Skip to main content
Top
Published in: Radiation Oncology 1/2015

Open Access 01-12-2015 | Research

In-vivo dosimetry with Gafchromic films for multi-isocentric VMAT irradiation of total marrow lymph-nodes: a feasibility study

Authors: Pietro Mancosu, Pierina Navarria, Giacomo Reggiori, Luca Cozzi, Antonella Fogliata, Anna Gaudino, Francesca Lobefalo, Lucia Paganini, Valentina Palumbo, Barbara Sarina, Antonella Stravato, Luca Castagna, Stefano Tomatis, Marta Scorsetti

Published in: Radiation Oncology | Issue 1/2015

Login to get access

Abstract

Background

Total marrow (lymph-nodes) irradiation (TMI-TMLI) by volumetric modulated arc therapy (VMAT) was shown to be feasible by dosimetric feasibility studies. It was demonstrated that several partially overlapping arcs with different isocenters are required to achieve the desired coverage of the hematopoietic or lymphoid tissues targets and to spare the neighbouring healthy tissues. The effect of isocenter shifts was investigated with the treatment planning system but an in- vivo verification of the procedure was not carried out. The objective of this study was the in-vivo verification of the consistency between the delivered and planned doses using bi-dimensional GafChromic EBT3 films.

Methods

In a first phase a phantom study was carried out to quantify the uncertainties under controlled conditions. In a second phase three patients treated with TMLI were enrolled for in-vivo dosimetry. The dose prescription was 2Gy in single fraction. Ten arcs paired on 4-6 isocenters were used to cover the target. Cone Beam Computed Tomography (CBCT) was used to verify the patient positioning at each isocenter. GafChromic EBT3 films were placed below the patient on the top of a dedicated immobilization system specifically designed. The dose maps measured with the EBT3 films were compared with the corresponding calculations along the patient support couch. Gamma Agreement Index (GAI) with dose difference of 5% and distance to agreement of 5 mm was computed.

Results

In the phantom study, optimal target coverage and healthy tissue sparing was observed. GAI(5%,5 mm) was 99.4%. For the patient-specific measurements, GAI(5%,5 mm) was greater than 95% and GAI (5%,3 mm) > 90% for all patients.

Conclusions

In vivo measurements demonstrated the delivered dose to be in good agreement with the planned one for the TMI-TMLI protocol where partially overlapping arcs with different isocenters are required.
Literature
1.
go back to reference Bieri S, Helg C, Chapuis B, Miralbell R. Total body irradiation before allogenic bone marrow transplantation: is more dose better? Int J radiat Oncol Biol Phys. 2001;49:1071–7.CrossRefPubMed Bieri S, Helg C, Chapuis B, Miralbell R. Total body irradiation before allogenic bone marrow transplantation: is more dose better? Int J radiat Oncol Biol Phys. 2001;49:1071–7.CrossRefPubMed
2.
go back to reference Clif RA, Buckner CD, Appelbaum FA, Bryant E, Bearman SI, Petersen FB, et al. Allogenic marrow transplantation in patients with chronic myeloid-leukemia in the chronic phase – a randomized trial of 2 irradiation regiments. Blood. 1991;77:1660–5. Clif RA, Buckner CD, Appelbaum FA, Bryant E, Bearman SI, Petersen FB, et al. Allogenic marrow transplantation in patients with chronic myeloid-leukemia in the chronic phase – a randomized trial of 2 irradiation regiments. Blood. 1991;77:1660–5.
3.
go back to reference Kim JH, Stein A, Tsai N, Schultheiss TE, Palmer J, Liu A, et al. Extramedullary relapse following total marrow and lymphoid irradiation in patients undergoing allogeneic hematopoietic cell transplantation. Int J Radiat Oncol Biol Phys. 2014;89(1):75–81. doi:10.1016/j.ijrobp.2014.01.036.CrossRefPubMed Kim JH, Stein A, Tsai N, Schultheiss TE, Palmer J, Liu A, et al. Extramedullary relapse following total marrow and lymphoid irradiation in patients undergoing allogeneic hematopoietic cell transplantation. Int J Radiat Oncol Biol Phys. 2014;89(1):75–81. doi:10.1016/j.ijrobp.2014.01.036.CrossRefPubMed
4.
go back to reference Wilkie JR, Tiryaki H, Smith BD, Roeske JC, Radosevich JA, Aydogan B. Feasibility study for linac-based intensity modulated total marrow irradiation. Med Phys. 2008;35:5609–18.CrossRefPubMed Wilkie JR, Tiryaki H, Smith BD, Roeske JC, Radosevich JA, Aydogan B. Feasibility study for linac-based intensity modulated total marrow irradiation. Med Phys. 2008;35:5609–18.CrossRefPubMed
5.
go back to reference Aydogan B, Mundt A, Roeske JC. Linac-based intensity modulated total marrow irradiation (IM-TMI). Technol Cancer Res Treat. 2006;5:513–9.CrossRefPubMed Aydogan B, Mundt A, Roeske JC. Linac-based intensity modulated total marrow irradiation (IM-TMI). Technol Cancer Res Treat. 2006;5:513–9.CrossRefPubMed
6.
go back to reference Hui SK, Kapatoes J, Fowler J, Henderson D, Olivera G, Manon RR, et al. Feasibility study of helical tomotherapy for total body or total marrow irradiation. Med Phys. 2005;32:3214–24.CrossRefPubMed Hui SK, Kapatoes J, Fowler J, Henderson D, Olivera G, Manon RR, et al. Feasibility study of helical tomotherapy for total body or total marrow irradiation. Med Phys. 2005;32:3214–24.CrossRefPubMed
7.
go back to reference Wong JY, Rosenthal J, Liu A, Schultheiss T, Forman S, Somlo G. Image-guided total- marrow irradiation using helical tomotherapy in patients with multiple myeloma and acute leukemia undergoing hematopoietic cell transplantation. Int J Radiat Oncol Biol Phys. 2009;73:273–9.CrossRefPubMedCentralPubMed Wong JY, Rosenthal J, Liu A, Schultheiss T, Forman S, Somlo G. Image-guided total- marrow irradiation using helical tomotherapy in patients with multiple myeloma and acute leukemia undergoing hematopoietic cell transplantation. Int J Radiat Oncol Biol Phys. 2009;73:273–9.CrossRefPubMedCentralPubMed
8.
go back to reference Fogliata A, Cozzi L, Clivio A, Ibatici A, Mancosu P, Navarria P, et al. Preclinical assessment of volumetric modulated arc therapy for total marrow irradiation. Int J Radiat Oncol Biol Phys. 2011;80:628–36.CrossRefPubMed Fogliata A, Cozzi L, Clivio A, Ibatici A, Mancosu P, Navarria P, et al. Preclinical assessment of volumetric modulated arc therapy for total marrow irradiation. Int J Radiat Oncol Biol Phys. 2011;80:628–36.CrossRefPubMed
9.
go back to reference Mancosu P, Navarria P, Castagna L, Roggio A, Pellegrini C, Reggiori G, et al. Anatomy driven optimisation strategy for total marrow irradiation with a volumetric modulated arc therapy technique. J Appl Clin Med Phys. 2012;13:3653. doi:10.1120/jacmp.v13i1.3653.PubMed Mancosu P, Navarria P, Castagna L, Roggio A, Pellegrini C, Reggiori G, et al. Anatomy driven optimisation strategy for total marrow irradiation with a volumetric modulated arc therapy technique. J Appl Clin Med Phys. 2012;13:3653. doi:10.1120/jacmp.v13i1.3653.PubMed
10.
go back to reference Aydogan B, Yeginer M, Kavak GO, Fan J, Radosevich JA, Gwe-Ya K. Total Marrow Irradiation with RapidArc Volumetric Arc Therapy. Int J Radiat Oncol Biol Phys. 2011;81:592–9.CrossRefPubMed Aydogan B, Yeginer M, Kavak GO, Fan J, Radosevich JA, Gwe-Ya K. Total Marrow Irradiation with RapidArc Volumetric Arc Therapy. Int J Radiat Oncol Biol Phys. 2011;81:592–9.CrossRefPubMed
11.
go back to reference Han C, Schultheiss TE, Wong JY. Dosimetric study of volumetric modulated arc therapy fields for total marrow irradiation. Radiother Oncol. 2011;112:315–20. Han C, Schultheiss TE, Wong JY. Dosimetric study of volumetric modulated arc therapy fields for total marrow irradiation. Radiother Oncol. 2011;112:315–20.
12.
go back to reference Sorucu M, Yeginer M, Kaval O, Radosevich JA, Aydogan B. Verification of dose distribution for volumetric modulated arc therapy total marrow irradiation in a humanlike phantom. Med Phys. 2012;39:281–8.CrossRef Sorucu M, Yeginer M, Kaval O, Radosevich JA, Aydogan B. Verification of dose distribution for volumetric modulated arc therapy total marrow irradiation in a humanlike phantom. Med Phys. 2012;39:281–8.CrossRef
13.
go back to reference Mancosu P, Navarria P, Castagna L, Reggiori G, Sarina B, Tomatis S, et al. Interplay effects between dose distribution quality and positioning accuracy in total marrow irradiation with volumetric modulated arc therapy. Med Phys. 2013;40(11):111713. doi:10.1118/1.4823767.CrossRefPubMed Mancosu P, Navarria P, Castagna L, Reggiori G, Sarina B, Tomatis S, et al. Interplay effects between dose distribution quality and positioning accuracy in total marrow irradiation with volumetric modulated arc therapy. Med Phys. 2013;40(11):111713. doi:10.1118/1.4823767.CrossRefPubMed
14.
go back to reference Bornhäuser M, Kienast J, Trenschel R, Burchert A, Hegenbart U, Stadler M, et al. Reduced-intensity conditioning versus standard conditioning before allogeneic haemopoietic cell transplantation in patients with acute myeloid leukaemia in first complete remission: a prospective, open-label randomised phase 3 trial. Lancet Oncol. 2012;13(10):1035–44. doi:10.1016/S1470-2045(12)70349-2.CrossRefPubMed Bornhäuser M, Kienast J, Trenschel R, Burchert A, Hegenbart U, Stadler M, et al. Reduced-intensity conditioning versus standard conditioning before allogeneic haemopoietic cell transplantation in patients with acute myeloid leukaemia in first complete remission: a prospective, open-label randomised phase 3 trial. Lancet Oncol. 2012;13(10):1035–44. doi:10.1016/S1470-2045(12)70349-2.CrossRefPubMed
15.
go back to reference Storb RF, Champlin R, Riddell SR, Murata M, Bryant S, Warren EH. Non-myeloablative transplants for malignant disease. Hematology Am Soc Hematol Educ Program. 2001;1:375–91.CrossRef Storb RF, Champlin R, Riddell SR, Murata M, Bryant S, Warren EH. Non-myeloablative transplants for malignant disease. Hematology Am Soc Hematol Educ Program. 2001;1:375–91.CrossRef
16.
go back to reference Baron F, Storb R, Storer BE, Maris MB, Niederwieser D, Shizuru JA, et al. Factors associated with outcomes in allogeneic hematopoietic cell transplantation with nonmyeloablative conditioning after failed myeloablative hematopoietic cell transplantation. J Clin Oncol. 2006;24(25):4150–7. Epub 2006 Aug 8.CrossRefPubMed Baron F, Storb R, Storer BE, Maris MB, Niederwieser D, Shizuru JA, et al. Factors associated with outcomes in allogeneic hematopoietic cell transplantation with nonmyeloablative conditioning after failed myeloablative hematopoietic cell transplantation. J Clin Oncol. 2006;24(25):4150–7. Epub 2006 Aug 8.CrossRefPubMed
17.
go back to reference Shimohigashi Y, Araki F, Tominaga H, Sakata J, Kawasaki K, Kanetake N, et al. Angular dependence correction of MatriXX and its application to composite dose verification. J Appl Clin Med Phys. 2012;13(5):3856. doi:10.1120/jacmp.v13i5.3856.PubMed Shimohigashi Y, Araki F, Tominaga H, Sakata J, Kawasaki K, Kanetake N, et al. Angular dependence correction of MatriXX and its application to composite dose verification. J Appl Clin Med Phys. 2012;13(5):3856. doi:10.1120/jacmp.v13i5.3856.PubMed
18.
go back to reference Fogliata A, Clivio A, Fenoglietto P, Hrbacek J, Kloeck S, Lattuada P, et al. Quality assurance of RapidArc in clinical practice using portal dosimetry. Br J Radiol. 2011;84(1002):534–45. doi:10.1259/bjr/72327299.CrossRefPubMedCentralPubMed Fogliata A, Clivio A, Fenoglietto P, Hrbacek J, Kloeck S, Lattuada P, et al. Quality assurance of RapidArc in clinical practice using portal dosimetry. Br J Radiol. 2011;84(1002):534–45. doi:10.1259/bjr/72327299.CrossRefPubMedCentralPubMed
19.
go back to reference Mijnheer B, Beddar S, Izewska J, Reft C. In vivo dosimetry in external beam radiotherapy. Med Phys. 2013;40(7):070903.CrossRefPubMed Mijnheer B, Beddar S, Izewska J, Reft C. In vivo dosimetry in external beam radiotherapy. Med Phys. 2013;40(7):070903.CrossRefPubMed
20.
go back to reference Ganapathy K, Kurup PGG, Murali V, Muthukumaran M, Bhuvaneshwari N, Velmurugan J. Patient dose analysis in total body irradiation through in vivo dosimetry. J Med Phys. 2012;37:214–8.CrossRefPubMedCentralPubMed Ganapathy K, Kurup PGG, Murali V, Muthukumaran M, Bhuvaneshwari N, Velmurugan J. Patient dose analysis in total body irradiation through in vivo dosimetry. J Med Phys. 2012;37:214–8.CrossRefPubMedCentralPubMed
21.
go back to reference Yao R, Bernard D, Turian J, Abrams RA, Sensakovic W, Fung HC, et al. A simplified technique for delivering total body irradiation (TBI) with improved dose homogeneity. Med Phys. 2012;39:2239.CrossRefPubMed Yao R, Bernard D, Turian J, Abrams RA, Sensakovic W, Fung HC, et al. A simplified technique for delivering total body irradiation (TBI) with improved dose homogeneity. Med Phys. 2012;39:2239.CrossRefPubMed
22.
go back to reference Akino Y, McMullen JP, Das IJ. Patterns of patient specific dosimetry in total body irradiation. Med Phys. 2013;40:041719.CrossRefPubMed Akino Y, McMullen JP, Das IJ. Patterns of patient specific dosimetry in total body irradiation. Med Phys. 2013;40:041719.CrossRefPubMed
23.
go back to reference Fiandra C, Ricardi U, Ragona R, Anglesio S, Giglioli FR, Calamia E, et al. Clinical use of EBT model Gafchromic film in radiotherapy. Med Phys. 2006;33:4314–9.CrossRefPubMed Fiandra C, Ricardi U, Ragona R, Anglesio S, Giglioli FR, Calamia E, et al. Clinical use of EBT model Gafchromic film in radiotherapy. Med Phys. 2006;33:4314–9.CrossRefPubMed
24.
go back to reference Ganapathy K, Kurup PGG, Murali V, Muthukumaran M, Balaji Subramanian S, Velmurugan J. A study on rectal dose measurement in phantom and in vivo using Gafchromic EBT3 film in IMRT and CyberKnife treatments of carcinoma of prostate. J Med Phys. 2013;38:132–8.CrossRefPubMedCentralPubMed Ganapathy K, Kurup PGG, Murali V, Muthukumaran M, Balaji Subramanian S, Velmurugan J. A study on rectal dose measurement in phantom and in vivo using Gafchromic EBT3 film in IMRT and CyberKnife treatments of carcinoma of prostate. J Med Phys. 2013;38:132–8.CrossRefPubMedCentralPubMed
25.
go back to reference Kim S, Yoo S, Yin FF, Samei E, Yoshizumi T. Kilovoltage cone-beam CT: comparativedose and image quality evaluations in partial and full-angle scan protocols. Med Phys. 2010;37(7):3648–59.CrossRefPubMed Kim S, Yoo S, Yin FF, Samei E, Yoshizumi T. Kilovoltage cone-beam CT: comparativedose and image quality evaluations in partial and full-angle scan protocols. Med Phys. 2010;37(7):3648–59.CrossRefPubMed
26.
go back to reference Low DA, Harms WB, Mutic S, Purdy JA. A technique for quantitative evaluation of dose distributions. Med Phys. 1998;25:656–6.CrossRefPubMed Low DA, Harms WB, Mutic S, Purdy JA. A technique for quantitative evaluation of dose distributions. Med Phys. 1998;25:656–6.CrossRefPubMed
Metadata
Title
In-vivo dosimetry with Gafchromic films for multi-isocentric VMAT irradiation of total marrow lymph-nodes: a feasibility study
Authors
Pietro Mancosu
Pierina Navarria
Giacomo Reggiori
Luca Cozzi
Antonella Fogliata
Anna Gaudino
Francesca Lobefalo
Lucia Paganini
Valentina Palumbo
Barbara Sarina
Antonella Stravato
Luca Castagna
Stefano Tomatis
Marta Scorsetti
Publication date
01-12-2015
Publisher
BioMed Central
Published in
Radiation Oncology / Issue 1/2015
Electronic ISSN: 1748-717X
DOI
https://doi.org/10.1186/s13014-015-0391-y

Other articles of this Issue 1/2015

Radiation Oncology 1/2015 Go to the issue