Published in: Critical Care 3/2009

Open Access 01-06-2009 | Research

High flow biphasic positive airway pressure by helmet – effects on pressurization, tidal volume, carbon dioxide accumulation and noise exposure

Authors: Onnen Moerer, Peter Herrmann, José Hinz, Paolo Severgnini, Edoardo Calderini, Michael Quintel, Paolo Pelosi

Published in: Critical Care | Issue 3/2009

Login to get access

Abstract

Introduction

Non-invasive ventilation (NIV) with a helmet device is often associated with poor patient-ventilator synchrony and impaired carbon dioxide (CO2) removal, which might lead to failure. A possible solution is to use a high free flow system in combination with a time-cycled pressure valve placed into the expiratory circuit (HF-BiPAP). This system would be independent from triggering while providing a high flow to eliminate CO2.

Methods

Conventional pressure support ventilation (PSV) and time-cycled biphasic pressure controlled ventilation (BiVent) delivered by an Intensive Care Unit ventilator were compared to HF-BiPAP in an in vitro lung model study. Variables included delta pressures of 5 and 15 cmH2O, respiratory rates of 15 and 30 breaths/min, inspiratory efforts (respiratory drive) of 2.5 and 10 cmH2O) and different lung characteristics. Additionally, CO2 removal and noise exposure were measured.

Results

Pressurization during inspiration was more effective with pressure controlled modes compared to PSV (P < 0.001) at similar tidal volumes. During the expiratory phase, BiVent and HF-BiPAP led to an increase in pressure burden compared to PSV. This was especially true at higher upper pressures (P < 0.001). At high level of asynchrony both HF-BiPAP and BiVent were less effective. Only HF-BiPAP ventilation effectively removed CO2 (P < 0.001) during all settings. Noise exposure was higher during HF-BiPAP (P < 0.001).

Conclusions

This study demonstrates that in a lung model, the efficiency of NIV by helmet can be improved by using HF-BiPAP. However, it imposes a higher pressure during the expiratory phase. CO2 was almost completely removed with HF-BiPAP during all settings.
Appendix
Available only for authorised users
Literature
1.
go back to reference Ferrer M, Valencia M, Nicolas JM, Bernadich O, Badia JR, Torres A: Early noninvasive ventilation averts extubation failure in patients at risk: a randomized trial. Am J Respir Crit Care Med 2006, 173: 164-170. 10.1164/rccm.200505-718OCCrossRefPubMed Ferrer M, Valencia M, Nicolas JM, Bernadich O, Badia JR, Torres A: Early noninvasive ventilation averts extubation failure in patients at risk: a randomized trial. Am J Respir Crit Care Med 2006, 173: 164-170. 10.1164/rccm.200505-718OCCrossRefPubMed
2.
go back to reference Ferrer M, Esquinas A, Arancibia F, Bauer TT, Gonzalez G, Carrillo A, Rodriguez-Roisin R, Torres A: Noninvasive ventilation during persistent weaning failure: a randomized controlled trial. Am J Respir Crit Care Med 2003, 168: 70-76. 10.1164/rccm.200209-1074OCCrossRefPubMed Ferrer M, Esquinas A, Arancibia F, Bauer TT, Gonzalez G, Carrillo A, Rodriguez-Roisin R, Torres A: Noninvasive ventilation during persistent weaning failure: a randomized controlled trial. Am J Respir Crit Care Med 2003, 168: 70-76. 10.1164/rccm.200209-1074OCCrossRefPubMed
3.
go back to reference Carlucci A, Richard JC, Wysocki M, Lepage E, Brochard L: Noninvasive versus conventional mechanical ventilation. An epidemiologic survey. Am J Respir Crit Care Med 2001, 163: 874-880.CrossRefPubMed Carlucci A, Richard JC, Wysocki M, Lepage E, Brochard L: Noninvasive versus conventional mechanical ventilation. An epidemiologic survey. Am J Respir Crit Care Med 2001, 163: 874-880.CrossRefPubMed
4.
go back to reference Brochard L, Mancebo J, Wysocki M, Lofaso F, Conti G, Rauss A, Simonneau G, Benito S, Gasparetto A, Lemaire F: Noninvasive ventilation for acute exacerbations of chronic obstructive pulmonary disease. N Engl J Med 1995, 333: 817-822. 10.1056/NEJM199509283331301CrossRefPubMed Brochard L, Mancebo J, Wysocki M, Lofaso F, Conti G, Rauss A, Simonneau G, Benito S, Gasparetto A, Lemaire F: Noninvasive ventilation for acute exacerbations of chronic obstructive pulmonary disease. N Engl J Med 1995, 333: 817-822. 10.1056/NEJM199509283331301CrossRefPubMed
5.
go back to reference Ram FS, Picot J, Lightowler J, Wedzicha JA: Non-invasive positive pressure ventilation for treatment of respiratory failure due to exacerbations of chronic obstructive pulmonary disease. Cochrane Database Syst Rev 2004, CD004104. Ram FS, Picot J, Lightowler J, Wedzicha JA: Non-invasive positive pressure ventilation for treatment of respiratory failure due to exacerbations of chronic obstructive pulmonary disease. Cochrane Database Syst Rev 2004, CD004104.
6.
go back to reference Winck JC, Azevedo LF, Costa-Pereira A, Antonelli M, Wyatt JC: Efficacy and safety of non-invasive ventilation in the treatment of acute cardiogenic pulmonary edema – a systematic review and meta-analysis. Crit Care 2006, 10: R69. 10.1186/cc4905PubMedCentralCrossRefPubMed Winck JC, Azevedo LF, Costa-Pereira A, Antonelli M, Wyatt JC: Efficacy and safety of non-invasive ventilation in the treatment of acute cardiogenic pulmonary edema – a systematic review and meta-analysis. Crit Care 2006, 10: R69. 10.1186/cc4905PubMedCentralCrossRefPubMed
7.
go back to reference Esteban A, Ferguson ND, Meade MO, Frutos-Vivar F, Apezteguia C, Brochard L, Raymondos K, Nin N, Hurtado J, Tomicic V, Gonzalez M, Elizalde J, Nightingale P, Abroug F, Pelosi P, Arabi Y, Moreno R, Jibaja M, D'Empaire G, Sandi F, Matamis D, Montanez AM, Anzueto A: Evolution of mechanical ventilation in response to clinical research. Am J Respir Crit Care Med 2008, 177: 170-177. 10.1164/rccm.200706-893OCCrossRefPubMed Esteban A, Ferguson ND, Meade MO, Frutos-Vivar F, Apezteguia C, Brochard L, Raymondos K, Nin N, Hurtado J, Tomicic V, Gonzalez M, Elizalde J, Nightingale P, Abroug F, Pelosi P, Arabi Y, Moreno R, Jibaja M, D'Empaire G, Sandi F, Matamis D, Montanez AM, Anzueto A: Evolution of mechanical ventilation in response to clinical research. Am J Respir Crit Care Med 2008, 177: 170-177. 10.1164/rccm.200706-893OCCrossRefPubMed
8.
go back to reference Conti G, Antonelli M, Navalesi P, Rocco M, Bufi M, Spadetta G, Meduri GU: Noninvasive vs. conventional mechanical ventilation in patients with chronic obstructive pulmonary disease after failure of medical treatment in the ward: a randomized trial. Intensive Care Med 2002, 28: 1701-1707. 10.1007/s00134-002-1478-0CrossRefPubMed Conti G, Antonelli M, Navalesi P, Rocco M, Bufi M, Spadetta G, Meduri GU: Noninvasive vs. conventional mechanical ventilation in patients with chronic obstructive pulmonary disease after failure of medical treatment in the ward: a randomized trial. Intensive Care Med 2002, 28: 1701-1707. 10.1007/s00134-002-1478-0CrossRefPubMed
9.
go back to reference Navalesi P, Fanfulla F, Frigerio P, Gregoretti C, Nava S: Physiologic evaluation of noninvasive mechanical ventilation delivered with three types of masks in patients with chronic hypercapnic respiratory failure. Crit Care Med 2000, 28: 1785-1790. 10.1097/00003246-200006000-00015CrossRefPubMed Navalesi P, Fanfulla F, Frigerio P, Gregoretti C, Nava S: Physiologic evaluation of noninvasive mechanical ventilation delivered with three types of masks in patients with chronic hypercapnic respiratory failure. Crit Care Med 2000, 28: 1785-1790. 10.1097/00003246-200006000-00015CrossRefPubMed
10.
go back to reference Kramer N, Meyer TJ, Meharg J, Cece RD, Hill NS: Randomized, prospective trial of noninvasive positive pressure ventilation in acute respiratory failure. Am J Respir Crit Care Med 1995, 151: 1799-1806.CrossRefPubMed Kramer N, Meyer TJ, Meharg J, Cece RD, Hill NS: Randomized, prospective trial of noninvasive positive pressure ventilation in acute respiratory failure. Am J Respir Crit Care Med 1995, 151: 1799-1806.CrossRefPubMed
11.
go back to reference Gregoretti C, Confalonieri M, Navalesi P, Squadrone V, Frigerio P, Beltrame F, Carbone G, Conti G, Gamna F, Nava S, Calderini E, Skrobik Y, Antonelli M: Evaluation of patient skin breakdown and comfort with a new face mask for non-invasive ventilation: a multi-center study. Intensive Care Med 2002, 28: 278-284. 10.1007/s00134-002-1208-7CrossRefPubMed Gregoretti C, Confalonieri M, Navalesi P, Squadrone V, Frigerio P, Beltrame F, Carbone G, Conti G, Gamna F, Nava S, Calderini E, Skrobik Y, Antonelli M: Evaluation of patient skin breakdown and comfort with a new face mask for non-invasive ventilation: a multi-center study. Intensive Care Med 2002, 28: 278-284. 10.1007/s00134-002-1208-7CrossRefPubMed
12.
go back to reference Antonelli M, Pennisi MA, Conti G: New advances in the use of noninvasive ventilation for acute hypoxaemic respiratory failure. Eur Respir J Suppl 2003, 42: 65s-71s. 10.1183/09031936.03.00421003CrossRefPubMed Antonelli M, Pennisi MA, Conti G: New advances in the use of noninvasive ventilation for acute hypoxaemic respiratory failure. Eur Respir J Suppl 2003, 42: 65s-71s. 10.1183/09031936.03.00421003CrossRefPubMed
13.
go back to reference Chiumello D, Pelosi P, Carlesso E, Severgnini P, Aspesi M, Gamberoni C, Antonelli M, Conti G, Chiaranda M, Gattinoni L: Noninvasive positive pressure ventilation delivered by helmet vs. standard face mask. Intensive Care Med 2003, 29: 1671-1679. 10.1007/s00134-003-1825-9CrossRefPubMed Chiumello D, Pelosi P, Carlesso E, Severgnini P, Aspesi M, Gamberoni C, Antonelli M, Conti G, Chiaranda M, Gattinoni L: Noninvasive positive pressure ventilation delivered by helmet vs. standard face mask. Intensive Care Med 2003, 29: 1671-1679. 10.1007/s00134-003-1825-9CrossRefPubMed
14.
go back to reference Antonelli M, Conti G, Pelosi P, Gregoretti C, Pennisi MA, Costa R, Severgnini P, Chiaranda M, Proietti R: New treatment of acute hypoxemic respiratory failure: noninvasive pressure support ventilation delivered by helmet – a pilot controlled trial. Crit Care Med 2002, 30: 602-608. 10.1097/00003246-200203000-00019CrossRefPubMed Antonelli M, Conti G, Pelosi P, Gregoretti C, Pennisi MA, Costa R, Severgnini P, Chiaranda M, Proietti R: New treatment of acute hypoxemic respiratory failure: noninvasive pressure support ventilation delivered by helmet – a pilot controlled trial. Crit Care Med 2002, 30: 602-608. 10.1097/00003246-200203000-00019CrossRefPubMed
15.
go back to reference Pelosi P, Severgnini P, Aspesi M, Gamberoni C, Chiumello D, Fachinetti C, Introzzi L, Antonelli M, Chiaranda M: Non-invasive ventilation delivered by conventional interfaces and helmet in the emergency department. Eur J Emerg Med 2003, 10: 79-86. 10.1097/00063110-200306000-00002CrossRefPubMed Pelosi P, Severgnini P, Aspesi M, Gamberoni C, Chiumello D, Fachinetti C, Introzzi L, Antonelli M, Chiaranda M: Non-invasive ventilation delivered by conventional interfaces and helmet in the emergency department. Eur J Emerg Med 2003, 10: 79-86. 10.1097/00063110-200306000-00002CrossRefPubMed
16.
go back to reference Patroniti N, Foti G, Manfio A, Coppo A, Bellani G, Pesenti A: Head helmet versus face mask for non-invasive continuous positive airway pressure: a physiological study. Intensive Care Med 2003, 29: 1680-1687. 10.1007/s00134-003-1931-8CrossRefPubMed Patroniti N, Foti G, Manfio A, Coppo A, Bellani G, Pesenti A: Head helmet versus face mask for non-invasive continuous positive airway pressure: a physiological study. Intensive Care Med 2003, 29: 1680-1687. 10.1007/s00134-003-1931-8CrossRefPubMed
17.
go back to reference Navalesi P, Costa R, Ceriana P, Carlucci A, Prinianakis G, Antonelli M, Conti G, Nava S: Non-invasive ventilation in chronic obstructive pulmonary disease patients: helmet versus facial mask. Intensive Care Med 2006, 33: 74-81. 10.1007/s00134-006-0391-3CrossRefPubMed Navalesi P, Costa R, Ceriana P, Carlucci A, Prinianakis G, Antonelli M, Conti G, Nava S: Non-invasive ventilation in chronic obstructive pulmonary disease patients: helmet versus facial mask. Intensive Care Med 2006, 33: 74-81. 10.1007/s00134-006-0391-3CrossRefPubMed
18.
go back to reference Moerer O, Fischer S, Hartelt M, Kuvaki B, Quintel M, Neumann P: Influence of two different interfaces for noninvasive ventilation compared to invasive ventilation on the mechanical properties and performance of a respiratory system: a lung model study. Chest 2006, 129: 1424-1431. 10.1378/chest.129.6.1424CrossRefPubMed Moerer O, Fischer S, Hartelt M, Kuvaki B, Quintel M, Neumann P: Influence of two different interfaces for noninvasive ventilation compared to invasive ventilation on the mechanical properties and performance of a respiratory system: a lung model study. Chest 2006, 129: 1424-1431. 10.1378/chest.129.6.1424CrossRefPubMed
19.
go back to reference Racca F, Appendini L, Gregoretti C, Stra E, Patessio A, Donner CF, Ranieri VM: Effectiveness of mask and helmet interfaces to deliver noninvasive ventilation in a human model of resistive breathing. J Appl Physiol 2005, 99: 1262-1271. 10.1152/japplphysiol.01363.2004CrossRefPubMed Racca F, Appendini L, Gregoretti C, Stra E, Patessio A, Donner CF, Ranieri VM: Effectiveness of mask and helmet interfaces to deliver noninvasive ventilation in a human model of resistive breathing. J Appl Physiol 2005, 99: 1262-1271. 10.1152/japplphysiol.01363.2004CrossRefPubMed
20.
go back to reference Mojoli F, Iotti GA, Gerletti M, Lucarini C, Braschi A: Carbon dioxide rebreathing during non-invasive ventilation delivered by helmet: a bench study. Intensive Care Med 2008, 34: 1454-1460. 10.1007/s00134-008-1109-5CrossRefPubMed Mojoli F, Iotti GA, Gerletti M, Lucarini C, Braschi A: Carbon dioxide rebreathing during non-invasive ventilation delivered by helmet: a bench study. Intensive Care Med 2008, 34: 1454-1460. 10.1007/s00134-008-1109-5CrossRefPubMed
21.
go back to reference Chiumello D, Chierichetti M, Tallarini F, Cozzi P, Cressoni M, Polli F, Colombo R, Castelli A, Gattinoni L: Effect of a heated humidifier during continuous positive airway pressure delivered by a helmet. Crit Care 2008, 12: R55. 10.1186/cc6875PubMedCentralCrossRefPubMed Chiumello D, Chierichetti M, Tallarini F, Cozzi P, Cressoni M, Polli F, Colombo R, Castelli A, Gattinoni L: Effect of a heated humidifier during continuous positive airway pressure delivered by a helmet. Crit Care 2008, 12: R55. 10.1186/cc6875PubMedCentralCrossRefPubMed
22.
go back to reference Moerer O, Beck J, Brander L, Quintel M, Brunet F, Sinderby C: Subject-ventilator synchrony during neural vs. pressure triggered non-invasive helmet ventilation. Intensive Care Med 2006, 32: 148-148. Moerer O, Beck J, Brander L, Quintel M, Brunet F, Sinderby C: Subject-ventilator synchrony during neural vs. pressure triggered non-invasive helmet ventilation. Intensive Care Med 2006, 32: 148-148.
23.
go back to reference Costa R, Moerer O, Beck J, Navalesi P, Brunet F, Sinderby C: Inspiratory effort during neural and pressure triggering of helmet ventilation in healthy subjects. Intensive Care Med 2006, 31: 148-148. Costa R, Moerer O, Beck J, Navalesi P, Brunet F, Sinderby C: Inspiratory effort during neural and pressure triggering of helmet ventilation in healthy subjects. Intensive Care Med 2006, 31: 148-148.
24.
go back to reference Henzler D, Pelosi P, Bensberg R, Dembinski R, Quintel M, Pielen V, Rossaint R, Kuhlen R: Effects of partial ventilatory support modalities on respiratory function in severe hypoxemic lung injury. Crit Care Med 2006, 34: 1738-1745. 10.1097/01.CCM.0000218809.49883.54CrossRefPubMed Henzler D, Pelosi P, Bensberg R, Dembinski R, Quintel M, Pielen V, Rossaint R, Kuhlen R: Effects of partial ventilatory support modalities on respiratory function in severe hypoxemic lung injury. Crit Care Med 2006, 34: 1738-1745. 10.1097/01.CCM.0000218809.49883.54CrossRefPubMed
25.
go back to reference Schettino GP, Chatmongkolchart S, Hess DR, Kacmarek RM: Position of exhalation port and mask design affect CO2 rebreathing during noninvasive positive pressure ventilation. Crit Care Med 2003, 31: 2178-2182. 10.1097/01.CCM.0000081309.71887.E9CrossRefPubMed Schettino GP, Chatmongkolchart S, Hess DR, Kacmarek RM: Position of exhalation port and mask design affect CO2 rebreathing during noninvasive positive pressure ventilation. Crit Care Med 2003, 31: 2178-2182. 10.1097/01.CCM.0000081309.71887.E9CrossRefPubMed
26.
go back to reference Lofaso F, Brochard L, Touchard D, Hang T, Harf A, Isabey D: Evaluation of carbon dioxide rebreathing during pressure support ventilation with airway management system (BiPAP) devices. Chest 1995, 108: 772-778. 10.1378/chest.108.3.772CrossRefPubMed Lofaso F, Brochard L, Touchard D, Hang T, Harf A, Isabey D: Evaluation of carbon dioxide rebreathing during pressure support ventilation with airway management system (BiPAP) devices. Chest 1995, 108: 772-778. 10.1378/chest.108.3.772CrossRefPubMed
27.
go back to reference Taccone P, Hess D, Caironi P, Bigatello LM: Continuous positive airway pressure delivered with a "helmet": effects on carbon dioxide rebreathing. Crit Care Med 2004, 32: 2090-2096. 10.1097/01.CCM.0000142577.63316.C0CrossRefPubMed Taccone P, Hess D, Caironi P, Bigatello LM: Continuous positive airway pressure delivered with a "helmet": effects on carbon dioxide rebreathing. Crit Care Med 2004, 32: 2090-2096. 10.1097/01.CCM.0000142577.63316.C0CrossRefPubMed
28.
go back to reference Delclaux C, L'Her E, Alberti C, Mancebo J, Abroug F, Conti G, Guerin C, Schortgen F, Lefort Y, Antonelli M, Lepage E, Lemaire F, Brochard L: Treatment of acute hypoxemic nonhypercapnic respiratory insufficiency with continuous positive airway pressure delivered by a face mask: A randomized controlled trial. JAMA 2000, 284: 2352-2360. 10.1001/jama.284.18.2352CrossRefPubMed Delclaux C, L'Her E, Alberti C, Mancebo J, Abroug F, Conti G, Guerin C, Schortgen F, Lefort Y, Antonelli M, Lepage E, Lemaire F, Brochard L: Treatment of acute hypoxemic nonhypercapnic respiratory insufficiency with continuous positive airway pressure delivered by a face mask: A randomized controlled trial. JAMA 2000, 284: 2352-2360. 10.1001/jama.284.18.2352CrossRefPubMed
29.
go back to reference Cavaliere F, Conti G, Costa R, Spinazzola G, Proietti R, Sciuto A, Masieri S: Exposure to noise during continuous positive airway pressure: influence of interfaces and delivery systems. Acta Anaesthesiol Scand 2008, 52: 52-56.CrossRefPubMed Cavaliere F, Conti G, Costa R, Spinazzola G, Proietti R, Sciuto A, Masieri S: Exposure to noise during continuous positive airway pressure: influence of interfaces and delivery systems. Acta Anaesthesiol Scand 2008, 52: 52-56.CrossRefPubMed
Metadata
Title
High flow biphasic positive airway pressure by helmet – effects on pressurization, tidal volume, carbon dioxide accumulation and noise exposure
Authors
Onnen Moerer
Peter Herrmann
José Hinz
Paolo Severgnini
Edoardo Calderini
Michael Quintel
Paolo Pelosi
Publication date
01-06-2009
Publisher
BioMed Central
Published in
Critical Care / Issue 3/2009
Electronic ISSN: 1364-8535
DOI
https://doi.org/10.1186/cc7907

Other articles of this Issue 3/2009

Critical Care 3/2009 Go to the issue