Open Access 01-12-2011 | Research

In vivo assessment of catheter positioning accuracy and prolonged irradiation time on liver tolerance dose after single-fraction 192Ir high-dose-rate brachytherapy

Authors: Lutz Lüdemann, Christian Wybranski, Max Seidensticker, Konrad Mohnike, Siegfried Kropf, Peter Wust, Jens Ricke

Published in: Radiation Oncology | Issue 1/2011

Login to get access

Abstract

Background

To assess brachytherapy catheter positioning accuracy and to evaluate the effects of prolonged irradiation time on the tolerance dose of normal liver parenchyma following single-fraction irradiation with 192 Ir.

Materials and methods

Fifty patients with 76 malignant liver tumors treated by computed tomography (CT)-guided high-dose-rate brachytherapy (HDR-BT) were included in the study. The prescribed radiation dose was delivered by 1 - 11 catheters with exposure times in the range of 844 - 4432 seconds. Magnetic resonance imaging (MRI) datasets for assessing irradiation effects on normal liver tissue, edema, and hepatocyte dysfunction, obtained 6 and 12 weeks after HDR-BT, were merged with 3D dosimetry data. The isodose of the treatment plan covering the same volume as the irradiation effect was taken as a surrogate for the liver tissue tolerance dose. Catheter positioning accuracy was assessed by calculating the shift between the 3D center coordinates of the irradiation effect volume and the tolerance dose volume for 38 irradiation effects in 30 patients induced by catheters implanted in nearly parallel arrangement. Effects of prolonged irradiation were assessed in areas where the irradiation effect volume and tolerance dose volume did not overlap (mismatch areas) by using a catheter contribution index. This index was calculated for 48 irradiation effects induced by at least two catheters in 44 patients.

Results

Positioning accuracy of the brachytherapy catheters was 5-6 mm. The orthogonal and axial shifts between the center coordinates of the irradiation effect volume and the tolerance dose volume in relation to the direction vector of catheter implantation were highly correlated and in first approximation identically in the T1-w and T2-w MRI sequences (p = 0.003 and p < 0.001, respectively), as were the shifts between 6 and 12 weeks examinations (p = 0.001 and p = 0.004, respectively). There was a significant shift of the irradiation effect towards the catheter entry site compared with the planned dose distribution (p < 0.005). Prolonged treatment time increases the normal tissue tolerance dose. Here, the catheter contribution indices indicated a lower tolerance dose of the liver parenchyma in areas with prolonged irradiation (p < 0.005).

Conclusions

Positioning accuracy of brachytherapy catheters is sufficient for clinical practice. Reduced tolerance dose in areas exposed to prolonged irradiation is contradictory to results published in the current literature. Effects of prolonged dose administration on the liver tolerance dose for treatment times of up to 60 minutes per HDR-BT session are not pronounced compared to effects of positioning accuracy of the brachytherapy catheters and are therefore of minor importance in treatment planning.
Appendix
Available only for authorised users
Literature
1.
go back to reference Ricke J, Mohnike K, Pech M, Seidensticker M, Rühl R, Wieners G, Gaffke G, Kropf S, Felix R, Wust P: Local response and impact on survival after local ablation of liver metastases from colorectal carcinoma by computed tomography-guided high-dose-rate brachytherapy. Int J Radiat Oncol Biol Phys 2010,78(2):479-485. 10.1016/j.ijrobp.2009.09.026CrossRefPubMed Ricke J, Mohnike K, Pech M, Seidensticker M, Rühl R, Wieners G, Gaffke G, Kropf S, Felix R, Wust P: Local response and impact on survival after local ablation of liver metastases from colorectal carcinoma by computed tomography-guided high-dose-rate brachytherapy. Int J Radiat Oncol Biol Phys 2010,78(2):479-485. 10.1016/j.ijrobp.2009.09.026CrossRefPubMed
2.
go back to reference Mohnike K, Wieners G, Schwartz F, Seidensticker M, Pech M, Ruehl R, Wust P, Lopez-Hänninen E, Gademann G, Peters N, Berg T, Malfertheiner P, Ricke J: Computed tomography-guided high-dose-rate brachytherapy in hepatocellular carcinoma: safety, efficacy, and effect on survival. Int J Radiat Oncol Biol Phys 2010, 78: 172-179. 10.1016/j.ijrobp.2009.07.1700CrossRefPubMed Mohnike K, Wieners G, Schwartz F, Seidensticker M, Pech M, Ruehl R, Wust P, Lopez-Hänninen E, Gademann G, Peters N, Berg T, Malfertheiner P, Ricke J: Computed tomography-guided high-dose-rate brachytherapy in hepatocellular carcinoma: safety, efficacy, and effect on survival. Int J Radiat Oncol Biol Phys 2010, 78: 172-179. 10.1016/j.ijrobp.2009.07.1700CrossRefPubMed
3.
go back to reference Wieners G, Mohnike K, Peters N, Bischoff J, Kleine-Tebbe A, Seidensticker R, Seidensticker M, Gademann G, Wust P, Pech M, Ricke J: Treatment of hepatic metastases of breast cancer with CT-guided interstitial brachytherapy - A phase II-study. Radiother Oncol 2011. Wieners G, Mohnike K, Peters N, Bischoff J, Kleine-Tebbe A, Seidensticker R, Seidensticker M, Gademann G, Wust P, Pech M, Ricke J: Treatment of hepatic metastases of breast cancer with CT-guided interstitial brachytherapy - A phase II-study. Radiother Oncol 2011.
4.
go back to reference Rühl R, Lüdemann L, Czarnecka A, Streitparth F, Seidensticker M, Mohnike K, Pech M, Wust P, Ricke J: Radiobiological restrictions and tolerance doses of repeated single-fraction hdr-irradiation of intersecting small liver volumes for recurrent hepatic metastases. Radiat Oncol 2010, 5: 44-44.PubMedCentralCrossRefPubMed Rühl R, Lüdemann L, Czarnecka A, Streitparth F, Seidensticker M, Mohnike K, Pech M, Wust P, Ricke J: Radiobiological restrictions and tolerance doses of repeated single-fraction hdr-irradiation of intersecting small liver volumes for recurrent hepatic metastases. Radiat Oncol 2010, 5: 44-44.PubMedCentralCrossRefPubMed
5.
go back to reference Seidensticker M, Wust P, Rühl R, Mohnike K, Pech M, Wieners G, Gademann G, Ricke J: Safety margin in irradiation of colorectal liver metastases: assessment of the control dose of micrometastases. Radiat Oncol 2010, 5: 24-24. 10.1186/1748-717X-5-24PubMedCentralCrossRefPubMed Seidensticker M, Wust P, Rühl R, Mohnike K, Pech M, Wieners G, Gademann G, Ricke J: Safety margin in irradiation of colorectal liver metastases: assessment of the control dose of micrometastases. Radiat Oncol 2010, 5: 24-24. 10.1186/1748-717X-5-24PubMedCentralCrossRefPubMed
6.
go back to reference Hall EJ: Weiss lecture. The dose-rate factor in radiation biology. Int J Radiat Biol 1991,59(3):595-610. 10.1080/09553009114550531CrossRefPubMed Hall EJ: Weiss lecture. The dose-rate factor in radiation biology. Int J Radiat Biol 1991,59(3):595-610. 10.1080/09553009114550531CrossRefPubMed
7.
go back to reference Fowler JF, Van Limbergen EF: Biological effect of pulsed dose rate brachytherapy with stepping sources if short half-times of repair are present in tissues. Int J Radiat Oncol Biol Phys 1997,37(4):877-883. 10.1016/S0360-3016(96)00565-2CrossRefPubMed Fowler JF, Van Limbergen EF: Biological effect of pulsed dose rate brachytherapy with stepping sources if short half-times of repair are present in tissues. Int J Radiat Oncol Biol Phys 1997,37(4):877-883. 10.1016/S0360-3016(96)00565-2CrossRefPubMed
8.
go back to reference Pop LA, Millar WT, van der Plas M, van der Kogel AJ: Radiation tolerance of rat spinal cord to pulsed dose rate (PDR-) brachytherapy: the impact of differences in temporal dose distribution. Radiother Oncol 2000,55(3):301-315. 10.1016/S0167-8140(00)00205-XCrossRefPubMed Pop LA, Millar WT, van der Plas M, van der Kogel AJ: Radiation tolerance of rat spinal cord to pulsed dose rate (PDR-) brachytherapy: the impact of differences in temporal dose distribution. Radiother Oncol 2000,55(3):301-315. 10.1016/S0167-8140(00)00205-XCrossRefPubMed
9.
go back to reference Wybranski C, Seidensticker M, Mohnike K, Kropf S, Wust P, Ricke J, Lüdemann L: In vivo assessment of dose volume and dose gradient effects on the tolerance dose of small liver volumes after single-fraction high-dose-rate 192Ir irradiation. Radiat Res 2009,172(5):598-606. 10.1667/RR1773.1CrossRefPubMed Wybranski C, Seidensticker M, Mohnike K, Kropf S, Wust P, Ricke J, Lüdemann L: In vivo assessment of dose volume and dose gradient effects on the tolerance dose of small liver volumes after single-fraction high-dose-rate 192Ir irradiation. Radiat Res 2009,172(5):598-606. 10.1667/RR1773.1CrossRefPubMed
10.
go back to reference Ricke J, Seidensticker M, Lüdemann L, Pech M, Wieners G, Hengst S, Mohnike K, Cho CH, Hanninen EL, Al-Abadi H, Felix R, Wust P: In vivo assessment of the tolerance dose of small liver volumes after single-fraction HDR irradiation. Int J Radiat Oncol Biol Phys 2005,62(3):776-84. 10.1016/j.ijrobp.2004.11.022CrossRefPubMed Ricke J, Seidensticker M, Lüdemann L, Pech M, Wieners G, Hengst S, Mohnike K, Cho CH, Hanninen EL, Al-Abadi H, Felix R, Wust P: In vivo assessment of the tolerance dose of small liver volumes after single-fraction HDR irradiation. Int J Radiat Oncol Biol Phys 2005,62(3):776-84. 10.1016/j.ijrobp.2004.11.022CrossRefPubMed
11.
go back to reference Clement O, Siauve N, Cuenod CA, Vuillemin-Bodaghi V, Leconte I, Frija G: Mechanisms of action of liver contrast agents: impact for clinical use. J Comput Assist Tomogr 1999,23(Suppl 1):S45-52.CrossRefPubMed Clement O, Siauve N, Cuenod CA, Vuillemin-Bodaghi V, Leconte I, Frija G: Mechanisms of action of liver contrast agents: impact for clinical use. J Comput Assist Tomogr 1999,23(Suppl 1):S45-52.CrossRefPubMed
12.
go back to reference Kirchin MA, Pirovano GP, Spinazzi A: Gadobenate dimeglumine (Gd-BOPTA). An overview. Invest Radiol 1998,33(11):798-809. 10.1097/00004424-199811000-00003CrossRefPubMed Kirchin MA, Pirovano GP, Spinazzi A: Gadobenate dimeglumine (Gd-BOPTA). An overview. Invest Radiol 1998,33(11):798-809. 10.1097/00004424-199811000-00003CrossRefPubMed
13.
go back to reference de Haen C, Ferla RL, Maggioni F: Gadobenate dimeglumine 0.5 M solution for injection (MultiHance) as contrast agent for magnetic resonance imaging of the liver: mechanistic studies in animals. J Comput Assist Tomogr 1999,23(Suppl 1):S169-79.CrossRefPubMed de Haen C, Ferla RL, Maggioni F: Gadobenate dimeglumine 0.5 M solution for injection (MultiHance) as contrast agent for magnetic resonance imaging of the liver: mechanistic studies in animals. J Comput Assist Tomogr 1999,23(Suppl 1):S169-79.CrossRefPubMed
14.
go back to reference Rohlfing T, West JB, Beier J, Liebig T, Taschner CA, Thomale UW: Registration of functional and anatomical MRI: accuracy assessment and application in navigated neurosurgery. Comput Aided Surg 2000,5(6):414-25. 10.3109/10929080009148901CrossRefPubMed Rohlfing T, West JB, Beier J, Liebig T, Taschner CA, Thomale UW: Registration of functional and anatomical MRI: accuracy assessment and application in navigated neurosurgery. Comput Aided Surg 2000,5(6):414-25. 10.3109/10929080009148901CrossRefPubMed
15.
go back to reference Burton P, Gurrin L, Sly P: Extending the simple linear regression model to account for correlated responses: an introduction to generalized estimating equations and multi-level mixed modelling. Stat Med 1998,17(11):1261-91. 10.1002/(SICI)1097-0258(19980615)17:11<1261::AID-SIM846>3.0.CO;2-ZCrossRefPubMed Burton P, Gurrin L, Sly P: Extending the simple linear regression model to account for correlated responses: an introduction to generalized estimating equations and multi-level mixed modelling. Stat Med 1998,17(11):1261-91. 10.1002/(SICI)1097-0258(19980615)17:11<1261::AID-SIM846>3.0.CO;2-ZCrossRefPubMed
16.
go back to reference Zeger SL, Liang KY: Longitudinal data analysis for discrete and continuous outcomes. Biometrics 1986, 42: 121-30. 10.2307/2531248CrossRefPubMed Zeger SL, Liang KY: Longitudinal data analysis for discrete and continuous outcomes. Biometrics 1986, 42: 121-30. 10.2307/2531248CrossRefPubMed
17.
go back to reference Wulf J, Hadinger U, Oppitz U, Olshausen B, Flentje M: Stereotactic radiotherapy of extracranial targets: CT-simulation and accuracy of treatment in the stereotactic body frame. Radiother Oncol 2000,57(2):225-36. 10.1016/S0167-8140(00)00226-7CrossRefPubMed Wulf J, Hadinger U, Oppitz U, Olshausen B, Flentje M: Stereotactic radiotherapy of extracranial targets: CT-simulation and accuracy of treatment in the stereotactic body frame. Radiother Oncol 2000,57(2):225-36. 10.1016/S0167-8140(00)00226-7CrossRefPubMed
18.
go back to reference Herfarth KK, Debus J, Lohr F, Bahner ML, Fritz P, Hoss A, Schlegel W, Wannenmacher MF: Extracranial stereotactic radiation therapy: set-up accuracy of patients treated for liver metastases. Int J Radiat Oncol Biol Phys 2000,46(2):329-35. 10.1016/S0360-3016(99)00413-7CrossRefPubMed Herfarth KK, Debus J, Lohr F, Bahner ML, Fritz P, Hoss A, Schlegel W, Wannenmacher MF: Extracranial stereotactic radiation therapy: set-up accuracy of patients treated for liver metastases. Int J Radiat Oncol Biol Phys 2000,46(2):329-35. 10.1016/S0360-3016(99)00413-7CrossRefPubMed
19.
go back to reference Wulf J, Hadinger U, Oppitz U, Thiele W, Flentje M: Impact of target reproducibility on tumor dose in stereotactic radiotherapy of targets in the lung and liver. Radiother Oncol 2003,66(2):141-50. 10.1016/S0167-8140(02)00372-9CrossRefPubMed Wulf J, Hadinger U, Oppitz U, Thiele W, Flentje M: Impact of target reproducibility on tumor dose in stereotactic radiotherapy of targets in the lung and liver. Radiother Oncol 2003,66(2):141-50. 10.1016/S0167-8140(02)00372-9CrossRefPubMed
20.
go back to reference Foster W, Cunha JA, Hsu IC, Weinberg V, Krishnamurthy D, Pouliot J: Dosimetric impact of interfraction catheter movement in high-dose rate prostate brachytherapy. Int J Radiat Oncol Biol Phys 2011, 80: 85-90. 10.1016/j.ijrobp.2010.01.016CrossRefPubMed Foster W, Cunha JA, Hsu IC, Weinberg V, Krishnamurthy D, Pouliot J: Dosimetric impact of interfraction catheter movement in high-dose rate prostate brachytherapy. Int J Radiat Oncol Biol Phys 2011, 80: 85-90. 10.1016/j.ijrobp.2010.01.016CrossRefPubMed
21.
go back to reference Simnor T, Li S, Lowe G, Ostler P, Bryant L, Chapman C, Inchley D, Hoskin PJ: Justification for inter-fraction correction of catheter movement in fractionated high dose-rate brachytherapy treatment of prostate cancer. Radiother Oncol 2009,93(2):253-258. 10.1016/j.radonc.2009.09.015CrossRefPubMed Simnor T, Li S, Lowe G, Ostler P, Bryant L, Chapman C, Inchley D, Hoskin PJ: Justification for inter-fraction correction of catheter movement in fractionated high dose-rate brachytherapy treatment of prostate cancer. Radiother Oncol 2009,93(2):253-258. 10.1016/j.radonc.2009.09.015CrossRefPubMed
22.
go back to reference Kim Y, Hsu IC, Pouliot J: Measurement of craniocaudal catheter displacement between fractions in computed tomography-based high dose rate brachytherapy of prostate cancer. J Appl Clin Med Phys 2007,8(4):2415-2415.PubMed Kim Y, Hsu IC, Pouliot J: Measurement of craniocaudal catheter displacement between fractions in computed tomography-based high dose rate brachytherapy of prostate cancer. J Appl Clin Med Phys 2007,8(4):2415-2415.PubMed
23.
go back to reference Brock KK, Dawson LA, Sharpe MB, Moseley DJ, Jaffray DA: Feasibility of a novel deformable image registration technique to facilitate classification, targeting, and monitoring of tumor and normal tissue. Int J Radiat Oncol Biol Phys 2006,64(4):1245-1254. 10.1016/j.ijrobp.2005.10.027CrossRefPubMed Brock KK, Dawson LA, Sharpe MB, Moseley DJ, Jaffray DA: Feasibility of a novel deformable image registration technique to facilitate classification, targeting, and monitoring of tumor and normal tissue. Int J Radiat Oncol Biol Phys 2006,64(4):1245-1254. 10.1016/j.ijrobp.2005.10.027CrossRefPubMed
24.
go back to reference Voroney JP, Brock KK, Eccles C, Haider M, Dawson LA: Prospective comparison of computed tomography and magnetic resonance imaging for liver cancer delineation using deformable image registration. Int J Radiat Oncol Biol Phys 2006,66(3):780-791. 10.1016/j.ijrobp.2006.05.035CrossRefPubMed Voroney JP, Brock KK, Eccles C, Haider M, Dawson LA: Prospective comparison of computed tomography and magnetic resonance imaging for liver cancer delineation using deformable image registration. Int J Radiat Oncol Biol Phys 2006,66(3):780-791. 10.1016/j.ijrobp.2006.05.035CrossRefPubMed
25.
go back to reference van Dalen JA, Vogel W, Huisman HJ, Oyen WJ, Jager GJ, Karssemeijer N: Accuracy of rigid CT-FDG-PET image registration of the liver. Phys Med Biol 2004,49(23):5393-5405. 10.1088/0031-9155/49/23/014CrossRefPubMed van Dalen JA, Vogel W, Huisman HJ, Oyen WJ, Jager GJ, Karssemeijer N: Accuracy of rigid CT-FDG-PET image registration of the liver. Phys Med Biol 2004,49(23):5393-5405. 10.1088/0031-9155/49/23/014CrossRefPubMed
26.
go back to reference Carrillo A, Duerk JL, Lewin JS, Wilson DL: Semiautomatic 3-D image registration as applied to interventional MRI liver cancer treatment. IEEE Trans Med Imaging 2000,19(3):175-185. 10.1109/42.845176CrossRefPubMed Carrillo A, Duerk JL, Lewin JS, Wilson DL: Semiautomatic 3-D image registration as applied to interventional MRI liver cancer treatment. IEEE Trans Med Imaging 2000,19(3):175-185. 10.1109/42.845176CrossRefPubMed
27.
go back to reference Christina Lee WC, Tublin ME, Chapman BE: Registration of MR and CT images of the liver: comparison of voxel similarity and surface based registration algorithms. Comput Methods Programs Biomed 2005,78(2):101-114. 10.1016/j.cmpb.2004.12.006CrossRefPubMed Christina Lee WC, Tublin ME, Chapman BE: Registration of MR and CT images of the liver: comparison of voxel similarity and surface based registration algorithms. Comput Methods Programs Biomed 2005,78(2):101-114. 10.1016/j.cmpb.2004.12.006CrossRefPubMed
28.
go back to reference Elhawary H, Oguro S, Tuncali K, Morrison PR, Tatli S, Shyn PB, Silverman SG, Hata N: Multimodality non-rigid image registration for planning, targeting and monitoring during CT-guided percutaneous liver tumor cryoablation. Acad Radiol 2010,17(11):1334-1344. 10.1016/j.acra.2010.06.004PubMedCentralCrossRefPubMed Elhawary H, Oguro S, Tuncali K, Morrison PR, Tatli S, Shyn PB, Silverman SG, Hata N: Multimodality non-rigid image registration for planning, targeting and monitoring during CT-guided percutaneous liver tumor cryoablation. Acad Radiol 2010,17(11):1334-1344. 10.1016/j.acra.2010.06.004PubMedCentralCrossRefPubMed
Metadata
Title
In vivo assessment of catheter positioning accuracy and prolonged irradiation time on liver tolerance dose after single-fraction 192Ir high-dose-rate brachytherapy
Authors
Lutz Lüdemann
Christian Wybranski
Max Seidensticker
Konrad Mohnike
Siegfried Kropf
Peter Wust
Jens Ricke
Publication date
01-12-2011
Publisher
BioMed Central
Published in
Radiation Oncology / Issue 1/2011
Electronic ISSN: 1748-717X
DOI
https://doi.org/10.1186/1748-717X-6-107

Other articles of this Issue 1/2011

Radiation Oncology 1/2011 Go to the issue