Abstract

Purpose

To compare the CFR-PEEK plates with conventional plates in fracture fixation with regards to clinical and radiological outcomes and complications.

Methods

A systematic literature search was conducted in four online databases independently by two reviewers using the Cochrane methodology for systematic reviews. The identified relevant studies were assessed against predetermined inclusion/exclusion criteria. Independent data extraction and assessment of risk of bias and study quality was carried out.

Results

Nine studies (patient n = 361) out of 6594 records were included for analysis: 2 RCTs (n = 63), 3 prospective cohort studies (n = 151), and 4 retrospective cohort studies (n = 147). Studies were grouped per anatomic area of fixation. Four studies (n = 200) examined fixation of proximal humerus fractures. Two studies (n = 74) examined fixation of distal radius fractures. Two studies (n = 53) assessed outcomes of fixation of distal femur fractures. One study (n = 87) assessed the outcomes of fixation of ankle fractures. All nine studies reported very high union rates (from 91% in distal femur to 100% in upper limb) for the CFR-PEEK plate groups and low complication rates. There was no significant difference in clinical outcomes, and rate of complications as compared to the conventional plate groups.

Conclusion

CFR-PEEK plates have high union rates in extremity fracture fixation similar to conventional plates with comparable good clinical outcomes and a very low and comparable rate of complications. Considering their advantages, CFR-PEEK plates seem to be valid alternative to conventional plating.
Appendix
Available only for authorised users
Literature
1.
go back to reference Bhatt P, Goe A. Carbon fibres: production, properties and potential use. Mater Sci Res India. 2017;14(1):52–7.CrossRef Bhatt P, Goe A. Carbon fibres: production, properties and potential use. Mater Sci Res India. 2017;14(1):52–7.CrossRef
2.
go back to reference Hillock R, Howard S. Utility of carbon fiber implants in orthopedic surgery: literature review. Reconstr Rev. 2014;4(1):23–32. Hillock R, Howard S. Utility of carbon fiber implants in orthopedic surgery: literature review. Reconstr Rev. 2014;4(1):23–32.
3.
go back to reference Li CS, Vannabouathong C, Sprague S, Bhandari M. The use of carbon-fiber-reinforced (CFR) PEEK material in orthopedic implants: a systematic review. Clin Med Insights Arthritis Musculoskelet Disord. 2015;8:33–45.CrossRef Li CS, Vannabouathong C, Sprague S, Bhandari M. The use of carbon-fiber-reinforced (CFR) PEEK material in orthopedic implants: a systematic review. Clin Med Insights Arthritis Musculoskelet Disord. 2015;8:33–45.CrossRef
4.
go back to reference Rotini R, Cavaciocchi M, Fabbri D, Bettelli G, Catani F, Campochiaro G, et al. Proximal humeral fracture fixation: multicenter study with carbon fiber peek plate. Musculoskelet Surg. 2015;99(1):1–8.CrossRef Rotini R, Cavaciocchi M, Fabbri D, Bettelli G, Catani F, Campochiaro G, et al. Proximal humeral fracture fixation: multicenter study with carbon fiber peek plate. Musculoskelet Surg. 2015;99(1):1–8.CrossRef
5.
go back to reference Lindeque BGP, Hak DJ, Mauffrey C, Seligson D, Lindeque B. Use of carbon–fiber-reinforced composite implants in orthopedic surgery. Orthopedics. 2014;37(12):825–30.CrossRef Lindeque BGP, Hak DJ, Mauffrey C, Seligson D, Lindeque B. Use of carbon–fiber-reinforced composite implants in orthopedic surgery. Orthopedics. 2014;37(12):825–30.CrossRef
6.
go back to reference Tarallo L, Mugnai R, Adani R, Zambianchi F, Catani F. A new volar plate made of carbon–fiber-reinforced polyetheretherketon for distal radius fracture: analysis of 40 cases. J Orthop Traumatol. 2014;15(4):277–83.CrossRef Tarallo L, Mugnai R, Adani R, Zambianchi F, Catani F. A new volar plate made of carbon–fiber-reinforced polyetheretherketon for distal radius fracture: analysis of 40 cases. J Orthop Traumatol. 2014;15(4):277–83.CrossRef
7.
go back to reference Goudriaan WA, Tordoir RL, Broekhuis D, van der Wal RJP. Early failure of a carbon–fiber-reinforced polyetheretherketone distal femur plate: a case report. JBJS Case Connect. 2020;10(3):e2000041.CrossRef Goudriaan WA, Tordoir RL, Broekhuis D, van der Wal RJP. Early failure of a carbon–fiber-reinforced polyetheretherketone distal femur plate: a case report. JBJS Case Connect. 2020;10(3):e2000041.CrossRef
8.
go back to reference Di Maggio B, Sessa P, Mantelli P, Maniscalco P, Rivera F, Calori GM, et al. PEEK radiolucent plate for distal radius fractures: multicentre clinical results at 12 months follow up. Injury. 2017;48:S34–8.CrossRef Di Maggio B, Sessa P, Mantelli P, Maniscalco P, Rivera F, Calori GM, et al. PEEK radiolucent plate for distal radius fractures: multicentre clinical results at 12 months follow up. Injury. 2017;48:S34–8.CrossRef
9.
go back to reference Tayton K, Johnson-Nurse C, McKibbin B, Bradley J, Hastings G. The use of semi-rigid carbon–fibre-reinforced plastic plates for fixation of human fractures. Results of preliminary trials. J Bone Jt Surg Br. 1982;64(1):105–11.CrossRef Tayton K, Johnson-Nurse C, McKibbin B, Bradley J, Hastings G. The use of semi-rigid carbon–fibre-reinforced plastic plates for fixation of human fractures. Results of preliminary trials. J Bone Jt Surg Br. 1982;64(1):105–11.CrossRef
10.
go back to reference Hak DJ, Mauffrey C, Seligson D, Lindeque B. Use of carbon–fiber-reinforced composite implants in orthopedic surgery. Orthopedics. 2014;37(12):825–30.CrossRef Hak DJ, Mauffrey C, Seligson D, Lindeque B. Use of carbon–fiber-reinforced composite implants in orthopedic surgery. Orthopedics. 2014;37(12):825–30.CrossRef
11.
12.
go back to reference Higgins JP, Altman DG, Gøtzsche PC, Jüni P, Moher D, Oxman AD, et al. The Cochrane Collaboration’s tool for assessing risk of bias in randomised trials. BMJ. 2011;343:d5928.CrossRef Higgins JP, Altman DG, Gøtzsche PC, Jüni P, Moher D, Oxman AD, et al. The Cochrane Collaboration’s tool for assessing risk of bias in randomised trials. BMJ. 2011;343:d5928.CrossRef
14.
go back to reference Slim K, Nini E, Forestier D, Kwiatkowski F, Panis Y, Chipponi J. Methodological index for non-randomized studies (minors): development and validation of a new instrument. ANZ J Surg. 2003;73(9):712–6.CrossRef Slim K, Nini E, Forestier D, Kwiatkowski F, Panis Y, Chipponi J. Methodological index for non-randomized studies (minors): development and validation of a new instrument. ANZ J Surg. 2003;73(9):712–6.CrossRef
15.
go back to reference Guyatt GH, Oxman AD, Vist GE, Kunz R, Falck-Ytter Y, Alonso-Coello P, et al. GRADE: an emerging consensus on rating quality of evidence and strength of recommendations. BMJ. 2008;336(7650):924–6.CrossRef Guyatt GH, Oxman AD, Vist GE, Kunz R, Falck-Ytter Y, Alonso-Coello P, et al. GRADE: an emerging consensus on rating quality of evidence and strength of recommendations. BMJ. 2008;336(7650):924–6.CrossRef
16.
go back to reference Moher D, Liberati A, Tetzlaff J, Altman DG. Preferred reporting items for systematic reviews and meta-analyses: the PRISMA statement. PLoS Med. 2009;6(7):e1000097.CrossRef Moher D, Liberati A, Tetzlaff J, Altman DG. Preferred reporting items for systematic reviews and meta-analyses: the PRISMA statement. PLoS Med. 2009;6(7):e1000097.CrossRef
17.
go back to reference Mathes T, Pieper D. Clarifying the distinction between case series and cohort studies in systematic reviews of comparative studies: potential impact on body of evidence and workload. BMC Med Res Methodol. 2017;17(1):107.CrossRef Mathes T, Pieper D. Clarifying the distinction between case series and cohort studies in systematic reviews of comparative studies: potential impact on body of evidence and workload. BMC Med Res Methodol. 2017;17(1):107.CrossRef
18.
go back to reference Perugia D, Guzzini M, Mazza D, Iorio C, Civitenga C, Ferretti A. Comparison between carbon-peek volar locking plates and titanium volar locking plates in the treatment of distal radius fractures. Injury. 2017;48:S24–9.CrossRef Perugia D, Guzzini M, Mazza D, Iorio C, Civitenga C, Ferretti A. Comparison between carbon-peek volar locking plates and titanium volar locking plates in the treatment of distal radius fractures. Injury. 2017;48:S24–9.CrossRef
19.
go back to reference Ziegler P, Maier S, Stockle U, Guhring M, Stuby FM. The treatment of proximal humerus fracture using internal fixation with fixed-angle plates. Dtsch Arztebl Int. 2019;116(45):757–63.PubMedPubMedCentral Ziegler P, Maier S, Stockle U, Guhring M, Stuby FM. The treatment of proximal humerus fracture using internal fixation with fixed-angle plates. Dtsch Arztebl Int. 2019;116(45):757–63.PubMedPubMedCentral
20.
go back to reference Guzzini M, Lanzetti RM, Lupariello D, Morelli F, Princi G, Perugia D, et al. Comparison between carbon-peek plate and conventional stainless steal plate in ankle fractures. A prospective study of two years follow up. Injury. 2017;48(6):1249–52.CrossRef Guzzini M, Lanzetti RM, Lupariello D, Morelli F, Princi G, Perugia D, et al. Comparison between carbon-peek plate and conventional stainless steal plate in ankle fractures. A prospective study of two years follow up. Injury. 2017;48(6):1249–52.CrossRef
21.
go back to reference Guzzini M, Lupariello D, Lanzetti RM, Mazza D, Ferretti A. Preliminary experience with triangular CarboFix “Piccolo” distal radius plate in wrist fractures. Clinical and radiological results. Acta Biomed. 2018;90(1S):61–6.PubMed Guzzini M, Lupariello D, Lanzetti RM, Mazza D, Ferretti A. Preliminary experience with triangular CarboFix “Piccolo” distal radius plate in wrist fractures. Clinical and radiological results. Acta Biomed. 2018;90(1S):61–6.PubMed
22.
go back to reference Katthagen JC, Ellwein A, Lutz O, Voigt C, Lill H. Outcomes of proximal humeral fracture fixation with locked CFR-PEEK plating. Eur J Orthop Surg Traumatol. 2017;27(3):351–8.CrossRef Katthagen JC, Ellwein A, Lutz O, Voigt C, Lill H. Outcomes of proximal humeral fracture fixation with locked CFR-PEEK plating. Eur J Orthop Surg Traumatol. 2017;27(3):351–8.CrossRef
23.
go back to reference Byun SE, Vintimilla DR, Bedeir YH, Dean CS, Parry JA, Hak DJ, et al. Evaluation of callus formation in distal femur fractures after carbon fiber composite versus stainless steel plate fixation. Eur J Orthop Surg Traumatol. 2020;30(6):1103–7.CrossRef Byun SE, Vintimilla DR, Bedeir YH, Dean CS, Parry JA, Hak DJ, et al. Evaluation of callus formation in distal femur fractures after carbon fiber composite versus stainless steel plate fixation. Eur J Orthop Surg Traumatol. 2020;30(6):1103–7.CrossRef
24.
go back to reference Mitchell PM, Lee AK, Collinge CA, Ziran BH, Hartley KG, Jahangir AA. Early comparative outcomes of carbon fiber-reinforced polymer plate in the fixation of distal femur fractures. J Orthop Trauma. 2018;32(8):386–90.CrossRef Mitchell PM, Lee AK, Collinge CA, Ziran BH, Hartley KG, Jahangir AA. Early comparative outcomes of carbon fiber-reinforced polymer plate in the fixation of distal femur fractures. J Orthop Trauma. 2018;32(8):386–90.CrossRef
25.
go back to reference Padolino A, Porcellini G, Guollo B, Fabbri E, Kiran Kumar GN, Paladini P, et al. Comparison of CFR-PEEK and conventional titanium locking plates for proximal humeral fractures: a retrospective controlled study of patient outcomes. Musculoskelet Surg. 2018;102(Suppl 1):49–56.CrossRef Padolino A, Porcellini G, Guollo B, Fabbri E, Kiran Kumar GN, Paladini P, et al. Comparison of CFR-PEEK and conventional titanium locking plates for proximal humeral fractures: a retrospective controlled study of patient outcomes. Musculoskelet Surg. 2018;102(Suppl 1):49–56.CrossRef
26.
go back to reference Schliemann B, Hartensuer R, Koch T, Theisen C, Raschke MJ, Kosters C, et al. Treatment of proximal humerus fractures with a CFR-PEEK plate: 2-year results of a prospective study and comparison to fixation with a conventional locking plate. J Shoulder Elbow Surg. 2015;24(8):1282–8.CrossRef Schliemann B, Hartensuer R, Koch T, Theisen C, Raschke MJ, Kosters C, et al. Treatment of proximal humerus fractures with a CFR-PEEK plate: 2-year results of a prospective study and comparison to fixation with a conventional locking plate. J Shoulder Elbow Surg. 2015;24(8):1282–8.CrossRef
27.
go back to reference Neer CS 2nd. Displaced proximal humeral fractures. I. Classification and evaluation. J Bone Jt Surg Am. 1970;52(6):1077–89.CrossRef Neer CS 2nd. Displaced proximal humeral fractures. I. Classification and evaluation. J Bone Jt Surg Am. 1970;52(6):1077–89.CrossRef
28.
go back to reference Al-Shawi AK, Smith SP, Anderson GH. The use of a carbon fiber plate for periprosthetic supracondylar femoral fractures. J Arthroplasty. 2002;17(3):320–4.CrossRef Al-Shawi AK, Smith SP, Anderson GH. The use of a carbon fiber plate for periprosthetic supracondylar femoral fractures. J Arthroplasty. 2002;17(3):320–4.CrossRef
29.
go back to reference Allemann F, Halvachizadeh S, Rauer T, Pape HC. Clinical outcomes after carbon-plate osteosynthesis in patients with distal radius fractures. Patient Saf Surg. 2019;13:30.CrossRef Allemann F, Halvachizadeh S, Rauer T, Pape HC. Clinical outcomes after carbon-plate osteosynthesis in patients with distal radius fractures. Patient Saf Surg. 2019;13:30.CrossRef
30.
go back to reference Baker D, Kadambande SS, Alderman PM. Carbon fibre plates in the treatment of femoral periprosthetic fractures. Injury. 2004;35(6):596–8.CrossRef Baker D, Kadambande SS, Alderman PM. Carbon fibre plates in the treatment of femoral periprosthetic fractures. Injury. 2004;35(6):596–8.CrossRef
31.
go back to reference Caforio M, Perugia D, Colombo M, Calori GM, Maniscalco P. Preliminary experience with Piccolo Composite, a radiolucent distal fibula plate, in ankle fractures. Injury. 2014;45(Suppl 6):S36–8.CrossRef Caforio M, Perugia D, Colombo M, Calori GM, Maniscalco P. Preliminary experience with Piccolo Composite, a radiolucent distal fibula plate, in ankle fractures. Injury. 2014;45(Suppl 6):S36–8.CrossRef
32.
go back to reference Pemberton DJ, Evans PD, Grant A, McKibbin B. Fractures of the distal femur in the elderly treated with a carbon fibre supracondylar plate. Injury. 1994;25(5):317–21.CrossRef Pemberton DJ, Evans PD, Grant A, McKibbin B. Fractures of the distal femur in the elderly treated with a carbon fibre supracondylar plate. Injury. 1994;25(5):317–21.CrossRef
33.
go back to reference Pinter ZW, Smith KS, Hudson PW, Jones CW, Hadden R, Elattar O, et al. A retrospective case series of carbon fiber plate fixation of ankle fractures. Foot Ankle Spec. 2018;11(3):223–9.CrossRef Pinter ZW, Smith KS, Hudson PW, Jones CW, Hadden R, Elattar O, et al. A retrospective case series of carbon fiber plate fixation of ankle fractures. Foot Ankle Spec. 2018;11(3):223–9.CrossRef
34.
go back to reference Tarallo L, Giorgini A, Novi M, Zambianchi F, Porcellini G, Catani F. Volar PEEK plate for distal radius fracture: analysis of adverse events. Eur J Orthop Surg Traumatol. 2020;30(7):1293–8.CrossRef Tarallo L, Giorgini A, Novi M, Zambianchi F, Porcellini G, Catani F. Volar PEEK plate for distal radius fracture: analysis of adverse events. Eur J Orthop Surg Traumatol. 2020;30(7):1293–8.CrossRef
35.
go back to reference Sproul RC, Iyengar JJ, Devcic Z, Feeley BT. A systematic review of locking plate fixation of proximal humerus fractures. Injury. 2011;42(4):408–13.CrossRef Sproul RC, Iyengar JJ, Devcic Z, Feeley BT. A systematic review of locking plate fixation of proximal humerus fractures. Injury. 2011;42(4):408–13.CrossRef
36.
go back to reference Thanasas C, Kontakis G, Angoules A, Limb D, Giannoudis P. Treatment of proximal humerus fractures with locking plates: a systematic review. J Shoulder Elbow Surg. 2009;18(6):837–44.CrossRef Thanasas C, Kontakis G, Angoules A, Limb D, Giannoudis P. Treatment of proximal humerus fractures with locking plates: a systematic review. J Shoulder Elbow Surg. 2009;18(6):837–44.CrossRef
37.
go back to reference Lowe JB, Monazzam S, Walton B, Nelson E, Wolinsky PR. How to use fluoroscopic imaging to prevent intraarticular screw perforation during locked plating of proximal humerus fractures: a cadaveric study. J Orthop Trauma. 2015;29(10):e401–7.CrossRef Lowe JB, Monazzam S, Walton B, Nelson E, Wolinsky PR. How to use fluoroscopic imaging to prevent intraarticular screw perforation during locked plating of proximal humerus fractures: a cadaveric study. J Orthop Trauma. 2015;29(10):e401–7.CrossRef
38.
go back to reference Mugnai R, Tarallo L, Capra F, Catani F. Biomechanical comparison between stainless steel, titanium and carbon–fiber reinforced polyetheretherketone volar locking plates for distal radius fractures. Orthop Traumatol Surg Res. 2018;104(6):877–82.CrossRef Mugnai R, Tarallo L, Capra F, Catani F. Biomechanical comparison between stainless steel, titanium and carbon–fiber reinforced polyetheretherketone volar locking plates for distal radius fractures. Orthop Traumatol Surg Res. 2018;104(6):877–82.CrossRef
39.
go back to reference Chloros GD, Wiesler ER, Poehling GG. Current concepts in wrist arthroscopy. Arthroscopy. 2008;24(3):343–54.CrossRef Chloros GD, Wiesler ER, Poehling GG. Current concepts in wrist arthroscopy. Arthroscopy. 2008;24(3):343–54.CrossRef
40.
go back to reference Ricci WM, Streubel PN, Morshed S, Collinge CA, Nork SE, Gardner MJ. Risk factors for failure of locked plate fixation of distal femur fractures: an analysis of 335 cases. J Orthop Trauma. 2014;28(2):83–9.CrossRef Ricci WM, Streubel PN, Morshed S, Collinge CA, Nork SE, Gardner MJ. Risk factors for failure of locked plate fixation of distal femur fractures: an analysis of 335 cases. J Orthop Trauma. 2014;28(2):83–9.CrossRef
Metadata
Title
Fracture fixation in extremity trauma with carbon fiber-reinforced polyetheretherketone (CFR-PEEK) plates: evidence today
Authors
George D. Chloros
Apostolos D. Prodromidis
Jo Wilson
Peter V. Giannoudis
Publication date
06-09-2021
Publisher
Springer Berlin Heidelberg
Published in
European Journal of Trauma and Emergency Surgery / Issue 3/2022
Print ISSN: 1863-9933
Electronic ISSN: 1863-9941
DOI
https://doi.org/10.1007/s00068-021-01778-x

Other articles of this Issue 3/2022

European Journal of Trauma and Emergency Surgery 3/2022 Go to the issue