Abstract

Introduction

Intraoperative technical complications are occasionally encountered while implanting INTERTAN™ nails for intertrochanteric fractures. Surgeons need to pay attention to the difficulties they may encounter during the implantation of an INTERTAN™ nail.

Case presentation

We report four cases with intraoperative difficulties during the implantation of INTERTAN™ nails among Han Chinese patients from mainland China. In Case 1, during the operation on a 75-year-old woman, an anatomical specificity of excessive femoral shaft curvation at the coronal and sagittal planes was observed; a relatively smooth implantation was achieved by adjusting the entry point. In Case 2, due to fat obstruction, an INTERTAN™ nail was implanted at an oblique angle in 64-year-old obese woman, which resulted in an iatrogenic fracture of the proximal femur. In Case 3, an iatrogenic fracture of the distal femur developed in an 83-year-old woman because of violent hammering and underestimating of bone fragility. In Case 4, an iatrogenic fracture occurred in a 40-year-old woman around the distal locking slot during the drilling process.

Conclusions

Preoperative evaluation should be considered as an important preparation for the implantation of an INTERTAN™ nail. Full-length anteroposterior and lateral radiographs of the injured femur are necessary to confirm the anatomical specificity. The vertical trajectory as well as sufficient reaming is important in reducing the possibility of iatrogenic fractures, particularly for obese patients. In older patients, violent hammering should be avoided and full reaming is recommended even if the canal seems to be wide enough. For cases where hard fracture reduction is predicted, the strategy of open reduction and fixation with a dynamic hip screw seems to be more rational and should be considered as an alternative method.
Appendix
Available only for authorised users
Literature
1.
go back to reference Leung KS, So WS, Shen WY, Hui PW: Gamma nails and dynamic hip screws for peritrochanteric fractures. A randomised prospective study in elderly patients. J Bone Joint Surg Br. 1992, 74: 345-351.PubMed Leung KS, So WS, Shen WY, Hui PW: Gamma nails and dynamic hip screws for peritrochanteric fractures. A randomised prospective study in elderly patients. J Bone Joint Surg Br. 1992, 74: 345-351.PubMed
2.
go back to reference Bridle SH, Patel AD, Bircher M, Calvert PT: Fixation of intertrochanteric fractures of the femur. A randomised prospective comparison of the gamma nail and the dynamic hip screw. J Bone Joint Surg Br. 1991, 73: 330-334.PubMed Bridle SH, Patel AD, Bircher M, Calvert PT: Fixation of intertrochanteric fractures of the femur. A randomised prospective comparison of the gamma nail and the dynamic hip screw. J Bone Joint Surg Br. 1991, 73: 330-334.PubMed
3.
go back to reference Simmermacher RK, Bosch AM, Van der Werken C: The AO/ASIF-proximal femoral nail (PFN): a new device for the treatment of unstable proximal femoral fractures. Injury. 1999, 30: 327-332. 10.1016/S0020-1383(99)00091-1.CrossRefPubMed Simmermacher RK, Bosch AM, Van der Werken C: The AO/ASIF-proximal femoral nail (PFN): a new device for the treatment of unstable proximal femoral fractures. Injury. 1999, 30: 327-332. 10.1016/S0020-1383(99)00091-1.CrossRefPubMed
4.
go back to reference Ruecker AH, Rupprecht M, Gruber M, Gebauer M, Barvencik F, Briem D, Rueger JM: The treatment of intertrochanteric fractures: results using an intramedullary nail with integrated cephalocervical screws and linear compression. J Orthop Trauma. 2009, 23: 22-30. 10.1097/BOT.0b013e31819211b2.CrossRefPubMed Ruecker AH, Rupprecht M, Gruber M, Gebauer M, Barvencik F, Briem D, Rueger JM: The treatment of intertrochanteric fractures: results using an intramedullary nail with integrated cephalocervical screws and linear compression. J Orthop Trauma. 2009, 23: 22-30. 10.1097/BOT.0b013e31819211b2.CrossRefPubMed
5.
go back to reference Ahrengart L, Tornkvist H, Fornander P, Thorngren KG, Pasanen L, Wahlstrom P, Honkonen S, Lindgren U: A randomized study of the compression hip screw and Gamma nail in 426 fractures. Clin Orthop Relat Res. 2002, 401: 209-222.CrossRefPubMed Ahrengart L, Tornkvist H, Fornander P, Thorngren KG, Pasanen L, Wahlstrom P, Honkonen S, Lindgren U: A randomized study of the compression hip screw and Gamma nail in 426 fractures. Clin Orthop Relat Res. 2002, 401: 209-222.CrossRefPubMed
6.
go back to reference Schipper IB, Marti RK, van der Werken C: Unstable trochanteric femoral fractures: extramedullary or intramedullary fixation. Review of literature. Injury. 2004, 35: 142-151. 10.1016/S0020-1383(03)00287-0.CrossRefPubMed Schipper IB, Marti RK, van der Werken C: Unstable trochanteric femoral fractures: extramedullary or intramedullary fixation. Review of literature. Injury. 2004, 35: 142-151. 10.1016/S0020-1383(03)00287-0.CrossRefPubMed
7.
go back to reference Hardy DC, Descamps PY, Krallis P, Fabeck L, Smets P, Bertens CL, Delince PE: Use of an intramedullary hip-screw compared with a compression hip-screw with a plate for intertrochanteric femoral fractures. A prospective, randomized study of one hundred patients. J Bone Joint Surg Am. 1998, 80: 618-630.PubMed Hardy DC, Descamps PY, Krallis P, Fabeck L, Smets P, Bertens CL, Delince PE: Use of an intramedullary hip-screw compared with a compression hip-screw with a plate for intertrochanteric femoral fractures. A prospective, randomized study of one hundred patients. J Bone Joint Surg Am. 1998, 80: 618-630.PubMed
8.
go back to reference Leung KS, Chen CM, So WS, Sato K, Lai CH, Machaisavariya B, Suntharalingam S: Multicenter trial of modified Gamma nail in East Asia. Clin Orthop Relat Res. 1996, 323: 146-154.CrossRefPubMed Leung KS, Chen CM, So WS, Sato K, Lai CH, Machaisavariya B, Suntharalingam S: Multicenter trial of modified Gamma nail in East Asia. Clin Orthop Relat Res. 1996, 323: 146-154.CrossRefPubMed
9.
go back to reference Leung KS, Procter P, Robioneck B, Behrens K: Geometric mismatch of the Gamma nail to the Chinese femur. Clin Orthop Relat Res. 1996, 323: 42-48.CrossRefPubMed Leung KS, Procter P, Robioneck B, Behrens K: Geometric mismatch of the Gamma nail to the Chinese femur. Clin Orthop Relat Res. 1996, 323: 42-48.CrossRefPubMed
10.
go back to reference Windolf J, Hollander DA, Hakimi M, Linhart W: Pitfalls and complications in the use of the proximal femoral nail. Langenbecks Arch Surg. 2005, 390: 59-65. 10.1007/s00423-004-0466-y.CrossRefPubMed Windolf J, Hollander DA, Hakimi M, Linhart W: Pitfalls and complications in the use of the proximal femoral nail. Langenbecks Arch Surg. 2005, 390: 59-65. 10.1007/s00423-004-0466-y.CrossRefPubMed
11.
go back to reference Hwang JH, Oh JK, Han SH, Shon WY, Oh CW: Mismatch between PFNa and medullary canal causing difficulty in nailing of the pertrochanteric fractures. Arch Orthop Trauma Surg. 2008, 128: 1443-1446. 10.1007/s00402-008-0736-1.CrossRefPubMed Hwang JH, Oh JK, Han SH, Shon WY, Oh CW: Mismatch between PFNa and medullary canal causing difficulty in nailing of the pertrochanteric fractures. Arch Orthop Trauma Surg. 2008, 128: 1443-1446. 10.1007/s00402-008-0736-1.CrossRefPubMed
12.
go back to reference Ostrum RF, Levy MS: Penetration of the distal femoral anterior cortex during intramedullary nailing for subtrochanteric fractures: a report of three cases. J Orthop Trauma. 2005, 19: 656-660. 10.1097/01.bot.0000154481.46693.69.CrossRefPubMed Ostrum RF, Levy MS: Penetration of the distal femoral anterior cortex during intramedullary nailing for subtrochanteric fractures: a report of three cases. J Orthop Trauma. 2005, 19: 656-660. 10.1097/01.bot.0000154481.46693.69.CrossRefPubMed
13.
go back to reference Koval KJ, Sala DA, Kummer FJ, Zuckerman JD: Postoperative weight-bearing after a fracture of the femoral neck or an intertrochanteric fracture. J Bone Joint Surg Am. 1998, 80: 352-356.PubMed Koval KJ, Sala DA, Kummer FJ, Zuckerman JD: Postoperative weight-bearing after a fracture of the femoral neck or an intertrochanteric fracture. J Bone Joint Surg Am. 1998, 80: 352-356.PubMed
14.
go back to reference Shen L, Zhang Y, Shen Y, Cui Z: Antirotation proximal femoral nail versus dynamic hip screw for intertrochanteric fractures: a meta-analysis of randomized controlled studies. Orthop Traumatol Surg Res. 2013, 99: 377-383. 10.1016/j.otsr.2012.12.019.CrossRefPubMed Shen L, Zhang Y, Shen Y, Cui Z: Antirotation proximal femoral nail versus dynamic hip screw for intertrochanteric fractures: a meta-analysis of randomized controlled studies. Orthop Traumatol Surg Res. 2013, 99: 377-383. 10.1016/j.otsr.2012.12.019.CrossRefPubMed
Metadata
Title
Implantation of INTERTAN™ nail in four patients with intertrochanteric fractures leading to single or comminute fractures: pitfalls and recommendations: a case series
Authors
Yong Jiang
Jie Li
Hassan H Dib
Yuan-Cheng Li
Publication date
01-12-2014
Publisher
BioMed Central
Published in
Journal of Medical Case Reports / Issue 1/2014
Electronic ISSN: 1752-1947
DOI
https://doi.org/10.1186/1752-1947-8-383

Other articles of this Issue 1/2014

Journal of Medical Case Reports 1/2014 Go to the issue