01-09-2021 | Incision | Original Article
Tissue effects of a newly developed diode pumped pulsed Thulium:YAG laser compared to continuous wave Thulium:YAG and pulsed Holmium:YAG laser
Authors:
Stephan Huusmann, Marcel Lafos, Ingo Meyenburg, Rolf Muschter, Heinrich-Otto Teichmann, Thomas Herrmann
Published in:
World Journal of Urology
|
Issue 9/2021
Login to get access
Abstract
Purpose
The objective of this study is to evaluate the laser-tissue effects of laser radiation emitted by a newly developed high frequency pulsed Tm:YAG laser in comparison to the continuous wave Tm:YAG laser and the pulsed Ho:YAG laser.
Methods
Ex-vivo experiments were performed on freshly slaughtered porcine kidneys in a physiological saline solution. Experiments were performed using two different laser devices in different settings: A Tm:YAG laser was operated in a pulsed mode up to 300 Hz and in a continuous wave (CW) mode. Results were compared with a 100 W standard pulsed Ho:YAG laser system. Comparative tissue experiments were performed at 5 W, 40 W and 80 W. The incision depth and the laser damage zone were measured under a microscope using a calibrated ocular scale.
Results
Increased laser power resulted in increased incision depth and increased laser damage zone for all investigated lasers in this set-up. The Ho:YAG created the largest combined tissue effect at the 5 W power setting and seems to be the least controllable laser at low power for soft tissue incisions. The CW Tm:YAG did not incise at all at 5 W, but created the largest laser damage zone. For the new pulsed Tm:YAG laser the tissue effect grew evenly with increasing power.
Conclusion
Among the investigated laser systems in this setting the pulsed Tm:YAG laser shows the most controllable behavior, insofar as both the incision depth and the laser damage zone increase evenly with increasing laser power.