Skip to main content
Top
Published in: International Journal of Computer Assisted Radiology and Surgery 7/2019

01-07-2019 | Original Article

High-precision evaluation of electromagnetic tracking

Authors: David Kügler, Henry Krumb, Judith Bredemann, Igor Stenin, Julia Kristin, Thomas Klenzner, Jörg Schipper, Robert Schmitt, Georgios Sakas, Anirban Mukhopadhyay

Published in: International Journal of Computer Assisted Radiology and Surgery | Issue 7/2019

Login to get access

Abstract

Purpose

Navigation in high-precision minimally invasive surgery (HP-MIS) demands high tracking accuracy in the absence of line of sight (LOS). Currently, no tracking technology can satisfy this requirement. Electromagnetic tracking (EMT) is the best tracking paradigm in the absence of LOS despite limited accuracy and robustness. Novel evaluation protocols are needed to ensure high-precision and robust EMT for navigation in HP-MIS.

Methods

We introduce a novel protocol for EMT measurement evaluation featuring a high-accuracy phantom based on LEGO\(^{\circledR }\), which is calibrated by a coordinate measuring machine to ensure accuracy. Our protocol includes relative sequential positions and an uncertainty estimation of positioning. We show effects on distortion compensation using a learned interpolation model.

Results

Our high-precision protocol clarifies properties of errors and uncertainties of EMT for high-precision use cases. For EMT errors reaching clinically relevant 0.2 mm, our design is 5–10 times more accurate than previous protocols with 95% confidence margins of 0.02 mm. This high-precision protocol ensures the performance improvement in compensated EMT by 0.05 mm.

Conclusion

Our protocol improves the reliability of EMT evaluations because of significantly lower protocol-inherent uncertainties. To reduce patient risk in HP-MIS and to evaluate magnetic field distortion compensation, more high-accuracy protocols such as the one proposed here are required.
Literature
1.
go back to reference Franz AM, Haidegger T, Birkfellner W, Cleary K, Peters TM, Maier-Hein L (2014) Electromagnetic tracking in medicine: a review of technology, validation, and applications. IEEE TMI 33(8):1702–1725 Franz AM, Haidegger T, Birkfellner W, Cleary K, Peters TM, Maier-Hein L (2014) Electromagnetic tracking in medicine: a review of technology, validation, and applications. IEEE TMI 33(8):1702–1725
3.
go back to reference Haidegger T, Fenyvesi G, Sirokai B, Kelemen M, Nagy M, Takács B, Kovács L, Benyó B, Benyó Z (2011) Towards unified electromagnetic tracking system assessment-static errors. EMBC Proc 2011:1905–1908 Haidegger T, Fenyvesi G, Sirokai B, Kelemen M, Nagy M, Takács B, Kovács L, Benyó B, Benyó Z (2011) Towards unified electromagnetic tracking system assessment-static errors. EMBC Proc 2011:1905–1908
4.
go back to reference Hummel J, Figl M, Birkfellner W, Bax MR, Shahidi R, Maurer CR, Bergmann H (2006) Evaluation of a new electromagnetic tracking system using a standardized assessment protocol. Phys Med Biol 51(10):N205–10CrossRefPubMed Hummel J, Figl M, Birkfellner W, Bax MR, Shahidi R, Maurer CR, Bergmann H (2006) Evaluation of a new electromagnetic tracking system using a standardized assessment protocol. Phys Med Biol 51(10):N205–10CrossRefPubMed
5.
go back to reference Hummel JB, Bax MR, Figl ML, Kang Y, Maurer C, Birkfellner WW, Bergmann H, Shahidi R (2005) Design and application of an assessment protocol for electromagnetic tracking systems. MedPhys 32(7):2371–2379 Hummel JB, Bax MR, Figl ML, Kang Y, Maurer C, Birkfellner WW, Bergmann H, Shahidi R (2005) Design and application of an assessment protocol for electromagnetic tracking systems. MedPhys 32(7):2371–2379
6.
go back to reference Joint Committee for Guides in Metrology (2008) Evaluation of measurement data—guide to the expression of uncertainty in measurement Joint Committee for Guides in Metrology (2008) Evaluation of measurement data—guide to the expression of uncertainty in measurement
7.
go back to reference Kindratenko VV (2000) A survey of electromagnetic position tracker calibration techniques. Virtual Real 5(3):169–182CrossRef Kindratenko VV (2000) A survey of electromagnetic position tracker calibration techniques. Virtual Real 5(3):169–182CrossRef
8.
10.
go back to reference Labadie RF, Balachandran R, Mitchell J, Noble JH, Majdani O, Haynes D, Bennett M, Dawant BM, Fitzpatrick JM (2010) Clinical validation study of percutaneous cochlear access using patient customized micro-stereotactic frames. Otol Neurotol Off Publ Am Otol Soc Am Neurotol Soc Eur Acad Otol Neurotol 31(1):94CrossRef Labadie RF, Balachandran R, Mitchell J, Noble JH, Majdani O, Haynes D, Bennett M, Dawant BM, Fitzpatrick JM (2010) Clinical validation study of percutaneous cochlear access using patient customized micro-stereotactic frames. Otol Neurotol Off Publ Am Otol Soc Am Neurotol Soc Eur Acad Otol Neurotol 31(1):94CrossRef
11.
go back to reference Lugez E, Sadjadi H, Pichora DR, Ellis RE, Akl SG, Fichtinger G (2015) Electromagnetic tracking in surgical and interventional environments: usability study. IJCARS 10(3):253–262 Lugez E, Sadjadi H, Pichora DR, Ellis RE, Akl SG, Fichtinger G (2015) Electromagnetic tracking in surgical and interventional environments: usability study. IJCARS 10(3):253–262
12.
go back to reference Maier-Hein L, Franz AM, Birkfellner W, Hummel J, Gergel I, Wegner I, Meinzer HP (2012) Standardized assessment of new electromagnetic field generators in an interventional radiology setting. MedPhys 39(6):3424–3434 Maier-Hein L, Franz AM, Birkfellner W, Hummel J, Gergel I, Wegner I, Meinzer HP (2012) Standardized assessment of new electromagnetic field generators in an interventional radiology setting. MedPhys 39(6):3424–3434
13.
go back to reference Reichl T, Gardiazabal J, Navab N (2013) Electromagnetic servoing-a new tracking paradigm. IEEE TMI 32(8):1526–1535 Reichl T, Gardiazabal J, Navab N (2013) Electromagnetic servoing-a new tracking paradigm. IEEE TMI 32(8):1526–1535
14.
go back to reference Rice SO (1945) Mathematical analysis of random noise. In: Wax N (ed) Selected papers on noise and stochastic processes. Dover Publications Inc., New York, pp 133–294 Rice SO (1945) Mathematical analysis of random noise. In: Wax N (ed) Selected papers on noise and stochastic processes. Dover Publications Inc., New York, pp 133–294
15.
go back to reference Sadjadi H, Hashtrudi-Zaad K, Fichtinger G (2016) Simultaneous electromagnetic tracking and calibration for dynamic field distortion compensation. IEEE TBME 63(8):1771–1781 Sadjadi H, Hashtrudi-Zaad K, Fichtinger G (2016) Simultaneous electromagnetic tracking and calibration for dynamic field distortion compensation. IEEE TBME 63(8):1771–1781
16.
go back to reference Schipper J, Aschendorff A, Arapakis I, Klenzner T, Teszler CB, Ridder GJ, Laszig R (2004) Navigation as a quality management tool in cochlear implant surgery. JLO 118(10):764–770CrossRef Schipper J, Aschendorff A, Arapakis I, Klenzner T, Teszler CB, Ridder GJ, Laszig R (2004) Navigation as a quality management tool in cochlear implant surgery. JLO 118(10):764–770CrossRef
17.
go back to reference Sirokai B, Kiss M, Kovács L, Benyó B, Benyó Z, Haidegger T (2012) Best practice in electromagnetic tracking system assessment. In: II Proceedings of the joint workshop on new technologies for CRAS, vol 12, pp 1–4 Sirokai B, Kiss M, Kovács L, Benyó B, Benyó Z, Haidegger T (2012) Best practice in electromagnetic tracking system assessment. In: II Proceedings of the joint workshop on new technologies for CRAS, vol 12, pp 1–4
18.
go back to reference Sładek JA (2016) Coordinate metrology: accuracy of systems and measurements, 1st edn. Springer, BerlinCrossRef Sładek JA (2016) Coordinate metrology: accuracy of systems and measurements, 1st edn. Springer, BerlinCrossRef
19.
go back to reference Vakharia VN, Sparks R, O’Keeffe AG, Rodionov R, Miserocchi A, McEvoy A, Ourselin S, Duncan J (2017) Accuracy of intracranial electrode placement for stereoencephalography: a systematic review and meta-analysis. Epilepsia 58(6):921–932CrossRefPubMed Vakharia VN, Sparks R, O’Keeffe AG, Rodionov R, Miserocchi A, McEvoy A, Ourselin S, Duncan J (2017) Accuracy of intracranial electrode placement for stereoencephalography: a systematic review and meta-analysis. Epilepsia 58(6):921–932CrossRefPubMed
20.
go back to reference Wilson E, Yaniv Z, Zhang H, Nafis C, Shen E, Shechter G, Wiles AD, Peters T, Lindisch D, Cleary K (2007) A hardware and software protocol for the evaluation of electromagnetic tracker accuracy in the clinical environment: a multi-center study. In: SPIE proceedings, SPIE, p 65092T Wilson E, Yaniv Z, Zhang H, Nafis C, Shen E, Shechter G, Wiles AD, Peters T, Lindisch D, Cleary K (2007) A hardware and software protocol for the evaluation of electromagnetic tracker accuracy in the clinical environment: a multi-center study. In: SPIE proceedings, SPIE, p 65092T
21.
go back to reference Wimmer W, Venail F, Williamson T, Akkari M, Gerber N, Weber S, Caversaccio M, Uziel A, Bell B (2014) Semiautomatic cochleostomy target and insertion trajectory planning for minimally invasive cochlear implantation. BioMed Res Int 2014:596498CrossRefPubMedPubMedCentral Wimmer W, Venail F, Williamson T, Akkari M, Gerber N, Weber S, Caversaccio M, Uziel A, Bell B (2014) Semiautomatic cochleostomy target and insertion trajectory planning for minimally invasive cochlear implantation. BioMed Res Int 2014:596498CrossRefPubMedPubMedCentral
Metadata
Title
High-precision evaluation of electromagnetic tracking
Authors
David Kügler
Henry Krumb
Judith Bredemann
Igor Stenin
Julia Kristin
Thomas Klenzner
Jörg Schipper
Robert Schmitt
Georgios Sakas
Anirban Mukhopadhyay
Publication date
01-07-2019
Publisher
Springer International Publishing
Published in
International Journal of Computer Assisted Radiology and Surgery / Issue 7/2019
Print ISSN: 1861-6410
Electronic ISSN: 1861-6429
DOI
https://doi.org/10.1007/s11548-019-01959-5

Other articles of this Issue 7/2019

International Journal of Computer Assisted Radiology and Surgery 7/2019 Go to the issue