Skip to main content
Top
Published in: Journal of Cardiothoracic Surgery 1/2021

Open Access 01-12-2021 | Heart Surgery | Research article

Effect of cannulation site on emboli travel during cardiac surgery

Authors: Mira Puthettu, Stijn Vandenberghe, Stefanos Demertzis

Published in: Journal of Cardiothoracic Surgery | Issue 1/2021

Login to get access

Abstract

Background

During cardiac surgery, micro-air emboli regularly enter the blood stream and can cause cognitive impairment or stroke. It is not clearly understood whether the most threatening air emboli are generated by the heart-lung machine (HLM) or by the blood-air contact when opening the heart. We performed an in vitro study to assess, for the two sources, air emboli distribution in the arterial tree, especially in the brain region, during cardiac surgery with different cannulation sites.

Methods

A model of the arterial tree was 3D printed and included in a hydraulic circuit, divided such that flow going to the brain was separated from the rest of the circuit. Air micro-emboli were injected either in the HLM (“ECC Bubbles”) or in the mock left ventricle (“Heart Bubbles”) to simulate the two sources. Emboli distribution was measured with an ultrasonic bubble counter. Five repetitions were performed for each combination of injection site and cannulation site, where air bubble counts and volumes were recorded. Air bubbles were separated in three categories based on size.

Results

For both injection sites, it was possible to identify statistically significant differences between cannulation sites. For ECC Bubbles, axillary cannulation led to a higher amount of air bubbles in the brain with medium-sized bubbles. For Heart Bubbles, aortic cannulation showed a significantly bigger embolic load in the brain with large bubbles.

Conclusions

These preliminary in vitro findings showed that air embolic load in the brain may be dependent on the cannulation site, which deserves further in vivo exploration.
Literature
7.
go back to reference Indja B, Woldendorp K, Vallely MP, Grieve SM. Silent brain infarcts following cardiac procedures: a systematic review and meta-analysis. J Am Heart Assoc. 2019;8:e010920.CrossRef Indja B, Woldendorp K, Vallely MP, Grieve SM. Silent brain infarcts following cardiac procedures: a systematic review and meta-analysis. J Am Heart Assoc. 2019;8:e010920.CrossRef
10.
go back to reference Nussmeier NA. A review of risk factors for adverse neurologic outcome after cardiac surgery. JECT. 2002;34:4–10. Nussmeier NA. A review of risk factors for adverse neurologic outcome after cardiac surgery. JECT. 2002;34:4–10.
11.
go back to reference Mukherjee D, Jani ND, Selvaganesan K, Weng CL, Shadden SC. Computational Assessment of the Relation Between Embolism Source and Embolus Distribution to the Circle of Willis for Improved Understanding of Stroke Etiology. J Biomech Eng. 2016;138(8):081008. https://doi.org/10.1115/1.4033986. Mukherjee D, Jani ND, Selvaganesan K, Weng CL, Shadden SC. Computational Assessment of the Relation Between Embolism Source and Embolus Distribution to the Circle of Willis for Improved Understanding of Stroke Etiology. J Biomech Eng. 2016;138(8):081008. https://​doi.​org/​10.​1115/​1.​4033986.
14.
go back to reference Schönburg M, Urbanek P, Erhardt G, Taborski U, Plechinger H, Hein S, et al. A dynamic bubble trap reduces microbubbles during cardiopulmonary bypass: a case study. J Extra Corpor Technol. 2000;32(3):165–9.PubMed Schönburg M, Urbanek P, Erhardt G, Taborski U, Plechinger H, Hein S, et al. A dynamic bubble trap reduces microbubbles during cardiopulmonary bypass: a case study. J Extra Corpor Technol. 2000;32(3):165–9.PubMed
17.
go back to reference Lei W, Hong WZ, Xin C, Wang LM. Comparison of outcomes between axillary and femoral artery cannulation for type a aortic dissection surgery. Clin Surg. 2019;4:2312. Lei W, Hong WZ, Xin C, Wang LM. Comparison of outcomes between axillary and femoral artery cannulation for type a aortic dissection surgery. Clin Surg. 2019;4:2312.
19.
go back to reference Bongert M, Gehron J, Geller M, Böning A, Grieshaber P. Cannula position and Bernoulli effect contribute to leg malperfusion during extracorporeal life support with femoral arterial cannulation-an in silico simulation study. Interact CardioVasc Thorac Surg. 2019. https://doi.org/10.1093/icvts/ivz048. Bongert M, Gehron J, Geller M, Böning A, Grieshaber P. Cannula position and Bernoulli effect contribute to leg malperfusion during extracorporeal life support with femoral arterial cannulation-an in silico simulation study. Interact CardioVasc Thorac Surg. 2019. https://​doi.​org/​10.​1093/​icvts/​ivz048.
21.
go back to reference Shadden SC, Carr IA, Schwartz RS. Predilections of cardioaortic embolic transport. WCCM XI, 2014; July 20–25, Barcelona, Spain. Shadden SC, Carr IA, Schwartz RS. Predilections of cardioaortic embolic transport. WCCM XI, 2014; July 20–25, Barcelona, Spain.
Metadata
Title
Effect of cannulation site on emboli travel during cardiac surgery
Authors
Mira Puthettu
Stijn Vandenberghe
Stefanos Demertzis
Publication date
01-12-2021
Publisher
BioMed Central
Published in
Journal of Cardiothoracic Surgery / Issue 1/2021
Electronic ISSN: 1749-8090
DOI
https://doi.org/10.1186/s13019-021-01564-1

Other articles of this Issue 1/2021

Journal of Cardiothoracic Surgery 1/2021 Go to the issue