Published in: Neuroradiology 9/2010

01-09-2010 | Diagnostic Neuroradiology

In vitro comparison of different carotid artery stents: a pixel-by-pixel analysis using CT angiography and contrast-enhanced MR angiography at 1.5 and 3 T

Authors: Michael Lettau, Annett Sauer, Sabine Heiland, Stefan Rohde, Julia Reinhardt, Martin Bendszus, Stefan Hähnel

Published in: Neuroradiology | Issue 9/2010

Login to get access

Abstract

Introduction

CT angiography (CTA) and MR angiography (MRA) are increasingly used methods for evaluation of stented vessel segments. The purpose of this study was to compare CTA, contrast-enhanced MRA (CEMRA) at 1.5 T, and CEMRA at 3 T for the visualization of carotid artery stents and to define the best noninvasive imaging technique for each stent.

Methods

CTA and CEMRA appearances of 18 carotid artery stents of different designs and sizes (4.0 to 10.0 mm) were investigated in vitro. The profile of the contrast-to-noise ratio (CNR) of the lumen of each stent was calculated semiautomatically by a pixel-by-pixel analysis using the medical imaging software OSIRIS®. For each stent, artificial lumen narrowing (ALN) was calculated.

Results

In all but one stents, ALN was lower on CEMRA at 3 T than at 1.5 T. With CEMRA at 3 T and at 1.5 T, ALN in most nitinol stents was lower than in the groups of stainless steel and cobalt alloy stents. In most nitinol stents, ALN on CEMRA at 3 T was lower than on CTA. In all stainless steel stents and cobalt alloy stents, ALN was lower on CTA than on CEMRA. With CTA and CEMRA, in most stents ALN decreased with increasing stent diameter.

Conclusion

CTA and CEMRA evaluation of vessel patency after stent placement is possible, but considerably impaired by ALN. Investigators should be informed about the method of choice for every stent.
Literature
1.
go back to reference Eckstein HH, Ringleb P, Allenberg JR, Berger J, Fraedrich G, Hacke W, Hennerici M, Stingele R, Fiehler J, Zeumer H, Jansen O (2008) Results of the stent-protected angioplasty versus carotid endarterectomy (SPACE) study to treat symptomatic stenoses at 2 years: a multinational, prospective, randomised trial. Lancet Neurol 7:893–902CrossRefPubMed Eckstein HH, Ringleb P, Allenberg JR, Berger J, Fraedrich G, Hacke W, Hennerici M, Stingele R, Fiehler J, Zeumer H, Jansen O (2008) Results of the stent-protected angioplasty versus carotid endarterectomy (SPACE) study to treat symptomatic stenoses at 2 years: a multinational, prospective, randomised trial. Lancet Neurol 7:893–902CrossRefPubMed
2.
go back to reference Grunwald IQ, Papanagiotou P, Roth C, Fassbender K, Karp K, Krick C, Schieber H, Müller M, Haass A, Reith W (2009) Lesion load in unprotected carotid artery stenting. Neuroradiology 51:313–317CrossRefPubMed Grunwald IQ, Papanagiotou P, Roth C, Fassbender K, Karp K, Krick C, Schieber H, Müller M, Haass A, Reith W (2009) Lesion load in unprotected carotid artery stenting. Neuroradiology 51:313–317CrossRefPubMed
3.
go back to reference Bendszus M, Koltzenburg M, Burger R, Warmuth-Metz M, Hofmann E, Solymosi L (1999) Silent embolism in diagnostic cerebral angiography and neurointerventional procedures: a prospective study. Lancet 354:1594–1597CrossRefPubMed Bendszus M, Koltzenburg M, Burger R, Warmuth-Metz M, Hofmann E, Solymosi L (1999) Silent embolism in diagnostic cerebral angiography and neurointerventional procedures: a prospective study. Lancet 354:1594–1597CrossRefPubMed
4.
go back to reference Dawkins AA, Evans AL, Wattam J, Romanowski CAJ, Connolly DJA, Hodgson TJ, Coley SC (2007) Complications of cerebral angiography: a prospective analysis of 2, 924 consecutive procedures. Neuroradiology 49:753–759CrossRefPubMed Dawkins AA, Evans AL, Wattam J, Romanowski CAJ, Connolly DJA, Hodgson TJ, Coley SC (2007) Complications of cerebral angiography: a prospective analysis of 2, 924 consecutive procedures. Neuroradiology 49:753–759CrossRefPubMed
5.
go back to reference Hähnel S, Trossbach M, Braun C, Heiland S, Knauth M, Sartor K, Hartmann M (2003) Small-vessel stents for intracranial angioplasty: in vitro comparison of different stent designs and sizes by using CT angiography. AJNR Am J Neuroradiol 24:1512–1516PubMed Hähnel S, Trossbach M, Braun C, Heiland S, Knauth M, Sartor K, Hartmann M (2003) Small-vessel stents for intracranial angioplasty: in vitro comparison of different stent designs and sizes by using CT angiography. AJNR Am J Neuroradiol 24:1512–1516PubMed
6.
go back to reference Lenhart M, Völk M, Manke C, Nitz WR, Strotzer M, Feuerbach S, Link J (2000) Stent appearance at contrast-enhanced MR angiography: in vitro examination with 14 stents. Radiology 217:173–178PubMed Lenhart M, Völk M, Manke C, Nitz WR, Strotzer M, Feuerbach S, Link J (2000) Stent appearance at contrast-enhanced MR angiography: in vitro examination with 14 stents. Radiology 217:173–178PubMed
7.
go back to reference Wall A, Kugel H, Bachman R, Matuszewski L, Krämer S, Heindel W, Maintz D (2005) 3.0 T vs. 1.5 T MR angiography: in vitro comparison of intravascular stent artifacts. J Magn Reson Imaging 22:772–779CrossRefPubMed Wall A, Kugel H, Bachman R, Matuszewski L, Krämer S, Heindel W, Maintz D (2005) 3.0 T vs. 1.5 T MR angiography: in vitro comparison of intravascular stent artifacts. J Magn Reson Imaging 22:772–779CrossRefPubMed
8.
go back to reference Anderson CM, Saloner D, Tsuruda JS, Shapeero LG, Lee RE (1990) Artifacts in maximum-intensity-projection display of MR angiograms. AJR Am J Roentgenol 154:623–629PubMed Anderson CM, Saloner D, Tsuruda JS, Shapeero LG, Lee RE (1990) Artifacts in maximum-intensity-projection display of MR angiograms. AJR Am J Roentgenol 154:623–629PubMed
9.
go back to reference Schenck JF (1996) The role of magnetic susceptibility in magnetic resonance imaging: MRI magnetic compatibility of the first and second kinds. Med Phys 23:815–850CrossRefPubMed Schenck JF (1996) The role of magnetic susceptibility in magnetic resonance imaging: MRI magnetic compatibility of the first and second kinds. Med Phys 23:815–850CrossRefPubMed
10.
go back to reference Bakker CJ, Bhagwandien R, Moerland MA, Ramos LM (1994) Simulation of susceptibility artifacts in 2D and 3D fourier transform spin-echo and gradient-echo magnetic resonance imaging. Magn Reson Imaging 12:767–774CrossRefPubMed Bakker CJ, Bhagwandien R, Moerland MA, Ramos LM (1994) Simulation of susceptibility artifacts in 2D and 3D fourier transform spin-echo and gradient-echo magnetic resonance imaging. Magn Reson Imaging 12:767–774CrossRefPubMed
11.
go back to reference Bartels LW, Bakker CJ, Viergever MA (2002) Improved lumen visualization in metallic vascular implants by reducing RF artifacts. Magn Reson Med 47:171–180CrossRefPubMed Bartels LW, Bakker CJ, Viergever MA (2002) Improved lumen visualization in metallic vascular implants by reducing RF artifacts. Magn Reson Med 47:171–180CrossRefPubMed
12.
go back to reference Wang Y, Truong TN, Yen C, Bilecen D, Watts R, Trost DW, Prince MR (2003) Quantitative evaluation of susceptibility and shielding effects of nitinol, platinum, cobalt-alloy, and stainless steel stents. Magn Reson Med 49:972–976CrossRefPubMed Wang Y, Truong TN, Yen C, Bilecen D, Watts R, Trost DW, Prince MR (2003) Quantitative evaluation of susceptibility and shielding effects of nitinol, platinum, cobalt-alloy, and stainless steel stents. Magn Reson Med 49:972–976CrossRefPubMed
13.
go back to reference Klemm T, Duda S, Machann J, Seekamp-Rahn K, Schnieder L, Claussen CD, Schick F (2000) MR imaging in the presence of vascular stents: a systematic assessment of artifacts for various stent orientations, sequence types, and field strengths. J Magn Reson Imaging 12:606–615CrossRefPubMed Klemm T, Duda S, Machann J, Seekamp-Rahn K, Schnieder L, Claussen CD, Schick F (2000) MR imaging in the presence of vascular stents: a systematic assessment of artifacts for various stent orientations, sequence types, and field strengths. J Magn Reson Imaging 12:606–615CrossRefPubMed
14.
go back to reference Graf H, Klemm T, Lauer UA, Duda S, Claussen CD, Schick F (2003) Systematics of imaging artifacts in MRT caused by metallic vascular implants (stents). Röfo 175:1711–1719PubMed Graf H, Klemm T, Lauer UA, Duda S, Claussen CD, Schick F (2003) Systematics of imaging artifacts in MRT caused by metallic vascular implants (stents). Röfo 175:1711–1719PubMed
15.
go back to reference Krämer SC, Wall A, Maintz D, Bachmann R, Kugel H, Heindel W (2004) 3.0 Tesla magnetic resonance angiography of endovascular aortic stent grafts: phantom measurements in comparison with 1.5 Tesla. Invest Radiol 39:413–417CrossRefPubMed Krämer SC, Wall A, Maintz D, Bachmann R, Kugel H, Heindel W (2004) 3.0 Tesla magnetic resonance angiography of endovascular aortic stent grafts: phantom measurements in comparison with 1.5 Tesla. Invest Radiol 39:413–417CrossRefPubMed
16.
go back to reference Hähnel S, Nguyen-Trong TH, Rohde S, Hartmann M, Braun C, Sartor K, Heiland S (2006) 3.0 Tesla contrast-enhanced MR angiography of carotid artery stents: in vitro measurements as compared with 1.5 Tesla. J Neuroradiol 33:75–80CrossRefPubMed Hähnel S, Nguyen-Trong TH, Rohde S, Hartmann M, Braun C, Sartor K, Heiland S (2006) 3.0 Tesla contrast-enhanced MR angiography of carotid artery stents: in vitro measurements as compared with 1.5 Tesla. J Neuroradiol 33:75–80CrossRefPubMed
17.
go back to reference Buecker A, Spuentrup E, Ruebben A, Günther RW (2002) Artifact-free in-stent lumen visualization by standard magnetic resonance angiography using a new metallic magnetic resonance imaging stent. Circulation 105:1772–1775CrossRefPubMed Buecker A, Spuentrup E, Ruebben A, Günther RW (2002) Artifact-free in-stent lumen visualization by standard magnetic resonance angiography using a new metallic magnetic resonance imaging stent. Circulation 105:1772–1775CrossRefPubMed
18.
go back to reference Bartels LW, Smits HF, Bakker CJ, Viergever MA (2001) MR imaging of vascular stents: effects of susceptibility, flow, and radiofrequency eddy currents. J Vasc Interv Radiol 12:365–371CrossRefPubMed Bartels LW, Smits HF, Bakker CJ, Viergever MA (2001) MR imaging of vascular stents: effects of susceptibility, flow, and radiofrequency eddy currents. J Vasc Interv Radiol 12:365–371CrossRefPubMed
19.
go back to reference Holton A, Walsh E, Anayiotos A, Pohost G, Venugopalan R (2002) Comparative MRI compatibility of 316 L stainless steel alloy and nickel-titanium alloy stents. J Cardiovasc Magn Reson 4:423–430CrossRefPubMed Holton A, Walsh E, Anayiotos A, Pohost G, Venugopalan R (2002) Comparative MRI compatibility of 316 L stainless steel alloy and nickel-titanium alloy stents. J Cardiovasc Magn Reson 4:423–430CrossRefPubMed
20.
go back to reference Meyer JM, Buecker A, Spuentrup E, Schuermann K, Huetten M, Hilgers RD, van Vaals JJ, Guenther RW (2001) Improved in-stent magnetic resonance angiography with high flip angle excitation. Invest Radiol 36:677–681CrossRefPubMed Meyer JM, Buecker A, Spuentrup E, Schuermann K, Huetten M, Hilgers RD, van Vaals JJ, Guenther RW (2001) Improved in-stent magnetic resonance angiography with high flip angle excitation. Invest Radiol 36:677–681CrossRefPubMed
21.
go back to reference Straube T, Wolf S, Flesser A, Deli M, Alfke K, Nabavi A, Jansen O (2005) MRI of carotid stents: influence of stent properties and sequence parameters on visualization of the carotid artery lumen. Röfo 177:375–380PubMed Straube T, Wolf S, Flesser A, Deli M, Alfke K, Nabavi A, Jansen O (2005) MRI of carotid stents: influence of stent properties and sequence parameters on visualization of the carotid artery lumen. Röfo 177:375–380PubMed
22.
go back to reference Strotzer M, Lenhart M, Butz B, Volk M, Manke C, Feuerbach S (2001) Appearance of vascular stents in computed tomographic angiography: in vitro examination of 14 different stent types. Invest Radiol 36:652–658CrossRefPubMed Strotzer M, Lenhart M, Butz B, Volk M, Manke C, Feuerbach S (2001) Appearance of vascular stents in computed tomographic angiography: in vitro examination of 14 different stent types. Invest Radiol 36:652–658CrossRefPubMed
23.
go back to reference Heuschmid M, Wiesinger B, Tepe G, Luz O, Kopp AF, Claussen CD, Duda SH (2007) Evaluation of various image reconstruction parameters in lower extremity stents using multidetector-row CT angiography: initial findings. Eur Radiol 17:265–271CrossRefPubMed Heuschmid M, Wiesinger B, Tepe G, Luz O, Kopp AF, Claussen CD, Duda SH (2007) Evaluation of various image reconstruction parameters in lower extremity stents using multidetector-row CT angiography: initial findings. Eur Radiol 17:265–271CrossRefPubMed
24.
go back to reference Maintz D, Seifarth H, Raupach R, Flohr T, Rink M, Sommer T, Ozgün M, Heindel W, Fischbach R (2006) 64-slice multidetector coronary CT angiography: in vitro evaluation of 68 different stents. Eur Radiol 16:818–826CrossRefPubMed Maintz D, Seifarth H, Raupach R, Flohr T, Rink M, Sommer T, Ozgün M, Heindel W, Fischbach R (2006) 64-slice multidetector coronary CT angiography: in vitro evaluation of 68 different stents. Eur Radiol 16:818–826CrossRefPubMed
25.
go back to reference Bladin CF, Alexandrova NA, Murphy J, Alexandrov AV, Maggisano R, Norris JW (1996) The clinical value of methods to measure carotid stenosis. Int Angiol 15:295–299PubMed Bladin CF, Alexandrova NA, Murphy J, Alexandrov AV, Maggisano R, Norris JW (1996) The clinical value of methods to measure carotid stenosis. Int Angiol 15:295–299PubMed
26.
go back to reference Watarai H, Kaku Y, Yamada M, Kokuzawa J, Tanaka T, Andoh T, Iwama T (2009) Follow-up study on in-stent thrombosis after carotid stenting using multidetector CT angiography. Neuroradiology 51:243–251CrossRefPubMed Watarai H, Kaku Y, Yamada M, Kokuzawa J, Tanaka T, Andoh T, Iwama T (2009) Follow-up study on in-stent thrombosis after carotid stenting using multidetector CT angiography. Neuroradiology 51:243–251CrossRefPubMed
27.
go back to reference Mordasini P, Al-Senani F, Gralla J, Do D-D, Brekenfeld C, Schroth G (2009) The use of flat panel angioCT (DynaCT) for navigation through a deformed and fractured carotid stent. Neuroradiology. doi:10.1007/s00234-009-0556-1 Mordasini P, Al-Senani F, Gralla J, Do D-D, Brekenfeld C, Schroth G (2009) The use of flat panel angioCT (DynaCT) for navigation through a deformed and fractured carotid stent. Neuroradiology. doi:10.​1007/​s00234-009-0556-1
28.
go back to reference Maintz D, Tombach B, Juergens KU, Weigel S, Heindel W, Fischbach R (2002) Revealing in-stent stenoses of the iliac arteries: comparison of multidetector CT with MR angiography and digital radiographic angiography in a phantom model. AJR Am J Roentgenol 179:1319–1322PubMed Maintz D, Tombach B, Juergens KU, Weigel S, Heindel W, Fischbach R (2002) Revealing in-stent stenoses of the iliac arteries: comparison of multidetector CT with MR angiography and digital radiographic angiography in a phantom model. AJR Am J Roentgenol 179:1319–1322PubMed
29.
go back to reference Amano Y, Ishihara M, Hayashi H, Gemma K, Kawamata H, Amano M, Kumazaki T (1999) Metallic artifacts of coronary and iliac arteries stents in MR angiography and contrast-enhanced CT. Clin Imaging 23:85–89CrossRefPubMed Amano Y, Ishihara M, Hayashi H, Gemma K, Kawamata H, Amano M, Kumazaki T (1999) Metallic artifacts of coronary and iliac arteries stents in MR angiography and contrast-enhanced CT. Clin Imaging 23:85–89CrossRefPubMed
30.
go back to reference Prince MR (1994) Gadolinium-enhanced MR aortography. Radiology 191:155–164PubMed Prince MR (1994) Gadolinium-enhanced MR aortography. Radiology 191:155–164PubMed
Metadata
Title
In vitro comparison of different carotid artery stents: a pixel-by-pixel analysis using CT angiography and contrast-enhanced MR angiography at 1.5 and 3 T
Authors
Michael Lettau
Annett Sauer
Sabine Heiland
Stefan Rohde
Julia Reinhardt
Martin Bendszus
Stefan Hähnel
Publication date
01-09-2010
Publisher
Springer-Verlag
Published in
Neuroradiology / Issue 9/2010
Print ISSN: 0028-3940
Electronic ISSN: 1432-1920
DOI
https://doi.org/10.1007/s00234-009-0625-5

Other articles of this Issue 9/2010

Neuroradiology 9/2010 Go to the issue