Abstract

Background

For cervical cancer patients treated with MR-guided high dose rate brachytherapy, the accuracy of radiation delivery depends on accurate localization of both tumors and the applicator, e.g. tandem and ovoid. Standard T2-weighted (T2W) MRI has good tumor-tissue contrast. However, it suffers from poor uterus-tandem contrast, which makes the tandem delineation very challenging. In this study, we evaluated the possibility of using proton density weighted (PDW) MRI to improve the definition of titanium tandems.

Methods

Both T2W and PDW MRI images were obtained from each cervical cancer patient. Imaging parameters were kept the same between the T2W and PDW sequences for each patient except the echo time (90 ms for T2W and 5.5 ms for PDW) and the slice thickness (0.5 cm for T2W and 0.25 cm for PDW). Uterus-tandem contrast was calculated by the equation C = (Su-St)/Su, where Su and St represented the average signal in the uterus and the tandem, respectively. The diameter of the tandem was measured 1.5 cm away from the tip of the tandem. The tandem was segmented by the histogram thresholding technique.

Results

PDW MRI could significantly improve the uterus-tandem contrast compared to T2W MRI (0.42±0.24 for T2W MRI, 0.77±0.14 for PDW MRI, p=0.0002). The average difference between the measured and physical diameters of the tandem was reduced from 0.20±0.15 cm by using T2W MRI to 0.10±0.11 cm by using PDW MRI (p=0.0003). The tandem segmented from the PDW image looked more uniform and complete compared to that from the T2W image.

Conclusions

Compared to the standard T2W MRI, PDW MRI has better uterus-tandem contrast. The information provided by PDW MRI is complementary to those provided by T2W MRI. Therefore, we recommend adding PDW MRI to the simulation protocol to assist tandem delineation process for cervical cancer patients.
Appendix
Available only for authorised users
Literature
1.
go back to reference Altekruse SF, Kosary CL, Krapcho M, Neyman N, Aminou R, Waldron W, Ruhl J, Howlader N, Tatalovich Z, Cho H, et al.: SEER Cancer Statistics Review, 1975–2007. MD: National Cancer Institute Bethesda; , based on November 2009 SEER data submission, posted to the SEER web site, 2010 http://seer.cancer.gov/csr/1975_2007/ Altekruse SF, Kosary CL, Krapcho M, Neyman N, Aminou R, Waldron W, Ruhl J, Howlader N, Tatalovich Z, Cho H, et al.: SEER Cancer Statistics Review, 1975–2007. MD: National Cancer Institute Bethesda; , based on November 2009 SEER data submission, posted to the SEER web site, 2010 http://​seer.​cancer.​gov/​csr/​1975_​2007/​
2.
go back to reference Nag S, Erickson B, Thomadsen E, Orton C, Demanes JD, Petereit D, Soc AB: The American Brachytherapy Society recommendations for high-dose-rate brachytherapy for carcinoma of the cervix. Int J Radiat Oncol Biol Phys 2000, 48: 201-211.CrossRefPubMed Nag S, Erickson B, Thomadsen E, Orton C, Demanes JD, Petereit D, Soc AB: The American Brachytherapy Society recommendations for high-dose-rate brachytherapy for carcinoma of the cervix. Int J Radiat Oncol Biol Phys 2000, 48: 201-211.CrossRefPubMed
3.
go back to reference Patel FD, Sharma SC, Negi PS, Ghoshal S, Gupta BD: Low-Dose Rate Vs High-Dose-Rate Brachytherapy in the Treatment of Carcinoma of the Uterine Cervix - a Clinical-Trial. Int J Radiat Oncol Biol Phys 1994, 28: 335-341. 10.1016/0360-3016(94)90055-8CrossRefPubMed Patel FD, Sharma SC, Negi PS, Ghoshal S, Gupta BD: Low-Dose Rate Vs High-Dose-Rate Brachytherapy in the Treatment of Carcinoma of the Uterine Cervix - a Clinical-Trial. Int J Radiat Oncol Biol Phys 1994, 28: 335-341. 10.1016/0360-3016(94)90055-8CrossRefPubMed
4.
go back to reference Hareyama M, Sakata K, Oouchi A, Nagakura H, Shido M, Someya M, Koito K: High-dose-rate versus low-dose-rate intracavitary therapy for carcinoma of the uterine cervix - a randomized trial. Cancer 2002, 94: 117-124. 10.1002/cncr.10207CrossRefPubMed Hareyama M, Sakata K, Oouchi A, Nagakura H, Shido M, Someya M, Koito K: High-dose-rate versus low-dose-rate intracavitary therapy for carcinoma of the uterine cervix - a randomized trial. Cancer 2002, 94: 117-124. 10.1002/cncr.10207CrossRefPubMed
5.
go back to reference Aird EGA, Conway J: CT simulation for radiotherapy treatment planning. Br J Radiol 2002, 75: 937-949.CrossRefPubMed Aird EGA, Conway J: CT simulation for radiotherapy treatment planning. Br J Radiol 2002, 75: 937-949.CrossRefPubMed
6.
go back to reference Hellebust TP, Tanderup K, Bergstrand ES, Knutsen BH, Roislien J, Olsen DR: Reconstruction of a ring applicator using CT imaging: impact of the reconstruction method and applicator orientation. Phys Med Biol 2007, 52: 4893-4904. 10.1088/0031-9155/52/16/012CrossRefPubMed Hellebust TP, Tanderup K, Bergstrand ES, Knutsen BH, Roislien J, Olsen DR: Reconstruction of a ring applicator using CT imaging: impact of the reconstruction method and applicator orientation. Phys Med Biol 2007, 52: 4893-4904. 10.1088/0031-9155/52/16/012CrossRefPubMed
7.
go back to reference Rasch C, Barillot I, Remeijer P, Touw A, van Herk M, Lebesque JV: Definition of the prostate in CT and MRI: a multi-observer study. Int J Radiat Oncol Biol Phys 1999, 43: 57-66. 10.1016/S0360-3016(98)00351-4CrossRefPubMed Rasch C, Barillot I, Remeijer P, Touw A, van Herk M, Lebesque JV: Definition of the prostate in CT and MRI: a multi-observer study. Int J Radiat Oncol Biol Phys 1999, 43: 57-66. 10.1016/S0360-3016(98)00351-4CrossRefPubMed
8.
go back to reference Li G, Citrin D, Camphausen K, Mueller B, Burman C, Mychalczak B, Miller RW, Song YL: Advances in 4D medical imaging and 4D radiation therapy. Technology in Cancer Research & Treatment 2008, 7: 67-81.CrossRef Li G, Citrin D, Camphausen K, Mueller B, Burman C, Mychalczak B, Miller RW, Song YL: Advances in 4D medical imaging and 4D radiation therapy. Technology in Cancer Research & Treatment 2008, 7: 67-81.CrossRef
9.
go back to reference Zhu J, Ma D, Lu W, Mutic S, Low D, Grigsby P: MRI-guided Radiotherapy of Cervical Carcinoma: Applications with Diffusion MRI and MR Spectroscopy. Int J Radiat Oncol Biol Phys 2009, 75: S371-S371.CrossRef Zhu J, Ma D, Lu W, Mutic S, Low D, Grigsby P: MRI-guided Radiotherapy of Cervical Carcinoma: Applications with Diffusion MRI and MR Spectroscopy. Int J Radiat Oncol Biol Phys 2009, 75: S371-S371.CrossRef
10.
go back to reference Berger D, Dimopoulos J, Potter R, Kirisits C: Direct reconstruction of the Vienna applicator on MR images. Radiother Oncol 2009, 93: 347-351. 10.1016/j.radonc.2009.07.002CrossRefPubMed Berger D, Dimopoulos J, Potter R, Kirisits C: Direct reconstruction of the Vienna applicator on MR images. Radiother Oncol 2009, 93: 347-351. 10.1016/j.radonc.2009.07.002CrossRefPubMed
11.
go back to reference De Leeuw AA, Moerland MA, Nomden C, Tersteeg RH, Roesink JM, Jurgenliemk-Schulz IM: Applicator reconstruction and applicator shifts in 3D MR-based PDR brachytherapy of cervical cancer. Radiother Oncol 2009, 93: 341-346. 10.1016/j.radonc.2009.05.003CrossRefPubMed De Leeuw AA, Moerland MA, Nomden C, Tersteeg RH, Roesink JM, Jurgenliemk-Schulz IM: Applicator reconstruction and applicator shifts in 3D MR-based PDR brachytherapy of cervical cancer. Radiother Oncol 2009, 93: 341-346. 10.1016/j.radonc.2009.05.003CrossRefPubMed
12.
go back to reference Perez-Calatayud J, Kuipers F, Ballester F, Granero D, Richart J, Rodriguez S, Tormo A, Santos M: Exclusive MRI-based tandem and colpostats reconstruction in gynaecological brachytherapy treatment planning. Radiother Oncol 2009, 91: 181-186. 10.1016/j.radonc.2008.09.004CrossRefPubMed Perez-Calatayud J, Kuipers F, Ballester F, Granero D, Richart J, Rodriguez S, Tormo A, Santos M: Exclusive MRI-based tandem and colpostats reconstruction in gynaecological brachytherapy treatment planning. Radiother Oncol 2009, 91: 181-186. 10.1016/j.radonc.2008.09.004CrossRefPubMed
13.
go back to reference Hellebust TP, Kirisits C, Berger D, Perez-Calatayud J, De Brabandere M, De Leeuw A, Dumas I, Hudej R, Lowe G, Wills R, Tanderup K: Recommendations from Gynaecological (GYN) GEC-ESTRO Working Group: considerations and pitfalls in commissioning and applicator reconstruction in 3D image-based treatment planning of cervix cancer brachytherapy. Radiother Oncol 2010, 96: 153-160. 10.1016/j.radonc.2010.06.004CrossRefPubMed Hellebust TP, Kirisits C, Berger D, Perez-Calatayud J, De Brabandere M, De Leeuw A, Dumas I, Hudej R, Lowe G, Wills R, Tanderup K: Recommendations from Gynaecological (GYN) GEC-ESTRO Working Group: considerations and pitfalls in commissioning and applicator reconstruction in 3D image-based treatment planning of cervix cancer brachytherapy. Radiother Oncol 2010, 96: 153-160. 10.1016/j.radonc.2010.06.004CrossRefPubMed
14.
go back to reference Haack S, Nielsen SK, Lindegaard JC, Gelineck J, Tanderup K: Applicator reconstruction in MRI 3D image-based dose planning of brachytherapy for cervical cancer. Radiother Oncol 2009, 91: 187-193. 10.1016/j.radonc.2008.09.002CrossRefPubMed Haack S, Nielsen SK, Lindegaard JC, Gelineck J, Tanderup K: Applicator reconstruction in MRI 3D image-based dose planning of brachytherapy for cervical cancer. Radiother Oncol 2009, 91: 187-193. 10.1016/j.radonc.2008.09.002CrossRefPubMed
15.
go back to reference Wills R, Lowe G, Inchley D, Anderson C, Beenstock V, Hoskin P: Applicator reconstruction for HDR cervix treatment planning using images from 0.35 T open MR scanner. Radiother Oncol 2010, 94: 346-352. 10.1016/j.radonc.2009.10.015CrossRefPubMed Wills R, Lowe G, Inchley D, Anderson C, Beenstock V, Hoskin P: Applicator reconstruction for HDR cervix treatment planning using images from 0.35 T open MR scanner. Radiother Oncol 2010, 94: 346-352. 10.1016/j.radonc.2009.10.015CrossRefPubMed
16.
go back to reference Kim Y, Muruganandham M, Modrick JM, Bayouth JE: Evaluation of artifacts and distortions of titanium applicators on 3.0-Tesla MRI: feasibility of titanium applicators in MRI-guided brachytherapy for gynecological cancer. Int J Radiat Oncol Biol Phys 2011, 80: 947-955. 10.1016/j.ijrobp.2010.07.1981CrossRefPubMed Kim Y, Muruganandham M, Modrick JM, Bayouth JE: Evaluation of artifacts and distortions of titanium applicators on 3.0-Tesla MRI: feasibility of titanium applicators in MRI-guided brachytherapy for gynecological cancer. Int J Radiat Oncol Biol Phys 2011, 80: 947-955. 10.1016/j.ijrobp.2010.07.1981CrossRefPubMed
17.
go back to reference Cho ZH, Kim DJ, Kim YK: Total inhomogeneity correction including chemical shifts and susceptibility by view angle tilting. Med Phys 1988, 15: 7-11. 10.1118/1.596162CrossRefPubMed Cho ZH, Kim DJ, Kim YK: Total inhomogeneity correction including chemical shifts and susceptibility by view angle tilting. Med Phys 1988, 15: 7-11. 10.1118/1.596162CrossRefPubMed
18.
go back to reference Kolind SH, MacKay AL, Munk PL, Xiang QS: Quantitative evaluation of metal artifact reduction techniques. J Magn Reson Imaging 2004, 20: 487-495. 10.1002/jmri.20144CrossRefPubMed Kolind SH, MacKay AL, Munk PL, Xiang QS: Quantitative evaluation of metal artifact reduction techniques. J Magn Reson Imaging 2004, 20: 487-495. 10.1002/jmri.20144CrossRefPubMed
19.
go back to reference Butts K, Pauly JM, Gold GE: Reduction of blurring in view angle tilting MRI. Magn Reson Med 2005, 53: 418-424. 10.1002/mrm.20375CrossRefPubMed Butts K, Pauly JM, Gold GE: Reduction of blurring in view angle tilting MRI. Magn Reson Med 2005, 53: 418-424. 10.1002/mrm.20375CrossRefPubMed
20.
go back to reference Lu W, Pauly KB, Gold GE, Pauly JM, Hargreaves BA: SEMAC: Slice Encoding for Metal Artifact Correction in MRI. Magn Reson Med 2009, 62: 66-76. 10.1002/mrm.21967CrossRefPubMedPubMedCentral Lu W, Pauly KB, Gold GE, Pauly JM, Hargreaves BA: SEMAC: Slice Encoding for Metal Artifact Correction in MRI. Magn Reson Med 2009, 62: 66-76. 10.1002/mrm.21967CrossRefPubMedPubMedCentral
21.
go back to reference Nag S, Cardenes H, Chang S, Das IJ, Erickson B, Ibbott GS, Lowenstein J, Roll J, Thomadsen B, Varia M: Proposed guidelines for image-based intracavitary brachytherapy for cervical carcinoma: Report from Image-Guided Brachytherapy Working Group. Int J Radiat Oncol Biol Phys 2004, 60: 1160-1172. 10.1016/j.ijrobp.2004.04.032CrossRefPubMed Nag S, Cardenes H, Chang S, Das IJ, Erickson B, Ibbott GS, Lowenstein J, Roll J, Thomadsen B, Varia M: Proposed guidelines for image-based intracavitary brachytherapy for cervical carcinoma: Report from Image-Guided Brachytherapy Working Group. Int J Radiat Oncol Biol Phys 2004, 60: 1160-1172. 10.1016/j.ijrobp.2004.04.032CrossRefPubMed
22.
go back to reference Tanderup K, Hellebust TP, Lang S, Granfeldt J, Potter R, Lindegaard JC, Kirisits C: Consequences of random and systematic reconstruction uncertainties in 3D image based brachytherapy in cervical cancer. Radiother Oncol 2008, 89: 156-163. 10.1016/j.radonc.2008.06.010CrossRefPubMed Tanderup K, Hellebust TP, Lang S, Granfeldt J, Potter R, Lindegaard JC, Kirisits C: Consequences of random and systematic reconstruction uncertainties in 3D image based brachytherapy in cervical cancer. Radiother Oncol 2008, 89: 156-163. 10.1016/j.radonc.2008.06.010CrossRefPubMed
23.
go back to reference Jackson EF, Ginsberg LE, Schomer DF, Leeds NE: A review of MRI pulse sequences and techniques in neuroimaging. Surg Neurol 1997, 47: 185-199. 10.1016/S0090-3019(96)00375-8CrossRefPubMed Jackson EF, Ginsberg LE, Schomer DF, Leeds NE: A review of MRI pulse sequences and techniques in neuroimaging. Surg Neurol 1997, 47: 185-199. 10.1016/S0090-3019(96)00375-8CrossRefPubMed
24.
go back to reference Sezgin M, Sankur B: Survey over image thresholding techniques and quantitative performance evaluation. Journal of Electronic Imaging 2004, 13: 146-168. 10.1117/1.1631315CrossRef Sezgin M, Sankur B: Survey over image thresholding techniques and quantitative performance evaluation. Journal of Electronic Imaging 2004, 13: 146-168. 10.1117/1.1631315CrossRef
25.
go back to reference Esthappan J, Garcia-Ramirez J, Hu Y, Grigsby PW: Transition from CT to MR Imaging of Titanium Applicators for 3D Image‐Based High Dose Rate Cervix Cancer Brachytherapy. Medical Physics 2011, 38: 3428.CrossRef Esthappan J, Garcia-Ramirez J, Hu Y, Grigsby PW: Transition from CT to MR Imaging of Titanium Applicators for 3D Image‐Based High Dose Rate Cervix Cancer Brachytherapy. Medical Physics 2011, 38: 3428.CrossRef
Metadata
Title
Improve definition of titanium tandems in MR-guided high dose rate brachytherapy for cervical cancer using proton density weighted MRI
Authors
Yanle Hu
Jacqueline Esthappan
Sasa Mutic
Susan Richardson
Hiram A Gay
Julie K Schwarz
Perry W Grigsby
Publication date
01-12-2013
Publisher
BioMed Central
Published in
Radiation Oncology / Issue 1/2013
Electronic ISSN: 1748-717X
DOI
https://doi.org/10.1186/1748-717X-8-16

Other articles of this Issue 1/2013

Radiation Oncology 1/2013 Go to the issue