Abstract

PEEP is regulated by the internal PEEP/maximum peak inspiratory pressure limit (Pmax) valve. Malfunctioning of the PEEP/Pmax valve can result in the creation of unintentional or unstable PEEP, and a reduction of inspired tidal volume. Some of our Dräger Fabius® anesthesia machines were noted to exhibit changes in expiratory waveforms and unstable PEEP during general anesthesia. We considered that the cause was associated with PEEP/Pmax valve malfunction, and then investigated the problems in collaboration with the manufacturer. Seven of the 22 Dräger Fabius® anesthesia workstations at our department exhibited problems with their PEEP/Pmax valves. We replaced the PEEP membrane and sealing washers in these seven anesthesia machines, and the problems were temporarily resolved. After a short interval, however, one of the seven machines began to show a similar phenomenon. We then asked the manufacturer to overhaul the PEEP/Pmax valve and the entire breathing circuit of the machine. On close investigation, we found that the valve components and the internal surface of the breathing circuit were contaminated with unexpected deposits. The build-up of deposits occurred within a year after the previous regular inspection. Our troubleshooting process determined the issue with the PEEP/Pmax valve, which could go unnoticed because the valve is encased inside the breathing circuit, and requires disassembly for close inspection. Our findings should raise awareness regarding the importance of the preventive maintenance cycle as a safety precaution to keep the anesthetic circuit free of unexpected contamination.
Literature
1.
go back to reference Hess DR, Kondili D, Burns E, Bittner EA, Schmidt UH. A 5-year observational study of lung-protective ventilation in the operating room: a single-center experience. J Crit Care. 2013;28(4):533.CrossRef Hess DR, Kondili D, Burns E, Bittner EA, Schmidt UH. A 5-year observational study of lung-protective ventilation in the operating room: a single-center experience. J Crit Care. 2013;28(4):533.CrossRef
2.
go back to reference Severgnini P, Selmo G, Lanza C, Chiesa A, Frigerio A, Bacuzzi A, Dionigi G, Novario R, Gregoretti C, de Abreu MG, Schultz MJ, Jaber S, Futier E, Chiaranda M, Pelosi P. Protective mechanical ventilation during general anesthesia for open abdominal surgery improves postoperative pulmonary function. Anesthesiology. 2013;118(6):1307–21.CrossRef Severgnini P, Selmo G, Lanza C, Chiesa A, Frigerio A, Bacuzzi A, Dionigi G, Novario R, Gregoretti C, de Abreu MG, Schultz MJ, Jaber S, Futier E, Chiaranda M, Pelosi P. Protective mechanical ventilation during general anesthesia for open abdominal surgery improves postoperative pulmonary function. Anesthesiology. 2013;118(6):1307–21.CrossRef
3.
go back to reference Barbosa FT, Castro AA, de Sousa-Rodrigues CF. Positive end-expiratory pressure (PEEP) during anaesthesia for prevention of mortality and postoperative pulmonary complications. Cochrane Database Syst Rev. 2014;12(6):CD007922. Barbosa FT, Castro AA, de Sousa-Rodrigues CF. Positive end-expiratory pressure (PEEP) during anaesthesia for prevention of mortality and postoperative pulmonary complications. Cochrane Database Syst Rev. 2014;12(6):CD007922.
4.
go back to reference Sulemanji D, Kacmarek RM, Jiang Y. Manual and mechanical ventilators. In: Sandberg W, Urman R, Ehrenfeld E, editors. MGH textbook of anesthetic equipment. 1st ed. Philadelphia: Elsevier Saunders; 2011. Sulemanji D, Kacmarek RM, Jiang Y. Manual and mechanical ventilators. In: Sandberg W, Urman R, Ehrenfeld E, editors. MGH textbook of anesthetic equipment. 1st ed. Philadelphia: Elsevier Saunders; 2011.
5.
go back to reference Ehrenwerth J, Eisenkraft JB, Berry JM. Anesthesia ventilators. Anesthesia equipment e-book. Principles and applications. 2nd ed. London: Saunders, Mosby Inc; 2013. p. 148–178. Ehrenwerth J, Eisenkraft JB, Berry JM. Anesthesia ventilators. Anesthesia equipment e-book. Principles and applications. 2nd ed. London: Saunders, Mosby Inc; 2013. p. 148–178.
6.
go back to reference Eng TS, Durieux ME. Case report: automated machine checkout leaves an internal gas leak undetected: the need for complete checkout procedures. Anesth Analg. 2012;114(1):144–6.CrossRef Eng TS, Durieux ME. Case report: automated machine checkout leaves an internal gas leak undetected: the need for complete checkout procedures. Anesth Analg. 2012;114(1):144–6.CrossRef
7.
go back to reference Liew WL, Jayamaha J. Anaesthetic machine leak from desflurane vaporiser. Anaesthesia. 2011;66(5):399–400.CrossRef Liew WL, Jayamaha J. Anaesthetic machine leak from desflurane vaporiser. Anaesthesia. 2011;66(5):399–400.CrossRef
8.
go back to reference Glen J, Marshall S. Gas leak related to Draeger Primus anaesthetic machine. Anaesthesia. 2010;65(7):750.CrossRef Glen J, Marshall S. Gas leak related to Draeger Primus anaesthetic machine. Anaesthesia. 2010;65(7):750.CrossRef
9.
go back to reference Patil V, Mackenzie IM. Hidden gas leak on a Datex-Ohmeda Aestiva/5 anesthetic machine. Anesth Analg. 2007;104(1):234–5.CrossRef Patil V, Mackenzie IM. Hidden gas leak on a Datex-Ohmeda Aestiva/5 anesthetic machine. Anesth Analg. 2007;104(1):234–5.CrossRef
10.
go back to reference Yamakage M, Takahashi K, Takahashi M, Satoh JI, Namiki A. Performance of four carbon dioxide absorbents in experimental and clinical settings. Anaesthesia. 2009;64(3):287–92.CrossRef Yamakage M, Takahashi K, Takahashi M, Satoh JI, Namiki A. Performance of four carbon dioxide absorbents in experimental and clinical settings. Anaesthesia. 2009;64(3):287–92.CrossRef
11.
go back to reference Solanki SL, Doctor JR, Patil VP, Rana M. Positive end-expiratory pressure valve malfunctioning detected by capnography and airway pressure waveform. Ann Card Anaesth. 2014;17(3):255–7.CrossRef Solanki SL, Doctor JR, Patil VP, Rana M. Positive end-expiratory pressure valve malfunctioning detected by capnography and airway pressure waveform. Ann Card Anaesth. 2014;17(3):255–7.CrossRef
Metadata
Title
Unexpected deposits in the anesthetic circuit: a possible cause of PEEP/Pmax valve malfunction
Authors
Takamitsu Ikeda
Ryo Orii
Masaki Iwakiri
Kanji Uchida
Yoshitsugu Yamada
Publication date
01-08-2021
Publisher
Springer Netherlands
Published in
Journal of Clinical Monitoring and Computing / Issue 4/2021
Print ISSN: 1387-1307
Electronic ISSN: 1573-2614
DOI
https://doi.org/10.1007/s10877-020-00562-3

Other articles of this Issue 4/2021

Journal of Clinical Monitoring and Computing 4/2021 Go to the issue