Published in: International Orthopaedics 7/2015

01-07-2015 | Original Paper

In vitro effects on mobile polyethylene insert under highly demanding daily activities: stair climbing

Authors: Sami Abdel Jaber, Paola Taddei, Silvia Tozzi, Alessandra Sudanese, Saverio Affatato

Published in: International Orthopaedics | Issue 7/2015

Login to get access

Abstract

Purpose

Wear and survival of total joint replacements do not depend on the duration of the implant in situ, but rather on the amount of its use, i.e. the patient’s activity level. With this in mind, the present study was driven by two questions: (1) How does total knee replacement (TKR) respond to the simulation of daily highly demanding activities? (2) Are certain activities to be advised against or, on the contrary, useful to implanted patients, in order to reduce wear of TKR and its related problems?

Methods

One set of the same total knee prosthesis (TKP), equal in design and size, was tested on a three-plus-one knee joint simulator for two million cycles using a highly demanding daily load waveform, replicating a stair-climbing movement. The results were compared with a set of TKP previously tested with the ISO level walking task. A digital microscope was used to characterise the superficial structure of all the TKPs. Gravimetric and micro-Raman spectroscopic analyses were carried out on the polyethylene inserts. Visual comparison with in vivo explants was carried out.

Results

The average volumetric mass loss after two million cycles was 44 ± 6 mm3. Microscope examinations showed some deep scratches along the flexion/extension movements for all the components. Also, the metallic backside surface showed intense non-linear scratches and the polyethylene counterface was characterised by some craters. A decrease in crystallinity, induced by mechanical stress was observed on all polyethylene components and was quantitatively confirmed by the orthorhombic fraction αo value.

Conclusions

The results of this study demonstrated that the forces and motion sustained by the knee are highly activity-dependent. Moreover, this test confirmed that under more severe conditions, the material properties change according to a different wear mechanism and a decrease in crystallinity occurs. Loading characteristics for specific activities should be considered for the design of functional and robust TKRs.
Literature
3.
go back to reference Vasileios N, Dimitrios C, Babis G (2014) Common controversies in total knee replacement surgery: current evidence. World J Orthop 5(4):460–468CrossRef Vasileios N, Dimitrios C, Babis G (2014) Common controversies in total knee replacement surgery: current evidence. World J Orthop 5(4):460–468CrossRef
7.
go back to reference Rohman L, Hadi S, Whitwell G (2014) Surgeons’ knowledge about the costs of orthopaedic implants. J Orthop Surg 22:221–223 Rohman L, Hadi S, Whitwell G (2014) Surgeons’ knowledge about the costs of orthopaedic implants. J Orthop Surg 22:221–223
10.
go back to reference Morlock M, Schneider E, Bluhm A et al (2001) Duration and frequency of every day activities in total hip patients. J Biomech 34:873–881PubMedCrossRef Morlock M, Schneider E, Bluhm A et al (2001) Duration and frequency of every day activities in total hip patients. J Biomech 34:873–881PubMedCrossRef
11.
go back to reference Windsor RE, Scuderi GR, Moran MC, Insall JN (1989) Mechanisms of failure of the femoral and tibial components in total knee arthroplasty. Clin Orthop Relat Res 1989(248):15–19. doi:10.1097/00003086-198911000-00005, discussion 19–20 Windsor RE, Scuderi GR, Moran MC, Insall JN (1989) Mechanisms of failure of the femoral and tibial components in total knee arthroplasty. Clin Orthop Relat Res 1989(248):15–19. doi:10.​1097/​00003086-198911000-00005, discussion 19–20
12.
go back to reference Cheng T, Pan X, Liu T, Zhang X (2012) Tibial component designs in primary total knee arthroplasty: should we reconsider all-polyethylene component? Knee Surg Sport Traumatol Arthrosc 20:1438–1449. doi:10.1007/s00167-011-1682-y CrossRef Cheng T, Pan X, Liu T, Zhang X (2012) Tibial component designs in primary total knee arthroplasty: should we reconsider all-polyethylene component? Knee Surg Sport Traumatol Arthrosc 20:1438–1449. doi:10.​1007/​s00167-011-1682-y CrossRef
13.
go back to reference Grupp TM, Kaddick C, Schwiesau J et al (2009) Fixed and mobile bearing total knee arthroplasty—influence on wear generation, corresponding wear areas, knee kinematics and particle composition. Clin Biomech 24:210–217CrossRef Grupp TM, Kaddick C, Schwiesau J et al (2009) Fixed and mobile bearing total knee arthroplasty—influence on wear generation, corresponding wear areas, knee kinematics and particle composition. Clin Biomech 24:210–217CrossRef
14.
go back to reference Schmalzried TP, Shepherd EF, Dorey FJ et al (2000) The John Charnley award. Wear is a function of use, not time. Clin Orthop Relat Res 381:36–46PubMedCrossRef Schmalzried TP, Shepherd EF, Dorey FJ et al (2000) The John Charnley award. Wear is a function of use, not time. Clin Orthop Relat Res 381:36–46PubMedCrossRef
15.
go back to reference Rowe PJ, Myles CM, Walker C, Nutton R (2000) Knee joint kinematics in gait and other functional activities measured using flexible electrogoniometry: how much knee motion is sufficient for normal daily life? Gait Posture 12:143–155. doi:10.1016/S0966-6362(00)00060-6 PubMedCrossRef Rowe PJ, Myles CM, Walker C, Nutton R (2000) Knee joint kinematics in gait and other functional activities measured using flexible electrogoniometry: how much knee motion is sufficient for normal daily life? Gait Posture 12:143–155. doi:10.​1016/​S0966-6362(00)00060-6 PubMedCrossRef
19.
go back to reference Desjardins JD, Walker PS, Haider H, Perry J (2000) The use of a force-controlled dynamic knee simulator to quantify the mechanical performance of total knee replacement designs during functional activity. J Biomech 33:1231–1242PubMedCrossRef Desjardins JD, Walker PS, Haider H, Perry J (2000) The use of a force-controlled dynamic knee simulator to quantify the mechanical performance of total knee replacement designs during functional activity. J Biomech 33:1231–1242PubMedCrossRef
20.
go back to reference Affatato S, Grillini L, Battaglia S et al (2013) Does knee implant size affect wear variability? Tribol Int 66:174–181CrossRef Affatato S, Grillini L, Battaglia S et al (2013) Does knee implant size affect wear variability? Tribol Int 66:174–181CrossRef
21.
go back to reference Essner A, Herrera L, Hughes P, Kester M (2011) The influence of material and design on total knee replacement wear. J Knee Surg 24:9–17PubMedCrossRef Essner A, Herrera L, Hughes P, Kester M (2011) The influence of material and design on total knee replacement wear. J Knee Surg 24:9–17PubMedCrossRef
22.
go back to reference Haider H (2009) Tribological assessment of UHMWPE in the knee. In: Kurtz S (ed) UHMWPE biomaterials handbook, 2nd edn. Elsevier, Amsterdam, pp 381–408 Haider H (2009) Tribological assessment of UHMWPE in the knee. In: Kurtz S (ed) UHMWPE biomaterials handbook, 2nd edn. Elsevier, Amsterdam, pp 381–408
23.
go back to reference ISO DIS 14243-1 (2002) Implants for surgery—Wear of total knee joint prostheses—Part 1: loading and displacement paramenters for wear-testing machines with load control and corresponding enviromental conditions for tests. International Organization for Standardization, Geneva ISO DIS 14243-1 (2002) Implants for surgery—Wear of total knee joint prostheses—Part 1: loading and displacement paramenters for wear-testing machines with load control and corresponding enviromental conditions for tests. International Organization for Standardization, Geneva
24.
go back to reference ISO DIS 14243-3 (2004) Implants for surgery—Wear of total knee joint prostheses—Part 1: loading and displacement paramenters for wear-testing machines with displacement control and corresponding enviromental conditions for tests. International Organization for Standardization, Geneva ISO DIS 14243-3 (2004) Implants for surgery—Wear of total knee joint prostheses—Part 1: loading and displacement paramenters for wear-testing machines with displacement control and corresponding enviromental conditions for tests. International Organization for Standardization, Geneva
28.
31.
go back to reference Battaglia S, Belvedere C, Abdel Jaber S, Affatato S, D’Angeli V, Leradini A (2014) A new protocol from real joint motion data for wear simulation in total knee arthroplasty: stair climbing. Med Eng Phys 36:1605–1610PubMedCrossRef Battaglia S, Belvedere C, Abdel Jaber S, Affatato S, D’Angeli V, Leradini A (2014) A new protocol from real joint motion data for wear simulation in total knee arthroplasty: stair climbing. Med Eng Phys 36:1605–1610PubMedCrossRef
33.
go back to reference Affatato S, Vandelli C, Bordini B, Toni A (2001) Fluid absorption study in ultra-high molecular weight polyethylene (UHMWPE) sterilized and unsterilized acetabular cups. Proc Inst Mech Eng [H] 215:107–111CrossRef Affatato S, Vandelli C, Bordini B, Toni A (2001) Fluid absorption study in ultra-high molecular weight polyethylene (UHMWPE) sterilized and unsterilized acetabular cups. Proc Inst Mech Eng [H] 215:107–111CrossRef
34.
go back to reference ISO DIS 14243-2 (2009) Implants for surgery—Wear of total knee-joint prostheses—Part 2: methods of measurement. International Organization for Standardization ISO DIS 14243-2 (2009) Implants for surgery—Wear of total knee-joint prostheses—Part 2: methods of measurement. International Organization for Standardization
35.
go back to reference Belvedere C, Tamarri S, Notarangelo DP et al (2013) Three-dimensional motion analysis of the human knee joint: comparison between intra- and post-operative measurements. Knee Surg Sports Traumatol Arthrosc 21:2375–83. doi:10.1007/s00167-012-2271-4 PubMedCrossRef Belvedere C, Tamarri S, Notarangelo DP et al (2013) Three-dimensional motion analysis of the human knee joint: comparison between intra- and post-operative measurements. Knee Surg Sports Traumatol Arthrosc 21:2375–83. doi:10.​1007/​s00167-012-2271-4 PubMedCrossRef
41.
42.
go back to reference McHellop H, Lu BBP (1992) Friction, lubrication and wear of cobalt-chromium, alumina and zirconia hip prostheses compared on a joint simulator. In: Trans 38th Ann Meeting Orthop Res Soc, Washington D.C., p 402 McHellop H, Lu BBP (1992) Friction, lubrication and wear of cobalt-chromium, alumina and zirconia hip prostheses compared on a joint simulator. In: Trans 38th Ann Meeting Orthop Res Soc, Washington D.C., p 402
43.
go back to reference Mutter R, Stille W, Strobl G (1993) Transition regions and surface melting in partially crystalline polyethylene: a Raman spectroscopic study. J Polym Sci B Polym Phys 31:99–105CrossRef Mutter R, Stille W, Strobl G (1993) Transition regions and surface melting in partially crystalline polyethylene: a Raman spectroscopic study. J Polym Sci B Polym Phys 31:99–105CrossRef
45.
go back to reference Simis KS, Bistolfi A, Bellare A, Pruitt LA (2006) The combined effects of crosslinking and high crystallinity on the microstructural and mechanical properties of ultra high molecular weight polyethylene. Biomaterials 27:1688–94PubMedCrossRef Simis KS, Bistolfi A, Bellare A, Pruitt LA (2006) The combined effects of crosslinking and high crystallinity on the microstructural and mechanical properties of ultra high molecular weight polyethylene. Biomaterials 27:1688–94PubMedCrossRef
46.
go back to reference Karuppiah KSK, Bruck AL, Wang J, Xu ZH, Li X (2008) Friction and wear behavior of ultra-high molecular weight polyethylene as a function of polymer crystallinity. Acta Biomater 4:1401–1410CrossRef Karuppiah KSK, Bruck AL, Wang J, Xu ZH, Li X (2008) Friction and wear behavior of ultra-high molecular weight polyethylene as a function of polymer crystallinity. Acta Biomater 4:1401–1410CrossRef
Metadata
Title
In vitro effects on mobile polyethylene insert under highly demanding daily activities: stair climbing
Authors
Sami Abdel Jaber
Paola Taddei
Silvia Tozzi
Alessandra Sudanese
Saverio Affatato
Publication date
01-07-2015
Publisher
Springer Berlin Heidelberg
Published in
International Orthopaedics / Issue 7/2015
Print ISSN: 0341-2695
Electronic ISSN: 1432-5195
DOI
https://doi.org/10.1007/s00264-014-2622-5

Other articles of this Issue 7/2015

International Orthopaedics 7/2015 Go to the issue