Skip to main content
Top
Published in: Journal of Clinical Monitoring and Computing 6/2022

19-05-2022 | Heart Surgery | Original Research

Evaluation of indigenously developed closed-loop automated blood pressure control system (claps): a preliminary study

Authors: Sumit Kumar, Goverdhan Dutt Puri, Preethy J. Mathew, Banashree Mandal

Published in: Journal of Clinical Monitoring and Computing | Issue 6/2022

Login to get access

Abstract

Closed-loop systems have been designed to assist anesthetists in controlling anesthetic drugs and also maintaining the stability of various physiological variables in the normal range. In the present study, we describe and clinically evaluated a novel closed-loop automated blood pressure control system (CLAPS) in patients undergoing cardiac surgery under cardiopulmonary bypass. Forty ASA II–IV adult patients undergoing elective cardiac surgery were randomly allocated to receive adrenaline, noradrenaline, phenylephrine and nitroglycerine (NTG) adjusted either through CLAPS (CLAPS group) or manually (Manual group). The desired target mean arterial blood pressure (MAP) for each patient in both groups was set by the attending anesthesiologist. The hemodynamic performance was assessed based on the percentage duration of time the MAP remained within 20% of the set target. Automated controller performances were compared using performance error criteria of Varvel (MDPE, MDAPE, Wobble) and Global Score. MAP was maintained a significantly longer proportion of time within 20% of the target in the CLAPS group (79.4% vs. 65.5% p < 0.001, 't' test) as compared to the manual group. Median absolute performance error, wobble, and Global score was significantly lower in the CLAPS group. Hemodynamic stability was achieved with a significantly lower dose of Phenyepherine in the CLAPS group (1870 μg vs. 5400 μg, p < 0.05, 't' test). The dose of NTG was significantly higher in the CLAPS group (3070 μg vs. 1600 μg, p-value < 0.05, 't' test). The cardiac index and left ventricular end-diastolic area were comparable between the groups. Automated infusion of vasoactive drugs using CLAPS is feasible and also better than manual control for controlling hemodynamics during cardiac surgery. Trial registration number and date This trial was registered in the Clinical Trial Registry of India under Registration Number CTRI/2018/01/011487 (Retrospective; registration date; January 23, 2018).
Appendix
Available only for authorised users
Literature
3.
go back to reference Klein EG, Santora JA, Pascale PM, KItrenos JG. Medication cart-filling time, accuracy, and cost with an automated dispensing system. Am J Health Syst Pharmacy 1994;51:1193–6. Klein EG, Santora JA, Pascale PM, KItrenos JG. Medication cart-filling time, accuracy, and cost with an automated dispensing system. Am J Health Syst Pharmacy 1994;51:1193–6.
5.
go back to reference Olkolla KT, Schwilden H. Adaptive closed loop control of vecuronium induced neuromuscular relaxation. Eur J Anaesthesiol. 1991;8:7–12. Olkolla KT, Schwilden H. Adaptive closed loop control of vecuronium induced neuromuscular relaxation. Eur J Anaesthesiol. 1991;8:7–12.
6.
go back to reference Pernick NL, Rodbard D. Personal computer programs to assist with self- monitoring of blood glucose and self- adjustment of insulin dosage. Diabetes Care. 1986;9:61–9.CrossRefPubMed Pernick NL, Rodbard D. Personal computer programs to assist with self- monitoring of blood glucose and self- adjustment of insulin dosage. Diabetes Care. 1986;9:61–9.CrossRefPubMed
7.
go back to reference Liu N, Chazot T, Gently A, Landais A, Restoux A, McGee K, et al. Titration of propofol for anesthetic induction and maintenance guided by the bispectral index: closed loop versus manual control: a prospective, randomized, multi-center study. Anesthesiology. 2006;104:686–95.CrossRefPubMed Liu N, Chazot T, Gently A, Landais A, Restoux A, McGee K, et al. Titration of propofol for anesthetic induction and maintenance guided by the bispectral index: closed loop versus manual control: a prospective, randomized, multi-center study. Anesthesiology. 2006;104:686–95.CrossRefPubMed
8.
go back to reference Puri GD, Kumar B, Aveek J. Closed loop anaesthesia delivery (CLADS) using bispectral index: a performance assessment study. Anaesth Intensive Care. 2007;35:357–62.CrossRefPubMed Puri GD, Kumar B, Aveek J. Closed loop anaesthesia delivery (CLADS) using bispectral index: a performance assessment study. Anaesth Intensive Care. 2007;35:357–62.CrossRefPubMed
9.
go back to reference Rinehart J, Lilot M, Lee C, Joosten A, Huynh T, Canales C, et al. Closed loop assisted versus manual goal directed fluid therapy during high risk abdominal surgery: a case control study with propensity matching. Crit Care. 2015;19:94.CrossRefPubMedPubMedCentral Rinehart J, Lilot M, Lee C, Joosten A, Huynh T, Canales C, et al. Closed loop assisted versus manual goal directed fluid therapy during high risk abdominal surgery: a case control study with propensity matching. Crit Care. 2015;19:94.CrossRefPubMedPubMedCentral
10.
go back to reference Rinehart J, Liu N, Alexander B, Cannesson M. Closed loop systems in anesthesia: is there potential for closed loop fluid management and hemodynamic optimisation. Anesth Analg. 2012;114:130–43.CrossRefPubMed Rinehart J, Liu N, Alexander B, Cannesson M. Closed loop systems in anesthesia: is there potential for closed loop fluid management and hemodynamic optimisation. Anesth Analg. 2012;114:130–43.CrossRefPubMed
11.
go back to reference Sia ATH, Tan HS, Sng BL. Closed loop double vasopressor automated system to treat hypotension during spinal anaesthesia for caesarean section: a preliminary study. Anaesthesia. 2012;67:1348–55.CrossRefPubMed Sia ATH, Tan HS, Sng BL. Closed loop double vasopressor automated system to treat hypotension during spinal anaesthesia for caesarean section: a preliminary study. Anaesthesia. 2012;67:1348–55.CrossRefPubMed
12.
go back to reference Varvel JR, Donoho DL, Shaper SL. Measuring the predictive performance of computer controlled infusion pumps. J Pharm Biopharm. 1992;20:63–94.CrossRef Varvel JR, Donoho DL, Shaper SL. Measuring the predictive performance of computer controlled infusion pumps. J Pharm Biopharm. 1992;20:63–94.CrossRef
13.
go back to reference Ngan Kee WD, Khaw KS, Ng FF, Tam YH. Randomised comparison of closed loop feedback computer controlled with manual controlled infusion of phenylephrine for maintaining arterial pressure during spinal anaesthesia for Caesarean delivery. Br J Anaesth. 2012;110:59–65.CrossRefPubMed Ngan Kee WD, Khaw KS, Ng FF, Tam YH. Randomised comparison of closed loop feedback computer controlled with manual controlled infusion of phenylephrine for maintaining arterial pressure during spinal anaesthesia for Caesarean delivery. Br J Anaesth. 2012;110:59–65.CrossRefPubMed
14.
go back to reference Kim SH, Lilot M, Sidhu KS, Rinehart J, Yu Z, Canales C, et al. Accuracy and precision of continuous noninvasive arterial pressure monitoring compared with invasive arterial pressure: a systematic review and meta-analysis. Anesthesiology. 2014;120:1080–97.CrossRefPubMed Kim SH, Lilot M, Sidhu KS, Rinehart J, Yu Z, Canales C, et al. Accuracy and precision of continuous noninvasive arterial pressure monitoring compared with invasive arterial pressure: a systematic review and meta-analysis. Anesthesiology. 2014;120:1080–97.CrossRefPubMed
Metadata
Title
Evaluation of indigenously developed closed-loop automated blood pressure control system (claps): a preliminary study
Authors
Sumit Kumar
Goverdhan Dutt Puri
Preethy J. Mathew
Banashree Mandal
Publication date
19-05-2022
Publisher
Springer Netherlands
Published in
Journal of Clinical Monitoring and Computing / Issue 6/2022
Print ISSN: 1387-1307
Electronic ISSN: 1573-2614
DOI
https://doi.org/10.1007/s10877-022-00810-8

Other articles of this Issue 6/2022

Journal of Clinical Monitoring and Computing 6/2022 Go to the issue