01-08-2021 | ORIGINAL ARTICLE

Head-to-head comparison of a CZT-based all-purpose SPECT camera and a dedicated CZT cardiac device for myocardial perfusion and functional analysis

Authors: Alessia Gimelli, MD, Riccardo Liga, MD, Michelangelo Bertasi, RT, Annette Kusch, RT, Paolo Marzullo, MD

Published in: Journal of Nuclear Cardiology | Issue 4/2021

Login to get access

Abstract

Purpose

To compare the outputs of a novel all-purpose SPECT camera equipped with CZT detectors (Discovery NM/CT 670) with the state-of-the-art represented by a dedicated CZT (Alcyone, Discovery 530c) cardiac camera in patients submitted to myocardial perfusion imaging (MPI).

Methods

We included 19 patients that underwent sequential low-dose 99mTc-tetrofosmin (148-185 MBq during stress and 296-370 MBq at rest) MPI with Alcyone and Discovery 670 cameras. Quantitative (% tracer’s uptake) and semi-quantitative analyses of perfusion data were performed for each scan. Moreover, major left ventricular (LV) functional and structural parameters were derived from each camera and compared.

Results

The two cameras showed excellent correlation for segmental myocardial % uptake at stress (R = 0.90; P < 0.001) and at rest (R = 0.88; P < 0.001) with narrow Bland-Altman limits of agreement. The level of diagnostic agreement of Discovery 670 and Alcyone cameras regarding perfusion analysis was excellent (Cohen’s κ 0.85). Similarly, the two cameras showed excellent correlation in the evaluation of LV ejection fraction (R = 0.95), peak filling rate (R = 0.97), and mass (R = 0.98).

Conclusions

Our preliminary results suggest that MPI with an all-purpose Discovery 670 CZT-SPECT camera is feasible, comparing well with the current state-of-the-art technology.
Appendix
Available only for authorised users
Literature
1.
go back to reference Douglas PS, Hoffmann U, Patel MR, Mark DB, Al-Khalidi HR, Cavanaugh B, et al. Outcomes of anatomical versus functional testing for coronary artery disease. N Engl J Med 2015;372:1291-300.CrossRef Douglas PS, Hoffmann U, Patel MR, Mark DB, Al-Khalidi HR, Cavanaugh B, et al. Outcomes of anatomical versus functional testing for coronary artery disease. N Engl J Med 2015;372:1291-300.CrossRef
2.
go back to reference Abbott BG, Case JA, Dorbala S, Einstein AJ, Galt JR, Pagnanelli R, et al. Contemporary cardiac SPECT imaging-innovations and best practices: An information statement from the American Society of Nuclear Cardiology. Circ Cardiovasc Imaging 2018;11:e000020.CrossRef Abbott BG, Case JA, Dorbala S, Einstein AJ, Galt JR, Pagnanelli R, et al. Contemporary cardiac SPECT imaging-innovations and best practices: An information statement from the American Society of Nuclear Cardiology. Circ Cardiovasc Imaging 2018;11:e000020.CrossRef
3.
go back to reference Herzog BA, Buechel RR, Katz R, Brueckner M, Husmann L, Burger IA, et al. Nuclear myocardial perfusion imaging with a cadmium–zinc–telluride detector technique: Optimized protocol for scan time reduction. J Nucl Med 2010;51:46-51.CrossRef Herzog BA, Buechel RR, Katz R, Brueckner M, Husmann L, Burger IA, et al. Nuclear myocardial perfusion imaging with a cadmium–zinc–telluride detector technique: Optimized protocol for scan time reduction. J Nucl Med 2010;51:46-51.CrossRef
4.
go back to reference Nkoulou R, Pazhenkottil AP, Kuest SM, Ghadri JR, Wolfrum M, Husmann L, et al. Semiconductor detectors allow low-dose-low-dose 1-day SPECT myocardial perfusion imaging. J Nucl Med 2011;52:1204-9.CrossRef Nkoulou R, Pazhenkottil AP, Kuest SM, Ghadri JR, Wolfrum M, Husmann L, et al. Semiconductor detectors allow low-dose-low-dose 1-day SPECT myocardial perfusion imaging. J Nucl Med 2011;52:1204-9.CrossRef
5.
go back to reference Gimelli A, Bottai M, Giorgetti A, Genovesi D, Kusch A, Ripoli A, et al. Comparison between ultrafast and standard single-photon emission CT in patients with coronary artery disease: A pilot study. Circ Cardiovasc Imaging 2011;4:51-8.CrossRef Gimelli A, Bottai M, Giorgetti A, Genovesi D, Kusch A, Ripoli A, et al. Comparison between ultrafast and standard single-photon emission CT in patients with coronary artery disease: A pilot study. Circ Cardiovasc Imaging 2011;4:51-8.CrossRef
6.
go back to reference Gimelli A, Liga R, Duce V, Kusch A, Clemente A, Marzullo P. Accuracy of myocardial perfusion imaging in detecting multivessel coronary artery disease: A cardiac CZT study. J Nucl Cardiol 2017;24:687-95.CrossRef Gimelli A, Liga R, Duce V, Kusch A, Clemente A, Marzullo P. Accuracy of myocardial perfusion imaging in detecting multivessel coronary artery disease: A cardiac CZT study. J Nucl Cardiol 2017;24:687-95.CrossRef
7.
go back to reference Fiechter M, Gebhard C, Fuchs TA, Ghadri JR, Stehli J, Kazakauskaite E, et al. Cadmium-zinc-telluride myocardial perfusion imaging in obese patients. J Nucl Med 2012;53:1401-6.CrossRef Fiechter M, Gebhard C, Fuchs TA, Ghadri JR, Stehli J, Kazakauskaite E, et al. Cadmium-zinc-telluride myocardial perfusion imaging in obese patients. J Nucl Med 2012;53:1401-6.CrossRef
8.
go back to reference Keidar Z, Raysberg I, Lugassi R, Frenkel A, Israel O. Novel cadmium zinc telluride based detector general purpose gamma camera: Initial evaluation and comparison with a standard camera. J Nucl Med. 2016;57:259. Keidar Z, Raysberg I, Lugassi R, Frenkel A, Israel O. Novel cadmium zinc telluride based detector general purpose gamma camera: Initial evaluation and comparison with a standard camera. J Nucl Med. 2016;57:259.
9.
go back to reference Gregoire B, Pina-Jomir G, Bani-Sadr A, Moreau-Triby C, Janier M, Scheiber C. Four-minute bone SPECT using large-field cadmium–zinc–telluride camera. Clin Nucl Med 2018;43:389-95.CrossRef Gregoire B, Pina-Jomir G, Bani-Sadr A, Moreau-Triby C, Janier M, Scheiber C. Four-minute bone SPECT using large-field cadmium–zinc–telluride camera. Clin Nucl Med 2018;43:389-95.CrossRef
10.
11.
go back to reference Gimelli A, Liga R, Clemente A, Pasanisi EM, Favilli B, Marzullo P. Appropriate choice of stress modality in patients undergoing myocardial perfusion scintigraphy with a cardiac camera equipped with solid-state detectors: the role of diabetes mellitus. Eur Heart J Cardiovasc Imaging 2018;19:1268-75.CrossRef Gimelli A, Liga R, Clemente A, Pasanisi EM, Favilli B, Marzullo P. Appropriate choice of stress modality in patients undergoing myocardial perfusion scintigraphy with a cardiac camera equipped with solid-state detectors: the role of diabetes mellitus. Eur Heart J Cardiovasc Imaging 2018;19:1268-75.CrossRef
12.
go back to reference Gimelli A, Liga R, Pasanisi EM, Casagranda M, Marzullo P. Myocardial ischemia in the absence of obstructive coronary lesion: The role of post-stress diastolic dysfunction in detecting early coronary atherosclerosis. J Nucl Cardiol 2017;24:1542-50.CrossRef Gimelli A, Liga R, Pasanisi EM, Casagranda M, Marzullo P. Myocardial ischemia in the absence of obstructive coronary lesion: The role of post-stress diastolic dysfunction in detecting early coronary atherosclerosis. J Nucl Cardiol 2017;24:1542-50.CrossRef
13.
go back to reference Kapur A, Latus KA, Davies G, Dhawan RT, Eastick S, Jarritt PH, et al. A comparison of three radionuclide myocardial perfusion tracers in clinical practice: the ROBUST study. Eur J Nucl Med Mol Imaging 2002;29:1608-16.CrossRef Kapur A, Latus KA, Davies G, Dhawan RT, Eastick S, Jarritt PH, et al. A comparison of three radionuclide myocardial perfusion tracers in clinical practice: the ROBUST study. Eur J Nucl Med Mol Imaging 2002;29:1608-16.CrossRef
14.
go back to reference Gimelli A, Liga R, Startari U, Giorgetti A, Pieraccini L, Marzullo P. Evaluation of ischaemia in patients with atrial fibrillation: impact of stress protocol on myocardial perfusion imaging accuracy. Eur Heart J Cardiovasc Imaging 2015;16:781-7.CrossRef Gimelli A, Liga R, Startari U, Giorgetti A, Pieraccini L, Marzullo P. Evaluation of ischaemia in patients with atrial fibrillation: impact of stress protocol on myocardial perfusion imaging accuracy. Eur Heart J Cardiovasc Imaging 2015;16:781-7.CrossRef
16.
go back to reference Agostini D, Marie PY, Ben-Haim S, Rouzet F, Songy B, Giordano A, et al. Performance of cardiac cadmium-zinc-telluride gamma camera imaging in coronary artery disease: a review from the cardiovascular committee of the European Association of Nuclear Medicine (EANM). Eur J Nucl Med Mol Imaging 2016;43:2423-32.CrossRef Agostini D, Marie PY, Ben-Haim S, Rouzet F, Songy B, Giordano A, et al. Performance of cardiac cadmium-zinc-telluride gamma camera imaging in coronary artery disease: a review from the cardiovascular committee of the European Association of Nuclear Medicine (EANM). Eur J Nucl Med Mol Imaging 2016;43:2423-32.CrossRef
17.
go back to reference Hindorf C, Oddstig J, Hedeer F, Hansson MJ, Jögi J, Engblom H. Importance of correct patient positioning in myocardial perfusion SPECT when using a CZT camera. J Nucl Cardiol 2014;21:695-702.CrossRef Hindorf C, Oddstig J, Hedeer F, Hansson MJ, Jögi J, Engblom H. Importance of correct patient positioning in myocardial perfusion SPECT when using a CZT camera. J Nucl Cardiol 2014;21:695-702.CrossRef
Metadata
Title
Head-to-head comparison of a CZT-based all-purpose SPECT camera and a dedicated CZT cardiac device for myocardial perfusion and functional analysis
Authors
Alessia Gimelli, MD
Riccardo Liga, MD
Michelangelo Bertasi, RT
Annette Kusch, RT
Paolo Marzullo, MD
Publication date
01-08-2021
Publisher
Springer International Publishing
Published in
Journal of Nuclear Cardiology / Issue 4/2021
Print ISSN: 1071-3581
Electronic ISSN: 1532-6551
DOI
https://doi.org/10.1007/s12350-019-01835-4

Other articles of this Issue 4/2021

Journal of Nuclear Cardiology 4/2021 Go to the issue