Published in: Archives of Orthopaedic and Trauma Surgery 11/2017

01-11-2017 | Hip Arthroplasty

Geometric analysis and clinical outcome of two cemented stems for primary total hip replacement with and without modular necks

Authors: Marcel Haversath, Christine Wendelborn, Marcus Jäger, Boris Schmidt, Wojciech Kowalczyk, Stefan Landgraeber

Published in: Archives of Orthopaedic and Trauma Surgery | Issue 11/2017

Login to get access

Abstract

Introduction

Restoration of the physiological biomechanical principles of the hip is crucial in total hip replacement. The aim of this study was to compare an arthroplasty system with different offset options (a: Exeter®) with a dual-modular stem (b: Profemur Xm®).

Materials and methods

A local and an inertial coordinate system were used to assist the description of the components’ assembly in the prosthesis. A resection line of the femoral head in standard position was added to the arthroplasties and geometric parameters were measured. The outcomes of 93 patients were clinically evaluated (a: n = 50, b: n = 43). Preoperative planning was compared to postoperative radiographs (femoral offset, leg-length), and clinical scores (HHS, WOMAC, total range of motion) were assessed preoperatively, and then 1 and 2 years after surgery.

Results

The Exeter® offers an offset range from 32.1 to 56.9 mm and the Profemur Xm® a range from 29.3 to 55.3 mm. The leg-length variability of the Profemur Xm® has a range of 25.9 mm, the Exeter® a range of 13.7 mm. The Profemur Xm® offers more possible combinations of offset and leg-length reconstruction. The neck–stem angles of the Exeter® range from 125.2° to 126.3°, of the Profemur Xm® from 127.2° to 142.6°. There was no statistically significant difference in clinical outcome and radiological parameters.

Conclusions

We conclude that both stems offer a wide range of options for anatomical reconstruction of the hip resulting in similarly good clinical results. The Profemur Xm® stem has advantages for the reconstruction of hips that deviate from standard anatomy but has the drawback of additional corrosive wear at the stem/neck interface.
Literature
2.
go back to reference Asayama I, Naito M, Fujisawa M, Kambe T (2002) Relationship between radiographic measurements of reconstructed hip joint position and the Trendelenburg sign. J Arthroplasty 17(6):747–751CrossRefPubMed Asayama I, Naito M, Fujisawa M, Kambe T (2002) Relationship between radiographic measurements of reconstructed hip joint position and the Trendelenburg sign. J Arthroplasty 17(6):747–751CrossRefPubMed
3.
go back to reference McGrory BJ, Morrey BF, Cahalan TD, An KN, Cabanela ME (1995) Effect of femoral offset on range of motion and abductor muscle strength after total hip arthroplasty. J Bone Joint Surg Br 77(6):865–869PubMed McGrory BJ, Morrey BF, Cahalan TD, An KN, Cabanela ME (1995) Effect of femoral offset on range of motion and abductor muscle strength after total hip arthroplasty. J Bone Joint Surg Br 77(6):865–869PubMed
4.
go back to reference Yamaguchi T, Naito M, Asayama I, Ishiko T (2004) Total hip arthroplasty: the relationship between posterolateral reconstruction, abductor muscle strength, and femoral offset. J Orthop Surg (Hong Kong) 12(2):164–167CrossRef Yamaguchi T, Naito M, Asayama I, Ishiko T (2004) Total hip arthroplasty: the relationship between posterolateral reconstruction, abductor muscle strength, and femoral offset. J Orthop Surg (Hong Kong) 12(2):164–167CrossRef
5.
go back to reference Sakalkale DP, Sharkey PF, Eng K, Hozack WJ, Rothman RH (2001) Effect of femoral component offset on polyethylene wear in total hip arthroplasty. Clin Orthop Relat Res 388:125–134CrossRef Sakalkale DP, Sharkey PF, Eng K, Hozack WJ, Rothman RH (2001) Effect of femoral component offset on polyethylene wear in total hip arthroplasty. Clin Orthop Relat Res 388:125–134CrossRef
6.
go back to reference Howie DW (1990) Tissue response in relation to type of wear particles around failed hip arthroplasties. J Arthroplasty 5(4):337–348CrossRefPubMed Howie DW (1990) Tissue response in relation to type of wear particles around failed hip arthroplasties. J Arthroplasty 5(4):337–348CrossRefPubMed
7.
go back to reference Fujita Y, Takagi H, Hase T (1998) Cloning of the gene encoding a protochlorophyllide reductase: the physiological significance of the co-existence of light-dependent and -independent protochlorophyllide reduction systems in the cyanobacterium Plectonema boryanum. Plant Cell Physiol 39(2):177–185CrossRefPubMed Fujita Y, Takagi H, Hase T (1998) Cloning of the gene encoding a protochlorophyllide reductase: the physiological significance of the co-existence of light-dependent and -independent protochlorophyllide reduction systems in the cyanobacterium Plectonema boryanum. Plant Cell Physiol 39(2):177–185CrossRefPubMed
8.
go back to reference Davey JR, O’Connor DO, Burke DW, Harris WH (1993) Femoral component offset. Its effect on strain in bone-cement. J Arthroplasty 8(1):23–26CrossRefPubMed Davey JR, O’Connor DO, Burke DW, Harris WH (1993) Femoral component offset. Its effect on strain in bone-cement. J Arthroplasty 8(1):23–26CrossRefPubMed
11.
go back to reference Bobyn JD, Tanzer M, Krygier JJ, Dujovne AR, Brooks CE (1994) Concerns with modularity in total hip arthroplasty. Clin Orthop Relat Res 298:27–36 Bobyn JD, Tanzer M, Krygier JJ, Dujovne AR, Brooks CE (1994) Concerns with modularity in total hip arthroplasty. Clin Orthop Relat Res 298:27–36
12.
go back to reference Collier JP, Mayor MB, Jensen RE, Surprenant VA, Surprenant HP, McNamar JL, Belec L (1992) Mechanisms of failure of modular prostheses. Clin Orthop Relat Res 285:129–139 Collier JP, Mayor MB, Jensen RE, Surprenant VA, Surprenant HP, McNamar JL, Belec L (1992) Mechanisms of failure of modular prostheses. Clin Orthop Relat Res 285:129–139
13.
go back to reference Barrack RL, Burke DW, Cook SD, Skinner HB, Harris WH (1993) Complications related to modularity of total hip components. J Bone Joint Surg Br 75(5):688–692PubMed Barrack RL, Burke DW, Cook SD, Skinner HB, Harris WH (1993) Complications related to modularity of total hip components. J Bone Joint Surg Br 75(5):688–692PubMed
14.
go back to reference Helm CS, Greenwald AS (2005) The rationale and performance of modularity in total hip arthroplasty. Orthopedics 28(9 Suppl):s1113–s1115PubMed Helm CS, Greenwald AS (2005) The rationale and performance of modularity in total hip arthroplasty. Orthopedics 28(9 Suppl):s1113–s1115PubMed
15.
go back to reference Reigstad O, Siewers P, Rokkum M, Espehaug B (2008) Excellent long-term survival of an uncemented press-fit stem and screw cup in young patients: follow-up of 75 hips for 15–18 years. Acta Orthop 79(2):194–202. doi:10.1080/17453670710014978 CrossRefPubMed Reigstad O, Siewers P, Rokkum M, Espehaug B (2008) Excellent long-term survival of an uncemented press-fit stem and screw cup in young patients: follow-up of 75 hips for 15–18 years. Acta Orthop 79(2):194–202. doi:10.​1080/​1745367071001497​8 CrossRefPubMed
16.
go back to reference Clark JM, Freeman MA, Witham D (1987) The relationship of neck orientation to the shape of the proximal femur. J Arthroplasty 2(2):99–109CrossRefPubMed Clark JM, Freeman MA, Witham D (1987) The relationship of neck orientation to the shape of the proximal femur. J Arthroplasty 2(2):99–109CrossRefPubMed
18.
go back to reference Johnston RC, Brand RA, Crowninshield RD (1979) Reconstruction of the hip. A mathematical approach to determine optimum geometric relationships. J Bone Joint Surg Am 61(5):639–652CrossRefPubMed Johnston RC, Brand RA, Crowninshield RD (1979) Reconstruction of the hip. A mathematical approach to determine optimum geometric relationships. J Bone Joint Surg Am 61(5):639–652CrossRefPubMed
19.
go back to reference Simpson DJ, Little JP, Gray H, Murray DW, Gill HS (2009) Effect of modular neck variation on bone and cement mantle mechanics around a total hip arthroplasty stem. Clin Biomech (Bristol, Avon) 24(3):274–285CrossRef Simpson DJ, Little JP, Gray H, Murray DW, Gill HS (2009) Effect of modular neck variation on bone and cement mantle mechanics around a total hip arthroplasty stem. Clin Biomech (Bristol, Avon) 24(3):274–285CrossRef
20.
go back to reference Aljenaei F, Catelas I, Louati H, Beaule PE, Nganbe M (2016) Effects of hip implant modular neck material and assembly method on fatigue life and distraction force. J Orthop Res. doi:10.1002/jor.23481 PubMed Aljenaei F, Catelas I, Louati H, Beaule PE, Nganbe M (2016) Effects of hip implant modular neck material and assembly method on fatigue life and distraction force. J Orthop Res. doi:10.​1002/​jor.​23481 PubMed
22.
go back to reference Somers JF, Dedrye L, Goeminne S (2016) Metal ion levels in ceramic-on-ceramic THR with cobalt-chrome modular necks: analysis of cobalt and chromium serum levels in 23 healthy hip patients. Hip Int. doi:10.5301/hipint.5000430 Somers JF, Dedrye L, Goeminne S (2016) Metal ion levels in ceramic-on-ceramic THR with cobalt-chrome modular necks: analysis of cobalt and chromium serum levels in 23 healthy hip patients. Hip Int. doi:10.​5301/​hipint.​5000430
Metadata
Title
Geometric analysis and clinical outcome of two cemented stems for primary total hip replacement with and without modular necks
Authors
Marcel Haversath
Christine Wendelborn
Marcus Jäger
Boris Schmidt
Wojciech Kowalczyk
Stefan Landgraeber
Publication date
01-11-2017
Publisher
Springer Berlin Heidelberg
Published in
Archives of Orthopaedic and Trauma Surgery / Issue 11/2017
Print ISSN: 0936-8051
Electronic ISSN: 1434-3916
DOI
https://doi.org/10.1007/s00402-017-2785-9

Other articles of this Issue 11/2017

Archives of Orthopaedic and Trauma Surgery 11/2017 Go to the issue

Abstracts

Abstracts