30-07-2023
Publisher
Springer Berlin Heidelberg
Keywords
Knee-TEP
Knee-TEP
Published in
Knee Surgery, Sports Traumatology, Arthroscopy / Issue 11/2023
Print ISSN: 0942-2056
Electronic ISSN: 1433-7347
DOI
https://doi.org/10.1007/s00167-023-07509-6

Other articles of this Issue 11/2023

Knee Surgery, Sports Traumatology, Arthroscopy 11/2023 Go to the issue