01-07-2018 | Cardiac

High-pitch low-voltage CT coronary artery calcium scoring with tin filtration: accuracy and radiation dose reduction

Authors: Georg Apfaltrer, Moritz H. Albrecht, U. Joseph Schoepf, Taylor M. Duguay, Carlo N. De Cecco, John W. Nance, Domenico De Santis, Paul Apfaltrer, Marwen H. Eid, Chelsea D. Eason, Zachary M. Thompson, Maximilian J. Bauer, Akos Varga-Szemes, Brian E. Jacobs, Erich Sorantin, Christian Tesche

Published in: European Radiology | Issue 7/2018

Login to get access

Abstract

Objectives

To investigate diagnostic accuracy and radiation dose of high-pitch CT coronary artery calcium scoring (CACS) with tin filtration (Sn100kVp) versus standard 120kVp high-pitch acquisition.

Methods

78 patients (58% male, 61.5±9.1 years) were prospectively enrolled. Subjects underwent clinical 120kVp high-pitch CACS using third-generation dual-source CT followed by additional high-pitch Sn100kVp acquisition. Agatston scores, calcium volume scores, Agatston score categories, percentile-based risk categorization and radiation metrics were compared.

Results

61/78 patients showed coronary calcifications. Median Agatston scores were 34.9 [0.7–197.1] and 41.7 [0.7–207.2] and calcium volume scores were 34.1 [0.7–218.0] for Sn100kVp and 35.7 [1.1–221.0] for 120kVp acquisitions, respectively (both p<0.0001). Bland-Altman analysis revealed underestimated Agatston scores and calcium volume scores with Sn100kVp versus 120kVp acquisitions (mean difference: 16.4 and 11.5). However, Agatston score categories and percentile-based risk categories showed excellent agreement (ĸ=0.98 and ĸ=0.99). Image noise was 25.8±4.4HU and 16.6±2.9HU in Sn100kVp and 120kVp scans, respectively (p<0.0001). Dose-length-product was 9.9±4.8mGy*cm and 40.9±14.4mGy*cm with Sn100kVp and 120kVp scans, respectively (p<0.0001). This resulted in significant effective radiation dose reduction (0.13±0.07mSv vs. 0.57±0.2mSv, p<0.0001) for Sn100kVp acquisitions.

Conclusion

CACS using high-pitch low-voltage tin-filtered acquisitions demonstrates excellent agreement in Agatston score and percentile-based cardiac risk categorization with standard 120kVp high-pitch acquisitions. Furthermore, radiation dose was significantly reduced by 78% while maintaining accurate risk prediction.

Key points

• Coronary artery calcium scoring with tin filtration reduces radiation dose by 78%.
• There is excellent correlation between high-pitch Sn100kVp and standard 120kVp acquisitions.
• Excellent agreement regarding Agatston score categories and percentile-based risk categorization was achieved.
• No cardiac risk reclassifications were observed using Sn100kVp coronary artery calcium scoring.
Literature
1.
go back to reference Agatston AS, Janowitz WR, Hildner FJ, Zusmer NR, Viamonte M Jr, Detrano R (1990) Quantification of coronary artery calcium using ultrafast computed tomography. J Am Coll Cardiol 15:827–832CrossRefPubMed Agatston AS, Janowitz WR, Hildner FJ, Zusmer NR, Viamonte M Jr, Detrano R (1990) Quantification of coronary artery calcium using ultrafast computed tomography. J Am Coll Cardiol 15:827–832CrossRefPubMed
2.
go back to reference Bauer RW, Thilo C, Chiaramida SA, Vogl TJ, Costello P, Schoepf UJ (2009) Noncalcified atherosclerotic plaque burden at coronary CT angiography: a better predictor of ischemia at stress myocardial perfusion imaging than calcium score and stenosis severity. AJR Am J Roentgenol 193:410–418CrossRefPubMed Bauer RW, Thilo C, Chiaramida SA, Vogl TJ, Costello P, Schoepf UJ (2009) Noncalcified atherosclerotic plaque burden at coronary CT angiography: a better predictor of ischemia at stress myocardial perfusion imaging than calcium score and stenosis severity. AJR Am J Roentgenol 193:410–418CrossRefPubMed
3.
go back to reference Renker M, Geyer LL, Krazinski AW, Silverman JR, Ebersberger U, Schoepf UJ (2013) Iterative image reconstruction: a realistic dose-saving method in cardiac CT imaging? Expert Rev Cardiovasc Ther 11:403–409CrossRefPubMed Renker M, Geyer LL, Krazinski AW, Silverman JR, Ebersberger U, Schoepf UJ (2013) Iterative image reconstruction: a realistic dose-saving method in cardiac CT imaging? Expert Rev Cardiovasc Ther 11:403–409CrossRefPubMed
4.
go back to reference Yin WH, Lu B, Li N et al (2013) Iterative reconstruction to preserve image quality and diagnostic accuracy at reduced radiation dose in coronary CT angiography: an intraindividual comparison. JACC Cardiovasc Imaging 6:1239–1249CrossRefPubMed Yin WH, Lu B, Li N et al (2013) Iterative reconstruction to preserve image quality and diagnostic accuracy at reduced radiation dose in coronary CT angiography: an intraindividual comparison. JACC Cardiovasc Imaging 6:1239–1249CrossRefPubMed
5.
go back to reference Meyer M, Haubenreisser H, Schoepf UJ et al (2014) Closing in on the K edge: coronary CT angiography at 100, 80, and 70 kV-initial comparison of a second- versus a third-generation dual-source CT system. Radiology 273:373–382CrossRefPubMed Meyer M, Haubenreisser H, Schoepf UJ et al (2014) Closing in on the K edge: coronary CT angiography at 100, 80, and 70 kV-initial comparison of a second- versus a third-generation dual-source CT system. Radiology 273:373–382CrossRefPubMed
6.
go back to reference Tesche C, De Cecco CN, Vliegenthart R et al (2017) Accuracy and Radiation Dose Reduction Using Low-Voltage Computed Tomography Coronary Artery Calcium Scoring With Tin Filtration. Am J Cardiol 119:675–680CrossRefPubMed Tesche C, De Cecco CN, Vliegenthart R et al (2017) Accuracy and Radiation Dose Reduction Using Low-Voltage Computed Tomography Coronary Artery Calcium Scoring With Tin Filtration. Am J Cardiol 119:675–680CrossRefPubMed
7.
go back to reference Tesche C, De Cecco CN, Schoepf UJ et al (2017) CT coronary calcium scoring with tin filtration using iterative beam-hardening calcium correction reconstruction. Eur J Radiol 91:29–34CrossRefPubMed Tesche C, De Cecco CN, Schoepf UJ et al (2017) CT coronary calcium scoring with tin filtration using iterative beam-hardening calcium correction reconstruction. Eur J Radiol 91:29–34CrossRefPubMed
8.
go back to reference Tesche C, De Cecco CN, Schoepf UJ et al (2017) Iterative beam-hardening correction with advanced modeled iterative reconstruction in low voltage CT coronary calcium scoring with tin filtration: Impact on coronary artery calcium quantification and image quality. J Cardiovasc Comput Tomogr 11:354–359CrossRefPubMed Tesche C, De Cecco CN, Schoepf UJ et al (2017) Iterative beam-hardening correction with advanced modeled iterative reconstruction in low voltage CT coronary calcium scoring with tin filtration: Impact on coronary artery calcium quantification and image quality. J Cardiovasc Comput Tomogr 11:354–359CrossRefPubMed
9.
go back to reference Hausleiter J, Meyer T, Hermann F et al (2009) Estimated radiation dose associated with cardiac CT angiography. Jama 301:500–507CrossRefPubMed Hausleiter J, Meyer T, Hermann F et al (2009) Estimated radiation dose associated with cardiac CT angiography. Jama 301:500–507CrossRefPubMed
10.
11.
go back to reference Bland JM, Altman DG (1986) Statistical methods for assessing agreement between two methods of clinical measurement. Lancet 1:307–310CrossRefPubMed Bland JM, Altman DG (1986) Statistical methods for assessing agreement between two methods of clinical measurement. Lancet 1:307–310CrossRefPubMed
12.
go back to reference McClelland RL, Chung H, Detrano R, Post W, Kronmal RA (2006) Distribution of coronary artery calcium by race, gender, and age: results from the Multi-Ethnic Study of Atherosclerosis (MESA). Circulation 113:30–37CrossRefPubMed McClelland RL, Chung H, Detrano R, Post W, Kronmal RA (2006) Distribution of coronary artery calcium by race, gender, and age: results from the Multi-Ethnic Study of Atherosclerosis (MESA). Circulation 113:30–37CrossRefPubMed
13.
go back to reference Landis JR, Koch GG (1977) An application of hierarchical kappa-type statistics in the assessment of majority agreement among multiple observers. Biometrics 33:363–374CrossRefPubMed Landis JR, Koch GG (1977) An application of hierarchical kappa-type statistics in the assessment of majority agreement among multiple observers. Biometrics 33:363–374CrossRefPubMed
14.
go back to reference McQuiston AD, Muscogiuri G, Schoepf UJ et al (2016) Approaches to ultra-low radiation dose coronary artery calcium scoring based on 3rd generation dual-source CT: A phantom study. Eur J Radiol 85:39–47CrossRefPubMed McQuiston AD, Muscogiuri G, Schoepf UJ et al (2016) Approaches to ultra-low radiation dose coronary artery calcium scoring based on 3rd generation dual-source CT: A phantom study. Eur J Radiol 85:39–47CrossRefPubMed
15.
go back to reference Marwan M, Mettin C, Pflederer T et al (2013) Very low-dose coronary artery calcium scanning with high-pitch spiral acquisition mode: comparison between 120-kV and 100-kV tube voltage protocols. J Cardiovasc Comput Tomogr 7:32–38CrossRefPubMed Marwan M, Mettin C, Pflederer T et al (2013) Very low-dose coronary artery calcium scanning with high-pitch spiral acquisition mode: comparison between 120-kV and 100-kV tube voltage protocols. J Cardiovasc Comput Tomogr 7:32–38CrossRefPubMed
16.
go back to reference Hecht HS, de Siqueira ME, Cham M et al (2015) Low- vs. standard-dose coronary artery calcium scanning. Eur Heart J Cardiovasc Imaging 16:358–363CrossRefPubMed Hecht HS, de Siqueira ME, Cham M et al (2015) Low- vs. standard-dose coronary artery calcium scanning. Eur Heart J Cardiovasc Imaging 16:358–363CrossRefPubMed
17.
go back to reference Nakazato R, Dey D, Gutstein A et al (2009) Coronary artery calcium scoring using a reduced tube voltage and radiation dose protocol with dual-source computed tomography. J Cardiovasc Comput Tomogr 3:394–400CrossRefPubMed Nakazato R, Dey D, Gutstein A et al (2009) Coronary artery calcium scoring using a reduced tube voltage and radiation dose protocol with dual-source computed tomography. J Cardiovasc Comput Tomogr 3:394–400CrossRefPubMed
18.
go back to reference Willemink MJ, Vliegenthart R, Takx RA et al (2014) Coronary artery calcification scoring with state-of-the-art CT scanners from different vendors has substantial effect on risk classification. Radiology 273:695–702CrossRefPubMed Willemink MJ, Vliegenthart R, Takx RA et al (2014) Coronary artery calcification scoring with state-of-the-art CT scanners from different vendors has substantial effect on risk classification. Radiology 273:695–702CrossRefPubMed
Metadata
Title
High-pitch low-voltage CT coronary artery calcium scoring with tin filtration: accuracy and radiation dose reduction
Authors
Georg Apfaltrer
Moritz H. Albrecht
U. Joseph Schoepf
Taylor M. Duguay
Carlo N. De Cecco
John W. Nance
Domenico De Santis
Paul Apfaltrer
Marwen H. Eid
Chelsea D. Eason
Zachary M. Thompson
Maximilian J. Bauer
Akos Varga-Szemes
Brian E. Jacobs
Erich Sorantin
Christian Tesche
Publication date
01-07-2018
Publisher
Springer Berlin Heidelberg
Published in
European Radiology / Issue 7/2018
Print ISSN: 0938-7994
Electronic ISSN: 1432-1084
DOI
https://doi.org/10.1007/s00330-017-5249-2

Other articles of this Issue 7/2018

European Radiology 7/2018 Go to the issue