Open Access 01-12-2024 | Hip Arthroscopy | Research article

Experimental assessment of pelvis slipping during postless traction for orthopaedic applications

Authors: Marco Daghero, Simone Borrelli, Taian M. Vieira, Francesco Cannito, Alessandro Aprato, Andrea Audisio, Cristina Bignardi, Mara Terzini

Published in: Journal of Orthopaedic Surgery and Research | Issue 1/2024

Login to get access

Abstract

Background

The application of lower limb traction during hip arthroscopy and femur fractures osteosynthesis is commonplace in orthopaedic surgeries. Traditional methods utilize a perineal post on a traction table, leading to soft tissue damage and nerve neuropraxia. A postless technique, using high-friction pads, has been considered as a potential damage-free alternative. However, whether these pads sufficiently prevent patient displacement remains unknown. Thus, this study systematically assesses the efficacy of commercial high-friction pads (PinkPad and CarePad) in restraining subject displacement, for progressively increasing traction loads and different Trendelenburg angles.

Methods

Three healthy male subjects were recruited and tested in supine and Trendelenburg positions (5° and 10°), using a customized boot-pulley system. Ten load disks (5 kg) were dropped at 15s intervals, increasing gradually the traction load up to 50 kg. Pelvis displacement along the traction direction was measured with a motion capture system. The displacement at 50 kg of traction load was analyzed and compared across various pads and bed inclinations. Response to varying traction loads was statistically assessed with a quadratic function model.

Results

Pelvis displacement at 50 kg traction load was below 60 mm for all conditions. Comparing PinkPad and CarePad, no significant differences in displacement were observed. Finally, similar displacements were observed for the supine and Trendelenburg positions.

Conclusions

Both PinkPad and CarePad exhibited nearly linear behavior under increasing traction loads, limiting displacement to 60 mm at most for 50 kg loads. Contrary to expectations, placing subjects in the Trendelenburg position did not increase adhesion.
Appendix
Available only for authorised users
Literature
12.
Metadata
Title
Experimental assessment of pelvis slipping during postless traction for orthopaedic applications
Authors
Marco Daghero
Simone Borrelli
Taian M. Vieira
Francesco Cannito
Alessandro Aprato
Andrea Audisio
Cristina Bignardi
Mara Terzini
Publication date
01-12-2024
Publisher
BioMed Central
Keyword
Hip Arthroscopy
Published in
Journal of Orthopaedic Surgery and Research / Issue 1/2024
Electronic ISSN: 1749-799X
DOI
https://doi.org/10.1186/s13018-024-04704-0