Skip to main content
Top
Published in: Journal of NeuroEngineering and Rehabilitation 1/2014

Open Access 01-12-2014 | Research

Implanted functional electrical stimulation: case report of a paraplegic patient with complete SCI after 9 years

Authors: David Guiraud, Christine Azevedo Coste, Mourad Benoussaad, Charles Fattal

Published in: Journal of NeuroEngineering and Rehabilitation | Issue 1/2014

Login to get access

Abstract

Backgrounds

Experience of an implanted functional electrical stimulation neuroprosthesis (FES) associating 8-channel epimysial and 4-channel neural stimulations. The primary objective consisted in presenting clinical and technological experiences based on a 9-year follow-up of one patient implanted with this FES device. The secondary objective consisted in assessing resulting functional benefits.

Methods

One patient recruited in 1996 within the European Stand Up and Walk Project benefited from a 9-year follow-up with clinical and technological evaluations.

Results

The patient was still using the system nine years later making this a unique case, even when compared to other similar studies. The analysis of muscular response to FES underlined the great variability of stimulation thresholds evolution (−26% to +360%, mean +110%) and quality of the induced contraction. Three muscles out of five scored at least 4/5 on the Medical Research Council scale, all stimulated via neural pathways. The patient used the system once a week for 6 years, up to 2006, due to lack of use, the FES-induced muscular response worsened even though the implant was properly functioning, leading to significant decline in gait performances (best 3.45 m/s on 2.9 m), due to muscle fatigue and loss of muscle mass.

Conclusion

Two major issues arise: first the importance of muscle fatigue, underlining the relevance of muscle strength training, and second technological hurdles raising up the question of neural vs. epimysial FES. This advanced technology proves the concept of restoring lower limb motor functions in patients with spinal cord injury. The main features of the stimulation device remain stable even after long periods of inactivity, yet there is a real need for close clinical and technological monitoring.
Literature
1.
go back to reference Widerstrom-Noga EG, Felipe-Cuervo E, Broton JG, Duncan RC, Yezierski RP: Perceived difficulty in dealing with consequences of spinal cord injury. Arch Phys Med Rehabil 1999, 80: 580-586. 10.1016/S0003-9993(99)90203-4CrossRefPubMed Widerstrom-Noga EG, Felipe-Cuervo E, Broton JG, Duncan RC, Yezierski RP: Perceived difficulty in dealing with consequences of spinal cord injury. Arch Phys Med Rehabil 1999, 80: 580-586. 10.1016/S0003-9993(99)90203-4CrossRefPubMed
2.
go back to reference Snoek GJ, IJzerman MJ, Hermens HJ, Maxwell D, Biering-Sorensen F: Survey of the needs of patients with spinal cord injury: impact and priority for improvement in hand function in tetraplegics. Spinal Cord 2004, 42: 526-532. 10.1038/sj.sc.3101638CrossRefPubMed Snoek GJ, IJzerman MJ, Hermens HJ, Maxwell D, Biering-Sorensen F: Survey of the needs of patients with spinal cord injury: impact and priority for improvement in hand function in tetraplegics. Spinal Cord 2004, 42: 526-532. 10.1038/sj.sc.3101638CrossRefPubMed
3.
go back to reference Graupe D, Cerrel-Bazo H, Kern H, Carraro U: Walking performance, medical outcomes and patient training in FES of innervated muscles for ambulation by thoracic-level complete paraplegics. Neurol Res 2008, 30: 123-130. 10.1179/174313208X281136CrossRefPubMed Graupe D, Cerrel-Bazo H, Kern H, Carraro U: Walking performance, medical outcomes and patient training in FES of innervated muscles for ambulation by thoracic-level complete paraplegics. Neurol Res 2008, 30: 123-130. 10.1179/174313208X281136CrossRefPubMed
4.
go back to reference Agarwal S, Kobetic R, Nandurkar S, Marsolais EB: Functional electrical stimulation for walking in paraplegia: 17-year follow-up of 2 cases. J Spinal Cord Med 2003,26(1):86-91.PubMed Agarwal S, Kobetic R, Nandurkar S, Marsolais EB: Functional electrical stimulation for walking in paraplegia: 17-year follow-up of 2 cases. J Spinal Cord Med 2003,26(1):86-91.PubMed
5.
go back to reference Loeb GE, Peck RA, Moore WH, Hood K: BION system for distributed neural prosthetic interfaces. Med Eng Phys 2001, 23: 9-18. 10.1016/S1350-4533(01)00011-XCrossRefPubMed Loeb GE, Peck RA, Moore WH, Hood K: BION system for distributed neural prosthetic interfaces. Med Eng Phys 2001, 23: 9-18. 10.1016/S1350-4533(01)00011-XCrossRefPubMed
6.
go back to reference Smith B, Tang Z, Johnson MW, Pourmehdi S, Gazdik MM, Buckett JR, Peckham PH: An externally powered, multichannel, implantable stimulator-telemeter for control of paralyzed muscle. IEEE Trans Biomed Eng 1998, 45: 463-475. 10.1109/10.664202CrossRefPubMed Smith B, Tang Z, Johnson MW, Pourmehdi S, Gazdik MM, Buckett JR, Peckham PH: An externally powered, multichannel, implantable stimulator-telemeter for control of paralyzed muscle. IEEE Trans Biomed Eng 1998, 45: 463-475. 10.1109/10.664202CrossRefPubMed
7.
go back to reference Davis R, Houdayer T, Andrews B, Emmons S, Patrick J: Paraplegia: prolonged closed-loop standing with implanted nucleus FES-22 stimulator and Andrews’ foot-ankle orthosis. Stereotact Funct Neurosurg 1997, 69: 281-287. 10.1159/000099889CrossRefPubMed Davis R, Houdayer T, Andrews B, Emmons S, Patrick J: Paraplegia: prolonged closed-loop standing with implanted nucleus FES-22 stimulator and Andrews’ foot-ankle orthosis. Stereotact Funct Neurosurg 1997, 69: 281-287. 10.1159/000099889CrossRefPubMed
8.
go back to reference Donaldson ND, Perkins TA, Worley AC: Lumbar root stimulation for restoring leg function: stimulator and measurement of muscle actions. Artif Organs 1997, 21: 247-249. 10.1111/j.1525-1594.1997.tb04660.xCrossRefPubMed Donaldson ND, Perkins TA, Worley AC: Lumbar root stimulation for restoring leg function: stimulator and measurement of muscle actions. Artif Organs 1997, 21: 247-249. 10.1111/j.1525-1594.1997.tb04660.xCrossRefPubMed
9.
go back to reference Possover M, Schurch B, Henle KP: New strategies of pelvic nerves stimulation for recovery of pelvic visceral functions and locomotion in paraplegics. Neurourol Urody 2010, 29: 1433-1438. 10.1002/nau.20897CrossRef Possover M, Schurch B, Henle KP: New strategies of pelvic nerves stimulation for recovery of pelvic visceral functions and locomotion in paraplegics. Neurourol Urody 2010, 29: 1433-1438. 10.1002/nau.20897CrossRef
10.
go back to reference Harkema S, Gerasimenko Y, Hodes J, Burdick J, Angeli C, Chen Y, Ferreira C, Willhite A, Rejc E, Grossman RG, Edgerton VR: Effect of epidural stimulation of the lumbosacral spinal cord on voluntary movement, standing, and assisted stepping after motor complete paraplegia: a case study. Lancet 2011, 377: 1938-1947. 10.1016/S0140-6736(11)60547-3CrossRefPubMedPubMedCentral Harkema S, Gerasimenko Y, Hodes J, Burdick J, Angeli C, Chen Y, Ferreira C, Willhite A, Rejc E, Grossman RG, Edgerton VR: Effect of epidural stimulation of the lumbosacral spinal cord on voluntary movement, standing, and assisted stepping after motor complete paraplegia: a case study. Lancet 2011, 377: 1938-1947. 10.1016/S0140-6736(11)60547-3CrossRefPubMedPubMedCentral
11.
go back to reference Mushahwar VK, Jacobs PL, Normann RA, Triolo RJ, Kleitman N: New functional electrical stimulation approaches to standing and walking. J Neural Eng 2007, 4: S181-S197. 10.1088/1741-2560/4/3/S05CrossRefPubMed Mushahwar VK, Jacobs PL, Normann RA, Triolo RJ, Kleitman N: New functional electrical stimulation approaches to standing and walking. J Neural Eng 2007, 4: S181-S197. 10.1088/1741-2560/4/3/S05CrossRefPubMed
12.
go back to reference Johnston TE, Betz RR, Smith BT, Benda BJ, Mulcahey MJ, Davis R, Houdayer TP, Pontari MA, Barriskill A, Creasey GH: Implantable FES system for upright mobility and bladder and bowel function for individuals with spinal cord injury. Spinal Cord 2005, 43: 713-723. 10.1038/sj.sc.3101797CrossRefPubMed Johnston TE, Betz RR, Smith BT, Benda BJ, Mulcahey MJ, Davis R, Houdayer TP, Pontari MA, Barriskill A, Creasey GH: Implantable FES system for upright mobility and bladder and bowel function for individuals with spinal cord injury. Spinal Cord 2005, 43: 713-723. 10.1038/sj.sc.3101797CrossRefPubMed
13.
go back to reference Kobetic R, Triolo RJ, Uhlir JP, Bieri C, Wibowo M, Polando G, Marsolais EB, Davis JA Jr, Ferguson KA, Sharma M: Implanted functional electrical stimulation system for mobility in paraplegia: a follow-up case report rehabilitation engineering. IEEE Trans 1999,7(4):390-398. Kobetic R, Triolo RJ, Uhlir JP, Bieri C, Wibowo M, Polando G, Marsolais EB, Davis JA Jr, Ferguson KA, Sharma M: Implanted functional electrical stimulation system for mobility in paraplegia: a follow-up case report rehabilitation engineering. IEEE Trans 1999,7(4):390-398.
14.
go back to reference Guiraud D, Stieglitz T, Taroni G, Divoux JL: Original electronic design to perform epimysial and neural stimulation in paraplegia. J Neural Eng 2006, 3: 276-286. 10.1088/1741-2560/3/4/004CrossRefPubMed Guiraud D, Stieglitz T, Taroni G, Divoux JL: Original electronic design to perform epimysial and neural stimulation in paraplegia. J Neural Eng 2006, 3: 276-286. 10.1088/1741-2560/3/4/004CrossRefPubMed
15.
go back to reference Guiraud D, Stieglitz T, Koch KP, Divoux JL, Rabischong P: An implantable neuroprosthesis for standing and walking in paraplegia: 5-year patient follow-up. J Neural Eng 2006, 3: 268-275. 10.1088/1741-2560/3/4/003CrossRefPubMed Guiraud D, Stieglitz T, Koch KP, Divoux JL, Rabischong P: An implantable neuroprosthesis for standing and walking in paraplegia: 5-year patient follow-up. J Neural Eng 2006, 3: 268-275. 10.1088/1741-2560/3/4/003CrossRefPubMed
16.
go back to reference von Wild K, Rabischong P, Brunelli G, Benichou M, Krishnan K: Computer added locomotion by implanted electrical stimulation in paraplegic patients (SUAW). Acta Neurochir Suppl 2002, 79: 99-104.PubMed von Wild K, Rabischong P, Brunelli G, Benichou M, Krishnan K: Computer added locomotion by implanted electrical stimulation in paraplegic patients (SUAW). Acta Neurochir Suppl 2002, 79: 99-104.PubMed
17.
go back to reference Kobetic R, Triolo RJ, Marsolais EB: Muscle selection and walking performance of multichannel FES systems for ambulation in paraplegia. IEEE Trans Rehabil Eng 1997, 5: 23-29. 10.1109/86.559346CrossRefPubMed Kobetic R, Triolo RJ, Marsolais EB: Muscle selection and walking performance of multichannel FES systems for ambulation in paraplegia. IEEE Trans Rehabil Eng 1997, 5: 23-29. 10.1109/86.559346CrossRefPubMed
18.
go back to reference Hayashibe M, Zhang Q, Guiraud D, Fattal C: Evoked EMG-based torque prediction under muscle fatigue in implanted neural stimulation. J Neural Eng 2011,8(6):7.CrossRef Hayashibe M, Zhang Q, Guiraud D, Fattal C: Evoked EMG-based torque prediction under muscle fatigue in implanted neural stimulation. J Neural Eng 2011,8(6):7.CrossRef
19.
go back to reference Jovic J, Azevedo Coste C, Fraisse P, Fattal C: Upper and lower body coordination in FES-assisted sit-to-stand transfers in paraplegic subjects - a case study. Paladyn J Behav Robot 2012, 3: 2012. Jovic J, Azevedo Coste C, Fraisse P, Fattal C: Upper and lower body coordination in FES-assisted sit-to-stand transfers in paraplegic subjects - a case study. Paladyn J Behav Robot 2012, 3: 2012.
Metadata
Title
Implanted functional electrical stimulation: case report of a paraplegic patient with complete SCI after 9 years
Authors
David Guiraud
Christine Azevedo Coste
Mourad Benoussaad
Charles Fattal
Publication date
01-12-2014
Publisher
BioMed Central
Published in
Journal of NeuroEngineering and Rehabilitation / Issue 1/2014
Electronic ISSN: 1743-0003
DOI
https://doi.org/10.1186/1743-0003-11-15

Other articles of this Issue 1/2014

Journal of NeuroEngineering and Rehabilitation 1/2014 Go to the issue