01-10-2007 | Original Article

Fusion and failure following anterior cervical plating with dynamic or rigid plates: 6-months results of a multi-centric, prospective, randomized, controlled study

Authors: Jan Stulik, Tobias Rainer Pitzen, Jan Chrobok, Sabine Ruffing, Jörg Drumm, Laurentius Sova, Ravel Kucera, Tomas Vyskocil, Wolf Ingo Steudel

Published in: European Spine Journal | Issue 10/2007

Login to get access

Abstract

Anterior cervical plate fixation is an approved surgical technique for cervical spine stabilization in the presence of anterior cervical instability. Rigid plate design with screws rigidly locked to the plate is widely used and is thought to provide a better fixation for the treated spinal segment than a dynamic design in which the screws may slide when the graft is settling. Recent biomechanical studies showed that dynamic anterior plates provide a better graft loading possibly leading to accelerated spinal fusion with a lower incidence of implant complications. This, however, was investigated in vitro and does not necessarily mean to be the case in vivo, as well. Thus, the two major aspects of this study were to compare the speed of bone fusion and the rate of implant complications using either rigid- or dynamic plates. The study design is prospective, randomized, controlled, and multi-centric, having been approved by respective ethic committees of all participating sites. One hundred and thirty-two patients were included in this study and randomly assigned to one of the two groups, both undergoing routine level-1- or level-2 anterior cervical discectomy with autograft fusion receiving either a dynamic plate with screws being locked in ap - position (ABC, Aesculap, Germany), or a rigid plate (CSLP, Synthes, Switzerland). Segmental mobility and implant complications were compared after 3- and 6 months, respectively. All measurements were performed by an independent radiologist. Mobility results after 6 months were available for 77 patients (43 ABC/34 CSLP). Mean segmental mobility for the ABC group was 1.7 mm at the time of discharge, 1.4 mm after 3 months, and 0.8 mm after 6 months. For the CSLP- group the measurements were 1.0, 1.8, and 1.7 mm, respectively. The differences of mean segmental mobility were statistically significant between both groups after 6 months (P = 0.02). Four patients of the CSLP-group demonstrated surgical hardware complications, whereas no implant complications were observed within the ABC-group (P = 0.0375). Dynamic plate designs provided a faster fusion of the cervical spine compared with rigid plate designs after prior spinal surgery. Moreover, the rate of implant complications was lower within the group of patients receiving a dynamic plate. These interim results refer to a follow-up period of 6 months after prior spinal surgery. Further investigations will be performed 2 years postoperatively.
Literature
1.
go back to reference Bailey RW, Badgley CE (1960) Stabilization of the cervical spine by anterior fusion Am J Orthop 42-A:565–594 Bailey RW, Badgley CE (1960) Stabilization of the cervical spine by anterior fusion Am J Orthop 42-A:565–594
2.
go back to reference Böhler J, Gaudernak T (1980) Anterior plate stabilization for fracture-dislocations of the lower cervical spine J Trauma 20:203–205PubMedCrossRef Böhler J, Gaudernak T (1980) Anterior plate stabilization for fracture-dislocations of the lower cervical spine J Trauma 20:203–205PubMedCrossRef
3.
go back to reference Brodke DS, Gollogly S, Alexander Mohr R, Nguyen BK, Dailey AT, Bachus AK (2001) Dynamic cervical plates: biomechanical evaluation of load sharing and stiffness. Spine 26:1324–9PubMedCrossRef Brodke DS, Gollogly S, Alexander Mohr R, Nguyen BK, Dailey AT, Bachus AK (2001) Dynamic cervical plates: biomechanical evaluation of load sharing and stiffness. Spine 26:1324–9PubMedCrossRef
4.
go back to reference Brodsky AE, Khalil MA, Sassard WR, Newmann BP (1992) Repair of symptomatic pseudarthrosis of anterior cervical fusion. Posterior versus anterior repair. Spine 17:1137–1143PubMedCrossRef Brodsky AE, Khalil MA, Sassard WR, Newmann BP (1992) Repair of symptomatic pseudarthrosis of anterior cervical fusion. Posterior versus anterior repair. Spine 17:1137–1143PubMedCrossRef
5.
go back to reference Caspar W, Barbier D, Klara PM (1989) Anterior cervical fusion and Caspar plate stabilization for cervical trauma. Neurosurgery 25:491–502PubMedCrossRef Caspar W, Barbier D, Klara PM (1989) Anterior cervical fusion and Caspar plate stabilization for cervical trauma. Neurosurgery 25:491–502PubMedCrossRef
6.
go back to reference Caspar W, Geisler FH, Pitzen T, Johnson TA (1989) Anterior cervical plate stabilization in one- and two-level degenerative disease: overtreatment or benefit? J Spin Disord 11:1–11 Caspar W, Geisler FH, Pitzen T, Johnson TA (1989) Anterior cervical plate stabilization in one- and two-level degenerative disease: overtreatment or benefit? J Spin Disord 11:1–11
7.
go back to reference Chinoy MA, Parker MJ (1999) Fixed nail plates versus sliding hip systems for the treatment of trochanteric femoral fractures: a meta analysis of 14 studies. Injury 30:157–163PubMedCrossRef Chinoy MA, Parker MJ (1999) Fixed nail plates versus sliding hip systems for the treatment of trochanteric femoral fractures: a meta analysis of 14 studies. Injury 30:157–163PubMedCrossRef
8.
go back to reference Cloward RB (1958) The anterior approach for removal of ruptured discs. J Neurosurg 15:602–617PubMed Cloward RB (1958) The anterior approach for removal of ruptured discs. J Neurosurg 15:602–617PubMed
9.
go back to reference Dvorak MF, Pitzen T, Zhu Q, Gordon JD, Fisher CG, Oxland TR (2005) Anterior cervical plate fixation: a biomechanical study to evaluate the effects of plate design, endplate preparation, and bone mineral density. Spine 30:294–301PubMedCrossRef Dvorak MF, Pitzen T, Zhu Q, Gordon JD, Fisher CG, Oxland TR (2005) Anterior cervical plate fixation: a biomechanical study to evaluate the effects of plate design, endplate preparation, and bone mineral density. Spine 30:294–301PubMedCrossRef
10.
go back to reference Geisler FH, Caspar W, Pitzen T, Johnson TA (1998) Reoperation in patients after anterior cervical plate stabilization in degenerative disease. Spine 23:911–920PubMedCrossRef Geisler FH, Caspar W, Pitzen T, Johnson TA (1998) Reoperation in patients after anterior cervical plate stabilization in degenerative disease. Spine 23:911–920PubMedCrossRef
11.
go back to reference Hofmeister M, Bühren V (1999) Therapiekonzept für Verletzungen der unteren Halswirbelsäule. Orthopäde 5:401–413 Hofmeister M, Bühren V (1999) Therapiekonzept für Verletzungen der unteren Halswirbelsäule. Orthopäde 5:401–413
12.
go back to reference Kim S, Kim S, Moon S, Choi S (2005) Outcome analysis of patients fusion with locking screw-plate and with dynamic abc screw-plate in degenerative disc disease. Eur Spine J 14(Suppl):S65 Kim S, Kim S, Moon S, Choi S (2005) Outcome analysis of patients fusion with locking screw-plate and with dynamic abc screw-plate in degenerative disc disease. Eur Spine J 14(Suppl):S65
13.
go back to reference Lowery GL, McDonough RF (1998) The significance of hardware failure in anterior cervical plate fixation. Patients with 2- to 7-year follow-up. Spine 23:181–186PubMedCrossRef Lowery GL, McDonough RF (1998) The significance of hardware failure in anterior cervical plate fixation. Patients with 2- to 7-year follow-up. Spine 23:181–186PubMedCrossRef
14.
go back to reference Morscher E, Sutter F, Jenny H, Olerud S (1986 ) Anterior plating of the cervical spine with the hollow screw-plate system of titanium. Chirurg 11:702–707 Morscher E, Sutter F, Jenny H, Olerud S (1986 ) Anterior plating of the cervical spine with the hollow screw-plate system of titanium. Chirurg 11:702–707
15.
go back to reference Omeis I, DeMatta JA, Hillard VH, Murali R, Kaushik D (2004) History of instrumentation for stabilization of the subaxial cervical spine. Neurosurg Focus 16:1–6CrossRef Omeis I, DeMatta JA, Hillard VH, Murali R, Kaushik D (2004) History of instrumentation for stabilization of the subaxial cervical spine. Neurosurg Focus 16:1–6CrossRef
16.
go back to reference Orozco DR, Llovet TR (1971) Osteosintesis en las lesiones traumaticas y degeneratives de la columna vertebral. Revista Traumatol Chirurg Rehabil 1:45–52 Orozco DR, Llovet TR (1971) Osteosintesis en las lesiones traumaticas y degeneratives de la columna vertebral. Revista Traumatol Chirurg Rehabil 1:45–52
17.
go back to reference Paramore CG, Dickmann CA, Sonntag VK (1996) Radiographic and clinical follow-up review of Caspar plates in 49 patients. J Neurosurg 84:957–961PubMedCrossRef Paramore CG, Dickmann CA, Sonntag VK (1996) Radiographic and clinical follow-up review of Caspar plates in 49 patients. J Neurosurg 84:957–961PubMedCrossRef
18.
go back to reference Pitzen T, Wilke HJ, Caspar W, Claes L, Steudel WI (1999) Biomechanical evaluation of a new monocortical screw for anterior cervical fusion and plating by a combined biomechanical and clinical study. Eur Spine J 8:382–387PubMedCrossRef Pitzen T, Wilke HJ, Caspar W, Claes L, Steudel WI (1999) Biomechanical evaluation of a new monocortical screw for anterior cervical fusion and plating by a combined biomechanical and clinical study. Eur Spine J 8:382–387PubMedCrossRef
19.
go back to reference Robinson RA, Walker AE, Ferlic DC, Wieking DK (1962) The results of anterior interbody fusion of the cervical spine. J Bone Joint Surg [Am] 44:1569–1587 Robinson RA, Walker AE, Ferlic DC, Wieking DK (1962) The results of anterior interbody fusion of the cervical spine. J Bone Joint Surg [Am] 44:1569–1587
20.
go back to reference Smith GW, Robinson RA (1958) The treatment of cervical spine disorders by anterior removal of the intervertebral disc and interbody fusion. J Bone Joint Surg (Am) 40:607–624 Smith GW, Robinson RA (1958) The treatment of cervical spine disorders by anterior removal of the intervertebral disc and interbody fusion. J Bone Joint Surg (Am) 40:607–624
21.
go back to reference Ulrich C, Arand M; Nothwang J (2001) Internal fixation on the lower cervical spine—biomechanics and clinical practice of procedures and implants. Eur Spine J 10:88–100PubMedCrossRef Ulrich C, Arand M; Nothwang J (2001) Internal fixation on the lower cervical spine—biomechanics and clinical practice of procedures and implants. Eur Spine J 10:88–100PubMedCrossRef
22.
go back to reference White AA, Southwick WO, Deponte RJ, Gainor JW, Hardy R (1973) Relief of pain by anterior cervical spine fusion for spondylosis: A report of sixty-five patients. J Bone Joint Surg [Am] 55:525–534 White AA, Southwick WO, Deponte RJ, Gainor JW, Hardy R (1973) Relief of pain by anterior cervical spine fusion for spondylosis: A report of sixty-five patients. J Bone Joint Surg [Am] 55:525–534
23.
go back to reference Zdeblick TA, Ducker TB (1991) The use of freeze-dried allograft bone for anterior cervical fusions. Spine 16:726–729PubMedCrossRef Zdeblick TA, Ducker TB (1991) The use of freeze-dried allograft bone for anterior cervical fusions. Spine 16:726–729PubMedCrossRef
Metadata
Title
Fusion and failure following anterior cervical plating with dynamic or rigid plates: 6-months results of a multi-centric, prospective, randomized, controlled study
Authors
Jan Stulik
Tobias Rainer Pitzen
Jan Chrobok
Sabine Ruffing
Jörg Drumm
Laurentius Sova
Ravel Kucera
Tomas Vyskocil
Wolf Ingo Steudel
Publication date
01-10-2007
Publisher
Springer-Verlag
Published in
European Spine Journal / Issue 10/2007
Print ISSN: 0940-6719
Electronic ISSN: 1432-0932
DOI
https://doi.org/10.1007/s00586-007-0451-6

Other articles of this Issue 10/2007

European Spine Journal 10/2007 Go to the issue