Published in: Graefe's Archive for Clinical and Experimental Ophthalmology 7/2021

01-07-2021 | Keratoplasty | Basic Science

Ultra-thin DSAEK using an innovative artificial anterior chamber pressuriser: a proof-of-concept study

Authors: Alessandro Ruzza, Mohit Parekh, Luca Avoni, Gabriela Wojcik, Stefano Ferrari, Ludovic Desneux, Diego Ponzin, Hannah J. Levis, Vito Romano

Published in: Graefe's Archive for Clinical and Experimental Ophthalmology | Issue 7/2021

Login to get access

Abstract

Purpose

To report the impact of establishing and maintaining a high intracameral pressure (ICP) of 200 mmHg on UT-DSAEK graft preparation using an artificial anterior chamber pressuriser (ACP) control unit (Moria SA, Antony, France).

Method

Retrospective laboratory and clinical study. Four paired donor corneas were mounted on an artificial anterior chamber and subjected to 70 mmHg (“low”) and 200 mmHg (“high”) ICP using an ACP system. The central corneal thinning rate was measured after 5 min using AS-OCT and the endothelial cell viability was analysed using trypan blue and live/dead staining following 70 mmHg and 200 mmHg ICP. Visual outcomes and complications in a clinical case series of nine patients with bullous keratopathy who underwent UT-DSAEK using 200 mmHg ICP during graft preparation are reported.

Results

Laboratory outcomes showed 2 ± 1% and 2 ± 2% dead cells following 70 mmHg and 200 mmHg ICP respectively. Percentage viability in the 70 mmHg group (52.94 ± 5.88%) was not found to be significantly different (p = 0.7) compared to the 200 mmHg group (59.14 ± 10.43%). The mean corneal thinning rate after applying 200 mmHg ICP was 27 ± 13 μm/min centrally (7.2%/min). In the clinical case series, two cases were combined with cataract surgery. Re-bubbling rate was 11%. At the last follow-up (259 ± 109 days), graft thickness was 83 ± 22 μm centrally, endothelial cell density was 1175 ± 566 cell/mm2 and the BCVA of 0.08 ± 0.12 logMAR was recorded with no episodes of rejection.

Conclusion

ACP control unit for UT-DSAEK graft preparation helps in consistently obtaining UT-DSAEK grafts without compromising endothelial cell viability.
Literature
1.
go back to reference Stuart AJ, Romano V, Virgili G, Shortt AJ (2018) Descemet’s membrane endothelial keratoplasty (DMEK) versus Descemet’s stripping automated endothelial keratoplasty (DSAEK) for corneal endothelial failure. Cochrane Database Syst Rev 6:CD012097PubMed Stuart AJ, Romano V, Virgili G, Shortt AJ (2018) Descemet’s membrane endothelial keratoplasty (DMEK) versus Descemet’s stripping automated endothelial keratoplasty (DSAEK) for corneal endothelial failure. Cochrane Database Syst Rev 6:CD012097PubMed
2.
go back to reference Ruzza A, Parekh M, Salvalaio G et al (2015) (2015) Bubble technique for Descemet membrane endothelial keratoplasty tissue preparation in an eye bank: air or liquid? Acta Ophthalmol 93:e129-134CrossRef Ruzza A, Parekh M, Salvalaio G et al (2015) (2015) Bubble technique for Descemet membrane endothelial keratoplasty tissue preparation in an eye bank: air or liquid? Acta Ophthalmol 93:e129-134CrossRef
3.
go back to reference Busin M, Albé E (2014) Does thickness matter: ultrathin Descemet stripping automated endothelial keratoplasty. Curr Opin Ophthalmol 25:312–318CrossRef Busin M, Albé E (2014) Does thickness matter: ultrathin Descemet stripping automated endothelial keratoplasty. Curr Opin Ophthalmol 25:312–318CrossRef
4.
go back to reference Dickman MM, Cheng YY, Berendschot TT, van den Biggelaar FJ, Nuijts RM (2013) Effects of graft thickness and asymmetry on visual gain and aberrations after Descemet stripping automated endothelial keratoplasty. JAMA Ophthalmol 131:737–744. Erratum in: JAMA Ophthalmol 131:1102CrossRef Dickman MM, Cheng YY, Berendschot TT, van den Biggelaar FJ, Nuijts RM (2013) Effects of graft thickness and asymmetry on visual gain and aberrations after Descemet stripping automated endothelial keratoplasty. JAMA Ophthalmol 131:737–744. Erratum in: JAMA Ophthalmol 131:1102CrossRef
5.
go back to reference Gorovoy MS (2006) Descemet-stripping automated endothelial keratoplasty. Cornea 25:886–889CrossRef Gorovoy MS (2006) Descemet-stripping automated endothelial keratoplasty. Cornea 25:886–889CrossRef
6.
go back to reference Saunier V, Robinet Perrin A, Costet C, Touboul D (2016) Reproducibility of single-pass donor DSAEK tissue preparation with the MORIA single-use microkeratome. J Fr Ophtalmol 39:780–785CrossRef Saunier V, Robinet Perrin A, Costet C, Touboul D (2016) Reproducibility of single-pass donor DSAEK tissue preparation with the MORIA single-use microkeratome. J Fr Ophtalmol 39:780–785CrossRef
7.
go back to reference Parekh M, Ruzza A, Steger B et al (2019) Cross-country transportation efficacy and clinical outcomes of preloaded large-diameter ultra-thin Descemet stripping automated endothelial keratoplasty grafts. Cornea 38:30–34CrossRef Parekh M, Ruzza A, Steger B et al (2019) Cross-country transportation efficacy and clinical outcomes of preloaded large-diameter ultra-thin Descemet stripping automated endothelial keratoplasty grafts. Cornea 38:30–34CrossRef
8.
go back to reference Romano V, Steger B, Myneni J, Batterbury M, Willoughby CE, Kaye SB (2017) Preparation of ultrathin grafts for Descemet-stripping endothelial keratoplasty with a single microkeratome pass. J Cataract Refract Surg 43:12–15CrossRef Romano V, Steger B, Myneni J, Batterbury M, Willoughby CE, Kaye SB (2017) Preparation of ultrathin grafts for Descemet-stripping endothelial keratoplasty with a single microkeratome pass. J Cataract Refract Surg 43:12–15CrossRef
9.
go back to reference Parekh M, Peh G, Mehta JS et al (2019) Passaging capability of human corneal endothelial cells derived from old donors with and without accelerating cell attachment. Exp Eye Res 189:107814CrossRef Parekh M, Peh G, Mehta JS et al (2019) Passaging capability of human corneal endothelial cells derived from old donors with and without accelerating cell attachment. Exp Eye Res 189:107814CrossRef
10.
go back to reference Parekh M, Romano V, Ruzza A et al (2019) Culturing discarded peripheral human corneal endothelial cells from the tissues deemed for preloaded DMEK transplants. Cornea 38:1175–1181CrossRef Parekh M, Romano V, Ruzza A et al (2019) Culturing discarded peripheral human corneal endothelial cells from the tissues deemed for preloaded DMEK transplants. Cornea 38:1175–1181CrossRef
11.
go back to reference Romano V, Steger B, Chen JY et al (2015) Reliability of the effect of artificial anterior chamber pressure and corneal drying on corneal graft thickness. Cornea 34:866–869CrossRef Romano V, Steger B, Chen JY et al (2015) Reliability of the effect of artificial anterior chamber pressure and corneal drying on corneal graft thickness. Cornea 34:866–869CrossRef
12.
go back to reference Romano V, Parekh M, Ruzza A et al (2018) Comparison of preservation and transportation protocols for preloaded Descemet membrane endothelial keratoplasty. Br J Ophthalmol 102:549–555CrossRef Romano V, Parekh M, Ruzza A et al (2018) Comparison of preservation and transportation protocols for preloaded Descemet membrane endothelial keratoplasty. Br J Ophthalmol 102:549–555CrossRef
13.
go back to reference Romano V, Tey A, Hill NM et al (2015) Influence of graft size on graft survival following Descemet stripping automated endothelial keratoplasty. Br J Ophthalmol 99:784–788CrossRef Romano V, Tey A, Hill NM et al (2015) Influence of graft size on graft survival following Descemet stripping automated endothelial keratoplasty. Br J Ophthalmol 99:784–788CrossRef
14.
go back to reference Melamed S, Ben-Sira I, Ben-Shaul Y (1980) Corneal endothelial changes under induced intraocular pressure elevation: a scanning and transmission electron microscopic study in rabbits. Br J Ophthalmol 64:164–169CrossRef Melamed S, Ben-Sira I, Ben-Shaul Y (1980) Corneal endothelial changes under induced intraocular pressure elevation: a scanning and transmission electron microscopic study in rabbits. Br J Ophthalmol 64:164–169CrossRef
15.
go back to reference Bhogal MS, Allan BD (2012) Graft profile and thickness as a function of cut transition speed in Descemet-stripping automated endothelial keratoplasty. J Cataract Refract Surg 38:690–695CrossRef Bhogal MS, Allan BD (2012) Graft profile and thickness as a function of cut transition speed in Descemet-stripping automated endothelial keratoplasty. J Cataract Refract Surg 38:690–695CrossRef
16.
go back to reference Thomas PBM, Mukherjee AN, O’Donovan D, Rajan MS (2013) Preconditioned donor corneal thickness for microthin endothelial keratoplasty. Cornea 7:e173-178CrossRef Thomas PBM, Mukherjee AN, O’Donovan D, Rajan MS (2013) Preconditioned donor corneal thickness for microthin endothelial keratoplasty. Cornea 7:e173-178CrossRef
17.
go back to reference Ide T, Yoo SH, Kymionis GD et al (2008) Descemet-stripping automated endothelial keratoplasty: effect of anterior lamellar corneal tissue-on/-off storage condition on Descemet-stripping automated endothelial keratoplasty donor tissue. Cornea 27:754–757CrossRef Ide T, Yoo SH, Kymionis GD et al (2008) Descemet-stripping automated endothelial keratoplasty: effect of anterior lamellar corneal tissue-on/-off storage condition on Descemet-stripping automated endothelial keratoplasty donor tissue. Cornea 27:754–757CrossRef
18.
go back to reference Nishisako S, Aoki D, Sasaki C, Higa K, Shimazaki J (2018) Comparison of artificial anterior chamber internal pressures and cutting systems for Descemet’s stripping automated endothelial keratoplasty. Transl Vis Sci Technol 7:11CrossRef Nishisako S, Aoki D, Sasaki C, Higa K, Shimazaki J (2018) Comparison of artificial anterior chamber internal pressures and cutting systems for Descemet’s stripping automated endothelial keratoplasty. Transl Vis Sci Technol 7:11CrossRef
Metadata
Title
Ultra-thin DSAEK using an innovative artificial anterior chamber pressuriser: a proof-of-concept study
Authors
Alessandro Ruzza
Mohit Parekh
Luca Avoni
Gabriela Wojcik
Stefano Ferrari
Ludovic Desneux
Diego Ponzin
Hannah J. Levis
Vito Romano
Publication date
01-07-2021
Publisher
Springer Berlin Heidelberg
Keyword
Keratoplasty
Published in
Graefe's Archive for Clinical and Experimental Ophthalmology / Issue 7/2021
Print ISSN: 0721-832X
Electronic ISSN: 1435-702X
DOI
https://doi.org/10.1007/s00417-021-05194-0

Other articles of this Issue 7/2021

Graefe's Archive for Clinical and Experimental Ophthalmology 7/2021 Go to the issue