Abstract

Purpose

The aim of this study was to investigate the ability of a stemless shoulder prosthesis to restore shoulder anatomy in relation to premorbid anatomy.

Methods

This prospective study was performed between May 2007 and December 2013. The inclusion criteria were patients with primary osteoarthritis (OA) who had undergone stemless total anatomic shoulder arthroplasty. Radiographic measurements were done on anteroposterior X-ray views of the glenohumeral joint.

Results

Sixty-nine patients (70 shoulders) were included in the study. The mean difference between premorbid centre of rotation (COR) and post-operative COR was 1 ± 2 mm (range −3 to 5.8 mm). The mean difference between premorbid humeral head height (HH) and post-operative HH was −1 ± 3 mm (range −9.7 to 8.5 mm). The mean difference between premorbid neck-shaft angle (NSA) and post-operative NSA was −3 ± 12° (range −26 to 20°).

Conclusions

Stemless implants could be of help to reconstruct the shoulder anatomy. This study shows that there are some challenges to be addressed when attempting to ensure optimal implant positioning. The critical step is to determine the correct level of bone cut to avoid varus or valgus humeral head inclination and ensure correct head size.
Literature
1.
go back to reference Walch G, Boileau P (1999) Prosthetic adaptability: a new concept for shoulder arthroplasty. J Shoulder Elbow Surg 8(5):443–51CrossRefPubMed Walch G, Boileau P (1999) Prosthetic adaptability: a new concept for shoulder arthroplasty. J Shoulder Elbow Surg 8(5):443–51CrossRefPubMed
2.
go back to reference Robertson DD, Yuan J, Bigliani LU, Flatow EL, Yamaguchi K (2000) Three-dimensional analysis of the proximal part of the humerus: relevance to arthroplasty. J Bone Joint Surg Am 82-A(11):1594–602PubMed Robertson DD, Yuan J, Bigliani LU, Flatow EL, Yamaguchi K (2000) Three-dimensional analysis of the proximal part of the humerus: relevance to arthroplasty. J Bone Joint Surg Am 82-A(11):1594–602PubMed
5.
go back to reference Huguet D, DeClercq G, Rio B, Teissier J, Zipoli B, TESS Group (2010) Results of a new stemless shoulder prosthesis: radiologic proof of maintained fixation and stability after a minimum of three years' follow-up. J Shoulder Elbow Surg 19(6):847–52. doi:10.1016/j.jse.2009.12.009 CrossRefPubMed Huguet D, DeClercq G, Rio B, Teissier J, Zipoli B, TESS Group (2010) Results of a new stemless shoulder prosthesis: radiologic proof of maintained fixation and stability after a minimum of three years' follow-up. J Shoulder Elbow Surg 19(6):847–52. doi:10.​1016/​j.​jse.​2009.​12.​009 CrossRefPubMed
9.
go back to reference Mackenzie DB (1993) The antero-superior exposure for total shoulder replacement. Orthop Traumatol 2:690–700CrossRef Mackenzie DB (1993) The antero-superior exposure for total shoulder replacement. Orthop Traumatol 2:690–700CrossRef
11.
go back to reference Harryman DT, Sidles JA, Harris SL, Lippitt SB, Matsen FA (1995) The effect of articular conformity and the size of the humeral head component on laxity and motion after glenohumeral arthroplasty. A study in cadavera. J Bone Joint Surg Am 77(4):555–63PubMed Harryman DT, Sidles JA, Harris SL, Lippitt SB, Matsen FA (1995) The effect of articular conformity and the size of the humeral head component on laxity and motion after glenohumeral arthroplasty. A study in cadavera. J Bone Joint Surg Am 77(4):555–63PubMed
12.
go back to reference Takase K, Imakiire A, Burkhead WZ Jr (2002) Radiographic study of the anatomic relationships of the greater tuberosity. J Shoulder Elbow Surg 11(6):557–61CrossRefPubMed Takase K, Imakiire A, Burkhead WZ Jr (2002) Radiographic study of the anatomic relationships of the greater tuberosity. J Shoulder Elbow Surg 11(6):557–61CrossRefPubMed
18.
22.
go back to reference Pearl ML, Kurutz S (1999) Geometric analysis of commonly used prosthetic systems for proximal humeral replacement. J Bone Joint Surg Am 81(5):660–71PubMed Pearl ML, Kurutz S (1999) Geometric analysis of commonly used prosthetic systems for proximal humeral replacement. J Bone Joint Surg Am 81(5):660–71PubMed
23.
go back to reference Iannotti JP, Gabriel JP, Schneck SL, Evans BG, Misra S (1992) The normal glenohumeral relationships. An anatomical study of one hundred and forty shoulders. J Bone Joint Surg Am 74(4):491–500PubMed Iannotti JP, Gabriel JP, Schneck SL, Evans BG, Misra S (1992) The normal glenohumeral relationships. An anatomical study of one hundred and forty shoulders. J Bone Joint Surg Am 74(4):491–500PubMed
24.
go back to reference Nyffeler RW, Sheikh R, Jacob HA, Gerber C (2004) Influence of humeral prosthesis height on biomechanics of glenohumeral abduction, An in vitro study. J Bone Joint Surg Am 86-A(3):575–80PubMed Nyffeler RW, Sheikh R, Jacob HA, Gerber C (2004) Influence of humeral prosthesis height on biomechanics of glenohumeral abduction, An in vitro study. J Bone Joint Surg Am 86-A(3):575–80PubMed
25.
go back to reference Boileau P, Walch G (1997) The three-dimensional geometry of the proximal humerus. Implications for surgical technique and prosthetic design. J Bone Joint Surg (Br) 79(5):857–65CrossRef Boileau P, Walch G (1997) The three-dimensional geometry of the proximal humerus. Implications for surgical technique and prosthetic design. J Bone Joint Surg (Br) 79(5):857–65CrossRef
26.
27.
28.
go back to reference Rheault W, Albright B, Beyers C, Franta M, Johnson A, Skowronek M, Dougherty J (1992) Intertester reliability of the cervical range of motion device. J Orthop Sports Phys Ther 15(3):147–50CrossRefPubMed Rheault W, Albright B, Beyers C, Franta M, Johnson A, Skowronek M, Dougherty J (1992) Intertester reliability of the cervical range of motion device. J Orthop Sports Phys Ther 15(3):147–50CrossRefPubMed
29.
go back to reference Lindell O, Eriksson L, Strender L-E (2007) The reliability of a 10-test package for patients with prolonged back and neck pain: could an examiner without formal medical education be used without loss of quality? A methodological study. BMC Musculoskelet Disord 8:31. doi:10.1186/1471-2474-8-31 CrossRefPubMedPubMedCentral Lindell O, Eriksson L, Strender L-E (2007) The reliability of a 10-test package for patients with prolonged back and neck pain: could an examiner without formal medical education be used without loss of quality? A methodological study. BMC Musculoskelet Disord 8:31. doi:10.​1186/​1471-2474-8-31 CrossRefPubMedPubMedCentral
30.
go back to reference Kadum B, Sayed-Noor AS, Perisynakis N, Baea S, Sjödén G (2015) Radiologic assessment of glenohumeral relationship: reliability and reproducibility of lateral humeral offset. Surg Radiol Anat [Epub ahead of print] Kadum B, Sayed-Noor AS, Perisynakis N, Baea S, Sjödén G (2015) Radiologic assessment of glenohumeral relationship: reliability and reproducibility of lateral humeral offset. Surg Radiol Anat [Epub ahead of print]
31.
go back to reference Rozing PM, Obermann WR (1999) Osteometry of the glenohumeral joint. J Shoulder Elbow Surg 8(5):438–42CrossRefPubMed Rozing PM, Obermann WR (1999) Osteometry of the glenohumeral joint. J Shoulder Elbow Surg 8(5):438–42CrossRefPubMed
Metadata
Title
Geometrical analysis of stemless shoulder arthroplasty: a radiological study of seventy TESS total shoulder prostheses
Authors
Bakir Kadum
Hamid Hassany
Mats Wadsten
Arkan Sayed-Noor
Göran Sjödén
Publication date
01-04-2016
Publisher
Springer Berlin Heidelberg
Published in
International Orthopaedics / Issue 4/2016
Print ISSN: 0341-2695
Electronic ISSN: 1432-5195
DOI
https://doi.org/10.1007/s00264-015-2935-z

Other articles of this Issue 4/2016

International Orthopaedics 4/2016 Go to the issue