Inter-scan reproducibility of coronary calcium measurement using Multi Detector-Row Computed Tomography (MDCT)

Authors: Siamak Sabour, A. Rutten, Y. T. van der Schouw, F. Atsma, D. E. Grobbee, W. P. Mali, M. E. L. Bartelink, M. L. Bots, M. Prokop

Published in: European Journal of Epidemiology | Issue 4/2007

Login to get access

Abstract

Purpose

To assess inter-scan reproducibility of coronary calcium measurements obtained from Multi Detector-Row CT (MDCT) images and to evaluate whether this reproducibility is affected by different measurement protocols, slice thickness, cardiovascular risk factors and/or technical variables.

Design

Cross-sectional study with repeated measurements.

Materials and methods

The study population comprised 76 healthy women. Coronary calcium was assessed in these women twice in one session using 16-MDCT (Philips Mx 8000 IDT 16). Images were reconstructed with 1.5 mm slice thickness and 3.0 mm slice thickness. The 76 repeated scans were scored. The Agatston score, a volume measurement and a mass measurement were assessed. Reproducibility was determined by estimation of mean, absolute, relative difference, the weighted kappa value for agreement and the Intra-class correlation coefficient (ICCC).

Results

Fifty-five participants (72.4%) had a coronary calcification of more than zero in Agatston (1.5 mm slice thickness). The reproducibility of coronary calcium measurements between scans in terms of ranking was excellent with Intra-class correlation coefficients of >0.98, and kappa values above 0.80. The absolute difference in calcium score between scans increased with increasing calcium levels, indicating that measurement error increases with increasing calcium levels. However, no relation was found between the mean difference in scores and calcium levels, indicating that the increase in measurement error is likely to result in random misclassification in calcium score. Reproducibility results were similar for 1.5 mm slices and for 3.0 mm slices, and equal for Agatston, volume and mass measurements.

Conclusion

Inter-scan reproducibilility of measurement of coronary calcium using images from MDCT is excellent, irrespective of slice thickness and type of calcium parameter.
Literature
1.
go back to reference Masuda Y, Naito S, Aoyagi Y, Yamada Z, Uda T, Morooka N, et al. Coronary artery calcification detected by CT: clinical significance and angiographic correlates. Angiology 1990;41(12):1037–47PubMed Masuda Y, Naito S, Aoyagi Y, Yamada Z, Uda T, Morooka N, et al. Coronary artery calcification detected by CT: clinical significance and angiographic correlates. Angiology 1990;41(12):1037–47PubMed
2.
go back to reference Beadenkopf WG, Daoud AS, Love BM. Calcification in the coronary arteries and its relationship to arteriosclerosis and myocardial infarction. Am J Roentgenol Radium Ther Nucl Med 1964;92:865–71PubMed Beadenkopf WG, Daoud AS, Love BM. Calcification in the coronary arteries and its relationship to arteriosclerosis and myocardial infarction. Am J Roentgenol Radium Ther Nucl Med 1964;92:865–71PubMed
3.
go back to reference Warburton RK, Tampas JP, Soule AB, Taylor HC III. Coronary artery calcification: its relationship to coronary artery stenosis and myocardial infarction. Radiology 1968;91(1):109–15PubMed Warburton RK, Tampas JP, Soule AB, Taylor HC III. Coronary artery calcification: its relationship to coronary artery stenosis and myocardial infarction. Radiology 1968;91(1):109–15PubMed
4.
go back to reference Frink RJ, Achor RW, Brown AL Jr, Kincaid OW, Brandenburg RO. Significance of calcification of the coronary arteries. Am J Cardiol 1970;26(3):241–7PubMedCrossRef Frink RJ, Achor RW, Brown AL Jr, Kincaid OW, Brandenburg RO. Significance of calcification of the coronary arteries. Am J Cardiol 1970;26(3):241–7PubMedCrossRef
5.
go back to reference Wang S, Detrano RC, Secci A, Tang W, Doherty TM, Puentes G, et al. Detection of coronary calcification with electron-beam computed tomography: evaluation of interexamination reproducibility and comparison of three image-acquisition protocols. Am Heart J 1996;132(3):550–8PubMedCrossRef Wang S, Detrano RC, Secci A, Tang W, Doherty TM, Puentes G, et al. Detection of coronary calcification with electron-beam computed tomography: evaluation of interexamination reproducibility and comparison of three image-acquisition protocols. Am Heart J 1996;132(3):550–8PubMedCrossRef
6.
go back to reference Nasir K, Budoff MJ, Post WS, Fishman EK, Mahesh M, Lima JA, et al. Electron beam CT versus helical CT scans for assessing coronary calcification: current utility and future directions. Am Heart J 2003;146(6):969–77PubMedCrossRef Nasir K, Budoff MJ, Post WS, Fishman EK, Mahesh M, Lima JA, et al. Electron beam CT versus helical CT scans for assessing coronary calcification: current utility and future directions. Am Heart J 2003;146(6):969–77PubMedCrossRef
7.
go back to reference Thompson BH, Stanford W. Imaging of coronary calcium: a case for electron beam computed tomography. J Thorac Imaging 2001;16(1):8–15PubMedCrossRef Thompson BH, Stanford W. Imaging of coronary calcium: a case for electron beam computed tomography. J Thorac Imaging 2001;16(1):8–15PubMedCrossRef
8.
go back to reference Knez A, Becker C, Becker A, Leber A, White C, Reiser M, et al. Determination of coronary calcium with multi-slice spiral computed tomography: a comparative study with electron-beam CT. Int J Cardiovasc Imaging 2002;18(4):295–303PubMedCrossRef Knez A, Becker C, Becker A, Leber A, White C, Reiser M, et al. Determination of coronary calcium with multi-slice spiral computed tomography: a comparative study with electron-beam CT. Int J Cardiovasc Imaging 2002;18(4):295–303PubMedCrossRef
9.
go back to reference Stanford W, Thompson BH, Burns TL, Heery SD, Burr MC. Coronary artery calcium quantification at multi-detector row helical CT versus electron-beam CT. Radiology 2004;230(2):397–402PubMedCrossRef Stanford W, Thompson BH, Burns TL, Heery SD, Burr MC. Coronary artery calcium quantification at multi-detector row helical CT versus electron-beam CT. Radiology 2004;230(2):397–402PubMedCrossRef
10.
go back to reference Kopp AF, Ohnesorge B, Becker C, Schroder S, Heuschmid M, Kuttner A, et al. Reproducibility and accuracy of coronary calcium measurements with multi-detector row versus electron-beam CT. Radiology 2002;225(1):113–9PubMedCrossRef Kopp AF, Ohnesorge B, Becker C, Schroder S, Heuschmid M, Kuttner A, et al. Reproducibility and accuracy of coronary calcium measurements with multi-detector row versus electron-beam CT. Radiology 2002;225(1):113–9PubMedCrossRef
11.
go back to reference Ohnesorge B, Flohr T, Fischbach R, Kopp AF, Knez A, Schroder S, et al. Reproducibility of coronary calcium quantification in repeat examinations with retrospectively ECG-gated multisection spiral CT. Eur Radiol 2002;12(6):1532–40PubMedCrossRef Ohnesorge B, Flohr T, Fischbach R, Kopp AF, Knez A, Schroder S, et al. Reproducibility of coronary calcium quantification in repeat examinations with retrospectively ECG-gated multisection spiral CT. Eur Radiol 2002;12(6):1532–40PubMedCrossRef
12.
go back to reference Boker LK, van Noord PA, van der Schouw YT, Koot NV, Bueno de Mesquita HB, Riboli E, et al. Prospect-EPIC Utrecht: study design and characteristics of the cohort population. European Prospective Investigation into Cancer and Nutrition. Eur J Epidemiol 2001;17(11):1047–53PubMedCrossRef Boker LK, van Noord PA, van der Schouw YT, Koot NV, Bueno de Mesquita HB, Riboli E, et al. Prospect-EPIC Utrecht: study design and characteristics of the cohort population. European Prospective Investigation into Cancer and Nutrition. Eur J Epidemiol 2001;17(11):1047–53PubMedCrossRef
13.
go back to reference Agatston AS, Janowitz WR, Hildner FJ, Zusmer NR, Viamonte M Jr, Detrano R. Quantification of coronary artery calcium using ultrafast computed tomography. J Am Coll Cardiol 1990;15(4):827–32PubMedCrossRef Agatston AS, Janowitz WR, Hildner FJ, Zusmer NR, Viamonte M Jr, Detrano R. Quantification of coronary artery calcium using ultrafast computed tomography. J Am Coll Cardiol 1990;15(4):827–32PubMedCrossRef
14.
go back to reference Hong C, Bae KT, Pilgram TK. Coronary artery calcium: accuracy and reproducibility of measurements with multi-detector row CT––assessment of effects of different thresholds and quantification methods. Radiology 2003;227(3):795–801PubMedCrossRef Hong C, Bae KT, Pilgram TK. Coronary artery calcium: accuracy and reproducibility of measurements with multi-detector row CT––assessment of effects of different thresholds and quantification methods. Radiology 2003;227(3):795–801PubMedCrossRef
15.
go back to reference Rumberger JA, Brundage BH, Rader DJ, Kondos G. Electron beam computed tomographic coronary calcium scanning: a review and guidelines for use in asymptomatic persons. Mayo Clin Proc 1999;74(3):243–52PubMedCrossRef Rumberger JA, Brundage BH, Rader DJ, Kondos G. Electron beam computed tomographic coronary calcium scanning: a review and guidelines for use in asymptomatic persons. Mayo Clin Proc 1999;74(3):243–52PubMedCrossRef
16.
go back to reference Reilly MP, Wolfe ML, Localio AR, Rader DJ. Coronary artery calcification and cardiovascular risk factors: impact of the analytic approach. Atherosclerosis 2004;173(1):69–78PubMedCrossRef Reilly MP, Wolfe ML, Localio AR, Rader DJ. Coronary artery calcification and cardiovascular risk factors: impact of the analytic approach. Atherosclerosis 2004;173(1):69–78PubMedCrossRef
17.
go back to reference Rumberger JA, Kaufman L. A rosetta stone for coronary calcium risk stratification: agatston, volume, and mass scores in 11,490 individuals. AJR Am J Roentgenol 2003;181(3):743–8PubMed Rumberger JA, Kaufman L. A rosetta stone for coronary calcium risk stratification: agatston, volume, and mass scores in 11,490 individuals. AJR Am J Roentgenol 2003;181(3):743–8PubMed
18.
go back to reference Kaufmann RB, Sheedy PF, Breen JF, Kelzenberg JR, Kruger BL, Schwartz RS, et al. Detection of heart calcification with electron beam CT: interobserver and intraobserver reliability for scoring quantification. Radiology 1994;190(2):347–52PubMed Kaufmann RB, Sheedy PF, Breen JF, Kelzenberg JR, Kruger BL, Schwartz RS, et al. Detection of heart calcification with electron beam CT: interobserver and intraobserver reliability for scoring quantification. Radiology 1994;190(2):347–52PubMed
19.
go back to reference Hernigou A, Challande P, Boudeville JC, Sene V, Grataloup C, Plainfosse MC. Reproducibility of coronary calcification detection with electron-beam computed tomography. Eur Radiol 1996;6(2):210–6PubMedCrossRef Hernigou A, Challande P, Boudeville JC, Sene V, Grataloup C, Plainfosse MC. Reproducibility of coronary calcification detection with electron-beam computed tomography. Eur Radiol 1996;6(2):210–6PubMedCrossRef
20.
go back to reference Bielak LF, Kaufmann RB, Moll PP, McCollough CH, Schwartz RS, Sheedy PF. Small lesions in the heart identified at electron beam CT: calcification or noise? Radiology 1994;192(3):631–6PubMed Bielak LF, Kaufmann RB, Moll PP, McCollough CH, Schwartz RS, Sheedy PF. Small lesions in the heart identified at electron beam CT: calcification or noise? Radiology 1994;192(3):631–6PubMed
Metadata
Title
Inter-scan reproducibility of coronary calcium measurement using Multi Detector-Row Computed Tomography (MDCT)
Authors
Siamak Sabour
A. Rutten
Y. T. van der Schouw
F. Atsma
D. E. Grobbee
W. P. Mali
M. E. L. Bartelink
M. L. Bots
M. Prokop
Publication date
01-04-2007
Publisher
Kluwer Academic Publishers
Published in
European Journal of Epidemiology / Issue 4/2007
Print ISSN: 0393-2990
Electronic ISSN: 1573-7284
DOI
https://doi.org/10.1007/s10654-007-9123-4

Other articles of this Issue 4/2007

European Journal of Epidemiology 4/2007 Go to the issue