Background

CORKNOT® facilitates a reduction in cardiopulmonary bypass (CPB) time, aortic cross clamp (ACC) time and operative time, but reported to be associated with other complications. We aim to quantify the incidence of valvular complications related to CORKNOT® and determine the feasibility of its use between different valvular surgeries.

Methods

Patients who underwent heart valve repair or replacement surgery via the use of automated titanium suture fasteners (CORKNOT®) in a tertiary care hospital were included in the study. This single-centre retrospective study was conducted on 132 patients between January 2016 and June 2018.

Results

In our study, the overall mean operative time was 320.0 ± 97.0 min, mean CPB time was 171.4 ± 76.0 min and the calculated mean ACC time was 105.9 ± 54.0 min. Fifty-eight patients (43.9%) underwent minimally invasive valve replacement or repair surgery and 66 patients (50.0%) underwent concomitant procedures. A total of 157 valves were operated on, with 112 (84.8%) single valve surgeries, 15 (11.4%) double valve surgeries and 5 (3.8%) triple valve surgeries. After reviewed by the cardiologist blinded towards the study, we report trivial and/or mild paravalvular leak (PVL) in immediate post-operative echocardiography was found in 1 (1.01%) patients. There were no reported cases of valvular thrombosis, leaflet perforation, device dislodgement or embolization, moderate and/or severe PVL during hospitalization and follow-up echocardiography within 1 year. Single mitral valve and aortic surgeries had comparable incidences of post surgical complications.

Conclusion

We conclude the feasibility of CORKNOT® utilisation in mitral and aortic valve surgeries. Additionally, incidence of CORKNOT® related complications in heart valve repair or replacement surgery is less usual in our setting than previously reported. These results motivate the use of CORKNOT® as a valid alternative with complete commitment.
Appendix
Available only for authorised users
Literature
8.
go back to reference Loberman D, Mohr R, Pirundini PA, Yazdchi F, Rinewalt D, Ziv-Baran T. Automated fastener (Core-knot) versus manually tied knots in patients undergoing aortic valve replacement: impact on cross-clamp time and short-term echocardiographic results. Medicine (Baltimore). 2018;97(31):e11657. https://doi.org/10.1097/MD.0000000000011657.CrossRef Loberman D, Mohr R, Pirundini PA, Yazdchi F, Rinewalt D, Ziv-Baran T. Automated fastener (Core-knot) versus manually tied knots in patients undergoing aortic valve replacement: impact on cross-clamp time and short-term echocardiographic results. Medicine (Baltimore). 2018;97(31):e11657. https://​doi.​org/​10.​1097/​MD.​0000000000011657​.CrossRef
14.
go back to reference Salik I, Lee LS, Widrich J. Mitral Valve Repair. 2020 Dec 23. In: StatPearls. Treasure Island: StatPearls Publishing; 2021. PMID: 31751069. Salik I, Lee LS, Widrich J. Mitral Valve Repair. 2020 Dec 23. In: StatPearls. Treasure Island: StatPearls Publishing; 2021. PMID: 31751069.
Metadata
Title
Incidence of valvular regurgitation and leaflet perforation by using automated titanium fasteners (CORKNOT®) in heart valve repair or replacement: less usual than reported
Authors
Faizus Sazzad
Ong Zhi Xian
Ashlynn Ler
Chang Guohao
Kang Giap Swee
Theo Kofidis
Publication date
01-12-2021
Publisher
BioMed Central
Published in
Journal of Cardiothoracic Surgery / Issue 1/2021
Electronic ISSN: 1749-8090
DOI
https://doi.org/10.1186/s13019-021-01512-z

Other articles of this Issue 1/2021

Journal of Cardiothoracic Surgery 1/2021 Go to the issue