Iterative Algorithms Applied to Treated Intracranial Aneurysms

Authors: Aikaterini Fitsiori, Steve Philippe Martin, Alix Juillet De Saint Lager, Joanna Gariani, Karl-Olof Lovblad, Xavier Montet, Maria Isabel Vargas

Published in: Clinical Neuroradiology | Issue 4/2019

Login to get access

Abstract

Purpose

To investigate the impact of iterative metal artifact reduction (iMAR) on artifacts related to neurosurgical clips or endovascular coils when combined to filtered back projection (FBP) or advanced modelled iterative reconstruction (ADMIRE).

Material and Methods

In this study 21 unenhanced brain computed tomography (CT) examinations were reconstructed with FBP and level 2 of ADMIRE, both techniques with and without iMAR algorithm, resulting in 4 series per acquisition. Subjective assessment of artifact reduction was performed as a double-blinded evaluation with a 5-point-scale. Objective analysis was performed by comparing central tendencies and distributions of voxel densities. The central tendency was assessed as the mean voxel density in Hounsfield units. The distribution was assessed by evaluating the shape and asymmetry of the histograms of voxels densities with measures of kurtosis and skewness, respectively.

Results

Inter-reader agreement was excellent (>0.8). FBP and ADMIRE without iMAR were scored 4 and with iMAR 5. Unusual artifacts were noted in all of the series reconstructed with iMAR, especially when combined with ADMIRE. Kurtosis revealed statistical differences for all reconstruction techniques (p ≤ 0.0007) except for the association of FBP with iMAR (p = 0.2211) for the coiling population and skewness demonstrated no statistical difference in any population (p ≥ 0.0558), confirming the subjective analysis results, except for the ADMIRE algorithm with or without iMAR (p ≤ 0.0342) in the coiling population.

Conclusion

iMAR led to the reduction in artifacts due to intracranial metallic devices. However, it created a new artifact in the form of a halo of photon-starvation, especially when combined with ADMIRE. The combination of FBP and iMAR seems more suitable, combining the beneficial metal artifact reduction without the emergence of a halo of photon starvation just around the point of interest.
Literature
1.
go back to reference Bederson JB, Awad IA, Wiebers DO, Piepgras D, Haley EC Jr., Brott T, Hademenos G, Chyatte D, Rosenwasser R, Caroselli C. Recommendations for the management of patients with unruptured intracranial aneurysms: a statement for healthcare professionals from the Stroke Council of the American Heart Association. Stroke. 2000;31:2742–50.CrossRef Bederson JB, Awad IA, Wiebers DO, Piepgras D, Haley EC Jr., Brott T, Hademenos G, Chyatte D, Rosenwasser R, Caroselli C. Recommendations for the management of patients with unruptured intracranial aneurysms: a statement for healthcare professionals from the Stroke Council of the American Heart Association. Stroke. 2000;31:2742–50.CrossRef
2.
go back to reference Lanzino G, Murad MH, d’Urso PI, Rabinstein AA. Coil embolization versus clipping for ruptured intracranial aneurysms: a meta-analysis of prospective controlled published studies. AJNR Am J Neuroradiol. 2013;34:1764–8.CrossRef Lanzino G, Murad MH, d’Urso PI, Rabinstein AA. Coil embolization versus clipping for ruptured intracranial aneurysms: a meta-analysis of prospective controlled published studies. AJNR Am J Neuroradiol. 2013;34:1764–8.CrossRef
3.
go back to reference Li H, Pan R, Wang H, Rong X, Yin Z, Milgrom DP, Shi X, Tang Y, Peng Y. Clipping versus coiling for ruptured intracranial aneurysms: a systematic review and meta-analysis. Stroke. 2013;44:29–37.CrossRef Li H, Pan R, Wang H, Rong X, Yin Z, Milgrom DP, Shi X, Tang Y, Peng Y. Clipping versus coiling for ruptured intracranial aneurysms: a systematic review and meta-analysis. Stroke. 2013;44:29–37.CrossRef
4.
go back to reference Molyneux A, Kerr R, Stratton I, Sandercock P, Clarke M, Shrimpton J, Holman R. International Subarachnoid Aneurysm Trial Collaborative, International Subarachnoid Aneurysm Trial (ISAT) of neurosurgical clipping versus endovascular coiling in 2143 patients with ruptured intracranial aneurysms: a randomised trial. Lancet. 2002;360(9342):1267–74.CrossRef Molyneux A, Kerr R, Stratton I, Sandercock P, Clarke M, Shrimpton J, Holman R. International Subarachnoid Aneurysm Trial Collaborative, International Subarachnoid Aneurysm Trial (ISAT) of neurosurgical clipping versus endovascular coiling in 2143 patients with ruptured intracranial aneurysms: a randomised trial. Lancet. 2002;360(9342):1267–74.CrossRef
5.
go back to reference Molyneux AJ, Kerr RS, Birks J, Ramzi N, Yarnold J, Sneade M, Rischmiller J, ISAT Collaborators. Risk of recurrent subarachnoid haemorrhage, death, or dependence and standardised mortality ratios after clipping or coiling of an intracranial aneurysm in the International Subarachnoid Aneurysm Trial (ISAT): long-term follow-up. Lancet Neurol. 2009;8:427–33.CrossRef Molyneux AJ, Kerr RS, Birks J, Ramzi N, Yarnold J, Sneade M, Rischmiller J, ISAT Collaborators. Risk of recurrent subarachnoid haemorrhage, death, or dependence and standardised mortality ratios after clipping or coiling of an intracranial aneurysm in the International Subarachnoid Aneurysm Trial (ISAT): long-term follow-up. Lancet Neurol. 2009;8:427–33.CrossRef
6.
go back to reference Choudhri O, Mukerji N, Steinberg GK. Combined endovascular and microsurgical management of complex cerebral aneurysms. Front Neurol. 2013;4:108.CrossRef Choudhri O, Mukerji N, Steinberg GK. Combined endovascular and microsurgical management of complex cerebral aneurysms. Front Neurol. 2013;4:108.CrossRef
7.
go back to reference Glover GH, Pelc NJ. An algorithm for the reduction of metal clip artifacts in CT reconstructions. Med Phys. 1981;8:799–807.CrossRef Glover GH, Pelc NJ. An algorithm for the reduction of metal clip artifacts in CT reconstructions. Med Phys. 1981;8:799–807.CrossRef
8.
go back to reference Kalender WA, Hebel R, Ebersberger J. Reduction of CT artifacts caused by metallic implants. Radiology. 1987;164:576–7.CrossRef Kalender WA, Hebel R, Ebersberger J. Reduction of CT artifacts caused by metallic implants. Radiology. 1987;164:576–7.CrossRef
9.
go back to reference Klotz E, Kalender WA, Sokiransky R, Felsenberg D. Algorithms for the reduction of CT artifacts caused by metallic implants. Medical Imaging ′90, SPIE. 1990. p. 9. Klotz E, Kalender WA, Sokiransky R, Felsenberg D. Algorithms for the reduction of CT artifacts caused by metallic implants. Medical Imaging ′90, SPIE. 1990. p. 9.
10.
go back to reference Prell D, Kyriakou Y, Struffert T, Dörfler A, Kalender WA. Metal artifact reduction for clipping and coiling in interventional C‑arm CT. AJNR Am J Neuroradiol. 2010;31:634–9.CrossRef Prell D, Kyriakou Y, Struffert T, Dörfler A, Kalender WA. Metal artifact reduction for clipping and coiling in interventional C‑arm CT. AJNR Am J Neuroradiol. 2010;31:634–9.CrossRef
11.
go back to reference Chintalapani G, Chinnadurai P, Srinivasan V, Chen SR, Shaltoni H, Morsi H, Mawad ME, Kan P. Evaluation of C‑arm CT metal artifact reduction algorithm during intra-aneurysmal coil embolization: Assessment of brain parenchyma, stents and flow-diverters. Eur J Radiol. 2016;85:1312–21.CrossRef Chintalapani G, Chinnadurai P, Srinivasan V, Chen SR, Shaltoni H, Morsi H, Mawad ME, Kan P. Evaluation of C‑arm CT metal artifact reduction algorithm during intra-aneurysmal coil embolization: Assessment of brain parenchyma, stents and flow-diverters. Eur J Radiol. 2016;85:1312–21.CrossRef
12.
go back to reference Pjontek R, Önenköprülü B, Scholz B, Kyriakou Y, Schubert GA, Nikoubashman O, Othman A, Wiesmann M, Brockmann MA. Metal artifact reduction for flat panel detector intravenous CT angiography in patients with intracranial metallic implants after endovascular and surgical treatment. J Neurointerv Surg. 2016;8:824–9.CrossRef Pjontek R, Önenköprülü B, Scholz B, Kyriakou Y, Schubert GA, Nikoubashman O, Othman A, Wiesmann M, Brockmann MA. Metal artifact reduction for flat panel detector intravenous CT angiography in patients with intracranial metallic implants after endovascular and surgical treatment. J Neurointerv Surg. 2016;8:824–9.CrossRef
13.
go back to reference Psychogios MN, Scholz B, Rohkohl C, Kyriakou Y, Mohr A, Schramm P, Wachter D, Wasser K, Knauth M. Impact of a new metal artefact reduction algorithm in the noninvasive follow-up of intracranial clips, coils, and stents with flat-panel angiographic CTA: initial results. Neuroradiology. 2013;55:813–8.CrossRef Psychogios MN, Scholz B, Rohkohl C, Kyriakou Y, Mohr A, Schramm P, Wachter D, Wasser K, Knauth M. Impact of a new metal artefact reduction algorithm in the noninvasive follow-up of intracranial clips, coils, and stents with flat-panel angiographic CTA: initial results. Neuroradiology. 2013;55:813–8.CrossRef
14.
go back to reference van der Bom IM, Hou SY, Puri AS, Spilberg G, Ruijters D, van de Haar P, Carelsen B, Vedantham S, Gounis MJ, Wakhloo AK. Reduction of coil mass artifacts in high-resolution flat detector conebeam CT of cerebral stent-assisted coiling. AJNR Am J Neuroradiol. 2013;34:2163–70.CrossRef van der Bom IM, Hou SY, Puri AS, Spilberg G, Ruijters D, van de Haar P, Carelsen B, Vedantham S, Gounis MJ, Wakhloo AK. Reduction of coil mass artifacts in high-resolution flat detector conebeam CT of cerebral stent-assisted coiling. AJNR Am J Neuroradiol. 2013;34:2163–70.CrossRef
15.
go back to reference Wuest W, May MS, Brand M, Bayerl N, Krauss A, Uder M, Lell M. Improved image quality in head and neck CT using a 3D iterative approach to reduce metal artifact. AJNR Am J Neuroradiol. 2015;36:1988–93.CrossRef Wuest W, May MS, Brand M, Bayerl N, Krauss A, Uder M, Lell M. Improved image quality in head and neck CT using a 3D iterative approach to reduce metal artifact. AJNR Am J Neuroradiol. 2015;36:1988–93.CrossRef
16.
go back to reference Dolati P, Eichberg D, Wong JH, Goyal M. The utility of dual-energy computed tomographic angiography for the evaluation of brain aneurysms after surgical clipping: a prospective study. World Neurosurg. 2015;84:1362–71.CrossRef Dolati P, Eichberg D, Wong JH, Goyal M. The utility of dual-energy computed tomographic angiography for the evaluation of brain aneurysms after surgical clipping: a prospective study. World Neurosurg. 2015;84:1362–71.CrossRef
17.
go back to reference Fahrendorf DM, Goericke SL, Oezkan N, Breyer T, Hussain S, Sandalcioglu EI, Sure U, Forsting M, Gizewski ER. The value of dual-energy CTA for control of surgically clipped aneurysms. Eur Radiol. 2011;21:2193–201.CrossRef Fahrendorf DM, Goericke SL, Oezkan N, Breyer T, Hussain S, Sandalcioglu EI, Sure U, Forsting M, Gizewski ER. The value of dual-energy CTA for control of surgically clipped aneurysms. Eur Radiol. 2011;21:2193–201.CrossRef
18.
go back to reference Shirasaka T, Hiwatashi A, Yamashita K, Kondo M, Hamasaki H, Shimomiya Y, Nakamura Y, Funama Y, Honda H. Optimal scan timing for artery-vein separation at whole-brain CT angiography using a 320-row MDCT volume scanner. Br J Radiol. 2017;90(1070):20160634.CrossRef Shirasaka T, Hiwatashi A, Yamashita K, Kondo M, Hamasaki H, Shimomiya Y, Nakamura Y, Funama Y, Honda H. Optimal scan timing for artery-vein separation at whole-brain CT angiography using a 320-row MDCT volume scanner. Br J Radiol. 2017;90(1070):20160634.CrossRef
19.
go back to reference Sasaki M, Kudo K, Ogasawara K, Fujiwara S. Tracer delay-insensitive algorithm can improve reliability of CT perfusion imaging for cerebrovascular steno-occlusive disease: comparison with quantitative single-photon emission CT. AJNR Am J Neuroradiol. 2009;30:188–93.CrossRef Sasaki M, Kudo K, Ogasawara K, Fujiwara S. Tracer delay-insensitive algorithm can improve reliability of CT perfusion imaging for cerebrovascular steno-occlusive disease: comparison with quantitative single-photon emission CT. AJNR Am J Neuroradiol. 2009;30:188–93.CrossRef
20.
go back to reference Jia Y, Zhang J, Fan J, Li C, Sun Y, Li D, Xiao X. Gemstone spectral imaging reduced artefacts from metal coils or clips after treatment of cerebral aneurysms: a retrospective study of 35 patients. Br J Radiol. 2015;88(1055):20150222.CrossRef Jia Y, Zhang J, Fan J, Li C, Sun Y, Li D, Xiao X. Gemstone spectral imaging reduced artefacts from metal coils or clips after treatment of cerebral aneurysms: a retrospective study of 35 patients. Br J Radiol. 2015;88(1055):20150222.CrossRef
21.
go back to reference Shinohara Y, Sakamoto M, Iwata N, Kishimoto J, Kuya K, Fujii S, Kaminou T, Watanabe T, Ogawa T. Usefulness of monochromatic imaging with metal artifact reduction software for computed tomography angiography after intracranial aneurysm coil embolization. Acta Radiol. 2014;55:1015–23.CrossRef Shinohara Y, Sakamoto M, Iwata N, Kishimoto J, Kuya K, Fujii S, Kaminou T, Watanabe T, Ogawa T. Usefulness of monochromatic imaging with metal artifact reduction software for computed tomography angiography after intracranial aneurysm coil embolization. Acta Radiol. 2014;55:1015–23.CrossRef
22.
go back to reference Brown JH, Lustrin ES, Lev MH, Ogilvy CS, Taveras JM. Reduction of aneurysm clip artifacts on CT angiograms: a technical note. AJNR Am J Neuroradiol. 1999;20:694–6.PubMed Brown JH, Lustrin ES, Lev MH, Ogilvy CS, Taveras JM. Reduction of aneurysm clip artifacts on CT angiograms: a technical note. AJNR Am J Neuroradiol. 1999;20:694–6.PubMed
Metadata
Title
Iterative Algorithms Applied to Treated Intracranial Aneurysms
Authors
Aikaterini Fitsiori
Steve Philippe Martin
Alix Juillet De Saint Lager
Joanna Gariani
Karl-Olof Lovblad
Xavier Montet
Maria Isabel Vargas
Publication date
01-12-2019
Publisher
Springer Berlin Heidelberg
Published in
Clinical Neuroradiology / Issue 4/2019
Print ISSN: 1869-1439
Electronic ISSN: 1869-1447
DOI
https://doi.org/10.1007/s00062-018-0701-5

Other articles of this Issue 4/2019

Clinical Neuroradiology 4/2019 Go to the issue

Information

Information