Published in: Journal of Interventional Cardiac Electrophysiology 3/2014

01-04-2014

Incidence of nonphysiologic short VV intervals detected by the sensing integrity counter with integrated bipolar compared with true bipolar leads: clinically inconsequential or cause for concern?

Authors: Justin Ng, Nagesh Chopra, Chirag Barbhaiya, Tobias Reichlin, Eyal Nof, Koichi Nagashima, Thomas Tadros, Bruce A Koplan

Published in: Journal of Interventional Cardiac Electrophysiology | Issue 3/2014

Login to get access

Abstract

Introduction

The need to detect impending implantable cardiac defibrillator (ICD) lead failure has grown. Automated sensing diagnostics have been developed for this reason. The sensing integrity counter (SIC) is one such oversensing diagnostic, which forms an integral part of the Medtronic™ lead integrity alert (LIA) feature on implantable defibrillators. It records nonphysiologic short VV intervals (NPSVVIs). It is unclear whether SIC data derived from integrated bipolar (IBP) leads need to be interpreted differently when compared to true bipolar (TBP) leads. We hypothesized that IBP ICD leads by virtue of a larger “antennae” may generate more NPSVVIs on than TBP leads, leading to more false-positive SIC counts.

Methods

Equal durations of remote monitoring records of 44 patients (mean age of 65.9 ± 2.2 years, 52 % female) with IBP ICD leads and Medtronic (MDT) generators (IBP group) were compared with those of 44 randomly selected patients (64.0 ± 2.2 years, 24 % female) who had TBP ICD leads and MDT generators (TBP group). Mean surveillance time, defined as the time over which the cumulative SIC count was acquired, was 614 ± 44 days (TBP group) vs. 620 ± 49 days (IBP group, p = ns). The mean time of follow-up following the first documented short VV interval was 115.2 months in the integrated bipolar group and 66.9 months in the true bipolar group. Leads on advisory were excluded from the study.

Results

A total of 26/44 patients in the IBP group displayed NPSVVI compared to 11/44 patients in the TBP group (59 vs. 25 %; p = 0.002, Fisher exact test). When adjusted for gender and lead age, the difference was still significant (p = 0.008). When evaluating the clinical consequence of NPSVVI in this cohort, 3/11 TBP leads with NPSVVI of >0 were eventually extracted due to additional abnormalities vs. 0/26 IBP leads with NPSVVI (p = 0.02, Fisher exact test). None of the IBP group patients with NPSVVI have developed inappropriate therapy from lead noise or a need for abandonment or extraction.

Conclusion

Integrated bipolar ICD leads are more likely to have elevated SIC counts than true bipolar leads despite revealing no other evidence of lead failure. There does not appear to be a need for heightened surveillance in IBP leads with observed elevated SIC counts that have no other findings to suggest lead malfunction.
Literature
1.
go back to reference Sweeney, M. O. (2010). The contradiction of appropriate shocks in primary prevention ICDs: increasing and decreasing the risk of death. Circulation, 122, 2638–2641.PubMedCrossRef Sweeney, M. O. (2010). The contradiction of appropriate shocks in primary prevention ICDs: increasing and decreasing the risk of death. Circulation, 122, 2638–2641.PubMedCrossRef
2.
go back to reference Stevenson, W., & John, R. (2011). Ventricular arrhythmias in patients with implanted defibrillators. Circulation, 124, e411–e414.PubMedCrossRef Stevenson, W., & John, R. (2011). Ventricular arrhythmias in patients with implanted defibrillators. Circulation, 124, e411–e414.PubMedCrossRef
3.
go back to reference Swerdlow, C. D., Gunderson, B. D., Ousdigian, K. T., Abeyratne, A., Sachanandani, H., & Ellenbogen, K. A. (2010). Downloadable software algorithm reduces inappropriate shocks caused by implantable cardioverter-defibrillator lead fractures: a prospective study. Circulation, 122, 1449–1455.PubMedCrossRef Swerdlow, C. D., Gunderson, B. D., Ousdigian, K. T., Abeyratne, A., Sachanandani, H., & Ellenbogen, K. A. (2010). Downloadable software algorithm reduces inappropriate shocks caused by implantable cardioverter-defibrillator lead fractures: a prospective study. Circulation, 122, 1449–1455.PubMedCrossRef
4.
go back to reference Shah, M., Urban, J., Buck, K., Vetter, V., & Tanel, R. (2008). Increase in sensing integrity counts is the best marker for impending fracture in the Sprint Fidelis ICD leads. Heart Rhythm, 5, S250. Abstract. Shah, M., Urban, J., Buck, K., Vetter, V., & Tanel, R. (2008). Increase in sensing integrity counts is the best marker for impending fracture in the Sprint Fidelis ICD leads. Heart Rhythm, 5, S250. Abstract.
5.
go back to reference Kallinen, L., Hauser, R., Tang, C., Melby, D., Almquist, A., Katsiyiannis, W., et al. (2010). Lead integrity alert algorithm decreases inappropriate shocks in patients who have Sprint Fidelis pace-sense conductor fractures. Heart Rhythm, 7(8), 1048–1055.PubMedCrossRef Kallinen, L., Hauser, R., Tang, C., Melby, D., Almquist, A., Katsiyiannis, W., et al. (2010). Lead integrity alert algorithm decreases inappropriate shocks in patients who have Sprint Fidelis pace-sense conductor fractures. Heart Rhythm, 7(8), 1048–1055.PubMedCrossRef
6.
go back to reference Swerdlow C, Gunderson B, Stromberg K, Ellenbogen K. (2012). Abstract 11384: performance of ICD lead integrity alert for diagnosis of Riata lead failures. AHA 126:A11384. Swerdlow C, Gunderson B, Stromberg K, Ellenbogen K. (2012). Abstract 11384: performance of ICD lead integrity alert for diagnosis of Riata lead failures. AHA 126:A11384.
7.
go back to reference Ellenbogen, K. A., Gunderson, B. D., Patel, A. S., Ousdigian, K., Abeyratne, A., & Swerdlow, C. (2009). Lead integrity alert performance for non-Sprint Fidelis ICD lead fractures. Heart Rhythm, 6, S248. Abstract. Ellenbogen, K. A., Gunderson, B. D., Patel, A. S., Ousdigian, K., Abeyratne, A., & Swerdlow, C. (2009). Lead integrity alert performance for non-Sprint Fidelis ICD lead fractures. Heart Rhythm, 6, S248. Abstract.
8.
go back to reference Kneller, J., Delacey, W., Wood, M. A., & Ellenbogen, K. A. (2012). Detection of a riata™ insulation failure by the Medtronic lead integrity alert™. Europace, 14(8), 1215–6.PubMedCrossRef Kneller, J., Delacey, W., Wood, M. A., & Ellenbogen, K. A. (2012). Detection of a riata™ insulation failure by the Medtronic lead integrity alert™. Europace, 14(8), 1215–6.PubMedCrossRef
9.
go back to reference Swerdlow, C. D., Gunderson, B. D., Stromberg, K. D., & Ellenbogen, K. A. (2013). Oversensing in true-bipolar vs. Integrated-bipolar ICD leads: is there a difference? Heart Rhythm, S63, AB28–01. Swerdlow, C. D., Gunderson, B. D., Stromberg, K. D., & Ellenbogen, K. A. (2013). Oversensing in true-bipolar vs. Integrated-bipolar ICD leads: is there a difference? Heart Rhythm, S63, AB28–01.
10.
go back to reference Gunderson, B. D., Swerdlow, C. D., Wilcox, J. M., Hayman, J. E., Ousdigian, K. T., & Ellenbogen, K. A. (2010). Causes of ventricular oversensing in implantable cardioverter-defibrillators: implications for diagnosis of lead fracture. Heart Rhythm, 7, 626–633.PubMedCrossRef Gunderson, B. D., Swerdlow, C. D., Wilcox, J. M., Hayman, J. E., Ousdigian, K. T., & Ellenbogen, K. A. (2010). Causes of ventricular oversensing in implantable cardioverter-defibrillators: implications for diagnosis of lead fracture. Heart Rhythm, 7, 626–633.PubMedCrossRef
11.
go back to reference Swerdlow, C. D., Gunderson, B. D., Ousdigian, K. T., Abeyratne, A., Stadler, R. W., Gillberg, J. M., et al. (2008). Downloadable algorithm to reduce inappropriate shocks caused by fractures of implantable cardioverter-defibrillator leads. Circulation, 118, 2122–2129.PubMedCrossRef Swerdlow, C. D., Gunderson, B. D., Ousdigian, K. T., Abeyratne, A., Stadler, R. W., Gillberg, J. M., et al. (2008). Downloadable algorithm to reduce inappropriate shocks caused by fractures of implantable cardioverter-defibrillator leads. Circulation, 118, 2122–2129.PubMedCrossRef
Metadata
Title
Incidence of nonphysiologic short VV intervals detected by the sensing integrity counter with integrated bipolar compared with true bipolar leads: clinically inconsequential or cause for concern?
Authors
Justin Ng
Nagesh Chopra
Chirag Barbhaiya
Tobias Reichlin
Eyal Nof
Koichi Nagashima
Thomas Tadros
Bruce A Koplan
Publication date
01-04-2014
Publisher
Springer US
Published in
Journal of Interventional Cardiac Electrophysiology / Issue 3/2014
Print ISSN: 1383-875X
Electronic ISSN: 1572-8595
DOI
https://doi.org/10.1007/s10840-013-9866-7

Other articles of this Issue 3/2014

Journal of Interventional Cardiac Electrophysiology 3/2014 Go to the issue