Abstract

Backgrounds

Hemolysis related to a kinked prosthetic graft or inner felt strip is a very rare complication after aortic surgery. We describe herein a case of hemolytic anemia that developed due to aortic flap of the dissection and inversion of an inner felt strip that was applied at the proximal anastomosis of a replaced ascending aorta 10 years previously.

Case presentation

A 74-year-old woman presented with consistent hemolytic anemia 10 years after replacement of the ascending aorta to treat Stanford type A acute aortic dissection. The cause of hemolysis was attributed to mechanical injury of red blood cells at a site of stenosis caused by aortic flap of the dissection and inversion of the felt strip used for the proximal anastomosis. Repeated resection of the strip and graft replacement of the ascending aorta resolved this problem.

Conclusions

We considered that blood flow disrupted by a jet of blood at the site of the proximal inner felt strip was the cause of severe hemolysis, we describe rare hemolytic anemia at the site of aortic flap and inverted felt strip after replacement of the ascending aorta.
Literature
1.
go back to reference Izumi S, Tano K, Horiike K, Kaihotsu N. Repeat surgery for hemolysis 6 years after replacement of the ascending aorta for acute aortic dissection. Jpn J Thorac Cardiovasc Surg. 2003;51:459–61.CrossRefPubMed Izumi S, Tano K, Horiike K, Kaihotsu N. Repeat surgery for hemolysis 6 years after replacement of the ascending aorta for acute aortic dissection. Jpn J Thorac Cardiovasc Surg. 2003;51:459–61.CrossRefPubMed
2.
go back to reference Matsuura K, Ogino H, Minatoya K, Sakaki H. Aortic stenosis caused by the felt strip used in repair for acute aortic dissection. Interact Cardiovasc Thorac Surg. 2004;3:41–3.CrossRefPubMed Matsuura K, Ogino H, Minatoya K, Sakaki H. Aortic stenosis caused by the felt strip used in repair for acute aortic dissection. Interact Cardiovasc Thorac Surg. 2004;3:41–3.CrossRefPubMed
3.
go back to reference Shingu Y, Aoki H, Ebuoka N, Eya K, Takigami K, Oba J, et al. A surgical case for hemolytic anemia after ascending and total arch replacement. Ann Thorac Cardiovasc Surg. 2005;11:416–8.PubMed Shingu Y, Aoki H, Ebuoka N, Eya K, Takigami K, Oba J, et al. A surgical case for hemolytic anemia after ascending and total arch replacement. Ann Thorac Cardiovasc Surg. 2005;11:416–8.PubMed
4.
go back to reference Nakamura Y, Ogino H, Matsuda H, Minatoya K, Sasaki H, Kitamura S. Hemolytic anemia after operation for aortic dissection using teflon felt strips. Ann Thorac Surg. 2008;85:1784–7.CrossRefPubMed Nakamura Y, Ogino H, Matsuda H, Minatoya K, Sasaki H, Kitamura S. Hemolytic anemia after operation for aortic dissection using teflon felt strips. Ann Thorac Surg. 2008;85:1784–7.CrossRefPubMed
5.
go back to reference Sayar H, Dietl CA, Helms A, Rabinowitz I. Fragmentation hemolytic anemia 8 years after replacement of ascending aorta with a sutureless intraluminal graft. Am J Hematol. 2006;81:175–7.CrossRefPubMed Sayar H, Dietl CA, Helms A, Rabinowitz I. Fragmentation hemolytic anemia 8 years after replacement of ascending aorta with a sutureless intraluminal graft. Am J Hematol. 2006;81:175–7.CrossRefPubMed
6.
go back to reference Hori D, Kawahito K, Tanaka M, Nagano H. Hemolysis after ascending and arch replacement of the aorta; report of a case. Kyobu Geka. 2007;60:1103–6.PubMed Hori D, Kawahito K, Tanaka M, Nagano H. Hemolysis after ascending and arch replacement of the aorta; report of a case. Kyobu Geka. 2007;60:1103–6.PubMed
7.
go back to reference Heper G, Yorukoglu Y, Korkmaz ME. Clinical and hemodynamic follow-up of a patient after operation for dissection of an ascending aortic aneurysm secondary to coarctation of the aorta. Int Heart. 2005;46:1123–31.CrossRef Heper G, Yorukoglu Y, Korkmaz ME. Clinical and hemodynamic follow-up of a patient after operation for dissection of an ascending aortic aneurysm secondary to coarctation of the aorta. Int Heart. 2005;46:1123–31.CrossRef
8.
go back to reference Kunihara T, Iizuka K, Sasaki S, Shiiya N, Sata F, Matsui Y. Optimal proportions of geratin-resorcin-formarin components in aortic surgery. Eur J Cardiovasc Surg. 2009;36:962–6.CrossRef Kunihara T, Iizuka K, Sasaki S, Shiiya N, Sata F, Matsui Y. Optimal proportions of geratin-resorcin-formarin components in aortic surgery. Eur J Cardiovasc Surg. 2009;36:962–6.CrossRef
Metadata
Title
Hemolytic anemia caused by aortic flap and inversion of felt strip after ascending aorta replacement
Authors
Masayuki Sakaguchi
Tamaki Takano
Publication date
01-12-2016
Publisher
BioMed Central
Published in
Journal of Cardiothoracic Surgery / Issue 1/2016
Electronic ISSN: 1749-8090
DOI
https://doi.org/10.1186/s13019-016-0520-1

Other articles of this Issue 1/2016

Journal of Cardiothoracic Surgery 1/2016 Go to the issue