Published in: BMC Infectious Diseases 1/2005

Open Access 01-12-2005 | Research article

Insufficient neutralization in testing a chlorhexidine-containing ethanol-based hand rub can result in a false positive efficacy assessment

Authors: Günter Kampf, Marc Shaffer, Corrine Hunte

Published in: BMC Infectious Diseases | Issue 1/2005

Login to get access

Abstract

Background

Effective neutralization in testing hand hygiene preparations is considered to be a crucial element to ensure validity of the test results, especially with the difficulty to neutralize chlorhexidine gluconate. Aim of the study was to measure the effect of chemical neutralization under practical test conditions according to EN 1500.

Methods

We have investigated two ethanol-based hand rubs (product A, based on 61% ethanol and 1% chlorhexidine gluconate; product B, based on 85% ethanol). The efficacy of products (application of 3 ml for 30 s) was compared to 2-propanol 60% (v/v) (two 3 ml rubs of 30 s each) on hands artificially contaminated with Escherichia coli using a cross-over design with 15 volunteers. Pre-values were obtained by rubbing fingertips for 1 minute in liquid broth. Post-values were determined by sampling immediately after disinfection in liquid broth with and without neutralizers (0.5% lecithin, 4% polysorbate 20).

Results

The neutralizers were found to be effective and non-toxic. Without neutralization in the sampling fluid, the reference disinfection reduced the test bacteria by 3.7 log10, product B by 3.3 log10 and product A by 4.8 log10 (P = 0.001; ANOVA). With neutralization the reference disinfection reduced the test bacteria by 3.5 log10, product B by 3.3 log10 and product A by 2.7 log10 (P = 0.011; ANOVA). In comparison to the reference treatment Product B lead to a lower mean reduction than the reference disinfection but the difference was not significant (P > 0.1; Wilcoxon-Wilcox test). Without neutralizing agents in the sampling fluid, product A yielded a significantly higher reduction of test bacteria (4.8; P = 0.02) as compared to the situation with neutralizing agents (2.7; P = 0.033).

Conclusion

The crucial step of neutralization lies in the sampling fluid itself in order to stop any residual bacteriostatic or bactericidal activity immediately after the application of the preparation, especially with chlorhexidine gluconate-containing preparations. This is particularly important at short application times such as the 30 s.
Literature
1.
go back to reference Boyce JM, Pittet D: Guideline for hand hygiene in health-care settings. Recommendations of the healthcare infection control practices advisory committee and the HICPAC/SHEA/APIC/IDSA hand hygiene task force. MMWR - Morbidity & Mortality Weekly Report. 2002, 51: 1-45. Boyce JM, Pittet D: Guideline for hand hygiene in health-care settings. Recommendations of the healthcare infection control practices advisory committee and the HICPAC/SHEA/APIC/IDSA hand hygiene task force. MMWR - Morbidity & Mortality Weekly Report. 2002, 51: 1-45.
2.
go back to reference Kampf G, Kramer A: Epidemiologic background of hand hygiene and evaluation of the most important agents for scrubs and rubs. Clinical Microbiology Reviews. 2004, 17: 863-893. 10.1128/CMR.17.4.863-893.2004.CrossRefPubMedPubMedCentral Kampf G, Kramer A: Epidemiologic background of hand hygiene and evaluation of the most important agents for scrubs and rubs. Clinical Microbiology Reviews. 2004, 17: 863-893. 10.1128/CMR.17.4.863-893.2004.CrossRefPubMedPubMedCentral
3.
go back to reference Kampf G, Rudolf M, Labadie JC, Barrett SP: Spectrum of antimicrobial activity and user acceptability of the hand disinfectant agent Sterillium Gel. Journal of Hospital Infection. 2002, 52: 141-147. 10.1053/jhin.2002.1281.CrossRefPubMed Kampf G, Rudolf M, Labadie JC, Barrett SP: Spectrum of antimicrobial activity and user acceptability of the hand disinfectant agent Sterillium Gel. Journal of Hospital Infection. 2002, 52: 141-147. 10.1053/jhin.2002.1281.CrossRefPubMed
4.
go back to reference Kramer A, Rudolph P, Kampf G, Pittet D: Limited efficacy of alcohol-based hand gels. The Lancet. 2002, 359: 1489-1490. 10.1016/S0140-6736(02)08426-X.CrossRef Kramer A, Rudolph P, Kampf G, Pittet D: Limited efficacy of alcohol-based hand gels. The Lancet. 2002, 359: 1489-1490. 10.1016/S0140-6736(02)08426-X.CrossRef
5.
go back to reference Kampf G, Meyer B, Goroncy-Bermes P: Comparison of two test methods for the determination of sufficient antimicrobial efficacy of three different alcohol-based hand rubs for hygienic hand disinfection. Journal of Hospital Infection. 2003, 55: 220-225. 10.1016/S0195-6701(03)00255-X.CrossRefPubMed Kampf G, Meyer B, Goroncy-Bermes P: Comparison of two test methods for the determination of sufficient antimicrobial efficacy of three different alcohol-based hand rubs for hygienic hand disinfection. Journal of Hospital Infection. 2003, 55: 220-225. 10.1016/S0195-6701(03)00255-X.CrossRefPubMed
6.
go back to reference Anonymous: EN 1500. Chemical disinfectants and antiseptics. Hygienic handrub. Test method and requirement (phase 2, step 2). 1997 Anonymous: EN 1500. Chemical disinfectants and antiseptics. Hygienic handrub. Test method and requirement (phase 2, step 2). 1997
7.
go back to reference Werner HP, Engelhardt C: Problematik der Inaktivierung am Beispiel des in vitro-Tests. Hygiene + Medizin. 1978, 3: 326-330. Werner HP, Engelhardt C: Problematik der Inaktivierung am Beispiel des in vitro-Tests. Hygiene + Medizin. 1978, 3: 326-330.
8.
go back to reference Sheikh W: Development and validation of a neutralizer system for in vitro evaluation of some antiseptics. Antimicrobial Agents and Chemotherapy. 1981, 19: 429-434.CrossRefPubMedPubMedCentral Sheikh W: Development and validation of a neutralizer system for in vitro evaluation of some antiseptics. Antimicrobial Agents and Chemotherapy. 1981, 19: 429-434.CrossRefPubMedPubMedCentral
9.
go back to reference Kampf G, Höfer M, Rüden H: Inaktivierung von Chlorhexidin bei der in vitro Desinfektionsmitteltestung. Zentralblatt für Hygiene und Umweltmedizin. 1998, 200: 457-464.PubMed Kampf G, Höfer M, Rüden H: Inaktivierung von Chlorhexidin bei der in vitro Desinfektionsmitteltestung. Zentralblatt für Hygiene und Umweltmedizin. 1998, 200: 457-464.PubMed
10.
go back to reference Kampf G, Höfer M, Wendt C: Efficacy of hand disinfectants against vancomycin-resistant enterococci in vitro. Journal of Hospital Infection. 1999, 42: 143-150. 10.1053/jhin.1998.0559.CrossRefPubMed Kampf G, Höfer M, Wendt C: Efficacy of hand disinfectants against vancomycin-resistant enterococci in vitro. Journal of Hospital Infection. 1999, 42: 143-150. 10.1053/jhin.1998.0559.CrossRefPubMed
11.
go back to reference Shimizu M, Okuzumi K, Yoneyama A, Kunisada T, Araake M, Ogawa H, Kimura S: In vitro antiseptic susceptibility of clinical isolates from nosocomial infections. Dermatology. 2002, 204: 21-27. 10.1159/000057720.CrossRefPubMed Shimizu M, Okuzumi K, Yoneyama A, Kunisada T, Araake M, Ogawa H, Kimura S: In vitro antiseptic susceptibility of clinical isolates from nosocomial infections. Dermatology. 2002, 204: 21-27. 10.1159/000057720.CrossRefPubMed
12.
go back to reference Kampf G, Ostermeyer C: Intra-laboratory reproducibility of the hand hygiene reference procedures of EN 1499 (hygienic hand wash) and EN 1500 (hygienic hand disinfection). Journal of Hospital Infection. 2002, 52: 219-224. 10.1053/jhin.2002.1299.CrossRefPubMed Kampf G, Ostermeyer C: Intra-laboratory reproducibility of the hand hygiene reference procedures of EN 1499 (hygienic hand wash) and EN 1500 (hygienic hand disinfection). Journal of Hospital Infection. 2002, 52: 219-224. 10.1053/jhin.2002.1299.CrossRefPubMed
13.
go back to reference Kampf G, Ostermeyer C: Inter-laboratory reproducibility of the EN 1500 reference hand disinfection. Journal of Hospital Infection. 2003, 53: 304-306. 10.1053/jhin.2002.1357.CrossRefPubMed Kampf G, Ostermeyer C: Inter-laboratory reproducibility of the EN 1500 reference hand disinfection. Journal of Hospital Infection. 2003, 53: 304-306. 10.1053/jhin.2002.1357.CrossRefPubMed
14.
go back to reference Anonymous: ASTM E 1054 - 02. Standard test method for evaluation of inactivators of antimicrobial agents. 2002 Anonymous: ASTM E 1054 - 02. Standard test method for evaluation of inactivators of antimicrobial agents. 2002
15.
go back to reference Kampf G, Ostermeyer C: Efficacy of two distinct ethanol-based hand rubs for surgical hand disinfection - a controlled trial according to prEN 12791. BMC Infectious Disease. 2005, 5: 17-10.1186/1471-2334-5-17.CrossRef Kampf G, Ostermeyer C: Efficacy of two distinct ethanol-based hand rubs for surgical hand disinfection - a controlled trial according to prEN 12791. BMC Infectious Disease. 2005, 5: 17-10.1186/1471-2334-5-17.CrossRef
16.
go back to reference Benson L, Bush L, LeBlanc D: Importance of neutralizers in the stripping fluid in a simulated healthcare personnel handwash. Infection Control and Hospital Epidemiology. 1990, 11: 595-599.CrossRefPubMed Benson L, Bush L, LeBlanc D: Importance of neutralizers in the stripping fluid in a simulated healthcare personnel handwash. Infection Control and Hospital Epidemiology. 1990, 11: 595-599.CrossRefPubMed
17.
go back to reference Anonymous: prEN 12054. Chemical disinfectants and antiseptics. Quantitative suspension test for the evaluation of bactericidal activity of products for hygienic and surgical handrub and handwash used in human medicine. Test method and requirements (phase 2, step 1). 1997 Anonymous: prEN 12054. Chemical disinfectants and antiseptics. Quantitative suspension test for the evaluation of bactericidal activity of products for hygienic and surgical handrub and handwash used in human medicine. Test method and requirements (phase 2, step 1). 1997
18.
go back to reference Kampf G, Hollingsworth A: Validity of the four European test strains of prEN 12054 for the determination of comprehensive bactericidal activity of an alcohol-based hand rub. Journal of Hospital Infection. 2003, 55: 226-231. 10.1016/S0195-6701(03)00298-6.CrossRefPubMed Kampf G, Hollingsworth A: Validity of the four European test strains of prEN 12054 for the determination of comprehensive bactericidal activity of an alcohol-based hand rub. Journal of Hospital Infection. 2003, 55: 226-231. 10.1016/S0195-6701(03)00298-6.CrossRefPubMed
Metadata
Title
Insufficient neutralization in testing a chlorhexidine-containing ethanol-based hand rub can result in a false positive efficacy assessment
Authors
Günter Kampf
Marc Shaffer
Corrine Hunte
Publication date
01-12-2005
Publisher
BioMed Central
Published in
BMC Infectious Diseases / Issue 1/2005
Electronic ISSN: 1471-2334
DOI
https://doi.org/10.1186/1471-2334-5-48

Other articles of this Issue 1/2005

BMC Infectious Diseases 1/2005 Go to the issue
Live Webinar | 27-06-2024 | 18:00 (CEST)

Keynote webinar | Spotlight on medication adherence

Live: Thursday 27th June 2024, 18:00-19:30 (CEST)

WHO estimates that half of all patients worldwide are non-adherent to their prescribed medication. The consequences of poor adherence can be catastrophic, on both the individual and population level.

Join our expert panel to discover why you need to understand the drivers of non-adherence in your patients, and how you can optimize medication adherence in your clinics to drastically improve patient outcomes.

Prof. Kevin Dolgin
Prof. Florian Limbourg
Prof. Anoop Chauhan
Developed by: Springer Medicine
Obesity Clinical Trial Summary

At a glance: The STEP trials

A round-up of the STEP phase 3 clinical trials evaluating semaglutide for weight loss in people with overweight or obesity.

Developed by: Springer Medicine