01-06-2017 | Retinal Disorders
Inter-method agreement in retinal blood vessels diameter analysis between Dynamic Vessel Analyzer and optical coherence tomography
Authors:
Lucia Benatti, Federico Corvi, Livia Tomasso, Stefano Mercuri, Lea Querques, Fulvio Ricceri, Francesco Bandello, Giuseppe Querques
Published in:
Graefe's Archive for Clinical and Experimental Ophthalmology
|
Issue 6/2017
Login to get access
Abstract
Purpose
To analyze the inter-methods agreement in arteriovenous ratio (AVR) evaluation between spectral-domain optical coherence tomography (SD-OCT) and Dynamic Vessel Analyzer (DVA).
Methods
Healthy volunteers underwent DVA and SD-OCT examination. AVR was measured by SD-OCT using the four external lines of the optic nerve head-centered 7-line cube and by DVA using an automated AVR estimation. The mean AVR was calculated, twice, separately by two independent readers for each tool.
Results
Twenty-two eyes of 11 healthy subjects (five women and six men, mean age 35) were included. AVR analysis by DVA showed high inter-observer agreement between reader 1 and 2, and high intra-observer agreement for both reader 1 and reader 2. With regard to AVR analysis on SD-OCT, we found high inter-observer agreement between reader 1 and 2, and low intra-observer agreement for reader 2 but high intra-observer agreement for reader 1. Overall, the mean AVR measured on SD-OCT turned out to be significantly higher than mean AVR measured through DVA (reader 1, 0.9023 ± 0.06 vs 0.8036 ± 0.08; p < 0.001, and reader 2, 0.9067 ± 0.06 vs 0.8083 ± 0.05; p= 0.003).
Conclusions
No inter-method agreement in AVR could be detected in the present study due to bias in measurements (shift between DVA and SD-OCT). We found significant difference in the two noninvasive methods for AVR measurement, with a tendency for SD-OCT to overestimate retinal vascular caliber in comparison to DVA. This may be useful for achieving greater accuracy in the evaluation of retinal vessel in ocular as well as systemic diseases.