Immediate and 12 months follow up of function and lead integrity after cranial MRI in 356 patients with conventional cardiac pacemakers

Authors: Olaf M Muehling, Reza Wakili, Martin Greif, Franz von Ziegler, Dominik Morhard, Hartmut Brueckmann, Alexander Becker

Published in: Journal of Cardiovascular Magnetic Resonance | Issue 1/2014

Login to get access

Abstract

Background

Conventional cardiac pacemakers are still often regarded as a contraindication to magnetic resonance imaging (MRI). We conducted this study to support the hypothesis that it is safe to scan patients with cardiac pacemakers in a 1.5 Tesla MRI, if close supervision and monitoring as well as adequate pre- and postscan programming is provided.

Methods

We followed up 356 patients (age 61.3 ± 9.1 yrs., 229 men) with single (n = 132) or dual chamber (n = 224) cardiac pacemakers and urgent indication for a cranial MRI for 12 months. The scans were performed at 1.5T. During the scan patients were monitored with a 3-lead ECG and pulse oximetry. Prior to the scan pacemakers were programmed according to our own protocol.

Results

All 356 scans were completed without complications. No arrhythmias were induced, programmed parameters remained unchanged. No pacemaker dysfunction was identified. Follow-up examinations were performed immediately, 2 weeks, 2, 6, and 12 months after the scan. There was no significant change of pacing capture threshold (ventricular 0.9 ± 0.4 V@0.4 ms, atrial 0.9 ± 0.3 V@0.4 ms) immediately (ventricular 1.0 ± 0.3 V@0.4 ms, atrial 0.9 ± 0.4 V@0.4 ms) or at 12 months follow-up examinations (ventricular 0.9 ± 0.2 V@0.4 ms, atrial 0.9 ± 0.3 V@0.4 ms). There was no significant change in sensing threshold (8.0 ± 4.0 mV vs. 8.1 ± 4.2 mV ventricular lead, 2.0 ± 0.9 mV vs. 2.1 ± 1.0 mV atrial lead) or lead impedance (ventricular 584 ± 179 Ω vs. 578 ± 188 Ω, atrial 534 ± 176 Ω vs. 532 ± 169 Ω) after 12 months.

Conclusions

This supports the evidence that patients with conventional pacemakers can safely undergo cranial MRI in a 1.5T system with suitable preparation, supervision and precautions. Long term follow-up did not reveal significant changes in pacing capture nor sensing threshold.
Appendix
Available only for authorised users
Literature
1.
go back to reference Kalin R, Stanton MS: Current clinical issues for mri scanning of pacemaker and defibrillator patients. Pacing Clin Electrophysiol. 2005, 28: 326-8. 10.1111/j.1540-8159.2005.50024.x.CrossRefPubMed Kalin R, Stanton MS: Current clinical issues for mri scanning of pacemaker and defibrillator patients. Pacing Clin Electrophysiol. 2005, 28: 326-8. 10.1111/j.1540-8159.2005.50024.x.CrossRefPubMed
2.
go back to reference Wollmann CG, Thudt K, Vock P, Globits S, Mayr H: Clinical routine implantation of a dual chamber pacemaker system designed for safe use with mri: a single center, retrospective study on lead performance of medtronic lead 5086mri in comparison to medtronic leads 4592-53 and 4092-58. Herzschrittmacherther Elektrophysiol. 2011, 22: 233-6. 10.1007/s00399-011-0161-y. 239-242CrossRefPubMed Wollmann CG, Thudt K, Vock P, Globits S, Mayr H: Clinical routine implantation of a dual chamber pacemaker system designed for safe use with mri: a single center, retrospective study on lead performance of medtronic lead 5086mri in comparison to medtronic leads 4592-53 and 4092-58. Herzschrittmacherther Elektrophysiol. 2011, 22: 233-6. 10.1007/s00399-011-0161-y. 239-242CrossRefPubMed
3.
go back to reference Sutton R, Kanal E, Wilkoff BL, Bello D, Luechinger R, Jenniskens I, Hull M, Sommer T: Safety of magnetic resonance imaging of patients with a new medtronic enrhythm mri surescan pacing system: Clinical study design. Trials. 2008, 9: 68-10.1186/1745-6215-9-68.PubMedCentralCrossRefPubMed Sutton R, Kanal E, Wilkoff BL, Bello D, Luechinger R, Jenniskens I, Hull M, Sommer T: Safety of magnetic resonance imaging of patients with a new medtronic enrhythm mri surescan pacing system: Clinical study design. Trials. 2008, 9: 68-10.1186/1745-6215-9-68.PubMedCentralCrossRefPubMed
4.
go back to reference Faris OP, Shein MJ, Government viewpoint: U.S. Food & drug administration: pacemakers, icds and mri. Pacing Clin Electrophysiol. 2005, 28: 268-9. 10.1111/j.1540-8159.2005.50035.x.CrossRefPubMed Faris OP, Shein MJ, Government viewpoint: U.S. Food & drug administration: pacemakers, icds and mri. Pacing Clin Electrophysiol. 2005, 28: 268-9. 10.1111/j.1540-8159.2005.50035.x.CrossRefPubMed
5.
go back to reference Stanton MS: Industry viewpoint: medtronic: pacemakers, icds, and mri. Pacing Clin Electrophysiol. 2005, 28: 265-10.1111/j.1540-8159.2005.50025.x.CrossRefPubMed Stanton MS: Industry viewpoint: medtronic: pacemakers, icds, and mri. Pacing Clin Electrophysiol. 2005, 28: 265-10.1111/j.1540-8159.2005.50025.x.CrossRefPubMed
6.
go back to reference Smith JM: Industry viewpoint: guidant: pacemakers, icds, and mri. Pacing Clin Electrophysiol. 2005, 28: 264-10.1111/j.1540-8159.2005.09591.x.CrossRefPubMed Smith JM: Industry viewpoint: guidant: pacemakers, icds, and mri. Pacing Clin Electrophysiol. 2005, 28: 264-10.1111/j.1540-8159.2005.09591.x.CrossRefPubMed
7.
go back to reference Levine PA: Industry viewpoint: St. Jude medical: pacemakers, icds and mri. Pacing Clin Electrophysiol. 2005, 28: 266-7. 10.1111/j.1540-8159.2005.50027.x.CrossRefPubMed Levine PA: Industry viewpoint: St. Jude medical: pacemakers, icds and mri. Pacing Clin Electrophysiol. 2005, 28: 266-7. 10.1111/j.1540-8159.2005.50027.x.CrossRefPubMed
8.
go back to reference Martin ET, Sandler DA: Mri in patients with cardiac devices. Curr Cardiol Rep. 2007, 9: 63-71. 10.1007/s11886-007-0012-y.CrossRefPubMed Martin ET, Sandler DA: Mri in patients with cardiac devices. Curr Cardiol Rep. 2007, 9: 63-71. 10.1007/s11886-007-0012-y.CrossRefPubMed
9.
go back to reference Gimbel JR: Magnetic resonance imaging of implantable cardiac rhythm devices at 3.0 tesla. Pacing Clin Electrophysiol. 2008, 31: 795-801. 10.1111/j.1540-8159.2008.01117.x.CrossRefPubMed Gimbel JR: Magnetic resonance imaging of implantable cardiac rhythm devices at 3.0 tesla. Pacing Clin Electrophysiol. 2008, 31: 795-801. 10.1111/j.1540-8159.2008.01117.x.CrossRefPubMed
10.
go back to reference Roguin A, Schwitter J, Vahlhaus C, Lombardi M, Brugada J, Vardas P, Auricchio A, Priori S, Sommer T: Magnetic resonance imaging in individuals with cardiovascular implantable electronic devices. Europace. 2008, 10: 336-46. 10.1093/europace/eun021.CrossRefPubMed Roguin A, Schwitter J, Vahlhaus C, Lombardi M, Brugada J, Vardas P, Auricchio A, Priori S, Sommer T: Magnetic resonance imaging in individuals with cardiovascular implantable electronic devices. Europace. 2008, 10: 336-46. 10.1093/europace/eun021.CrossRefPubMed
11.
go back to reference Martin ET, Coman JA, Shellock FG, Pulling CC, Fair R, Jenkins K: Magnetic resonance imaging and cardiac pacemaker safety at 1.5-tesla. J Am Coll Cardiol. 2004, 43: 1315-24. 10.1016/j.jacc.2003.12.016.CrossRefPubMed Martin ET, Coman JA, Shellock FG, Pulling CC, Fair R, Jenkins K: Magnetic resonance imaging and cardiac pacemaker safety at 1.5-tesla. J Am Coll Cardiol. 2004, 43: 1315-24. 10.1016/j.jacc.2003.12.016.CrossRefPubMed
12.
go back to reference Nazarian S, Roguin A, Zviman MM, Lardo AC, Dickfeld TL, Calkins H, Weiss RG, Berger RD, Bluemke DA, Halperin HR: Clinical utility and safety of a protocol for noncardiac and cardiac magnetic resonance imaging of patients with permanent pacemakers and implantable-cardioverter defibrillators at 1.5 tesla. Circulation. 2006, 114: 1277-84. 10.1161/CIRCULATIONAHA.105.607655.PubMedCentralCrossRefPubMed Nazarian S, Roguin A, Zviman MM, Lardo AC, Dickfeld TL, Calkins H, Weiss RG, Berger RD, Bluemke DA, Halperin HR: Clinical utility and safety of a protocol for noncardiac and cardiac magnetic resonance imaging of patients with permanent pacemakers and implantable-cardioverter defibrillators at 1.5 tesla. Circulation. 2006, 114: 1277-84. 10.1161/CIRCULATIONAHA.105.607655.PubMedCentralCrossRefPubMed
13.
go back to reference Nazarian S, Hansford R, Roguin A, Goldsher D, Zviman MM, Lardo AC, Caffo BS, Frick KD, Kraut MA, Kamel IR, Calkins H, Berger RD, Bluemke DA, Halperin HR: A prospective evaluation of a protocol for magnetic resonance imaging of patients with implanted cardiac devices. Ann Intern Med. 2011, 155: 415-24. 10.7326/0003-4819-155-7-201110040-00004.PubMedCentralCrossRefPubMed Nazarian S, Hansford R, Roguin A, Goldsher D, Zviman MM, Lardo AC, Caffo BS, Frick KD, Kraut MA, Kamel IR, Calkins H, Berger RD, Bluemke DA, Halperin HR: A prospective evaluation of a protocol for magnetic resonance imaging of patients with implanted cardiac devices. Ann Intern Med. 2011, 155: 415-24. 10.7326/0003-4819-155-7-201110040-00004.PubMedCentralCrossRefPubMed
14.
go back to reference Muehling OM, Huber A, Friedrich D, Nabauer M, Reiser M, Schoenberg SO: Code blue in the mr suite–a drill to rescue a patients with cardiac arrest from the mr scanner. Int J Cardiovasc Imaging. 2006, 22: 257-61. 10.1007/s10554-005-9028-9.CrossRefPubMed Muehling OM, Huber A, Friedrich D, Nabauer M, Reiser M, Schoenberg SO: Code blue in the mr suite–a drill to rescue a patients with cardiac arrest from the mr scanner. Int J Cardiovasc Imaging. 2006, 22: 257-61. 10.1007/s10554-005-9028-9.CrossRefPubMed
15.
go back to reference Cohen JD, Costa HS, Russo RJ: Determining the risks of magnetic resonance imaging at 1.5 tesla for patients with pacemakers and implantable cardioverter defibrillators. Am J Cardiol. 2012, 110: 1631-6. 10.1016/j.amjcard.2012.07.030.CrossRefPubMed Cohen JD, Costa HS, Russo RJ: Determining the risks of magnetic resonance imaging at 1.5 tesla for patients with pacemakers and implantable cardioverter defibrillators. Am J Cardiol. 2012, 110: 1631-6. 10.1016/j.amjcard.2012.07.030.CrossRefPubMed
16.
go back to reference Naehle CP, Kreuz J, Strach K, Schwab JO, Pingel S, Luechinger R, Fimmers R, Schild H, Thomas D: Safety, feasibility, and diagnostic value of cardiac magnetic resonance imaging in patients with cardiac pacemakers and implantable cardioverters/defibrillators at 1.5 t. Am Heart J. 2011, 161: 1096-105. 10.1016/j.ahj.2011.03.007.CrossRefPubMed Naehle CP, Kreuz J, Strach K, Schwab JO, Pingel S, Luechinger R, Fimmers R, Schild H, Thomas D: Safety, feasibility, and diagnostic value of cardiac magnetic resonance imaging in patients with cardiac pacemakers and implantable cardioverters/defibrillators at 1.5 t. Am Heart J. 2011, 161: 1096-105. 10.1016/j.ahj.2011.03.007.CrossRefPubMed
17.
go back to reference Roguin A, Zviman MM, Meininger GR, Rodrigues ER, Dickfeld TM, Bluemke DA, Lardo A, Berger RD, Calkins H, Halperin HR: Modern pacemaker and implantable cardioverter/defibrillator systems can be magnetic resonance imaging safe: in vitro and in vivo assessment of safety and function at 1.5 t. Circulation. 2004, 110: 475-82. 10.1161/01.CIR.0000137121.28722.33.PubMedCentralCrossRefPubMed Roguin A, Zviman MM, Meininger GR, Rodrigues ER, Dickfeld TM, Bluemke DA, Lardo A, Berger RD, Calkins H, Halperin HR: Modern pacemaker and implantable cardioverter/defibrillator systems can be magnetic resonance imaging safe: in vitro and in vivo assessment of safety and function at 1.5 t. Circulation. 2004, 110: 475-82. 10.1161/01.CIR.0000137121.28722.33.PubMedCentralCrossRefPubMed
18.
go back to reference Irnich W: Risks to pacemaker patients undergoing magnetic resonance imaging examinations. Europace. 2010, 12: 918-20. 10.1093/europace/euq144.CrossRefPubMed Irnich W: Risks to pacemaker patients undergoing magnetic resonance imaging examinations. Europace. 2010, 12: 918-20. 10.1093/europace/euq144.CrossRefPubMed
19.
go back to reference Luechinger R, Duru F, Zeijlemaker VA, Scheidegger MB, Boesiger P, Candinas R: Pacemaker reed switch behavior in 0.5, 1.5, and 3.0 tesla magnetic resonance imaging units: are reed switches always closed in strong magnetic fields?. Pacing Clin Electrophysiol. 2002, 25: 1419-23. 10.1046/j.1460-9592.2002.01419.x.CrossRefPubMed Luechinger R, Duru F, Zeijlemaker VA, Scheidegger MB, Boesiger P, Candinas R: Pacemaker reed switch behavior in 0.5, 1.5, and 3.0 tesla magnetic resonance imaging units: are reed switches always closed in strong magnetic fields?. Pacing Clin Electrophysiol. 2002, 25: 1419-23. 10.1046/j.1460-9592.2002.01419.x.CrossRefPubMed
20.
go back to reference Gimbel JR: The safety of mri scanning of pacemakers and icds: what are the critical elements of safe scanning? Ask me again at 10,000. Europace. 2010, 12: 915-7. 10.1093/europace/euq174.CrossRefPubMed Gimbel JR: The safety of mri scanning of pacemakers and icds: what are the critical elements of safe scanning? Ask me again at 10,000. Europace. 2010, 12: 915-7. 10.1093/europace/euq174.CrossRefPubMed
21.
go back to reference Sommer T, Naehle CP, Yang A, Zeijlemaker V, Hackenbroch M, Schmiedel A, Meyer C, Strach K, Skowasch D, Vahlhaus C, Litt H, Schild H: Strategy for safe performance of extrathoracic magnetic resonance imaging at 1.5 tesla in the presence of cardiac pacemakers in non-pacemaker-dependent patients: a prospective study with 115 examinations. Circulation. 2006, 114: 1285-92. 10.1161/CIRCULATIONAHA.105.597013.CrossRefPubMed Sommer T, Naehle CP, Yang A, Zeijlemaker V, Hackenbroch M, Schmiedel A, Meyer C, Strach K, Skowasch D, Vahlhaus C, Litt H, Schild H: Strategy for safe performance of extrathoracic magnetic resonance imaging at 1.5 tesla in the presence of cardiac pacemakers in non-pacemaker-dependent patients: a prospective study with 115 examinations. Circulation. 2006, 114: 1285-92. 10.1161/CIRCULATIONAHA.105.597013.CrossRefPubMed
22.
go back to reference Sommer T, Vahlhaus C, Lauck G, von Smekal A, Reinke M, Hofer U, Block W, Traber F, Schneider C, Gieseke J, Jung W, Schild H: Mr imaging and cardiac pacemakers: in-vitro evaluation and in-vivo studies in 51 patients at 0.5 t. Radiology. 2000, 215: 869-79. 10.1148/radiology.215.3.r00jn08869.CrossRefPubMed Sommer T, Vahlhaus C, Lauck G, von Smekal A, Reinke M, Hofer U, Block W, Traber F, Schneider C, Gieseke J, Jung W, Schild H: Mr imaging and cardiac pacemakers: in-vitro evaluation and in-vivo studies in 51 patients at 0.5 t. Radiology. 2000, 215: 869-79. 10.1148/radiology.215.3.r00jn08869.CrossRefPubMed
23.
go back to reference Nordbeck P, Weiss I, Ehses P, Ritter O, Warmuth M, Fidler F, Herold V, Jakob PM, Ladd ME, Quick HH, Bauer WR: Measuring rf-induced currents inside implants: Impact of device configuration on mri safety of cardiac pacemaker leads. Magn Reson Med. 2009, 61: 570-8. 10.1002/mrm.21881.CrossRefPubMed Nordbeck P, Weiss I, Ehses P, Ritter O, Warmuth M, Fidler F, Herold V, Jakob PM, Ladd ME, Quick HH, Bauer WR: Measuring rf-induced currents inside implants: Impact of device configuration on mri safety of cardiac pacemaker leads. Magn Reson Med. 2009, 61: 570-8. 10.1002/mrm.21881.CrossRefPubMed
24.
go back to reference Luechinger R, Zeijlemaker VA, Pedersen EM, Mortensen P, Falk E, Duru F, Candinas R, Boesiger P: In vivo heating of pacemaker leads during magnetic resonance imaging. Eur Heart J. 2005, 26: 376-83. 10.1093/eurheartj/ehi009. discussion 325-377CrossRefPubMed Luechinger R, Zeijlemaker VA, Pedersen EM, Mortensen P, Falk E, Duru F, Candinas R, Boesiger P: In vivo heating of pacemaker leads during magnetic resonance imaging. Eur Heart J. 2005, 26: 376-83. 10.1093/eurheartj/ehi009. discussion 325-377CrossRefPubMed
25.
go back to reference Shellock FG, Fischer L, Fieno DS: Cardiac pacemakers and implantable cardioverter defibrillators: in vitro magnetic resonance imaging evaluation at 1.5-tesla. J Cardiovasc Magn Reson. 2007, 9: 21-31. 10.1080/10976640600897237.CrossRefPubMed Shellock FG, Fischer L, Fieno DS: Cardiac pacemakers and implantable cardioverter defibrillators: in vitro magnetic resonance imaging evaluation at 1.5-tesla. J Cardiovasc Magn Reson. 2007, 9: 21-31. 10.1080/10976640600897237.CrossRefPubMed
26.
go back to reference Calcagnini G, Triventi M, Censi F, Mattei E, Bartolini P, Kainz W, Bassen HI: In vitro investigation of pacemaker lead heating induced by magnetic resonance imaging: role of implant geometry. J Magn Reson Imaging. 2008, 28: 879-86. 10.1002/jmri.21536.CrossRefPubMed Calcagnini G, Triventi M, Censi F, Mattei E, Bartolini P, Kainz W, Bassen HI: In vitro investigation of pacemaker lead heating induced by magnetic resonance imaging: role of implant geometry. J Magn Reson Imaging. 2008, 28: 879-86. 10.1002/jmri.21536.CrossRefPubMed
27.
go back to reference Shellock FG, Fieno DS, Thomson LJ, Talavage TM, Berman DS: Cardiac pacemaker: in vitro assessment at 1.5 t. Am Heart J. 2006, 151: 436-43. 10.1016/j.ahj.2005.04.015.CrossRefPubMed Shellock FG, Fieno DS, Thomson LJ, Talavage TM, Berman DS: Cardiac pacemaker: in vitro assessment at 1.5 t. Am Heart J. 2006, 151: 436-43. 10.1016/j.ahj.2005.04.015.CrossRefPubMed
28.
go back to reference Langman DA, Goldberg IB, Finn JP, Ennis DB: Pacemaker lead tip heating in abandoned and pacemaker-attached leads at 1.5 tesla mri. J Magn Reson Imaging. 2011, 33: 426-31. 10.1002/jmri.22463.CrossRefPubMed Langman DA, Goldberg IB, Finn JP, Ennis DB: Pacemaker lead tip heating in abandoned and pacemaker-attached leads at 1.5 tesla mri. J Magn Reson Imaging. 2011, 33: 426-31. 10.1002/jmri.22463.CrossRefPubMed
29.
go back to reference Vahlhaus C, Sommer T, Lewalter T, Schimpf R, Schumacher B, Jung W, Luderitz B: Interference with cardiac pacemakers by magnetic resonance imaging: are there irreversible changes at 0.5 tesla?. Pacing Clin Electrophysiol. 2001, 24: 489-95. 10.1046/j.1460-9592.2001.00489.x.CrossRefPubMed Vahlhaus C, Sommer T, Lewalter T, Schimpf R, Schumacher B, Jung W, Luderitz B: Interference with cardiac pacemakers by magnetic resonance imaging: are there irreversible changes at 0.5 tesla?. Pacing Clin Electrophysiol. 2001, 24: 489-95. 10.1046/j.1460-9592.2001.00489.x.CrossRefPubMed
30.
go back to reference Achenbach S, Moshage W, Diem B, Bieberle T, Schibgilla V, Bachmann K: Effects of magnetic resonance imaging on cardiac pacemakers and electrodes. Am Heart J. 1997, 134: 467-73. 10.1016/S0002-8703(97)70083-8.CrossRefPubMed Achenbach S, Moshage W, Diem B, Bieberle T, Schibgilla V, Bachmann K: Effects of magnetic resonance imaging on cardiac pacemakers and electrodes. Am Heart J. 1997, 134: 467-73. 10.1016/S0002-8703(97)70083-8.CrossRefPubMed
31.
32.
go back to reference Delacretaz E: Asynchronous ventricular pacing triggering ventricular fibrillation. J Cardiovasc Electrophysiol. 2004, 15: 963-4. 10.1046/j.1540-8167.2004.04053.x.CrossRefPubMed Delacretaz E: Asynchronous ventricular pacing triggering ventricular fibrillation. J Cardiovasc Electrophysiol. 2004, 15: 963-4. 10.1046/j.1540-8167.2004.04053.x.CrossRefPubMed
33.
go back to reference Buendia F, Cano O, Sanchez-Gomez JM, Igual B, Osca J, Sancho-Tello MJ, Olague J, Salvador A: Cardiac magnetic resonance imaging at 1.5 t in patients with cardiac rhythm devices. Europace. 2011, 13: 533-8. 10.1093/europace/euq501.CrossRefPubMed Buendia F, Cano O, Sanchez-Gomez JM, Igual B, Osca J, Sancho-Tello MJ, Olague J, Salvador A: Cardiac magnetic resonance imaging at 1.5 t in patients with cardiac rhythm devices. Europace. 2011, 13: 533-8. 10.1093/europace/euq501.CrossRefPubMed
34.
go back to reference Irnich W, Irnich B, Bartsch C, Stertmann WA, Gufler H, Weiler G: Do we need pacemakers resistant to magnetic resonance imaging?. Europace. 2005, 7: 353-65. 10.1016/j.eupc.2005.02.120.CrossRefPubMed Irnich W, Irnich B, Bartsch C, Stertmann WA, Gufler H, Weiler G: Do we need pacemakers resistant to magnetic resonance imaging?. Europace. 2005, 7: 353-65. 10.1016/j.eupc.2005.02.120.CrossRefPubMed
35.
go back to reference Kanal E, Barkovich AJ, Bell C, Borgstede JP, Bradley WG, Froelich JW, Gilk T, Gimbel JR, Gosbee J, Kuhni-Kaminski E, Lester JW, Nyenhuis J, Parag Y, Schaefer DJ, Sebek-Scoumis EA, Weinreb J, Zaremba LA, Wilcox P, Lucey L, Sass N: Acr guidance document for safe mr practices: 2007. AJR Am J Roentgenol. 2007, 188: 1447-74. 10.2214/AJR.06.1616.CrossRefPubMed Kanal E, Barkovich AJ, Bell C, Borgstede JP, Bradley WG, Froelich JW, Gilk T, Gimbel JR, Gosbee J, Kuhni-Kaminski E, Lester JW, Nyenhuis J, Parag Y, Schaefer DJ, Sebek-Scoumis EA, Weinreb J, Zaremba LA, Wilcox P, Lucey L, Sass N: Acr guidance document for safe mr practices: 2007. AJR Am J Roentgenol. 2007, 188: 1447-74. 10.2214/AJR.06.1616.CrossRefPubMed
36.
go back to reference Levine GN, Gomes AS, Arai AE, Bluemke DA, Flamm SD, Kanal E, Manning WJ, Martin ET, Smith JM, Wilke N, Shellock FS: Safety of magnetic resonance imaging in patients with cardiovascular devices: an american heart association scientific statement from the committee on diagnostic and interventional cardiac catheterization, council on clinical cardiology, and the council on cardiovascular radiology and intervention: endorsed by the American college of cardiology foundation, the north American society for cardiac imaging, and the society for cardiovascular magnetic resonance. Circulation. 2007, 116: 2878-91. 10.1161/CIRCULATIONAHA.107.187256.CrossRefPubMed Levine GN, Gomes AS, Arai AE, Bluemke DA, Flamm SD, Kanal E, Manning WJ, Martin ET, Smith JM, Wilke N, Shellock FS: Safety of magnetic resonance imaging in patients with cardiovascular devices: an american heart association scientific statement from the committee on diagnostic and interventional cardiac catheterization, council on clinical cardiology, and the council on cardiovascular radiology and intervention: endorsed by the American college of cardiology foundation, the north American society for cardiac imaging, and the society for cardiovascular magnetic resonance. Circulation. 2007, 116: 2878-91. 10.1161/CIRCULATIONAHA.107.187256.CrossRefPubMed
37.
go back to reference Boilson BA, Wokhlu A, Acker NG, Felmlee JP, Watson RE, Julsrud PR, Friedman PA, Cha YM, Rea RF, Hayes DL, Shen WK: Safety of magnetic resonance imaging in patients with permanent pacemakers: a collaborative clinical approach. J Interv Card Electrophysiol. 2012, 33: 59-67. 10.1007/s10840-011-9615-8.CrossRefPubMed Boilson BA, Wokhlu A, Acker NG, Felmlee JP, Watson RE, Julsrud PR, Friedman PA, Cha YM, Rea RF, Hayes DL, Shen WK: Safety of magnetic resonance imaging in patients with permanent pacemakers: a collaborative clinical approach. J Interv Card Electrophysiol. 2012, 33: 59-67. 10.1007/s10840-011-9615-8.CrossRefPubMed
38.
go back to reference Tandri H, Zviman MM, Wedan SR, Lloyd T, Berger RD, Halperin H: Determinants of gradient field-induced current in a pacemaker lead system in a magnetic resonance imaging environment. Heart Rhythm. 2008, 5: 462-8. 10.1016/j.hrthm.2007.12.022.CrossRefPubMed Tandri H, Zviman MM, Wedan SR, Lloyd T, Berger RD, Halperin H: Determinants of gradient field-induced current in a pacemaker lead system in a magnetic resonance imaging environment. Heart Rhythm. 2008, 5: 462-8. 10.1016/j.hrthm.2007.12.022.CrossRefPubMed
39.
go back to reference Mattei E, Calcagnini G, Triventi M, Censi F, Bartolini P, Kainz W, Bassen H: Mri induced heating of pacemaker leads: effect of temperature probe positioning and pacemaker placement on lead tip heating and local sar. Conf Proc IEEE Eng Med Biol Soc. 2006, 1: 1889-92.CrossRefPubMed Mattei E, Calcagnini G, Triventi M, Censi F, Bartolini P, Kainz W, Bassen H: Mri induced heating of pacemaker leads: effect of temperature probe positioning and pacemaker placement on lead tip heating and local sar. Conf Proc IEEE Eng Med Biol Soc. 2006, 1: 1889-92.CrossRefPubMed
40.
go back to reference Naehle CP, Meyer C, Thomas D, Remerie S, Krautmacher C, Litt H, Luechinger R, Fimmers R, Schild H, Sommer T: Safety of brain 3-t mr imaging with transmit-receive head coil in patients with cardiac pacemakers: pilot prospective study with 51 examinations. Radiology. 2008, 249: 991-1001. 10.1148/radiol.2493072195.CrossRefPubMed Naehle CP, Meyer C, Thomas D, Remerie S, Krautmacher C, Litt H, Luechinger R, Fimmers R, Schild H, Sommer T: Safety of brain 3-t mr imaging with transmit-receive head coil in patients with cardiac pacemakers: pilot prospective study with 51 examinations. Radiology. 2008, 249: 991-1001. 10.1148/radiol.2493072195.CrossRefPubMed
41.
go back to reference Lauck G, von Smekal A, Wolke S, Seelos KC, Jung W, Manz M, Luderitz B: Effects of nuclear magnetic resonance imaging on cardiac pacemakers. Pacing Clin Electrophysiol. 1995, 18: 1549-55. 10.1111/j.1540-8159.1995.tb06741.x.CrossRefPubMed Lauck G, von Smekal A, Wolke S, Seelos KC, Jung W, Manz M, Luderitz B: Effects of nuclear magnetic resonance imaging on cardiac pacemakers. Pacing Clin Electrophysiol. 1995, 18: 1549-55. 10.1111/j.1540-8159.1995.tb06741.x.CrossRefPubMed
Metadata
Title
Immediate and 12 months follow up of function and lead integrity after cranial MRI in 356 patients with conventional cardiac pacemakers
Authors
Olaf M Muehling
Reza Wakili
Martin Greif
Franz von Ziegler
Dominik Morhard
Hartmut Brueckmann
Alexander Becker
Publication date
01-12-2014
Publisher
BioMed Central
Published in
Journal of Cardiovascular Magnetic Resonance / Issue 1/2014
Electronic ISSN: 1532-429X
DOI
https://doi.org/10.1186/1532-429X-16-39

Other articles of this Issue 1/2014

Journal of Cardiovascular Magnetic Resonance 1/2014 Go to the issue