Open Access 01-12-2020 | Research article

Influence of measurement principle on total hemoglobin value

Authors: Keisuke Hayashi, Takashi Hitosugi, Yoshifumi Kawakubo, Norihisa Kitamoto, Takeshi Yokoyama

Published in: BMC Anesthesiology | Issue 1/2020

Login to get access

Abstract

Background

Total hemoglobin (tHb) measurement is indispensable for determining the patient’s condition (hemorrhagic vs. ischemic) and need for blood transfusion. Conductivity- and absorbance-based measurement methods are used for blood gas analysis of tHb. For conductivity-based measurement, tHb is calculated after converting blood conductivity into a hematocrit value, whereas absorbance measurement is based on light absorbance after red blood cell hemolysis. Due to changes in plasma electrolytes and hemolysis, there is a possibility that conductivity- and absorbance-based measurement methods may cause a difference in tHb.

Methods

In this study, test samples with controlled electrolyte changes and hemolysis were created by adding sodium chloride, distilled water or hemolytic blood to blood samples collected from healthy volunteers, and tHb values were compared between both methods.

Results

Conductivity-based measurement revealed reduced tHb value (from 15.49 to 13.05 g/dl) following the addition of 10% sodium chloride, which was also reduced by the addition of hemolysate. Conversely, the addition of distilled water significantly increased tHb value than the expected value. In the absorbance method, there was no significant change in tHb value due to electrolyte change or hemolysis.

Conclusions

We have to recognize unexpected conductivity changes occur at all times when tHb is measured via conductivity- and absorbance-based measurement methods. The absorbance method should be used when measuring tHb in patients with expected blood conductivity changes. However, when using this method, the added contribution of hemoglobin from hemolytic erythrocytes lacking oxygen carrying capacity must be considered. We recognize that discrepancy can occur between conductivity- and absorbance-based measurement methods when tHb is measured.
Literature
1.
go back to reference Neelakantaswamy PS, Aspar KF, Rajaratnam A, Das NP. Conductimetric method of determining the hematocrit value of blood. Rev Sci Instrum. 1983;54:1186–90.CrossRef Neelakantaswamy PS, Aspar KF, Rajaratnam A, Das NP. Conductimetric method of determining the hematocrit value of blood. Rev Sci Instrum. 1983;54:1186–90.CrossRef
2.
go back to reference Ewing GW. Instrumental methods of chemical analysis. 5th ed. New York: McGraw-Hill; 1985. Ewing GW. Instrumental methods of chemical analysis. 5th ed. New York: McGraw-Hill; 1985.
3.
go back to reference Papadea C, Foster J, Grant S, Ballard SA, Cate JC 4th, Southgate WM, et al. Evaluation of the i-STAT portable clinical analyzer for point-of-care blood testing in the intensive care units of a university children's hospital. Ann Clin Lab Sci. 2002;32:231–43.PubMed Papadea C, Foster J, Grant S, Ballard SA, Cate JC 4th, Southgate WM, et al. Evaluation of the i-STAT portable clinical analyzer for point-of-care blood testing in the intensive care units of a university children's hospital. Ann Clin Lab Sci. 2002;32:231–43.PubMed
4.
go back to reference Wu P, Morey TE, Harris NS, Gravenstein N, Rice MJ. Intravenous fluids cause systemic bias in a conductivity-based point-of-care hematocrit meter. Anesth Analg. 2012;114:314–21.CrossRef Wu P, Morey TE, Harris NS, Gravenstein N, Rice MJ. Intravenous fluids cause systemic bias in a conductivity-based point-of-care hematocrit meter. Anesth Analg. 2012;114:314–21.CrossRef
5.
go back to reference Prichard JS, French JS, Alvar N. Clinical evaluation of the ABL-77 for point-of-care analysis in the cardiovascular operating room. J Extra Corpor Technol. 2006;38:128–33.PubMedPubMedCentral Prichard JS, French JS, Alvar N. Clinical evaluation of the ABL-77 for point-of-care analysis in the cardiovascular operating room. J Extra Corpor Technol. 2006;38:128–33.PubMedPubMedCentral
6.
go back to reference Steinfelder-Visscher J, Weerwind PW, Teerenstra S, Pop GA, Brouwer RM. Conductivity-based hematocrit measurement during cardiopulmonary bypass. J Clin Monit Comput. 2007;21:7–12.CrossRef Steinfelder-Visscher J, Weerwind PW, Teerenstra S, Pop GA, Brouwer RM. Conductivity-based hematocrit measurement during cardiopulmonary bypass. J Clin Monit Comput. 2007;21:7–12.CrossRef
7.
go back to reference Briggs C, Kimber S, Green L. Where are we at with point-of-care testing in haematology? Br J Haematol. 2012;158:679–90.CrossRef Briggs C, Kimber S, Green L. Where are we at with point-of-care testing in haematology? Br J Haematol. 2012;158:679–90.CrossRef
8.
go back to reference Stott RA, Hortin GL, Wilhite TR, Miller SB, Smith CH, Landt M. Analytical artifacts in hematocrit measurements by whole-blood chemistry analyzers. Clin Chem. 1995;41:306–11.CrossRef Stott RA, Hortin GL, Wilhite TR, Miller SB, Smith CH, Landt M. Analytical artifacts in hematocrit measurements by whole-blood chemistry analyzers. Clin Chem. 1995;41:306–11.CrossRef
9.
go back to reference Hopfer SM, Nadeau FL, Sundra M, Makowski GS. Effect of protein on hemoglobin and hematocrit assays with a conductivity-based point-of-care testing device: comparison with optical methods. Ann Clin Lab Sci. 2004;34:75–82.PubMed Hopfer SM, Nadeau FL, Sundra M, Makowski GS. Effect of protein on hemoglobin and hematocrit assays with a conductivity-based point-of-care testing device: comparison with optical methods. Ann Clin Lab Sci. 2004;34:75–82.PubMed
10.
go back to reference Martens H. Multivariate calibration: quantitative interpretation of non-selective chemical data. Norwegian Institute of Technology; 1985. Martens H. Multivariate calibration: quantitative interpretation of non-selective chemical data. Norwegian Institute of Technology; 1985.
11.
go back to reference Krzeminski A. Why correct for fetal hemoglobin in blood oximetry measurements? Radiometer publication info. No. 1992–3. Copenhagen: Radiometer Medical A/S; 1992. Krzeminski A. Why correct for fetal hemoglobin in blood oximetry measurements? Radiometer publication info. No. 1992–3. Copenhagen: Radiometer Medical A/S; 1992.
12.
go back to reference Lampert R, Brandt L. The effect of hyperbilirubinemia on the measurement of oxygenated hemoglobin (O2Hb), carboxyhemoglobin (COHb) and methemoglobin (MetHb) using multiwavelength oximeters in mixed venous blood. Anaesthesist. 1993;42:702–9.PubMed Lampert R, Brandt L. The effect of hyperbilirubinemia on the measurement of oxygenated hemoglobin (O2Hb), carboxyhemoglobin (COHb) and methemoglobin (MetHb) using multiwavelength oximeters in mixed venous blood. Anaesthesist. 1993;42:702–9.PubMed
13.
go back to reference Performance Characteristics. In: ABLTM700 Series Reference Manual. Ed. C. Copenhagen: Radiometer Medical A/S, 1999: Chapter 8, 13-22. Code No 989–312. Performance Characteristics. In: ABLTM700 Series Reference Manual. Ed. C. Copenhagen: Radiometer Medical A/S, 1999: Chapter 8, 13-22. Code No 989–312.
Metadata
Title
Influence of measurement principle on total hemoglobin value
Authors
Keisuke Hayashi
Takashi Hitosugi
Yoshifumi Kawakubo
Norihisa Kitamoto
Takeshi Yokoyama
Publication date
01-12-2020
Publisher
BioMed Central
Published in
BMC Anesthesiology / Issue 1/2020
Electronic ISSN: 1471-2253
DOI
https://doi.org/10.1186/s12871-020-00991-2

Other articles of this Issue 1/2020

BMC Anesthesiology 1/2020 Go to the issue