High-Pitch Carbon Dioxide Contrasted CT Angiography: Pilot Study

Authors: Tobias Penzkofer, Karin Slebocki, Jochen Grommes, Philipp Bruners, Peter Isfort, Thomas Schmitz-Rode, Stephan Langer, Christiane K. Kuhl, Andreas H. Mahnken

Published in: CardioVascular and Interventional Radiology | Issue 2/2014

Login to get access

Abstract

Purpose

This study was designed to evaluate CO2 computed tomography angiography (CO2-CTA) of the iliac and lower limb arteries in patients with contraindications for iodinated contrast agent (ICA).

Methods

Five patients with contraindications for ICA were examined using CO2-fluoroscopic enhanced angiography (CO2-FLA) and CO2-CTA using a high-pitch examination protocol. Objective (vessel diameter) and subjective (visual score) image quality parameters were evaluated. Pathological findings in both modalities were compared.

Results

CO2-CTA was feasible in all patients without adverse side effects, except for CO2 injection-associated pain. Objective vessel analysis revealed no significant difference in diameters as determined by CO2-CTA and CO2-FLA (0.44 ± 0.4 and 0.46 ± 0.41 mm, p = 0.93). CO2-CTA had on average a higher image-quality score (2.6 ± 1.0 vs. 2.3 ± 1.0, p = 0.009). While for pelvic and upper leg CO2-CTA advantageous (3.1 ± 0.74 vs. 2.7 ± 0.9, p = 0.0014) at good quality scores, for calf vessels no significant improvement was visible (1.9 ± 1.0 vs. 1.7 ± 0.9, p = 0.49) and scores were poorer.

Conclusions

CO2-CTA with high-pitch CT was feasible in a limited number of patients. Image-quality scores were on average higher for CO2-CTA than for CO2-FLA, while limited imaging quality in the vessels below the knee needs further work on the CT protocol. An added value of cross-sectional imaging was apparent but needs further quantification.
Literature
1.
go back to reference Akaeda T, Isaka K, Nakaji T, Kakizaki D, Abe K (2005) Clinical application of virtual hysteroscopy by CO2-multidetector-row computed tomography to submucosal myomas. J Minim Invasive Gynecol 12(3):261–266PubMedCrossRef Akaeda T, Isaka K, Nakaji T, Kakizaki D, Abe K (2005) Clinical application of virtual hysteroscopy by CO2-multidetector-row computed tomography to submucosal myomas. J Minim Invasive Gynecol 12(3):261–266PubMedCrossRef
2.
go back to reference Bae KT, Hong C, Becker CR et al (2003) CO2-enhanced CT imaging for the detection of pulmonary emboli: feasibility study in a porcine model. Acad Radiol 10(3):313–320PubMedCrossRef Bae KT, Hong C, Becker CR et al (2003) CO2-enhanced CT imaging for the detection of pulmonary emboli: feasibility study in a porcine model. Acad Radiol 10(3):313–320PubMedCrossRef
3.
go back to reference Barrett BJ, Katzberg RW, Thomsen HS et al (2006) Contrast-induced nephropathy in patients with chronic kidney disease undergoing computed tomography: a double-blind comparison of iodixanol and iopamidol. Investig Radiol 41(11):815–821CrossRef Barrett BJ, Katzberg RW, Thomsen HS et al (2006) Contrast-induced nephropathy in patients with chronic kidney disease undergoing computed tomography: a double-blind comparison of iodixanol and iopamidol. Investig Radiol 41(11):815–821CrossRef
4.
go back to reference Cohnen M, Vogt C, Beck A et al (2004) Feasibility of MDCT colonography in ultra-low-dose technique in the detection of colorectal lesions: comparison with high-resolution video colonoscopy. Am J Roentgenol 183(5):1355–1359CrossRef Cohnen M, Vogt C, Beck A et al (2004) Feasibility of MDCT colonography in ultra-low-dose technique in the detection of colorectal lesions: comparison with high-resolution video colonoscopy. Am J Roentgenol 183(5):1355–1359CrossRef
5.
go back to reference Dahm JB, Vogelgesang D, Hummel A, Staudt A, Volzke H, Felix SB (2002) A randomized trial of 5 vs. 6 French transradial percutaneous coronary interventions. Catheter Cardiovasc Interv 57(2):172–176PubMedCrossRef Dahm JB, Vogelgesang D, Hummel A, Staudt A, Volzke H, Felix SB (2002) A randomized trial of 5 vs. 6 French transradial percutaneous coronary interventions. Catheter Cardiovasc Interv 57(2):172–176PubMedCrossRef
6.
go back to reference Dill KE, Rybicki FJ, Desjardins B et al (2012) American College of Radiology appropriateness criteria: claudication-suspected vascular etiology acr appropriateness criteria. American College of Radiology Dill KE, Rybicki FJ, Desjardins B et al (2012) American College of Radiology appropriateness criteria: claudication-suspected vascular etiology acr appropriateness criteria. American College of Radiology
7.
go back to reference Fleischmann D (2010) CT angiography: injection and acquisition technique. Radiol Clin N Am 48(2):237–247, vii Fleischmann D (2010) CT angiography: injection and acquisition technique. Radiol Clin N Am 48(2):237–247, vii
8.
go back to reference Greenspon AJ, Patel JD, Lau E et al (2012) Trends in permanent pacemaker implantation in the United States from 1993 to 2009: increasing complexity of patients and procedures. J Am Coll Cardiol 60(16):1540–1545PubMedCrossRef Greenspon AJ, Patel JD, Lau E et al (2012) Trends in permanent pacemaker implantation in the United States from 1993 to 2009: increasing complexity of patients and procedures. J Am Coll Cardiol 60(16):1540–1545PubMedCrossRef
9.
go back to reference Hawkins IF, Caridi JG (1998) Carbon dioxide (CO2) digital subtraction angiography: 26-year experience at the University of Florida. Eur Radiol 8(3):391–402PubMedCrossRef Hawkins IF, Caridi JG (1998) Carbon dioxide (CO2) digital subtraction angiography: 26-year experience at the University of Florida. Eur Radiol 8(3):391–402PubMedCrossRef
10.
go back to reference Hope TA, Herfkens RJ, Denianke KS, LeBoit PE, Hung YY, Weil E (2009) Nephrogenic systemic fibrosis in patients with chronic kidney disease who received gadopentetate dimeglumine. Investig Radiol 44(3):135–139CrossRef Hope TA, Herfkens RJ, Denianke KS, LeBoit PE, Hung YY, Weil E (2009) Nephrogenic systemic fibrosis in patients with chronic kidney disease who received gadopentetate dimeglumine. Investig Radiol 44(3):135–139CrossRef
11.
go back to reference Kessel DO, Robertson I, Patel JT et al (2002) Carbon-dioxide-guided vascular interventions: technique and pitfalls. Cardiovasc Interv Radiol 25(6):476–483CrossRef Kessel DO, Robertson I, Patel JT et al (2002) Carbon-dioxide-guided vascular interventions: technique and pitfalls. Cardiovasc Interv Radiol 25(6):476–483CrossRef
12.
go back to reference Kian K, Wyatt C, Schon D, Packer J, Vassalotti J, Mishler R (2006) Safety of low-dose radiocontrast for interventional AV fistula salvage in stage 4 chronic kidney disease patients. Kidney Int 69(8):1444–1449PubMed Kian K, Wyatt C, Schon D, Packer J, Vassalotti J, Mishler R (2006) Safety of low-dose radiocontrast for interventional AV fistula salvage in stage 4 chronic kidney disease patients. Kidney Int 69(8):1444–1449PubMed
13.
go back to reference Laskowski I, Verhagen HJ, Gagne PJ, Moll FL, Muhs BE (2007) Current state of dynamic imaging in endovascular aortic aneurysm repair. J Endovasc Ther 14(6):807–812PubMedCrossRef Laskowski I, Verhagen HJ, Gagne PJ, Moll FL, Muhs BE (2007) Current state of dynamic imaging in endovascular aortic aneurysm repair. J Endovasc Ther 14(6):807–812PubMedCrossRef
14.
go back to reference Mahnken AH, Bruners P, Mommertz G et al (2008) Carbon dioxide contrast agent for CT arteriography: results in a porcine model. J Vasc Interv Radiol 19(7):1055–1064PubMedCrossRef Mahnken AH, Bruners P, Mommertz G et al (2008) Carbon dioxide contrast agent for CT arteriography: results in a porcine model. J Vasc Interv Radiol 19(7):1055–1064PubMedCrossRef
15.
go back to reference Mahnken AH, Chalabi K, Jalali F, Gunther RW, Buecker A (2004) Magnetic resonance-guided placement of aortic stents grafts: feasibility with real-time magnetic resonance fluoroscopy. J Vasc Interv Radiol 15(2 Pt 1):189–195PubMedCrossRef Mahnken AH, Chalabi K, Jalali F, Gunther RW, Buecker A (2004) Magnetic resonance-guided placement of aortic stents grafts: feasibility with real-time magnetic resonance fluoroscopy. J Vasc Interv Radiol 15(2 Pt 1):189–195PubMedCrossRef
16.
go back to reference Mahnken AH, Penzkofer T, Grommes J et al (2010) Carbon dioxide-enhanced CT-guided placement of aortic stent-grafts: feasibility in an animal model. J Endovasc Ther 17(3):332–339PubMedCrossRef Mahnken AH, Penzkofer T, Grommes J et al (2010) Carbon dioxide-enhanced CT-guided placement of aortic stent-grafts: feasibility in an animal model. J Endovasc Ther 17(3):332–339PubMedCrossRef
17.
go back to reference Marenzi G, Assanelli E, Campodonico J et al (2009) Contrast volume during primary percutaneous coronary intervention and subsequent contrast-induced nephropathy and mortality. Ann Intern Med 150(3):170–177PubMedCrossRef Marenzi G, Assanelli E, Campodonico J et al (2009) Contrast volume during primary percutaneous coronary intervention and subsequent contrast-induced nephropathy and mortality. Ann Intern Med 150(3):170–177PubMedCrossRef
18.
go back to reference Ouwendijk R, de Vries M, Pattynama PMT et al (2005) Imaging peripheral arterial disease: a randomized controlled trial comparing contrast-enhanced MR angiography and multi-detector row CT angiography. Radiology 236(3):1094–1103PubMedCrossRef Ouwendijk R, de Vries M, Pattynama PMT et al (2005) Imaging peripheral arterial disease: a randomized controlled trial comparing contrast-enhanced MR angiography and multi-detector row CT angiography. Radiology 236(3):1094–1103PubMedCrossRef
19.
go back to reference Pannu N, Wiebe N, Tonelli M, Alberta Kidney Disease Network (2006) Prophylaxis strategies for contrast-induced nephropathy. JAMA 295(23):2765–2779PubMedCrossRef Pannu N, Wiebe N, Tonelli M, Alberta Kidney Disease Network (2006) Prophylaxis strategies for contrast-induced nephropathy. JAMA 295(23):2765–2779PubMedCrossRef
20.
go back to reference Penzkofer T, Slebocki K, Grommes J et al (2013) Carbon dioxide-contrasted computed tomography angiography: high pitch protocols and adapted injection parameters improve imaging quality. Rofo-Fortschr. Gebiet Rontgenstrahlen Bildgeb Verfahr 185(2):128–135 Penzkofer T, Slebocki K, Grommes J et al (2013) Carbon dioxide-contrasted computed tomography angiography: high pitch protocols and adapted injection parameters improve imaging quality. Rofo-Fortschr. Gebiet Rontgenstrahlen Bildgeb Verfahr 185(2):128–135
21.
go back to reference Renapurkar RD, Setser RM, O’Donnell TP et al (2012) Aortic volume as an indicator of disease progression in patients with untreated infrarenal abdominal aneurysm. Eur J Radiol 81(2):e87–e93PubMedCrossRef Renapurkar RD, Setser RM, O’Donnell TP et al (2012) Aortic volume as an indicator of disease progression in patients with untreated infrarenal abdominal aneurysm. Eur J Radiol 81(2):e87–e93PubMedCrossRef
22.
go back to reference Rydahl C, Thomsen HS, Marckmann P (2008) High prevalence of nephrogenic systemic fibrosis in chronic renal failure patients exposed to gadodiamide, a gadolinium-containing magnetic resonance contrast agent. Investig Radiol 43(2):141–144CrossRef Rydahl C, Thomsen HS, Marckmann P (2008) High prevalence of nephrogenic systemic fibrosis in chronic renal failure patients exposed to gadodiamide, a gadolinium-containing magnetic resonance contrast agent. Investig Radiol 43(2):141–144CrossRef
23.
go back to reference Saam T, Ferguson MS, Yarnykh VL et al (2005) Quantitative evaluation of carotid plaque composition by in vivo MRI. Arterioscler Thromb Vasc Biol 25(1):234–239PubMed Saam T, Ferguson MS, Yarnykh VL et al (2005) Quantitative evaluation of carotid plaque composition by in vivo MRI. Arterioscler Thromb Vasc Biol 25(1):234–239PubMed
24.
go back to reference Schroeder S, Kopp AF, Baumbach A et al (2001) Noninvasive detection and evaluation of atherosclerotic coronary plaques with multislice computed tomography. J Am Coll Cardiol 37(5):1430–1435PubMedCrossRef Schroeder S, Kopp AF, Baumbach A et al (2001) Noninvasive detection and evaluation of atherosclerotic coronary plaques with multislice computed tomography. J Am Coll Cardiol 37(5):1430–1435PubMedCrossRef
25.
go back to reference Solomon R, Dumouchel W (2006) Contrast media and nephropathy: findings from systematic analysis and Food and Drug Administration reports of adverse effects. Investig Radiol 41(8):651–660CrossRef Solomon R, Dumouchel W (2006) Contrast media and nephropathy: findings from systematic analysis and Food and Drug Administration reports of adverse effects. Investig Radiol 41(8):651–660CrossRef
26.
go back to reference Spuentrup E, Ruebben A, Mahnken A et al (2005) Artifact-free coronary magnetic resonance angiography and coronary vessel wall imaging in the presence of a new, metallic, coronary magnetic resonance imaging stent. Circulation 111(8):1019–1026PubMedCrossRef Spuentrup E, Ruebben A, Mahnken A et al (2005) Artifact-free coronary magnetic resonance angiography and coronary vessel wall imaging in the presence of a new, metallic, coronary magnetic resonance imaging stent. Circulation 111(8):1019–1026PubMedCrossRef
27.
go back to reference Sun ZH (2006) Diagnostic accuracy of multislice CT angiography in peripheral arterial disease. J Vasc Interv Radiol 17(12):1915–1921PubMedCrossRef Sun ZH (2006) Diagnostic accuracy of multislice CT angiography in peripheral arterial disease. J Vasc Interv Radiol 17(12):1915–1921PubMedCrossRef
28.
go back to reference Thomsen HS, Morcos SK (2009) Risk of contrast-medium-induced nephropathy in high-risk patients undergoing MDCT—a pooled analysis of two randomized trials. Eur Radiol 19(4):891–897PubMedCrossRef Thomsen HS, Morcos SK (2009) Risk of contrast-medium-induced nephropathy in high-risk patients undergoing MDCT—a pooled analysis of two randomized trials. Eur Radiol 19(4):891–897PubMedCrossRef
Metadata
Title
High-Pitch Carbon Dioxide Contrasted CT Angiography: Pilot Study
Authors
Tobias Penzkofer
Karin Slebocki
Jochen Grommes
Philipp Bruners
Peter Isfort
Thomas Schmitz-Rode
Stephan Langer
Christiane K. Kuhl
Andreas H. Mahnken
Publication date
01-04-2014
Publisher
Springer US
Published in
CardioVascular and Interventional Radiology / Issue 2/2014
Print ISSN: 0174-1551
Electronic ISSN: 1432-086X
DOI
https://doi.org/10.1007/s00270-013-0834-5

Other articles of this Issue 2/2014

CardioVascular and Interventional Radiology 2/2014 Go to the issue