01-12-2020 | Hip-TEP | Original Paper

The early- to medium-term results of a hemispherical, porous coated acetabular shell with multiple different bearing combinations are excellent with the exception of metal-on-metal

Authors: Zoe Thompson, Amir Khoshbin, Sarah Ward, James P. Waddell, Amit Atrey

Published in: International Orthopaedics | Issue 12/2020

Login to get access

Abstract

Purpose

This study prospectively reports survivorship and radiographic and clinical outcomes following primary elective total hip arthroplasty (THA) using a novel single hemispherical, porous-coated acetabular cup with five different bearing combinations and a minimum of five year follow-up.

Methods

Continuing post-market release monitoring of this cup, we prospectively enrolled 108 patients (121 THA) between 2009 and 2015. We followed this cohort by examining survivorship, in addition to clinical and radiological outcomes for metal-on-metal (MoM) compared with non-MoM bearing combinations (ceramic-on-ceramic, oxinium-on-polyethylene, ceramic-on-metal, and metal-on-polyethylene).

Results

All 108 (121 hips) patients were followed up. Average age at time of surgery was 45.1 years (range 19 to 71 years) of which 42.1% were males. A total of seven (5.8%) cups were revised, all of which were MoM.
No osteolysis was observed in any of the patients at the latest visit with a mean follow-up of 9.1 ± 1.7 years (range 4.4–10.7 years). With MoM excluded, survivorship of the cup at five  years is 97.8%. Survivorship for MoM implants was 90.0%. Validated hip scores showed significant improvements for all bearing types and no significant difference between groups at latest follow-up (p = 0.614). There was no cup migration with any bearing surface.

Conclusion

This cup showed excellent survivorship at five year follow-up, except for patients receiving a MoM articulation. While there were concerns over the early survivorship of this cup, our cohort and joint registry data confirm excellent outcomes.
Appendix
Available only for authorised users
Literature
1.
go back to reference Bourne RB, McCalden RW, Naudie D, Charron KD, Yuan X, Holdsworth DW (2008) The next generation of acetabular shell design and bearing surfaces. Orthopaedics 31(12 Suppl 2) Bourne RB, McCalden RW, Naudie D, Charron KD, Yuan X, Holdsworth DW (2008) The next generation of acetabular shell design and bearing surfaces. Orthopaedics 31(12 Suppl 2)
2.
go back to reference Nilskon K, Theodoulou A, Mercer G (2017) Mid-term migration of a cementless, porous acetabular cup; a 5 year radiostereometric analysis. J Orthop 14(4):454–460CrossRef Nilskon K, Theodoulou A, Mercer G (2017) Mid-term migration of a cementless, porous acetabular cup; a 5 year radiostereometric analysis. J Orthop 14(4):454–460CrossRef
3.
go back to reference Ilo KC, Dereby EJ, Whittaker RK (2017) Fretting and corrosion between a metal shell and metal liner may explain the high rates of R3 modular metal-on-metal hips. J Arthroplast 32(5):1679–1683CrossRef Ilo KC, Dereby EJ, Whittaker RK (2017) Fretting and corrosion between a metal shell and metal liner may explain the high rates of R3 modular metal-on-metal hips. J Arthroplast 32(5):1679–1683CrossRef
4.
go back to reference Dramis A, Clatworthy E, Jones SA (2014) High failure rates of the R3 metal-on-metal total hip arthroplasty. Hip Int 24(5):442–447CrossRef Dramis A, Clatworthy E, Jones SA (2014) High failure rates of the R3 metal-on-metal total hip arthroplasty. Hip Int 24(5):442–447CrossRef
6.
go back to reference Teoh KH, Whithma RD, Golding DM (2018) R3 cup does not have a high failure rate in conventional bearings: a minimum of 5 year follow up. J Arthroplast 33(2):460–463CrossRef Teoh KH, Whithma RD, Golding DM (2018) R3 cup does not have a high failure rate in conventional bearings: a minimum of 5 year follow up. J Arthroplast 33(2):460–463CrossRef
7.
go back to reference Wilson IR, Turgeon TR, Gascoyne TC, Della Valle CJ, McCalden RW (2020) Midterm results of a contemporary, porous-coated acetabular system in patients undergoing primary total hip replacement for degenerative hip disease: a prospective, multicenter study. J Arthroplast 35(7):1862–1867. https://doi.org/10.1016/j.arth.2020.02.068 Wilson IR, Turgeon TR, Gascoyne TC, Della Valle CJ, McCalden RW (2020) Midterm results of a contemporary, porous-coated acetabular system in patients undergoing primary total hip replacement for degenerative hip disease: a prospective, multicenter study. J Arthroplast 35(7):1862–1867. https://​doi.​org/​10.​1016/​j.​arth.​2020.​02.​068
8.
go back to reference Atrey A, Wolfstadt J, Hussain N (2018) The ideal total hip replacement bearing surface in the young patient: a prospective randomized trial comparing alumina ceramic-on-ceramic with ceramic-on-conventional polyethylene: 15-year follow-up. J Arthroplast 33(6):1752–1756CrossRef Atrey A, Wolfstadt J, Hussain N (2018) The ideal total hip replacement bearing surface in the young patient: a prospective randomized trial comparing alumina ceramic-on-ceramic with ceramic-on-conventional polyethylene: 15-year follow-up. J Arthroplast 33(6):1752–1756CrossRef
9.
go back to reference DeLee JG, Charnley J (1976) Radiological demarcation of cemented sockets in total hip replacement. Clin Orthop Relat Res 121:20–32 DeLee JG, Charnley J (1976) Radiological demarcation of cemented sockets in total hip replacement. Clin Orthop Relat Res 121:20–32
10.
go back to reference Gruen TA, McNeice GM, Amstutz HC (1979) “Modes of failure” of cemented stem-type femoral components: a radiographic analysis of loosening. Clin Orthop Relat Res 141:17–27 Gruen TA, McNeice GM, Amstutz HC (1979) “Modes of failure” of cemented stem-type femoral components: a radiographic analysis of loosening. Clin Orthop Relat Res 141:17–27
11.
go back to reference Campbell P, Ebramzadeh E, Nelson S (2010) Histological features of pseudotumor-like tissues from metal-on-metal hips. Clin Orthop Relat Res 468(9):2321–2327CrossRef Campbell P, Ebramzadeh E, Nelson S (2010) Histological features of pseudotumor-like tissues from metal-on-metal hips. Clin Orthop Relat Res 468(9):2321–2327CrossRef
12.
go back to reference Pisanu F, Doria C, Andreozzi M et al (2019) Pleomorphic clinical spectrum of metallosis in total hip arthroplasty. Int Orthop 43(1):85–96CrossRef Pisanu F, Doria C, Andreozzi M et al (2019) Pleomorphic clinical spectrum of metallosis in total hip arthroplasty. Int Orthop 43(1):85–96CrossRef
13.
go back to reference Kasparek MF, Renner L, Faschingbauer M et al (2018) Salvage of a monoblock metal-on-metal cup using a dual mobility liner: a two-year MRI follow-up study. Int Orthop 42(5):1035–1041CrossRef Kasparek MF, Renner L, Faschingbauer M et al (2018) Salvage of a monoblock metal-on-metal cup using a dual mobility liner: a two-year MRI follow-up study. Int Orthop 42(5):1035–1041CrossRef
14.
go back to reference Learmoth ID, Young C, Rorabeck C (2007) The operation of the century: total hip replacement. Lancet. 370(9597):1508–1519CrossRef Learmoth ID, Young C, Rorabeck C (2007) The operation of the century: total hip replacement. Lancet. 370(9597):1508–1519CrossRef
16.
go back to reference Matharu GS, Judge A, Pandit H et al (2018) Follow-up for patients with metal-on-metal hip replacements: are the new MHRA recommendations justified? BMJ. 13:360 Matharu GS, Judge A, Pandit H et al (2018) Follow-up for patients with metal-on-metal hip replacements: are the new MHRA recommendations justified? BMJ. 13:360
18.
go back to reference Bouveau V, Haen TX, Poupon J, Nich C (2018) Outcomes after revision of metal on metal hip resurfacing to total arthroplasty using the direct anterior approach. Int Orthop 42(11):2543–2548CrossRef Bouveau V, Haen TX, Poupon J, Nich C (2018) Outcomes after revision of metal on metal hip resurfacing to total arthroplasty using the direct anterior approach. Int Orthop 42(11):2543–2548CrossRef
19.
go back to reference Tao R, Liu F, Liu YK et al (2018) A prospective comparative study of hip resurfacing arthroplasty and large-diameter head metal-on-metal total hip arthroplasty in younger patients-a minimum of five year follow-up. Int Orthop 42(10):2323–2327CrossRef Tao R, Liu F, Liu YK et al (2018) A prospective comparative study of hip resurfacing arthroplasty and large-diameter head metal-on-metal total hip arthroplasty in younger patients-a minimum of five year follow-up. Int Orthop 42(10):2323–2327CrossRef
20.
go back to reference Benelli G, Maritato M, Cerulli Mariani P, Sasso F (2019) Revision of ASR hip arthroplasty: analysis of two hundred and ninety six recalled patients at seven years. Int Orthop 43(1):97–101CrossRef Benelli G, Maritato M, Cerulli Mariani P, Sasso F (2019) Revision of ASR hip arthroplasty: analysis of two hundred and ninety six recalled patients at seven years. Int Orthop 43(1):97–101CrossRef
21.
go back to reference Grammatopoulos G, Munemoto M, Inagaki Y (2016) The diagnosis of infection in metal-on-metal hip arthroplasties. J Arthroplast 31(11):2569–2573CrossRef Grammatopoulos G, Munemoto M, Inagaki Y (2016) The diagnosis of infection in metal-on-metal hip arthroplasties. J Arthroplast 31(11):2569–2573CrossRef
22.
go back to reference Prieto HA, Berbari EF, Sierra RJ (2014) Acute delayed infection: increased risk in failed metal on metal total hip arthroplasty. J Arthroplast 29(9):1808–1812CrossRef Prieto HA, Berbari EF, Sierra RJ (2014) Acute delayed infection: increased risk in failed metal on metal total hip arthroplasty. J Arthroplast 29(9):1808–1812CrossRef
23.
go back to reference Drummond J, Tran P, Fary C (2015) Metal-on-metal hip arthroplasty: a review of adverse reactions and patient management. J Funct Biomater 6(3):486–499CrossRef Drummond J, Tran P, Fary C (2015) Metal-on-metal hip arthroplasty: a review of adverse reactions and patient management. J Funct Biomater 6(3):486–499CrossRef
Metadata
Title
The early- to medium-term results of a hemispherical, porous coated acetabular shell with multiple different bearing combinations are excellent with the exception of metal-on-metal
Authors
Zoe Thompson
Amir Khoshbin
Sarah Ward
James P. Waddell
Amit Atrey
Publication date
01-12-2020
Publisher
Springer Berlin Heidelberg
Keywords
Hip-TEP
Hip-TEP
Published in
International Orthopaedics / Issue 12/2020
Print ISSN: 0341-2695
Electronic ISSN: 1432-5195
DOI
https://doi.org/10.1007/s00264-020-04817-1

Other articles of this Issue 12/2020

International Orthopaedics 12/2020 Go to the issue