Early failure for wear after ceramic-on-highly cross-linked polyethylene total hip arthroplasty: a case report

Authors: Chenkai Li, Haining Zhang

Published in: BMC Musculoskeletal Disorders | Issue 1/2020

Login to get access

Abstract

Background

Highly cross-linked polyethylene (HXLPE) enhances the anti-wear characteristics of the conventional polyethylene (PE). Early failure for wear after ceramic-on-highly cross-linked polyethylene (CoHXLPE) total hip arthroplasty (THA) is extremely rare.

Case presentation

We described the case of a 60-year-old man who underwent right CoHXLPE THA because of the developmental dysplasia hip (DDH) complained pain 32 months after this procedure. Plain radiographs showed that eccentric wear existed at the polyethylene insert. However, the patient refused surgery at that time and did not stop weight-bearing. The right hip pain continued for 7 months. Plain radiographs of the pelvis showed that the HXLPE liner was penetrated and partial inner wall of acetabular shell was worn. Acetabular cup revision was performed, and the ceramic head and HXLPE were exchanged.

Conclusions

Difficult reduction during primary THA, especially for DDH, can result in higher abductor tension, which may lead to early eccentric wear of the prosthesis. Whenever eccentric wear of HXLPE liner was found, weight-bearing must be stopped to avoid the accelerated wear and adverse reactions to metal debris (ARMD).
Literature
2.
go back to reference Hall A, Eilers M, Hansen R, et al. Advances in acetabular reconstruction in revision total hip arthroplasty: maximizing function and outcomes after treatment of periacetabular osteolysis around the well-fixed shell. Instr Course Lect. 2014;63:209–18.PubMed Hall A, Eilers M, Hansen R, et al. Advances in acetabular reconstruction in revision total hip arthroplasty: maximizing function and outcomes after treatment of periacetabular osteolysis around the well-fixed shell. Instr Course Lect. 2014;63:209–18.PubMed
5.
go back to reference Ji Z, Wei G, Yixin Z. Analysis of failure mechanisms of total hip anhroplasty. Chin J Bone Joint Surg. 2012;5(3):217–9. Ji Z, Wei G, Yixin Z. Analysis of failure mechanisms of total hip anhroplasty. Chin J Bone Joint Surg. 2012;5(3):217–9.
7.
go back to reference Longbo K. Retrospective clinical application of ceramic-on-polyethylene total hip arthroplasty. Int J Orthop. 2009;30(2):84–8. Longbo K. Retrospective clinical application of ceramic-on-polyethylene total hip arthroplasty. Int J Orthop. 2009;30(2):84–8.
16.
go back to reference Rubash HE, Sinha RK, Maloney WJ, Paprosky GW. Osteolysis: surgical treatment. Instr Course Lect. 1998;47:321–9.PubMed Rubash HE, Sinha RK, Maloney WJ, Paprosky GW. Osteolysis: surgical treatment. Instr Course Lect. 1998;47:321–9.PubMed
17.
go back to reference Rubash HE, Sinha RK, Paprosky W, Engh AC, Maloney JW. A new classification system for the management of acetabular osteolysis after total hip arthroplasty. Instr Course Lect. 1999;48:37–42.PubMed Rubash HE, Sinha RK, Paprosky W, Engh AC, Maloney JW. A new classification system for the management of acetabular osteolysis after total hip arthroplasty. Instr Course Lect. 1999;48:37–42.PubMed
Metadata
Title
Early failure for wear after ceramic-on-highly cross-linked polyethylene total hip arthroplasty: a case report
Authors
Chenkai Li
Haining Zhang
Publication date
01-12-2020
Publisher
BioMed Central
Keywords
Hip-TEP
Hip-TEP
Published in
BMC Musculoskeletal Disorders / Issue 1/2020
Electronic ISSN: 1471-2474
DOI
https://doi.org/10.1186/s12891-020-03697-1

Other articles of this Issue 1/2020

BMC Musculoskeletal Disorders 1/2020 Go to the issue